Reliability of Point-of-Care Venous Blood Gas Testing for Potassium and Sodium Compared to Central Laboratory Results in Emergency Patients: A Clinical Audit

Open access
Lasfer, Chafika, Lahib, Amal, Soliman, Kariem, Gouda, Shahinaz·2025·Cureus

Abstract

Background Point-of-care testing (POCT) is widely employed in emergency departments (EDs) for rapid clinical decision-making. However, the reliability of POCT for assessing critical electrolytes like potassium (K⁺) and sodium (Na⁺), compared to central laboratory methods, remains under scrutiny. Objective This quality improvement (QI) study aimed to evaluate and improve the reliability of POC venous blood gas (VBG) testing for K⁺ and Na⁺ by comparing results with central laboratory values. Structured within a Plan-Do-Study-Act (PDSA) cycle, the objectives included quantifying the analytical concordance using Pearson correlation and Bland-Altman methods, identifying systematic bias or clinical outliers, and assessing turnaround time (TAT) to inform iterative improvements in POCT protocol integration within the ED. Methods A retrospective audit was conducted on 120 patients (N = 120, 100%) at Fakeeh University Hospital, Dubai, from March to May 2023. Paired K⁺ and Na⁺ values from POCT VBG analyzers and laboratory reports were compared. Statistical analysis included scatter plots, Pearson correlation, and Bland-Altman agreement. Results Of the 120 patients, 65 (54.2%) were male and 55 (45.8%) were female, with a mean age of 47.3 years (SD = 16.1). Chief presentations included chest pain (N = 38, 31.7%), dyspnea (N = 32, 26.7%), vomiting (N = 23, 19.2%), and signs of dehydration (N = 27, 22.5%). POCT K⁺ and Na⁺ values showed strong correlation with laboratory values (r = 0.91 and r = 0.88, respectively). The mean bias for K⁺ was +0.12 mmol/L, and for Na⁺ it was -1.2 mmol/L. TAT was significantly shorter for POCT (mean = 2.4 minutes) versus laboratory results (mean = 12.3 minutes), with a median delay of 10 minutes. Conclusion POCT demonstrated high reliability for K⁺ and Na⁺ evaluation in EDs. However, caution is advised when interpreting extreme values, especially hyperkalemia (> 5.5 mmol/L). Central laboratory confirmation should be considered in borderline or critical cases.

AI snapshot

Written from the paper’s full text.

TL;DR

This retrospective QI audit of 120 adult ED patients found POC venous blood gas K⁺ and Na⁺ strongly correlated with central laboratory results (r = 0.91 and r = 0.88) and reduced turnaround time by about 10 minutes. However, POCT slightly overestimated K⁺ and underestimated Na⁺, including false-positive hyperkalemia in 3 patients, supporting confirmatory lab testing for borderline or critical values.

Key findings

  • Potassium: POCT vs lab correlation was r = 0.91 (p < 0.001), with mean bias +0.12 mmol/L and 95% limits of agreement −0.89 to +1.13 mmol/L; 3/120 patients (2.5%) had POCT K⁺ > 5.5 mmol/L not confirmed by lab.
  • Sodium: correlation was r = 0.88 (p < 0.001), with mean bias −1.2 mmol/L and 95% limits of agreement −5.4 to +3.0 mmol/L; 11/120 patients (9.2%) had low-normal Na⁺ (135–138 mmol/L) underestimated by POCT.
  • Turnaround time: POCT mean was 2.4 minutes (SD 0.6) versus central lab 12.3 minutes (SD 3.1), with a median delay of approximately 10 minutes.
  • Cohort: mean age 47.3 years (SD 16.1); 65/120 (54.2%) male and 55/120 (45.8%) female.

Methods

  • Retrospective clinical audit within a PDSA/QI framework at Fakeeh University Hospital, Dubai, March–May 2023; 136 cases screened, 16 excluded, final cohort of 120 adults (≥18 years) with paired K⁺ and Na⁺ POCT and central lab results collected within 30 minutes.
  • POCT used the Cobas b 221 blood gas analyzer; central lab used standard ion-selective electrode technology; data retrieved from electronic medical records (YASASII), with TAT calculated from electronic timestamps.
  • Analysis included descriptive statistics, Pearson correlation, Bland-Altman agreement with 95% limits of agreement, and outlier analysis near clinical thresholds; significance set at p < 0.05 using SPSS v28 and Python.

