The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the “defibrillation threshold” concept
Open accessThe DOI matches its Crossref record; 13 of the 13 references it cites were found in a registry or an index.
Abstract
The traditional assessment of the energy required for successful ventricular defibrillation involves the measurement of a “defibrillation threshold” (DFT), implying a clear-cut distinction between effective and ineffective energies. We examined the relationship between delivered energy and the likelihood of successful defibrillation in 10 open-chest pentobarbital anesthetized dogs, with the use of an internal spring/patch electrode system. An initial DFT was determined by decreasing the energy discharged until a failure first occurred (10.3 ± 3.4 J). Six energy levels in 1 to 2 J increments were then selected surrounding this value and each was administered eight times in balanced random order (total 48 trials). The relationship between energy and percent success in defibrillation exhibited a shallow slope, with a gradual increase in success from 0% to 100% over several energy increments. The initial DFTs showed actual success rates varying from 25% to 87.5% (mean 71 ± 26%). The results were fitted to a sigmoidal dose-response curve by logistic regression analysis and the energy associated with 50% success (E50) and 80% success (E80) was determined, as no single value for DFT could be defined in any animal. In 12 other dogs, a defibrillation curve was similarly constructed at baseline and was repeated after 90 minutes. No significant change in E50 (5.0 ± 2.1 J vs 5.2 ± 2.7 J) or E80 (6.3 ± 2.5 J vs 6.6 ± 3.2 J) was observed. We conclude that no unique DFT exists, and that the relation between the likelihood of successful defibrilliation and dellvered energy is best described by a “dose-response” curve requiring multiple trials for evaluation. Such curves are reproducible and establish a sensitive method for evaluating changes in defibrillation energy requirements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Cardiac electrophysiology and arrhythmias
Zipes, D. P., Jalife, J., & Moe, G. K. (1985) · Grune & Stratton eBooks · Cited by 944
Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardium
Zipes, D. P., Fischer, J., King, R. M., deB. Nicoll, A., & Jolly, W. W. (1975) · The American Journal of Cardiology · doi:10.1016/0002-9149(75)90865-6 · Cited by 487
SAS supplemental library user's guide
Reinhardt, P. S. (1980) · Medical Entomology and Zoology · Cited by 247
Elevation of ventricular defibrillation threshold in dogs by antiarrhythmic drugs
Babbs, C. F., Yim, G. K. W., Whistler, S. J., Tacker, W. A., & Geddes, L. A. (1979) · American Heart Journal · doi:10.1016/0002-8703(79)90047-4 · Cited by 127
Effect of pentobarbital anesthesia on ventricular defibrillation threshold in dogs
Babbs, C. F. (1978) · American Heart Journal · doi:10.1016/0002-8703(78)90364-2 · Cited by 87
Connell, P. N., Ewy, G. A., Dahl, C. F., & Ewy, M. D. (1973) · Journal of Electrocardiology · doi:10.1016/s0022-0736(73)80053-6 · Cited by 83
Lésigne, C., Levy, B. I., Saumont, R., et al. (1976) · Medical & Biological Engineering & Computing · doi:10.1007/bf02477038 · Cited by 68
Accepted, controversial, and speculative aspects of ventricular defibrillation
Crampton, R. S. (1980) · Progress in Cardiovascular Diseases · doi:10.1016/0033-0620(80)90020-1 · Cited by 67
Deeb, G. M., Hardesty, R. L., Griffith, B. P., Thompson, M. E., Heilman, M. S., & Myerowitz, R. L. (1983) · The Annals of Thoracic Surgery · doi:10.1016/s0003-4975(10)61585-8 · Cited by 59
Determinants of intracardiac current in defibrillation. Experimental studies in dogs.
HOYT, R. H., Grayzel, J., & Kerber, R. E. (1981) · Circulation · doi:10.1161/01.cir.64.4.818 · Cited by 49
Cited by
Echt, D. S., Black, J. N., Barbey, J. T., Coxe, D. R., & Cato, E. (1989) · Circulation · doi:10.1161/01.cir.79.5.1106 · Cited by 199
Guidelines for advanced life support
Chamberlain, D., Bossaert, L., Carli, P., Edgren, E., Ekstrom, L., Hapnes, S., Holmberg, S., Koster, R., Lindner, K., Pasqualucci, V., et al. (1992) · Resuscitation · doi:10.1016/0300-9572(92)90016-6 · Cited by 167
Pires, L. A., Lehmann, M. H., Steinman, R. T., Baga, J. J., & Schuger, C. D. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(98)00519-1 · Cited by 101
“Despite the statistical nature of defibrillation success (27), a purely random basis for all the first-shock therapy failures in this series is unlikely given the low implant DFTs, the high proportion of patients known to have received appropriate and successful ICD therapies (before death) and the one to three normal…”
Background
Metadata
- Authors
- Davy, J.-M., Fain, E. S., Dorian, P., & Winkle, R. A.
- Venue
- American Heart Journal
- Publisher
- Elsevier BV
- Published
- 1987-01
- Volume
- 113
- Issue
- 1
- Pages
- 77-84
- ISSN
- 0002-8703 (ISSN-L)
- OpenAlex
- W2035053023
- Cited by
- 217
- References
- 13
- FWCI
- 11.55
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the “defibrillation threshold” concept. https://app.cochat.ai/papers/10.1016/0002-8703(87)90012-3
- The paper
- Davy, J.-M., Fain, E. S., Dorian, P., & Winkle, R. A. (1987). The relationship between successful defibrillation and delivered energy in open-chest dogs: Reappraisal of the “defibrillation threshold” concept. American Heart Journal. https://doi.org/10.1016/0002-8703(87)90012-3
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.