Sequential epiretinal stimulation improves discrimination in simple shape discrimination tasks only

Open access
Christie, Breanne, Sadeghi, Roksana, Kartha, Arathy, Caspi, Avi, Tenore, Francesco V, Klatzky, Roberta L, Dagnelie, Gislin, Billings, Seth·2022·Journal of Neural Engineering

Published version of a preprint

Abstract

Objective . Electrical stimulation of the retina can elicit flashes of light called phosphenes, which can be used to restore rudimentary vision for people with blindness. Functional sight requires stimulation of multiple electrodes to create patterned vision, but phosphenes tend to merge together in an uninterpretable way. Sequentially stimulating electrodes in human visual cortex has recently demonstrated that shapes could be ‘drawn’ with better perceptual resolution relative to simultaneous stimulation. The goal of this study was to evaluate if sequential stimulation would also form clearer shapes when the retina is the neural target. Approach . Two human participants with retinitis pigmentosa who had Argus ® II epiretinal prostheses participated in this study. We evaluated different temporal parameters for sequential stimulation and performed phosphene shape mapping and forced choice discrimination tasks. For the discrimination tasks, performance was compared between stimulating electrodes simultaneously versus sequentially. Main results . Phosphenes elicited by different electrodes were reported as vastly different shapes. For sequential stimulation, the optimal pulse train duration was 200 ms when stimulating at 20 Hz and the optimal gap interval was tied between 0 and 50 ms. Sequential electrode stimulation outperformed simultaneous stimulation in simple discrimination tasks, in which shapes were created by stimulating 3–4 electrodes, but not in more complex discrimination tasks involving ≥5 electrodes. The efficacy of sequential stimulation depended strongly on selecting electrodes that elicited phosphenes with similar shapes and sizes. Significance . An epiretinal prosthesis can produce coherent simple shapes with a sequential stimulation paradigm, which can be used as rudimentary visual feedback. However, success in creating more complex shapes, such as letters of the alphabet, is still limited. Sequential stimulation may be most beneficial for epiretinal prostheses in simple tasks, such as basic navigation, rather than complex tasks such as novel object identification.

AI snapshot

Written from the paper’s full text.

TL;DR

Sequential epiretinal stimulation improved discrimination over simultaneous stimulation for simple shapes but not for more complex letter shapes. It helped prevent phosphenes from merging, yet its benefit declined as more electrodes were used and as phosphene shape/size variability increased.

Key findings

  • In two Argus II participants with retinitis pigmentosa, sequential stimulation significantly improved arrow-direction discrimination: both were correct on 19/20 trials versus 4/20 for S1 and 8/20 for S2 with simultaneous stimulation (p<0.001 for both); simultaneous performance was not better than chance.
  • For square-size discrimination, sequential stimulation yielded 27/36 correct for both participants, significantly above 50% chance (p=0.001); the improvement over simultaneous stimulation was statistically significant only for S1 (p=0.01). Simultaneous stimulation was 17/36 for S1 (not above chance) and 25/36 for S2 (p=0.001).
  • For complex letter discrimination, only S2 completed the task and only with sequential stimulation: 34/80 correct (significantly above 25% chance, p<0.001), with 40% for K, 35% for O, 50% for V, and 45% for Z.
  • Optimal sequential parameters: 200 ms train duration at 20 Hz gave highest direction-classification accuracy (92% correct); gap interval was tied between 0 and 50 ms, with 78%–81% correct across gap conditions.

