Long-term Repeatability and Reproducibility of Phosphene Characteristics in Chronically Implanted Argus II Retinal Prosthesis Subjects
Abstract
Purpose Previously published literatures of acute studies on few subjects have shown contradictory evidence on the reproducibility and characteristics of the elicited phosphenes, despite using the same stimulating parameters with epiretinal electrode arrays. In this study, we set out to investigate the long-term repeatilibity and reproducibility of phosphenes in subjects chronically implanted with the Argus II retinal prosthesis (Second Sight Medical Products, Inc., Sylmar, CA, USA).Design Retrospective interventional case series and reliability study.Methods Six Argus II subjects of >5 years implantation from a single site participated. The 4-electrode cluster ("quad") closest to fovea was stimulated in each subject with a fixed biphasic current. Perceived phosphenes were depicted relative to subjective visual field center. The stimulus was applied at reducing time intervals from 20 minutes to 1 second. Two sets of stimulations were performed on the same day and 2 further sets repeated on a separate visit >1 week apart.Results Each subject depicted phosphenes of consistent shapes and sizes, and reported seeing the same colors with the fixed stimulating parameters, irrespective of the interstimuli intervals. However, there is a wide intersubject variation in the phosphene characteristics. Four subjects drew phosphenes in the same visual field quadrant, as predicted by the quad-fovea location. Two subjects depicted phosphenes in the same hemifield as the expected locations.Conclusion Phosphenes for each subject were consistently reproducible in all our chronically implanted subjects. This has important implications in the development of long-term pixelated prosthetic vision for future devices.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Since the goal is not primarily to create natural vision, it suffices that phosphene characteristics are distinct and stable over time, which is the case for current implants.
“However, since the goal is not primarily to create natural vision, it suffices that phosphene characteristics are distinct and stable over time, which is the case for current implants (Luo et al., 2016; Fernández et al., 2021).”
From Towards a <i>Smart Bionic Eye</i> : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Luo et al., 2016
Phosphene shapes and sizes differ between electrodes and across participants.
“Since phosphene shapes and sizes differ between electrodes and across participants (D. Nanduri et al. 2008; Beyeler et al. 2019; Luo et al. 2016), which could influence the overall image produced by multi-electrode stimulation, we performed a systematic mapping experiment.”
From Sequential epiretinal stimulation improves discrimination in simple shape discrimination tasks only (2022) · cites this paper as Luo et al. 2016
Phosphene appearance varied greatly across electrodes and participants.
“The appearance of phosphenes was highly variable across electrodes and participants, which is consistent with prior studies (D. Nanduri et al. 2008; Beyeler et al. 2019; Luo et al. 2016).”
From Sequential epiretinal stimulation improves discrimination in simple shape discrimination tasks only (2022) · cites this paper as Luo et al. 2016
The citation lists haven’t been fetched yet
References 13 · Cited by 47
- Authors
- Luo, Y. H.-L., Zhong, J. J., Clemo, M., & da Cruz, L.
- Affiliations
- Moorfields Eye Hospital NHS Foundation Trust (GB)
- Corresponding author
- Luo, Yvonne H-L.
- Venue
- American Journal of Ophthalmology
- Publisher
- Elsevier BV
- Published
- 2016-10
- Volume
- 170
- Pages
- 100-109
- ISSN
- 0002-9394 (ISSN-L)
- OpenAlex
- W2485864648
- Cited by
- 47
- References
- 13
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 3 · 2023: 7 · 2022: 10 · 2021: 7 · 2020: 3 · 2019: 4 · 2018: 2 · 2017: 5
- FWCI
- 2.06
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Ophthalmology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
- Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
- Keywords
- Argus II retinal prosthesis, phosphenes, reproducibility, chronic implantation, visual field mapping, interstimulus interval, interindividual variability
- MeSH terms
- Aged, Electrodes, Implanted, Humans, Middle Aged, Phosphenes, Retinitis Pigmentosa, Retrospective Studies, Vision Disorders, Visual Perception, Reproducibility of Results, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review