Initial Experimental Characterization of a New Ultra-High Resolution FTICR Cell with Dynamic Harmonization
The DOI matches its Crossref record; 17 of the 17 references it cites were found in a registry or an index.
Abstract
A new Fourier transform ion cyclotron resonance (FTICR) cell based on completely new principles of formation of the effective electric potential distribution in Penning type traps, Boldin and Nikolaev (Proceedings of the 58th ASMS Conference, 2010), Boldin and Nikolaev (Rapid Commun Mass Spectrom 25:122-126, 2011) is constructed and tested experimentally. Its operation is based on the concept of electric potential space-averaging via charged particle cyclotron motion. Such an averaging process permits an effective electric force distribution in the entire volume of a cylindrical Penning trap to be equal to its distribution in the field created by hyperbolic electrodes in an ideal Penning trap. The excitation and detection electrodes of this new cell are shaped for generating a quadratic dependence on axial coordinates of an averaged (along cyclotron motion orbit) electric potential at any radius of the cyclotron motion. These electrodes together with the trapping segments form a cylindrical surface like in a conventional cylindrical cell. In excitation mode this cell being elongated behaves almost like an open cylindrical cell of the same length. It is more effective in ion motion harmonization at larger cyclotron radii than a Gabrielse et al.-type (Int J Mass Spectrom Ion Processes 88:319-332, 1989) cylindrical cell with four compensation sections. A mass resolving power of more than twenty millions of reserpine (m/z 609) and more than one million of highly charged BSA molecular ions (m/z 1357) has been obtained in a 7T magnetic field.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Electrodynamics of Continuous Media
Landau, L. D., Lifshitz, E. M., & King, A. L. (1961) · American Journal of Physics · doi:10.1119/1.1937882 · Cited by 10,997
Introduction to Electrodynamics
Griffiths, D. J. (2017) · Cambridge University Press eBooks · doi:10.1017/9781108333511 · Cited by 3,408 · cited 1×
“The method we used to obtain these theoretical spectra is based on Green’s reciprocity theorem [ 19 ].”
MethodFourier transform ion cyclotron resonance mass spectrometry: A primer
Marshall, A. G., Hendrickson, C. L., & Jackson, G. S. (1998) · Mass Spectrometry Reviews · doi:10.1002/(sici)1098-2787(1998)17:1<1::aid-mas1>3.0.co;2-k · Cited by 1,990 · cited 3×
“The theoretical resolving power at zero pressure limit can be calculated using the following numerical-value equation [9]:”
BackgroundMethodInfluentialLANDAU, L. D., & LIFSHITZ, E. M. (1984) · Electrodynamics of Continuous Media · doi:10.1016/b978-0-08-030275-1.50007-2 · Cited by 879
Open-endcap Penning traps for high precision experiments
Gabrielse, G., Haarsma, L., & Rolston, S. L. (1989) · International Journal of Mass Spectrometry and Ion Processes · doi:10.1016/0168-1176(89)85027-x · Cited by 288 · cited 1×
“implemented this approach by inserting compensation rings into an open cylindrical cell [3] between the center section and the trapping sections so that it became a five-section compensated cell.”
MethodBoldin, I. A., & Nikolaev, E. N. (2011) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.4838 · Cited by 128
Self-citationTrapped-ion cell with improved dc potential harmonicity for FT-ICR MS
Tolmachev, A. V., Robinson, E. W., Wu, S., Kang, H., Lourette, N. M., Paša-Tolić, L., & Smith, R. D. (2008) · Journal of the American Society for Mass Spectrometry · doi:10.1016/j.jasms.2008.01.006 · Cited by 108 · cited 8×
“The optimum compensation electrode voltage settings for the measurements evaluated (in broadband detection mode) as shown in Figure 5 have been found to be the same as described in reference [ 12 ].”
BackgroundMethodInfluentialBorts, D. J., & Bowers, L. D. (2000) · Journal of Mass Spectrometry · doi:10.1002/(sici)1096-9888(200001)35:1<85::aid-jms910>3.0.co;2-7 · Cited by 98 · cited 1×
“[15], Barlow and Tinkle [16] at Pacific Northwest National Laboratory (PNNL).”
BackgroundHigh mass resolution with a new variable anharmonicity Penning trap
Van Dyck, R. S., Wineland, D. J., Ekstrom, P. A., & Dehmelt, H. G. (1976) · Applied Physics Letters · doi:10.1063/1.88793 · Cited by 87
Parametric mode operation of a hyperbolic Penning trap for Fourier transform mass spectrometry
Rempel, D. L., Ledford, E. B., Huang, S. K., & Gross, M. L. (1987) · Analytical Chemistry · doi:10.1021/ac00147a018 · Cited by 83 · cited 1×
“A nearly ideal hyperbolic field is provided by a 3D hyperbolic trap (see for instance [10, 11]), but it has the major disadvantage of not effective usage of room temperature bore space of high magnetic field homogeneity, i.”