Methods

  • Two human participants (83-year-old male S1, 71-year-old female S2) with retinitis pigmentosa and unilateral Argus II retinal prostheses participated; the protocol was approved by the Johns Hopkins Medicine Institutional Review Board.
  • Phosphene shape mapping tested 25 electrodes per participant with 12 repetitions each; participants drew perceived phosphenes on a tablet. Pulse frequency was initially 20 Hz but lowered to 6 Hz due to rapid perceptual adaptation; train length was 200 ms. Shapes were quantified with MATLAB regionprops for area and elongation, with ANOVAs, linear regression, and resampling analyses.
  • Forced-choice tasks compared sequential versus simultaneous stimulation. Parameter evaluation varied train durations (50, 100, 150, 200 ms) and gaps (0, 50, 100 ms) over 48 trials per participant. Size discrimination used four electrodes and 36 trials per paradigm; simple arrow discrimination used three electrodes and 20 trials per paradigm; complex letter discrimination used 5–8 electrodes to form Sloan letters O, V, K, Z, with 20 trials per paradigm. Statistical tests included tests of proportions.

Limitations

  • Stated: only two Argus II recipients were tested, limiting generalizability; Argus II users are becoming rarer. The authors also note that phosphene shape/size differences made tasks difficult, that no control task quantified drawing accuracy, and that brightness differences, perceptual adaptation, gaze direction, and phosphene mislocalization may have influenced results.
  • Evident from the text: the complex letter task was completed only by S2 and only under sequential stimulation, so no direct sequential-versus-simultaneous comparison was possible for letters. Trial counts were small (e.g., 20 trials per paradigm for arrow and letter tasks), and electrode selection for larger shapes was constrained because some electrodes fell outside the mapped set.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1088/1741-2552/ac7326

Cites

43 of 43 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 28 cited sources checked · 0 supported · 0 flagged

  • Findingown data

    For participant S1, phosphene area and elongation differed significantly between electrodes (p<0.001).

    “Phosphene area and elongation were significantly different between electrodes for participant S1 (p<0.001) but not S2 (Figure 3b,c).”
    Quote in the paperNot a substantial claim
    • The authors state this as an assumption: When more than one phosphene was perceived in a trial, the first drawn phosphene was assumed to be the most noticeable.
  • Findingown data

    Participant S1 could not initially perform the arrow task until electrodes with similar phosphene shapes and sizes were selected.

    “Participant S1 could not originally perform this task until we selected electrodes that elicited similar phosphene shapes and sizes.”
    Quote in the paperNot a substantial claim
  • Findingown data

    When three electrodes were stimulated simultaneously and participants drew arrow shapes, the three phosphenes merged in an unpredictable way.

    “When the participants were asked to draw the shapes of the arrows, the three phosphenes merged together in an unpredictable manner during simultaneous stimulation (Figure 5).”
    Quote in the paperSubstantial claim
  • Findingcitation

    Phosphene appearance varied greatly across electrodes and participants.

    “The appearance of phosphenes was highly variable across electrodes and participants, which is consistent with prior studies (D. Nanduri et al. 2008; Beyeler et al. 2019; Luo et al. 2016).”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites D. Nanduri et al. 2008 source

      Not checked in the free run.

    • cites Beyeler et al. 2019 source

      Not checked in the free run.

    • cites Luo et al. 2016 source

      Not checked in the free run.

  • Findingown data

    In the arrow discrimination task, performance was significantly better with sequential than simultaneous stimulation for both participants (p<0.001).

    “The participants’ performance on the arrow discrimination task was significantly better when using sequential compared to simultaneous stimulation paradigms (p<0.001 for both participants, Table 2).”
    Quote in the paperSubstantial claim
  • May combine two claims. 32
  • Classified as probably not a substantial claim. 13
  • May combine two claims.

    Findingown data

    Participant S2 selected the correct alphabet letter in 34 of 80 trials, significantly above the 25% chance level (p<0.001).

    “Participant S2 was able to correctly choose which alphabet letter was presented in 34/80 trials, which was significantly higher than the 25% chance (p<0.001), but not optimal for functional use.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For participant S2, phosphene area and elongation did not differ significantly between electrodes.

