Performance of Peltier elements as a cryogenic heat flux sensor at temperatures down to 60 K
The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index; none of the 1 sources its claims cite is checked for support yet.
Overview
CoChat Summary
TL;DR
Commercially available Peltier elements were evaluated as cryogenic heat flux sensors from 200 down to 60 K, showing high output voltage and usable sensitivity. The 127-junction element retained sensitivity around 20 µV/(W/m²) at 60 K and thermal resistance about 2–3 K/W, supporting practical cryogenic use.
Key findings
- At 80 K, sensitivity was 28 µV/(W/m²) for the 127-junction CP1.0-127-05L and 18 µV/(W/m²) for the 71-junction CP0.8-71-06L; sensitivity depends on the number of PN junctions. The summary states 27 µV/(W/m²) at 80 K for the 127-junction element.
- Sensitivity remained around 20 µV/(W/m²) at 60 K (the summary states 17 µV/(W/m²)), nearly 10 times higher than that of a conventional heat flux sensor.
- Thermal resistance across the Peltier element was about 2–3 K/W from 200 to 60 K; the indium sheet thermal resistance was 0.09 K/W and almost constant over the same range.
- Empirical fits gave sensitivity temperature dependence with A = 0.073 for CP1.0-127-05L and A = 0.047 for CP0.8-71-06L, and thermal resistance with B = 0.52 for CP1.0-127-05L; uncertainties were less than 5% and less than 10%, respectively.
Methods
- Two Melcor Peltier elements were tested: CP1.0-127-05L (127 PN junctions, 30 × 30 × 3.3 mm) and CP0.8-71-06L (71 PN junctions, 17 × 17 × 3.5 mm). They were mounted on a copper sample holder attached to the cold head of a pulse tube refrigerator (PR111, AISIN; 800 W G–M compressor), using 0.1 mm indium sheets, four plastic screws at about 0.05 N m, and no grease; the cryostat was evacuated to 10⁻² Pa.
- Heat flux was generated with a sheet heater; temperatures were measured with Pt100 thermometers (NFR-CF2, 5 × 5 × 0.2 mm) calibrated against a Lake Shore PR101, and output voltages were measured with a Keithley 2000 scanner (0.1 µV resolution, 0.01% accuracy). Measurements were made at 60, 80, 101, 151, and 201 K.
- Analysis used linear fits of output voltage versus heat flux to obtain sensitivity, and ΔT versus heat input to obtain thermal resistance, followed by empirical temperature fits.
Limitations
- No explicit limitations section is present. The paper notes small offset voltages at each temperature, which were compensated to zero at no heat flux.
- Evident from the text: only two commercial Peltier elements from one supplier were tested, over 60–201 K; the empirical equations are limited to this range and to the stated uncertainties (<5% for sensitivity, <10% for thermal resistance).
- Evident from the text: the measured thermal resistance includes the BiTe semiconductors and ceramic plates, and contact was made through indium sheets and plastic screws, so mounting and contact effects are not separated from the element’s intrinsic behavior.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 0 Not valid · 1 Not checked · 27 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“For the element CP0.8-71-06L, the temperature dependence of the sensitivity can also be expressed by Eq. C402) with A ˆ 0 : 047.”
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“At 60 K, the sensitivity is still as large as 17 l V/C40W/m 2 ) demonstrating promise for practical cryogenic use.”
boosted- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“where B ˆ 0 : 52 for CP1.0-127-05L. Eq. C404) represents the experimental results with an uncertainty of less than 10%.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The thermal resistance across the Peltier element is also measured as about 2±3 K/W at temperatures from 200 to 60 K.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The sensitivity of each element depends on a number of PN junctions.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
8 more claims
Create a free account to see all 28 claims and to verify the 1 reference still open.
- It may state two separate claims as one. 6
- Probably not a substantial claim. 9
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 2. Peltier heat ¯ux sensor · p. 2
“C401) CP1.0-127-05L | Melcor C40USA) | 127 | 30 30 3 : 3”
§ 2. Peltier heat ¯ux sensor · p. 2
“Fig. 1. Peltier element C40CP1.0-127-05L: 30 mm -30 mm -3 : 5 mm) on a copper block.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 4. Results · p. 3
“The thermoelectric power of each element per unit heat ¯ux, de®ned as sensitivity, is 28 l V/C40W/m 2 ) for 127 junctions and 18 l V/C40W/m 2 ) for 71 junctions, respectively.”