Limitations

  • Stated: single urban tertiary care center, modest cohort of 120 adults; pediatric, obstetric, and oncology patients were excluded, limiting generalizability; only K⁺ and Na⁺ were assessed, not other VBG parameters such as chloride, bicarbonate, or lactate.
  • Stated: retrospective design did not measure downstream clinical outcomes of POCT-lab discrepancies, such as treatment changes, length of stay, or adverse events; possible spectrum bias because the cohort comprised only ED patients.
  • Evident: no formal sample size calculation was performed, and the paper notes absence of systematic blinding to lab values during POCT interpretation as a risk of bias.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.7759/cureus.89106

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

13 of 19 cited sources checked · 1 supported · 2 flagged · accuracy not yet measured

  • Premisecitation

    Whole-blood POCT analyzers tend to report slightly elevated potassium levels compared to serum-based laboratory systems due to differences in sample handling and electrode sensitivity.

    “This minor overestimation bias is consistent with earlier work by Chhapola et al. [3] and Jain et al. [4], who observed that whole-blood POCT analyzers tend to report slightly elevated K⁺ levels compared to serum-based lab systems due to differences in sample handling and electrode sensitivity.”
    Quote in the paperSubstantial claim1 of 2 sources checked
    • cites Chhapola et al. [3] Chhapola V, Kanwal SK, Sharma R, Kumar V. A comparative study on reliability of point of care sodium and potassium estimation in a pediatric intensive care unit. Indian J Pediatr. 2013, 80:731-5. 10.1007/s12098-013-0977-z source
    • cites Jain et al. [4] Jain A, Subhan I, Joshi M. Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes. Int J Emerg Med. 2009, 2:117-20. 10.1007/s12245-009-0091-1 source

      The Results report the opposite direction: mean potassium was 3.7 mmol/l on the blood gas analyzer versus 3.9 mmol/l on the laboratory auto-analyzer, and 71.5% of blood gas readings were lower. The Introduction describes differences in sample type and electrode technology, but does not attribute an upward potassium bias to them. The paper’s conclusion says the potassium measurements did not differ significantly. (LLM, conf=0.90)

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
  • Premisecitation

    Delays in identifying electrolyte abnormalities in critically ill patients can significantly increase the risk of morbidity and mortality.

    “In critically ill patients, delays in identifying electrolyte abnormalities can significantly increase the risk of morbidity and mortality, making rapid and accurate testing essential [1].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [1] Balcı AK, Koksal O, Kose A, Armagan E, Ozdemir F, Inal T, Oner N. General characteristics of patients with electrolyte imbalance admitted to emergency department. World J Emerg Med. 2013, 4:113-6. 10.5847/wjem.j.issn.1920-8642.2013.02.005 source

      The Discussion (page unavailable) says sodium imbalances are associated with increased mortality in patients needing intensive care and notes that hypomagnesemia is often overlooked. It does not show that delays in identifying electrolyte abnormalities increase morbidity or mortality, or establish a need for rapid testing. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • Premisecitation

    Multiple factors contribute to discrepancies between POCT and laboratory values: variation in sample matrix, pre-analytical handling, device calibration intervals, and analytical methodology.

    “Multiple factors contribute to discrepancies between POCT and laboratory values: variation in sample matrix (whole blood vs. serum/plasma), pre-analytical handling, device calibration intervals, and analytical methodology [4,5].”
    Quote in the paperSubstantial claim2 of 2 sources checked
    • cites [4] Jain A, Subhan I, Joshi M. Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes. Int J Emerg Med. 2009, 2:117-20. 10.1007/s12245-009-0091-1 source

      The Introduction (Table 1; page not provided) contrasts whole blood with serum, direct with indirect electrode methods, and different sample dilutions; it also explains how heparin dilution can lower measured electrolyte values. Materials and methods (page not provided) describes sample handling and says both analyzers were calibrated before each measurement cycle. It does not identify variation in device calibration intervals as a cause of discrepancies. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
    • cites [5] Nigro M, Valli G, Marchionne ML, et al.. Is there a risk of misinterpretation of potassium concentration from undetectable hemolysis using a POCT blood gas analyzer in the emergency department?. Medicina (Kaunas). 2022, 59:66. 10.3390/medicina59010066 source

      The Introduction and Discussion support pre-analytical handling as a source of error: collection, transport, storage, and hemolysis can alter measured potassium. The Introduction also notes that central laboratories routinely check for hemolysis whereas blood-gas POCT systems do not. Our copy does not identify sample-matrix variation or device calibration intervals as causes of POCT–laboratory discrepancies, or establish analytical methodology more generally as a cause. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
  • Premisecitation

    POCT often underestimates sodium concentrations by 1-4 mmol/L due to methodological differences in ion-selective electrodes.