    “Phosphene area and elongation were significantly different between electrodes for participant S1 (p<0.001) but not S2 (Figure 3b,c).”
    Quote in the paperNot a substantial claim
    • The authors state this as an assumption: When more than one phosphene was perceived in a trial, the first drawn phosphene was assumed to be the most noticeable.
  • May combine two claims.

    Findingown data

    Across the discrimination tasks, sequential stimulation was not consistently better than simultaneous stimulation, contrary to the authors' original hypothesis.

    “Contrary to our original hypothesis, sequential stimulation did not consistently outperform simultaneous stimulation in these discrimination tasks.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Changing the gap interval between electrodes produced only negligible performance differences, with accuracy ranging from 78% correct for 100 ms gaps to 81% correct for 0 or 50 ms gaps.

    “There were only negligible differences in performance as a result of changing the gap intervals between electrodes: classification accuracy had a minimum of 78% correct for 100 ms gaps and a maximum of 81% correct for 0 or 50 ms gaps.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For sequential stimulation at 20 Hz, the optimal pulse train duration was 200 ms and the optimal gap interval was tied between 0 and 50 ms.

    “For sequential stimulation, the optimal pulse train duration was 200 ms when stimulating at 20 Hz and the optimal gap interval was tied between 0 and 50 ms.”
    • Also stated in: § Sequential stimulation parameter evaluation · plain
      “The best combination of stimulation parameters across both participants was a 200 ms train duration with either a 0 ms or 50 ms gap.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    For participant S2, the phosphene shapes across trials were much less reliable and showed no clear patterns.

    “The shapes across trials for participant S2 were far less reliable and did not follow any noticeable patterns (Figure 2b);”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    In the arrow task, both participants correctly identified direction in 19 of 20 trials with sequential stimulation, significantly above 50% chance (p<0.001).

    “Both participants correctly identified arrow direction in 19/20 trials when using sequential stimulation, which was significantly higher than 50% chance (p<0.001).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    In the size discrimination task, both participants correctly answered 27 of 36 trials with sequential stimulation, significantly above 50% chance (p=0.001).

    “Both participants achieved 27 out of 36 trials correct using sequential stimulation, which was significantly higher than 50% chance (p=0.001).”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    In the square size discrimination task, performance was better with sequential than simultaneous stimulation, but the improvement was statistically significant only for participant S1 (p=0.01).

    “The participants’ performance on the square size discrimination task was better when using sequential versus simultaneous stimulation paradigms (Table 2), but the improvement was statistically significant only for participant S1 (p=0.01).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Only participant S2 completed the alphabet letter discrimination task, and she did so only with sequential stimulation.

    “Only participant S2 was successfully able to complete the alphabet letter discrimination task, and only when using the sequential stimulation paradigm.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Participant S2's identification accuracy was 40% for K, 35% for O, 50% for V, and 45% for Z.

    “She had a 40% accuracy for identifying letter ‘K,’ 35% for ‘O,’ 50% for ‘V,’ and 45% for ‘Z.’”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    The shapes most often reported were circles, ovals, lines, and rectangles, although many were irregular.

    “The most commonly elicited shapes were circles, ovals, lines, and rectangles, though many shapes were irregular.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    When three electrodes were stimulated sequentially, direction classification accuracy was highest at 92% correct with a 200 ms train duration per electrode and lowest at 71% correct with 100 ms.

    “Classification accuracy of phosphene direction was highest (92% correct) when the train duration was set to 200 ms per electrode and lowest (71% correct) at 100 ms (Table 1).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    With simultaneous stimulation of the four electrodes in the size task, the phosphenes merged and participants reported only one or two large phosphenes.

    “When the four electrodes were simultaneously stimulated, the phosphenes merged and participants reported seeing only one or two large phosphenes.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    Participants most often reported one phosphene per electrode.

    “It was most common for participants to report one phosphene per electrode (Figure 3a).”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Participant S1 performed far below chance in the arrow task with simultaneous stimulation; when shown a right-pointing arrow, he reported left in 9 of 10 trials.

