Dexamethasone/PLGA microspheres for continuous delivery of an anti-inflammatory drug for implantable medical devices
Abstract
The purpose of this research was to develop polylactic-co-glycolic acid (PLGA) microspheres for continuous delivery of dexamethasone for over a 1-month period, in an effort to suppress the acute and chronic inflammatory reactions to implants such as biosensors, which interfere with their functionality. The microspheres were prepared using an oil-in-water emulsion technique. The oil phase was composed of 9:1 dichloromethane to methanol with dissolved PLGA and dexamethasone. Some microspheres were predegraded for 1 or 2 weeks. Ten percent of polyethylene glycol was added to the oil phase in alternative formulations to delay drug release. The in vitro release studies were performed in a constant temperature (37 C) warm room, in phosphate-buffered saline at sink conditions. Drug loading and release rates were determined by HPLC-UV analysis. The standard microsphere systems did not provide the desired release profile since, following an initial burst release, a delay of 2 weeks occurred prior to continuous drug release. Predegraded microspheres started to release dexamethasone immediately but the rate of release decreased after only 2 weeks. A mixed standard and predegraded microsphere system was used to avoid this delay and to provide continuous release of dexamethasone for 1 month.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
PLGA particles provided approximately zero-order dexamethasone release over one month after an initial burst of about 15%.
“While an initial burst release of ~15% of total DX was observed, the PLGA particles provided approximate zero-order kinetic release of DX over one month, indicating potential for long-term delivery.”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [184]
The citation lists haven’t been fetched yet
References 47 · Cited by 268
- Authors
- Hickey, T., Kreutzer, D., Burgess, D. J., & Moussy, F.
- Affiliations
- UConn Health (US)
- University of Connecticut (US)
- Corresponding author
- Moussy, F
- ORCID
- Burgess, D.J
- Venue
- Biomaterials
- Publisher
- Elsevier BV
- Published
- 2002-04
- Volume
- 23
- Issue
- 7
- Pages
- 1649-1656
- ISSN
- 0142-9612 (ISSN-L)
- OpenAlex
- W1986826942
- Cited by
- 268
- References
- 47
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 7 · 2023: 10 · 2022: 9 · 2021: 8 · 2020: 11 · 2019: 15 · 2018: 13 · 2017: 12 · 2016: 11 · 2015: 16 · 2014: 20 · 2013: 17 · 2012: 23
- FWCI
- 6.40
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Biomaterials (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (5r01rr014171-03)
- National Center for Research Resources (1R01RR14171)
- Topics
- Life Sciences › Pharmacology, Toxicology and Pharmaceutics › Pharmaceutical Science › Advanced Drug Delivery Systems
- Physical Sciences › Materials Science › Biomaterials › Nanoparticle-Based Drug Delivery
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Medicine › Hydrogels: synthesis, properties, applications
- Keywords
- dexamethasone, PLGA microspheres, controlled drug release, implant-associated inflammation, implantable biosensors, polyethylene glycol, oil-in-water emulsions, sustained drug release, anti-inflammatory drug delivery
- MeSH terms
- Polylactic Acid-Polyglycolic Acid Copolymer, Anti-Inflammatory Agents, Biocompatible Materials, Chromatography, High Pressure Liquid, Dexamethasone, Microscopy, Electron, Scanning, Microspheres, Polyglycolic Acid, Polymers, Time Factors, Ultraviolet Rays, Drug Delivery Systems, Lactic Acid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review