    “Na⁺ measurement discrepancies are also well documented, with several researchers noting that POCT often underestimates Na⁺ concentrations by 1-4 mmol/L due to methodological differences in ion-selective electrodes [9,10].”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [9] Goldstein L, Wells M, Vincent-Lambert C. A randomized controlled trial to assess the impact of upfront point-of-care testing on emergency department treatment time. Am J Clin Pathol. 2018, 150:224-34. 10.1093/ajcp/aqy042 source

      The 'Abbott Point of Care i-STAT System' section (page not provided) lists sodium among the tests performed, but gives no sodium comparison with laboratory results. Searches for sodium underestimation, a 1–4 mmol/L difference, and ion-selective electrode methods found no passage addressing the claimed discrepancy or its cause. (LLM, conf=0.76)

      Experimental — accuracy not yet measured
    • cites [10] Nichols JH, Alter D, Chen Y, et el.. AACC guidance document on management of point-of-care testing. J Appl Lab Med. 2020, 1-20. source
  • Premisecitation

    Sodium underestimation patterns are supported by studies attributing discrepancies to differences in ion-selective electrode calibration, matrix effects, and ambient temperature and humidity.

    “These patterns are supported by studies such as Kapoor et al. [7] and Allardet-Servent et al. [8], who attribute these discrepancies to differences in ISE calibration, matrix effects, and the influence of ambient temperature and humidity on POCT devices.”
    Quote in the paperNot a substantial claim2 of 2 sources checked
    • cites Kapoor et al. [7] Kapoor D, Srivastava M, Singh P. Point of care blood gases with electrolytes and lactates in adult emergencies. Int J Crit Illn Inj Sci. 2014, 4:216-22. 10.4103/2229-5151.141411 source

      In the Electrolytes—Sodium, Potassium, Chloride section (page unavailable), the review reports sodium underestimation in a pediatric ICU study and says electrolyte discrepancies may reflect device calibrators, sample differences, and direct versus indirect ion-selective electrodes affected by proteins. It does not attribute the discrepancies to ambient temperature or humidity, and it presents the proposed causes cautiously rather than as established explanations. (LLM, conf=0.87)

      Experimental — accuracy not yet measured
    • cites Allardet-Servent et al. [8] Allardet-Servent J, Lebsir M, Dubroca C, et al.. Point-of-care versus central laboratory measurements of hemoglobin, hematocrit, glucose, bicarbonate and electrolytes: a prospective observational study in critically ill patients. PLoS One. 2017, 12:e0169593. 10.1371/journal.pone.0169593 source

      The Discussion (page not provided) attributes the sodium difference to direct versus indirect ISE measurement and the effect of protein content on dilution; it reports sodium averaging 1.3 mmol/L higher with indirect ISE. This supports a sample-matrix explanation for lower point-of-care readings relative to the laboratory, but the paper does not attribute the discrepancy to ISE calibration or ambient temperature and humidity. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
  • Classified as probably not a substantial claim. 12
  • May combine two claims. 10
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Some earlier studies noted positive bias in POCT potassium readings, especially in hemolyzed samples.

    “However, some of these studies also noted positive bias in POCT readings, especially in hemolyzed samples [7,8].”
    Quote in the paperNot a substantial claim2 of 2 sources checked
    • cites [7] Kapoor D, Srivastava M, Singh P. Point of care blood gases with electrolytes and lactates in adult emergencies. Int J Crit Illn Inj Sci. 2014, 4:216-22. 10.4103/2229-5151.141411 source

      The Electrolytes > Sodium, Potassium, Chloride section reports that earlier studies found differences between POCT and central-laboratory potassium readings. It does not report positive bias or identify hemolyzed samples as a particular cause; the specific pediatric ICU study it discusses found underestimation instead. No page number is available in our copy. (LLM, conf=0.79)

      Experimental — accuracy not yet measured
    • cites [8] Allardet-Servent J, Lebsir M, Dubroca C, et al.. Point-of-care versus central laboratory measurements of hemoglobin, hematocrit, glucose, bicarbonate and electrolytes: a prospective observational study in critically ill patients. PLoS One. 2017, 12:e0169593. 10.1371/journal.pone.0169593 source

      The Results (Potassium; page not available) report a mean potassium difference of 0.2 mmol/L, with the laboratory reading higher than POC. The Discussion likewise reports more high-potassium readings by the laboratory than by POC. The paper does not link potassium bias to hemolyzed samples; its discussion of hemolysis concerns hemoglobin measurements. (LLM, conf=0.86)

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
  • Classified as probably not a substantial claim.

    Premisecitation

    POCT has emerged as a central component in modern emergency diagnostics primarily due to its rapid turnaround time and ability to provide near-immediate data.