    “In fact, participant S1’s accuracy was far below chance; when presented with a right-pointing arrow using simultaneous stimulation, he reported that it was pointing to the left in 9/10 trials.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Participants perceived three separate phosphenes under sequential stimulation but only one or two phosphenes when the same three electrodes were stimulated simultaneously.

    “Though the participants saw three discrete phosphenes during the sequential stimulation paradigm, they reported seeing only 1–2 phosphenes when the three electrodes were simultaneously stimulated.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Participant S2 tended to have larger phosphene areas and lower detection thresholds than participant S1.

    “Compared to participant S1, participant S2 tended to have larger phosphene areas and lower detection thresholds.”
    Quote in the paperNot a substantial claim
    • The authors state this as an assumption: When more than one phosphene was perceived in a trial, the first drawn phosphene was assumed to be the most noticeable.
  • May combine two claims.

    Findingown data

    In the arrow task with simultaneous stimulation, participant S1 correctly answered 4 of 20 trials and participant S2 correctly answered 8 of 20 trials, neither significantly above chance (p>0.05).

    “When using simultaneous stimulation, participant S1 achieved 4/20 trials correct and participant S2 achieved 8/20 trials correct, neither of which were significantly better than chance (p>0.05).”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    In the size discrimination task with simultaneous stimulation, participant S1 correctly answered 17 of 36 trials, which was not significantly above chance.

    “When performing the task using simultaneous stimulation, participant S1 got 17/36 trials correct, which was not significantly higher than chance”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    The authors also found large trial-to-trial variability within individual electrodes, particularly for participant S2, which differed from prior studies.

    “Unlike prior studies, we also noticed a substantial amount of variation between trials within a single electrode, especially for participant S2.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    The composite image from several electrodes was sometimes intuitive or quickly understood, yet most cases needed practice trials, particularly for participant S1.

    “The overall image created by multiple electrodes was sometimes intuitive or explained and quickly understood, but most instances required practice trials, especially for participant S1.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Sequential stimulation performed better than simultaneous stimulation for size discrimination and simple shape discrimination, but not for complex shape discrimination.

    “Sequential electrode stimulation outperformed simultaneous stimulation on the size discrimination task and the simple shape discrimination task, but not on the complex shape discrimination task.”
    • Also stated in: § Conclusion · plain
      “In conclusion, sequential electrode stimulation outperformed simultaneous stimulation in simple discrimination tasks, but not in more complex discrimination tasks.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Findingown data

    Stimulation produced irregular phosphene shapes as well as circles, ovals, lines, and rectangles.

    “Stimulation elicited irregular shapes in addition to circles, ovals, lines, and rectangles.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationown data

    An epiretinal prosthesis can produce coherent simple shapes using a sequential stimulation paradigm, which can serve as rudimentary visual feedback.

    “An epiretinal prosthesis can produce coherent simple shapes with a sequential stimulation paradigm, which can be used as rudimentary visual feedback.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Argus II devices allowed only about half of test subjects to reliably read large-print letters, and single-letter recognition took between six seconds and 3.5 minutes.

    “For example, Argus II devices enabled only ~50% of test subjects to reliably read large-print letters (41° or ~22.6 cm letters viewed from 30 cm away), and single-letter recognition took anywhere between six seconds to 3.5 minutes (da Cruz et al. 2013).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites da Cruz et al. 2013 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductnone given

    Significance levels in all statistical analyses were set to α=0.05, and Bonferroni corrections were applied.

    “Significance levels in all statistical analyses were set to α=0.05 and Bonferroni corrections were applied.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductown data

    The study protocol was approved by the Johns Hopkins Medicine Institutional Review Board, and informed consent was obtained before participation.

    “The study protocol was approved by the Johns Hopkins Medicine Institutional Review Board. Informed consent was obtained prior to participation in research-related activities.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning

    Participant S1 may have interpreted the larger merged phosphene as the arrow's wider base and the smaller phosphene as its narrower tip, which could explain his tendency to report right-pointing arrows as left-pointing during simultaneous trials.

