Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring bone
Open accessAbstract
AIM: Combined PET/MR systems have now become available for clinical use. Given the lack of integrated standard transmission (TX) sources in these systems, attenuation and scatter correction (AC) must be performed using the available MR-images. Since bone tissue cannot easily be accounted for during MR-AC, PET quantification can be biased, in particular, in the vicinity of the skull. Here, we assess PET quantification in PET/MR imaging of patients using phantoms and patient data. MATERIALS AND METHODS: Nineteen patients referred to our clinic for a PET/CT exam as part of the diagnostic evaluation of suspected dementia were included in our study. The patients were injected with 200MBq [(18)F]FDG and imaged with PET/CT and PET/MR in random sequence within 1h. Both, PET/CT and PET/MR were performed as single-bed acquisitions without contrast administration. PET/CT and PET/MR data were reconstructed following CT-based and MR-based AC, respectively. MR-AC was performed based on: (A) standard Dixon-Water-Fat segmentation (DWFS), (B) DWFS with co-registered and segmented CT bone values superimposed, and (C) with co-registered full CT-based attenuation image. All PET images were reconstructed using AW-OSEM, with neither resolution recovery nor time-of-flight option employed. PET/CT (D) or PET/MR (A-C) images were decay-corrected to the start time of the first examination. PET images following AC were evaluated visually and quantitatively using 10 homeomorphic regions of interest drawn on a transaxial T1w-MR image traversing the central basal ganglia. We report the relative difference (%) of the mean ROI values for (A)-(C) in reference to PET/CT (D). In a separate phantom experiment a 2L plastic bottle was layered with approximately 12mm of Gypsum plaster to mimic skull bone. The phantom was imaged on PET/CT only and standard MR-AC was performed by replacing hyperdense CT attenuation values corresponding to bone (plaster) with attenuation values of water. PET image reconstruction was performed with CT-AC (D) and CT-AC using the modified CT images corresponding to MR-AC using DWFS (A). RESULTS: PET activity values in patients following MR-AC (A) showed a substantial radial dependency when compared to PET/CT. In all patients cortical PET activity was lower than the activity in the central region of the brain (10-15%). When adding bone attenuation values to standard MR-AC (B and C) the radial gradient of PET activity values was removed. Further evaluation of PET/MR activity following MR-AC (A) relative to MR-AC (C) using the full CT for attenuation correction showed an underestimation of 25% in the cortical regions and 5-10% in the central regions of the brain. Observations in patients were replicated by observations from the phantom study. CONCLUSION: Our phantom and patient data demonstrate a spatially varying bias of the PET activity in PET/MR images of the brain when bone tissue is not accounted for during attenuation correction. This has immediate implications for PET/MR imaging of the brain. Therefore, refinements to existing MR-AC methods or alternative strategies need to be found prior to adopting PET/MR imaging of the brain in clinical routine and research.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Attenuation correction for a combined 3D PET/CT scanner
Kinahan, P. E., Townsend, D. W., Beyer, T., & Sashin, D. (1998) · Medical Physics · doi:10.1118/1.598392 · Cited by 946 · cited 2×
“In PET/CT, the X-ray transmission source (up to 140 kVp) is used to generate CT images that can be used to calculatemaps of attenuation coefficients at 511 keV (Burger et al., 2002; Carney et al., 2006; Kinahan et al., 1998).”
Self-citationClark, C. M., Pontecorvo, M. J., Beach, T. G., Bedell, B. J., Coleman, R. E., Doraiswamy, P. M., Fleisher, A. S., Reiman, E. M., Sabbagh, M. N., Sadowsky, C. H., et al. (2012) · The Lancet Neurology · doi:10.1016/s1474-4422(12)70142-4 · Cited by 855
Mechanism of amyloid removal in patients with Alzheimer disease treated with gantenerumab
Ostrowitzki, S. (2012) · Archives of Neurology · doi:10.1001/archneurol.2011.1538 · Cited by 431 · cited 1×
“The brain imaging data acquired with the present state of combined PET/MR technology is likely to fail the site qualification necessary to be accepted in multicenter studies such as ADNI and for the clinical trials running using brain amyloid modifying agents (Ostrowitzki et al., 2011).”
PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging Applications
Pichler, B. J., Kolb, A., Nägele, T., & Schlemmer, H.-P. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.061853 · Cited by 418 · cited 1×
“Therefore, combined PET/MR imaging is expected to provide diagnostic information similar or better than that of PET/CT, while soft tissue contrast is improved and patient exposure to ionizing radiation is reduced in PET/MR (Pichler et al., 2010).”
Cited by
Law, I., Albert, N. L., Arbizu, J., Boellaard, R., Drzezga, A., Galldiks, N., la Fougère, C., Langen, K.-J., Lopci, E., Lowe, V., et al. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4207-9 · Cited by 556
Self-citationAdvances in multimodal data fusion in neuroimaging: Overview, challenges, and novel orientation
Zhang, Y.-D., Dong, Z., Wang, S.-H., Yu, X., Yao, X., Zhou, Q., Hu, H., Li, M., Jiménez-Mesa, C., Ramirez, J., et al. (2020) · Information Fusion · doi:10.1016/j.inffus.2020.07.006 · Cited by 453
Attenuation Correction Synthesis for Hybrid PET-MR Scanners: Application to Brain Studies
Burgos, N., Cardoso, M. J., Thielemans, K., Modat, M., Pedemonte, S., Dickson, J., Barnes, A., Ahmed, R., Mahoney, C. J., Schott, J. M., et al. (2014) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2014.2340135 · Cited by 364
Mehranian, A., Arabi, H., & Zaidi, H. (2016) · Medical Physics · doi:10.1118/1.4941014 · Cited by 152
Renner, A., Rausch, I., Cal Gonzalez, J., Laistler, E., Moser, E., Jochimsen, T., Sattler, T., Sabri, O., Beyer, T., Figl, M., et al. (2022) · Medical Physics · doi:10.1002/mp.15586 · Cited by 3
Self-citation
Metadata
- Authors
- Andersen, F. L., Ladefoged, C. N., Beyer, T., Keller, S. H., Hansen, A. E., Højgaard, L., Kjær, A., Law, I., & Holm, S.
- Venue
- NeuroImage
- Publisher
- Elsevier BV
- Published
- 2014-01
- Volume
- 84
- Pages
- 206-216
- ISSN
- 1053-8119 (ISSN-L)
- OpenAlex
- W2027981446
- Cited by
- 190
- References
- 4
- FWCI
- 18.78
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.