A low cost MEMS based NDIR system for the monitoring of carbon dioxide in breath analysis at ppm levels
Open accessAbstract
The molecules in our breath can provide a wealth of information about the health and well-being of a person. The level of carbon dioxide (CO 2 ) is not only a sign of life but also when combined with the level of exhaled oxygen provides valuable health information in the form of our metabolic rate. We report upon the development of a MEMS-based non-dispersive infrared CO 2 sensor for inclusion in a hand held portable breath analyser. Our novel sensor system comprises a thermopile detector and low power MEMS silicon on insulator (SOI) wideband infrared (IR) emitter. A lock-in amplifier design permits a CO 2 concentration of 50 ppm to be detected on gas bench rig. Different IR path lengths were studied with gases in dry and humid (25% and 50% RH) in order to design a sensor suitable for detecting CO 2 in breath with concentrations in the range of 4–5%. A breath analyser was constructed from acetal and in part 3D printed with a side-stream sampling mechanism and tested on a range of subjects with two data-sets presented here. The performance of the novel MEMS based sensor was validated using a reference commercial breath-by-breath sensor and produced comparable results and gave a response time of 1.3 s. Further work involves the detection of other compounds on breath for further metabolic analysis and reducing the overall resolution of our MEMS sensor system from ca . 250 ppm to 10 ppm.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Impact of obesity on respiratory function
LITTLETON, S. W. (2012) · Respirology · doi:10.1111/j.1440-1843.2011.02096.x · Cited by 398
Cited by
Towards Integrated Mid-Infrared Gas Sensors
Popa, D., & Udrea, F. (2019) · Sensors · doi:10.3390/s19092076 · Cited by 311
Non-dispersive infrared multi-gas sensing via nanoantenna integrated narrowband detectors
Tan, X., Zhang, H., Li, J., Wan, H., Guo, Q., Zhu, H., Liu, H., & Yi, F. (2020) · Nature Communications · doi:10.1038/s41467-020-19085-1 · Cited by 311
Zhou, H., Hui, X., Li, D., Hu, D., Chen, X., He, X., Gao, L., Huang, H., Lee, C., & Mu, X. (2020) · Advanced Science · doi:10.1002/advs.202001173 · Cited by 136
Metadata
- Authors
- Vincent, T. A., & Gardner, J. W.
- Venue
- Sensors and Actuators B: Chemical
- Publisher
- Elsevier BV
- Published
- 2016-11
- Volume
- 236
- Pages
- 954-964
- ISSN
- 0925-4005 (ISSN-L)
- OpenAlex
- W2314350043
- Cited by
- 171
- References
- 1
- FWCI
- 4.90
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A low cost MEMS based NDIR system for the monitoring of carbon dioxide in breath analysis at ppm levels. https://app.cochat.ai/papers/10.1016/j.snb.2016.04.016
- The paper
- Vincent, T. A., & Gardner, J. W. (2016). A low cost MEMS based NDIR system for the monitoring of carbon dioxide in breath analysis at ppm levels. Sensors and Actuators B: Chemical. https://doi.org/10.1016/j.snb.2016.04.016
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.