    “It is possible that participant S1 interpreted the larger phosphene as being the wider base of the arrow, and the smaller phosphene as the narrower tip of the arrow, and was inclined to report right-pointing arrows as left-pointing arrows during simultaneous stimulation trials.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Phosphene shapes and sizes differ between electrodes and across participants.

    “Since phosphene shapes and sizes differ between electrodes and across participants (D. Nanduri et al. 2008; Beyeler et al. 2019; Luo et al. 2016), which could influence the overall image produced by multi-electrode stimulation, we performed a systematic mapping experiment.”
    Quote in the paperSubstantial claim0 of 3 sources checked
    • cites D. Nanduri et al. 2008 source

      Not checked in the free run.

    • cites Beyeler et al. 2019 source

      Not checked in the free run.

    • cites Luo et al. 2016 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisecitation

    Visual neuroprostheses have been implanted in the retina, optic nerve, and visual cortex of human subjects.

    “To date, visual neuroprostheses have been implanted in the retina (Humayun et al. 2012; Stingl et al. 2013; Fujikado et al. 2016; Petoe et al. 2021; Barnes et al. 2016; Palanker et al. 2022; Xu et al. 2021), optic nerve (Veraart et al. 2003), and visual cortex of human subjects (Beauchamp et al. 2020; Fernández et al. 2021).”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Humayun et al. 2012 source

      Not checked in the free run.

    • cites Veraart et al. 2003 source

      Not checked in the free run.

    • cites Beauchamp et al. 2020 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationcitation

    Phosphene shape variation across trials could also have been affected by perceptual adaptation and gaze direction.

    “Phosphene shape variation across trials also could have been influenced by perceptual adaptation (Pérez Fornos et al. 2012) and gaze direction (Barry and Dagnelie 2016; Sabbah et al. 2014; Caspi et al. 2017; 2021).”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Pérez Fornos et al. 2012 source

      Not checked in the free run.

    • cites Barry and Dagnelie 2016

      Not checked in the free run.

    • cites Sabbah et al. 2014 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Study design and conductreasoning

    The authors took frequent breaks to reduce the chance of perceptual adaptation and instructed participants to keep the same head and eye position between trials.

    “However, we took frequent breaks to minimize the likelihood of adaptation, and we asked the participants to maintain the same head and eye position between trials.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationreasoning

    Because daily-life visual scenes are complicated, sequential stimulation may be most useful for epiretinal prostheses in basic navigation rather than difficult tasks like object identification.

    “Because our surroundings in daily life are visually complicated, sequential stimulation may be most beneficial for epiretinal prostheses in basic navigational applications rather than difficult tasks such as object identification.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationcitation

    Even though each electrode was stimulated in proportion to its detection threshold and pulse frequency was constant, perceived brightness may have differed between electrodes.

    “Additionally, though we stimulated each electrode in proportion to its detection threshold and held pulse frequency constant, it is possible that there were differences in the perceived brightness of phosphenes elicited by different electrodes (Devyani Nanduri et al. 2012).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Devyani Nanduri et al. 2012

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationreasoning

    Small trial-to-trial variations were likely caused by participants drawing without visual feedback of their finger traces on the tablet screen.

    “In general, small variations between trials were likely due to the fact that the participants were drawing without any visual feedback of traces made by their finger on the tablet screen.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Interpretationreasoning

    Because inadvertent axon bundle recruitment occurs, targeting the visual cortex may be advantageous for visual neuroprostheses, since its cellular makeup differs from that of the retina.

    “The inadvertent recruitment of axon bundles is one reason why it may be advantageous to target the visual cortex for visual neuroprostheses, because it has a different cellular makeup than the retina.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationcitation

    Mislocalization may be due to eye movements affecting phosphene location or to inconsistencies in phosphene size and brightness.