    “Point-of-care testing (POCT) has emerged as a central component in modern emergency diagnostics, primarily due to its rapid turnaround time (TAT) and its ability to provide near-immediate data to guide time-sensitive interventions [2].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [2] Price CP. Point of care testing. BMJ. 2001, 322:1285-8. 10.1136/bmj.322.7297.1285 source

      The opening section says POCT aims to generate results quickly so appropriate treatment can be implemented, and the “Accident and emergency” section says it has potential to speed decisions and triage. But that section does not establish POCT as a central component of emergency diagnostics; it says the main emergency studies found little clinical benefit compared with laboratory testing. (LLM, conf=0.88)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Premisecitation

    Hemolysis, often undetectable during POCT, and sampling inconsistencies can result in elevated potassium readings.

    “These discrepancies are also aligned with findings from Nigro et al. [5], who reported that hemolysis, often undetectable during POCT, and sampling inconsistencies can result in elevated K⁺ readings, potentially triggering unnecessary interventions.”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites Nigro et al. [5] Nigro M, Valli G, Marchionne ML, et al.. Is there a risk of misinterpretation of potassium concentration from undetectable hemolysis using a POCT blood gas analyzer in the emergency department?. Medicina (Kaunas). 2022, 59:66. 10.3390/medicina59010066 source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation

    Decentralized POCT systems, when implemented under controlled conditions with appropriate quality assurance, can alleviate ED crowding and improve triage flow.

    “Plebani et al. [11] emphasized that, when implemented under controlled conditions with appropriate quality assurance, decentralized POCT systems can alleviate ED crowding and improve triage flow.”
    Quote in the paperNot a substantial claim1 of 1 sources checked
    • cites Plebani et al. [11] Plebani M, Nichols JH, Luppa PB, et al.. Point-of-care testing: state-of-the art and perspectives. Clin Chem Lab Med. 2025, 63:35-51. 10.1515/cclm-2024-0675 source

      The section “POCT and cardiovascular diseases: a moving target” (page unavailable) reports shorter emergency-department stays and more discharges with certain POCTs, while calling for quality systems. The disaster-response section (page unavailable) discusses faster triage and balancing ambulance versus emergency-room workload. These passages support the general potential benefit, but do not establish that quality-assured decentralized POCT alleviates ED crowding and improves triage flow within the ED. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
  • Classified as probably not a substantial claim.

    Premisecitation

    Clinical guidelines generally recommend confirmatory laboratory testing when POCT values fall near critical therapeutic thresholds.

    “Consequently, clinical guidelines generally recommend confirmatory laboratory testing when POCT values fall near critical therapeutic thresholds [11].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [11] Plebani M, Nichols JH, Luppa PB, et al.. Point-of-care testing: state-of-the art and perspectives. Clin Chem Lab Med. 2025, 63:35-51. 10.1515/cclm-2024-0675 source

      The “Quality Indicators for POCT” section says critically high POCT glucose results should be repeated before clinical action. It does not say that clinical guidelines generally recommend confirmatory laboratory testing for POCT values near therapeutic thresholds; neither the testing method nor the threshold condition is established by that passage. (LLM, conf=0.82)

      Experimental — accuracy not yet measured
  • May combine two claims.

    Findingown data

    In this audit, the median time difference between POCT and central laboratory turnaround times was approximately 10 minutes, with POCT being faster.

    “TAT was significantly shorter for POCT (mean = 2.4 minutes) versus laboratory results (mean = 12.3 minutes), with a median delay of 10 minutes.”
    • Also stated in: § Results · plain
      “The median time difference of approximately 10 minutes highlights the practical advantage of POCT in emergent settings, where minutes can influence treatment pathways.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Expected laboratory analytical variation is ±0.1-0.2 mmol/L for K+ and ±1-2 mmol/L for Na+ according to CLIA and manufacturer standards.

    “According to Clinical Laboratory Improvement Amendments (CLIA) and manufacturer standards, the expected laboratory analytical variation is ±0.1-0.2 mmol/L for K⁺ and ±1-2 mmol/L for Na⁺ [10].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [10] Nichols JH, Alter D, Chen Y, et el.. AACC guidance document on management of point-of-care testing. J Appl Lab Med. 2020, 1-20. source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    Only K+ and Na+ were assessed; other VBG-derived parameters such as chloride, bicarbonate, and lactate were not evaluated.

    “Only K⁺ and Na⁺ were assessed; other VBG-derived parameters, such as chloride, bicarbonate, and lactate, frequently used in emergency care, were not evaluated here.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data

    The observed agreement and bias for K+ and Na+ are clinically acceptable.

