Student or educator? Sign up with your school email and get 6 months of Pro free.
Cited by 348

The DOI matches its Crossref record; 24 of the 24 references it cites were found in a registry or an index.

Does Rubidium-82 PET Have Superior Accuracy to SPECT Perfusion Imaging for the Diagnosis of Obstructive Coronary Disease?

Open access
Mc Ardle, Brian A., Dowsley, Taylor F., deKemp, Robert A., Wells, George A., Beanlands, Rob S. · 2012 · Journal of the American College of Cardiology

Abstract

OBJECTIVES The purpose of this study was to evaluate the accuracy of rubidium (Rb)-82 positron emission tomography (PET) for the diagnosis of obstructive coronary artery disease (CAD) in comparison to single-photon emission tomography (SPECT). BACKGROUND Myocardial perfusion imaging is widely used in the assessment of patients with known or suspected CAD. PET using Rb-82 has potential advantages over SPECT that may make it more accurate and that reduce radiation exposure compared with SPECT but has increased costs. Comparisons of these technologies are highly relevant for policy makers and practice guidelines. However, studies directly comparing Rb-82 PET with contemporary SPECT have been limited. METHOD The authors therefore undertook a systematic review of studies where either Rb-82 PET or technetium-99m SPECT with both attenuation correction and electrocardiography-gating were used as a diagnostic test for obstructive CAD with invasive coronary angiogram as a reference standard. These technologies were then compared. RESULTS Fifteen PET and 8 SPECT studies (1,344 and 1,755 patients, respectively) met inclusion criteria and pooled accuracy using weighted averages according to the size of the patient population was determined for PET and SPECT with sensitivities of 90% (confidence interval [CI]: 0.88 to 0.92) and 85% (CI: 0.82 to 0.87) and specificities of 88% (CI: 0.85 to 0.91) and 85% (CI: 0.82 to 0.87), respectively. Summary receiver-operating characteristic curves were computed: area under the curve was 0.95 and 0.90 for PET and SPECT, respectively (p < 0.0001). There was heterogeneity among study populations and some studies were limited by referral bias. CONCLUSIONS Rb-82 PET is accurate for the detection of obstructive CAD and, despite advances in SPECT technology, remains superior. More widespread use of Rb-82 PET may be beneficial to improve CAD detection.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/j.jacc.2012.07.038

Cites

24 of 24 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 24 references

Cited by

See all 348

Metadata

Authors
Mc Ardle, B. A., Dowsley, T. F., deKemp, R. A., Wells, G. A., & Beanlands, R. S.
Venue
Journal of the American College of Cardiology
Publisher
Elsevier BV
Published
2012-10
Volume
60
Issue
18
Pages
1828-1837
ISSN
0735-1097 (ISSN-L)
OpenAlex
W1741041
Cited by
348
References
24
FWCI
20.92
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Does Rubidium-82 PET Have Superior Accuracy to SPECT Perfusion Imaging for the Diagnosis of Obstructive Coronary Disease? https://app.cochat.ai/papers/10.1016/j.jacc.2012.07.038
The paper
Mc Ardle, B. A., Dowsley, T. F., deKemp, R. A., Wells, G. A., & Beanlands, R. S. (2012). Does Rubidium-82 PET Have Superior Accuracy to SPECT Perfusion Imaging for the Diagnosis of Obstructive Coronary Disease? Journal of the American College of Cardiology. https://doi.org/10.1016/j.jacc.2012.07.038

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.