    “Mislocalization can be attributed to eye movements that influence the location of the phosphene (Sabbah et al. 2014; Caspi et al. 2017; 2021), or inconsistencies in phosphene size and brightness.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Sabbah et al. 2014 source

      Not checked in the free run.

    • cites Caspi et al. 2017 source

      Not checked in the free run.

    • cites [2021]

      Not checked in the free run.

38 of 38
  • The authors state this as an assumption.

    When more than one phosphene was perceived in a trial, the first drawn phosphene was assumed to be the most noticeable.

    “If two or more phosphenes were perceived in a single trial, only the first drawn phosphene was included when calculating elongation; it was assumed to be the most noticeable, and the computation cannot account for multiple phosphenes.”
  • The paper relies on this without support.

    The information from a single phosphene is minimal, so patterned vision is needed to maximize real-time visual understanding of the environment.

    “The information delivered by a single phosphene is minimal, so patterned vision is used to maximize real-time visual understanding of one’s environment.”
  • The paper relies on this without support.

    Participants' drawings and reports accurately reflect their phosphene percepts.

    “When the participants were asked to draw the shapes of the arrows, the three phosphenes merged together in an unpredictable manner during simultaneous stimulation (Figure 5).”
    “Though the participants saw three discrete phosphenes during the sequential stimulation paradigm, they reported seeing only 1–2 phosphenes when the three electrodes were simultaneously stimulated.”
3 of 3

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Sequential stimulation does not always lead to perceptually intelligible forms or objects; participants often can only trace an outline of the perceived shape, and their interpretation relies heavily on this basic outline.

    “However, sequential stimulation does not always lead to perceptually intelligible forms or objects; often participants are only able to trace an outline of the perceived shape, and their interpretation of the shape relies heavily on this basic outline (Christie et al 2022).”
  • Multiple phosphene patterns may not automatically group into perceptually intelligible objects.

    “However, it should be noted that multiple phosphene patterns may not automatically group into perceptually intelligible objects (Stingl et al 2015, Shivdasani et al 2017, Christie et al 2022).”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

  • Trial count does not match the stated repetitions

    Minor

    Four pulse-train durations, three gap intervals, and three repetitions per combination give 36 trials per participant, not 48.

    Reported 48 · recomputed 36

1 of 1

Cites

Flagged first, then by citations

Research works printed but not linked: 1

  1. Sloan, 1952, Comparison of three types of test target for the measurement of visual acuity, Q. Rev. Ophthalmol., 8, 4

Cited by

Sorted by citations

Authors
Christie, B., Sadeghi, R., Kartha, A., Caspi, A., Tenore, F. V., Klatzky, R. L., Dagnelie, G., & Billings, S.
Affiliations
  • Johns Hopkins University (US)
  • Jerusalem College of Technology (IL)
  • Carnegie Mellon University (US)
Corresponding author
Christie, Breanne
Venue
Journal of Neural Engineering
Published
2022-06-01
Volume
19
Issue
3
Pages
036033
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W4281489008
Cited by
10
References
42
Citations by year
2026: 1 · 2025: 5 · 2024: 1 · 2023: 1 · 2022: 2
FWCI
0.99
Type
journal article
Language
en
Open access
Locations
  • Journal of Neural Engineering (publishedVersion, cc-by)
  • medRxiv (acceptedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion)
Funders
  • Johns Hopkins Hospital
  • National Institutes of Health (R01EY029741)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Tactile and Sensory Interactions
  • Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
Keywords
epiretinal prosthesis, sequential stimulation, phosphenes, retinitis pigmentosa, Argus II, shape discrimination, stimulation timing, electrode selection, vision restoration
MeSH terms
Blindness, Electric Stimulation, Electrodes, Implanted, Humans, Phosphenes, Retina, Retinitis Pigmentosa, Vision Disorders, Visual Prosthesis

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-08.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-08.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review