BackgroundНиколаев, Е. Н., Heeren, R. M. A., Popov, A. M., Pozdneev, A. V., & Chingin, K. (2007) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.3234 · Cited by 81 · cited 1×
“Signal duration does also significantly increase when the number of ions in a cell is sufficient for phase locking [4].”
BackgroundAn electrically compensated trap designed to eighth order for FT-ICR mass spectrometry
Brustkern, A. M., Rempel, D. L., & Gross, M. L. (2008) · Journal of the American Society for Mass Spectrometry · doi:10.1016/j.jasms.2008.05.016 · Cited by 55 · cited 3×
“The compensated cell described in reference [12] is a seven-section cell having two compensation rings on each side of the center section, whereas the one by Gross et al. [ 13 ] is a nine-section cell with three compensation rings on each side of the center section.”
BackgroundInfluentialJackson, G. S., White, F. M., Guan, S., & Marshall, A. G. (1999) · Journal of the American Society for Mass Spectrometry · doi:10.1016/s1044-0305(99)00048-3 · Cited by 30 · cited 2×
“using an ion motion simulation program [14] with ion–ion and ion–image charge interactions turned off.”
MethodImprovement of the electric field in the cylindrical trapped-ion cell
Naito, Y., Fujiwara, M., & Inoue, M. (1992) · International Journal of Mass Spectrometry and Ion Processes · doi:10.1016/0168-1176(92)85047-4 · Cited by 30 · cited 5×
“This indicates that in the new ICR cell type the ions are exposed to a much more uniform electric field over the complete cell area as predicted by theory [ 2 ].”
BackgroundMethodInfluentialAnalysis of harmonics for an elongated FTMS cell with multiple electrode detection
Николаев, Е. Н., Rakov, V. S., & Futrell, J. H. (1996) · International Journal of Mass Spectrometry and Ion Processes · doi:10.1016/s0168-1176(96)04457-6 · Cited by 29 · cited 1×
“The second method is based on measurements of the signal intensity distribution for the harmonics of ions with the same m/z, which contains information about the cyclotron radius [17, 18].”
MethodWeisbrod, C. R., Kaiser, N. K., Skulason, G. E., & Bruce, J. E. (2008) · Analytical Chemistry · doi:10.1021/ac800535e · Cited by 28 · cited 1×
“The first one is to reduce a radial electric field in the region of ion detection [5–8].”
Background“Linearizing” an ion cyclotron resonance cell
Barlow, S. E., & Tinkle, M. D. (2002) · Review of Scientific Instruments · doi:10.1063/1.1518787 · Cited by 23 · cited 2×
“Other segmentation methods involve a matrix cell by Marshall’s group [14] and segmented cells (based on cell segmentation in axial direction) developed by Bruce et al. [15], Barlow and Tinkle [16] at Pacific Northwest National Laboratory (PNNL).”
Background
Cited by
Mass spectrometry-based proteomics: existing capabilities and future directions
Angel, T. E., Aryal, U. K., Hengel, S. M., Baker, E. S., Kelly, R. T., Robinson, E. W., & Smith, R. D. (2012) · Chemical Society Reviews · doi:10.1039/c2cs15331a · Cited by 441
Recent developments in liquid chromatography–mass spectrometry and related techniques
Holčapek, M., Jirásko, R., & Lísa, M. (2012) · Journal of Chromatography A · doi:10.1016/j.chroma.2012.08.072 · Cited by 330
Hendrickson, C. L., Quinn, J. P., Kaiser, N. K., Smith, D. F., Blakney, G. T., Chen, T., Marshall, A. G., Weisbrod, C. R., & Beu, S. C. (2015) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-015-1182-2 · Cited by 278
Cho, Y., Ahmed, A., Islam, A., & Kim, S. (2015) · Mass Spectrometry Reviews · doi:10.1002/mas.21438 · Cited by 245
Global chemical analysis of biology by mass spectrometry
Aksenov, A. A., da Silva, R., Knight, R., Lopes, N. P., & Dorrestein, P. C. (2017) · Nature Reviews Chemistry · doi:10.1038/s41570-017-0054 · Cited by 230
Metadata
- Authors
- Nikolaev, E. N., Boldin, I. A., Jertz, R., & Baykut, G.
- Venue
- Journal of the American Society for Mass Spectrometry
- Publisher
- American Chemical Society (ACS)
- Published
- 2011-04-19
- Volume
- 22
- Issue
- 7
- Pages
- 1125-1133
- ISSN
- 1044-0305 (ISSN-L)
- OpenAlex
- W1965341129
- Cited by
- 164
- References
- 17
- FWCI
- 7.39
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.