§ 5. Summary · p. 4
“The 127 PN junction Peltier element with the surface area of 30 mm -30 mm shows a sensitivity of 27 l V/C40W/m 2 ) at 80 K”
Citations
Cites
Goldsmid, H. J. (1964) · Springer eBooks · doi:10.1007/978-1-4899-5723-8 · Cited by 1,058
Heat Leak Measurement Facility
Gonczy, J. D., Kuchnir, M., Nicol, T., Niemann, R. C., & Powers, R. J. (1986) · Springer eBooks · doi:10.1007/978-1-4613-2213-9_145 · Cited by 17
Peltier Current Lead Experiments with a Thermoelectric Semiconductor near 77 KPreprint
Yamaguchi, S., Nakamura, H., Ikeda, K., et al. (1997) · arXiv (Cornell University) · doi:10.48550/arxiv.cond-mat/9711042 · Cited by 6
Peltier Heat Flux Sensor for Cryogenic Use
Haruyama, T. (2000) · Advances in Cryogenic Engineering · doi:10.1007/978-1-4615-4215-5_120 · Cited by 4
Self-citationMeasuring Heat Leak with a Heatmeter
Kuchnir, M., Gonczy, J. D., & Tague, J. L. (1985) · doi:10.2172/1156269 · Cited by 4
Peltier current lead experiments with a thermoelectric semiconductor near 77 K [and HTSC]
Yamaguchi, S., Nakamura, H., Ikeda, K., et al. (n.d.) · XVI ICT '97. Proceedings ICT'97. 16th International Conference on Thermoelectrics (Cat. No.97TH8291) · doi:10.1109/ict.1997.667615
Cited by
Thermoelectrics: a review of present and potential applications
Riffat, S., & Ma, X. (2003) · Applied Thermal Engineering · doi:10.1016/s1359-4311(03)00012-7 · Cited by 1,640
Huang, L., Zheng, Y., Xing, L., & Hou, B. (2023) · Thermal Science and Engineering Progress · doi:10.1016/j.tsep.2023.102064 · Cited by 88
Kobari, T., Okajima, J., Komiya, A., & Maruyama, S. (2015) · International Journal of Heat and Mass Transfer · doi:10.1016/j.ijheatmasstransfer.2015.08.044 · Cited by 58
Improving the thermal performance of single-glazed windows using translucent granular aerogel
Dowson, M., Harrison, D. J., Craig, S., et al. (2011) · International Journal of Sustainable Engineering · doi:10.1080/19397038.2011.558931 · Cited by 42
“The voltage generated by this ‘Peltier effect’ corresponds to a heat flux when multiplied by a calibration factor (Haruyama 2001).”
BackgroundRecent developments of thermoelectric power generation
Luan, W. (2004) · Chinese Science Bulletin · doi:10.1360/04we0037 · Cited by 41
Metadata
- Authors
- Haruyama, T.
- Affiliations
- High Energy Accelerator Research Organization (JP)
- Corresponding author
- Haruyama, T.
- Venue
- Cryogenics
- Publisher
- Elsevier BV
- Published
- 2001-05
- Volume
- 41
- Issue
- 5-6
- Pages
- 335-339
- ISSN
- 0011-2275 (ISSN-L)
- OpenAlex
- W2134508431
- Semantic Scholar
- 53b5a0400d050ff198b4109d935146ac98392daa
- Cited by
- 25
- References
- 5
- Citations by year
- 2025: 2 · 2024: 3 · 2023: 1 · 2022: 2 · 2021: 3 · 2019: 1 · 2017: 1 · 2016: 2 · 2015: 5 · 2013: 1
- FWCI
- 1.33
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- Cryogenics (publishedVersion)
- Topics
- Physical Sciences › Physics and Astronomy › Instrumentation › Advanced Sensor Technologies Research
- Physical Sciences › Engineering › Aerospace Engineering › Spacecraft and Cryogenic Technologies
- Physical Sciences › Materials Science › Materials Chemistry › Advanced Thermoelectric Materials and Devices
- Keywords
- thermoelectric cooler, multilayer insulation, sensor sensitivity, thermal resistance, temperature dependence, cryogenic systems, cryogenic detectors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-11-17, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-11-17 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance of Peltier elements as a cryogenic heat flux sensor at temperatures down to 60 K. https://app.cochat.ai/papers/10.1016/s0011-2275(01)00081-9
- The paper
- Haruyama, T. (2001). Performance of Peltier elements as a cryogenic heat flux sensor at temperatures down to 60 K. Cryogenics. https://doi.org/10.1016/s0011-2275(01)00081-9
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.