    “Among the 120 patients (100%) assessed, statistically significant and clinically acceptable levels of agreement were observed for both electrolytes, thereby supporting the diagnostic utility of POCT in acute care settings.”
    • Also stated in: § Conclusions · boosted
      “This QI project demonstrated that POCT using VBG analyzers provides K⁺ and Na+ measurements with strong statistical agreement and clinically acceptable bias compared to central laboratory values in adult ED patients.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    The audit was conducted at a single urban tertiary care center and included a modest cohort of 120 adult patients.

    “This audit was conducted at a single urban tertiary care center and included a modest cohort of 120 adult patients.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductown data

    In this audit, 120 adult patients were included after screening 136 and excluding 16.

    “Out of 136 initially screened cases, a total of 16 patients were excluded, resulting in a final cohort of 120 adult patients (100%) included in the analysis.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Study design and conductnone given

    The audit lacked systematic blinding to laboratory values during POCT interpretation, posing a risk of bias.

    “The absence of systematic blinding to lab values during POCT interpretation poses a risk of bias, which should be addressed in future protocols.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Premisecitation

    Clinically acceptable bias thresholds are ≤0.3 mmol/L for K+ and ≤3 mmol/L for Na+ according to AACC and CLSI.

    “In accordance with guidelines from the American Association for Clinical Chemistry (AACC) and the Clinical and Laboratory Standards Institute (CLSI), clinically acceptable bias was defined as: ≤ 0.3 mmol/L for K⁺; ≤ 3 mmol/L for Na⁺ [10].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [10] Nichols JH, Alter D, Chen Y, et el.. AACC guidance document on management of point-of-care testing. J Appl Lab Med. 2020, 1-20. source
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    The Cobas b 221 blood gas analyzer was calibrated daily following manufacturer and institutional protocols: system calibration every 24 hours, one-point calibration every 60 minutes, and two-point calibration every 12 hours.

    “The analyzer was calibrated daily following manufacturer and institutional protocols: system calibration every 24 hours, one-point calibration every 60 minutes, and two-point calibration every 12 hours.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductreasoning

    This was a retrospective analysis and did not measure downstream clinical outcomes of POCT-lab discrepancies.

    “Furthermore, this was a retrospective analysis and did not measure the downstream clinical outcomes of POCT-lab discrepancies, such as changes in treatment, length of stay, or adverse events.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Study design and conductreasoning

    In this audit, no formal sample size calculation was performed.

    “Because this audit aimed to assess real-world process performance over a defined time window, no formal sample size calculation was performed.”
    Quote in the paperSubstantial claim
16 of 16
  • The paper relies on this without support.

    VBG-derived electrolyte readings are often used by clinicians to initiate treatment even before central laboratory results become available.

    “VBG-derived electrolyte readings are often used by clinicians to initiate treatment even before central laboratory results become available.”
  • The paper relies on this without support.

    Electrolyte imbalances such as dyskalemia and hyponatremia are among the most frequent laboratory abnormalities observed in emergency department presentations.

    “Electrolyte imbalances such as dyskalemia (potassium disturbances) and hyponatremia (low sodium levels) are among the most frequent laboratory abnormalities observed in emergency department (ED) presentations.”
  • The paper relies on this without support.

    The reliability of POCT for electrolytes compared to central laboratory analyzers has been questioned.

    “Despite its efficiency, the reliability of POCT for electrolytes compared to central laboratory analyzers has been questioned.”
3 of 3

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Potassium discrepancies exceed the stated acceptable threshold

    Major

    The paper says the observed discrepancies were within the ≤ 0.3 mmol/L potassium threshold, yet its reported 95% limits of agreement extend to −0.89 and +1.13 mmol/L.

1 of 1

Cites

Flagged first, then by citations

Authors
Lasfer, C., Lahib, A., Soliman, K., & Gouda, S.
Venue
Cureus
Published
2025-07-31
Volume
17
Issue
7
Pages
e89106-e89106
ISSN
2168-8184 (ISSN-L)
OpenAlex
W4412810177
Cited by
0
References
12
FWCI
0.00
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • Cureus (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Europe PMC (PubMed Central) (submittedVersion, cc-by)
  • PubMed Central (submittedVersion, cc-by)
Topics
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Bacterial Identification and Susceptibility Testing
Keywords
point-of-care testing, venous blood gas, potassium, sodium, emergency department, central laboratory comparison, test reliability, turnaround time, hyperkalemia, electrolyte measurement, quality improvement, Bland-Altman agreement, Pearson correlation, dehydration, chest pain, dyspnea, vomiting, PDSA cycle

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-14.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-14.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review