The reliability of the i-STAT clinical portable analyser
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
The purpose of this investigation was to assess the reliability of the i-STAT clinical portable analyser and CG(4)(+) cartridge measures of blood pH, partial pressures of O(2) (pO(2)) and CO(2) (pCO(2)), bicarbonate ([HCO(3)(-)]), base excess (BE), total carbon dioxide (TCO(2)), oxygen saturation (sO(2)) and blood lactate ([BLa(-)]) at various exercise intensities. A comparison between [BLa(-)] measured by the i-STAT and the Accusport lactate analysers during an intermittent treadmill run was also undertaken. The technical error of measurement (TEM%) at rest, at moderate (200W) and maximal exercise (V O(2)max) was acceptable (<15%) for all parameters. The intra-class correlation coefficients for each analyte ranged from weak-to-strong across resting (ICC=0.19-0.96) and moderate (ICC=0.30-0.96) exercise intensities. The ICC for all analytes were observed to be strong following maximal intensity exercise (ICC=0.77-0.95). The comparison of the [BLa(-)] measures between the i-STAT and Accusport showed that the difference between measures was acceptable at both low (<4mmolL(-1)) (-0.39+/-0.27mmolL(-1)), moderate to high concentrations (>4mmolL(-1)) (0.58+/-1.22mmolL(-1)), and across all [BLa(-)] data (0.36+/-1.13mmolL(-1)). In conclusion, the i-STAT clinical analyser and CG(4)(+) cartridge provides reliable measures of a number of blood parameters across exercise intensities. The [BLa(-)] measures from the i-STAT analyser are consistent with that of the Accusport lactate analyser.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Biochemistry of exercise-induced metabolic acidosis
Robergs, R. A., Ghiasvand, F., & Parker, D. (2004) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00114.2004 · Cited by 1,304
Cited by
The Effects of Chronic Sodium Bicarbonate Ingestion and Interval Training in Highly Trained Rowers
Driller, M. W., Gregory, J. R., Williams, A. D., & Fell, J. W. (2013) · International Journal of Sport Nutrition and Exercise Metabolism · doi:10.1123/ijsnem.23.1.40 · Cited by 43
Reliability and Effect of Sodium Bicarbonate: Buffering and 2000-m Rowing Performance
Carr, A. J., Slater, G. J., Gore, C. J., Dawson, B., & Burke, L. M. (2012) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.7.2.152 · Cited by 35
Scott, B. R., Slattery, K. M., Sculley, D. V., Lockie, R. G., & Dascombe, B. J. (2014) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2014.07.008 · Cited by 26
“The ICC was interpreted as per previous research (Dascombe et al., 2007): 0.”
MethodSelf-citationGurton, W. H., Matta, G. G., Gough, L. A., Ranchordas, M. K., King, D. G., & Hurst, P. (2023) · Sports Medicine - Open · doi:10.1186/s40798-023-00612-5 · Cited by 15
Evaluating a new wearable lactate threshold sensor in recreational to highly trained cyclists
Driller, M., Borges, N., & Plews, D. (2016) · Sports Engineering · doi:10.1007/s12283-016-0198-6 · Cited by 15
Metadata
- Authors
- Dascombe, B. J., Reaburn, P. R. J., Sirotic, A. C., & Coutts, A. J.
- Venue
- Journal of Science and Medicine in Sport
- Publisher
- Elsevier BV
- Published
- 2007-06
- Volume
- 10
- Issue
- 3
- Pages
- 135-140
- ISSN
- 1440-2440 (ISSN-L)
- OpenAlex
- W2097120531
- Cited by
- 66
- References
- 1
- FWCI
- 0.39
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.