Pavel, Andreea M, Rennie, Janet M, de Vries, Linda S, Blennow, Mats, Foran, Adrienne, Shah, Divyen K, Pressler, Ronit M, Kapellou, Olga, Dempsey, Eugene M, Mathieson, Sean R, Pavlidis, Elena, van Huffelen, Alexander C, Livingstone, Vicki, Toet, Mona C, Weeke, Lauren C, Finder, Mikael, Mitra, Subhabrata, Murray, Deirdre M, Marnane, William P, Boylan, Geraldine B · 2020 · The Lancet Child & Adolescent Health
Cited by 151

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Despite the availability of continuous conventional electroencephalography (cEEG), accurate diagnosis of neonatal seizures is challenging in clinical practice. Algorithms for decision support in the recognition of neonatal seizures could improve detection. We aimed to assess the diagnostic accuracy of an automated seizure detection algorithm called Algorithm for Neonatal Seizure Recognition (ANSeR). METHODS: This multicentre, randomised, two-arm, parallel, controlled trial was done in eight neonatal centres across Ireland, the Netherlands, Sweden, and the UK. Neonates with a corrected gestational age between 36 and 44 weeks with, or at significant risk of, seizures requiring EEG monitoring, received cEEG plus ANSeR linked to the EEG monitor displaying a seizure probability trend in real time (algorithm group) or cEEG monitoring alone (non-algorithm group). The primary outcome was diagnostic accuracy (sensitivity, specificity, and false detection rate) of health-care professionals to identify neonates with electrographic seizures and seizure hours with and without the support of the ANSeR algorithm. Neonates with data on the outcome of interest were included in the analysis. This study is registered with ClinicalTrials.gov, NCT02431780. FINDINGS: Between Feb 13, 2015, and Feb 7, 2017, 132 neonates were randomly assigned to the algorithm group and 132 to the non-algorithm group. Six neonates were excluded (four from the algorithm group and two from the non-algorithm group). Electrographic seizures were present in 32 (25·0%) of 128 neonates in the algorithm group and 38 (29·2%) of 130 neonates in the non-algorithm group. For recognition of neonates with electrographic seizures, sensitivity was 81·3% (95% CI 66·7-93·3) in the algorithm group and 89·5% (78·4-97·5) in the non-algorithm group; specificity was 84·4% (95% CI 76·9-91·0) in the algorithm group and 89·1% (82·5-94·7) in the non-algorithm group; and the false detection rate was 36·6% (95% CI 22·7-52·1) in the algorithm group and 22·7% (11·6-35·9) in the non-algorithm group. We identified 659 h in which seizures occurred (seizure hours): 268 h in the algorithm versus 391 h in the non-algorithm group. The percentage of seizure hours correctly identified was higher in the algorithm group than in the non-algorithm group (177 [66·0%; 95% CI 53·8-77·3] of 268 h vs 177 [45·3%; 34·5-58·3] of 391 h; difference 20·8% [3·6-37·1]). No significant differences were seen in the percentage of neonates with seizures given at least one inappropriate antiseizure medication (37·5% [95% CI 25·0 to 56·3] vs 31·6% [21·1 to 47·4]; difference 5·9% [-14·0 to 26·3]). INTERPRETATION: ANSeR, a machine-learning algorithm, is safe and able to accurately detect neonatal seizures. Although the algorithm did not enhance identification of individual neonates with seizures beyond conventional EEG, recognition of seizure hours was improved with use of ANSeR. The benefit might be greater in less experienced centres, but further study is required. FUNDING: Wellcome Trust, Science Foundation Ireland, and Nihon Kohden.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/s2352-4642(20)30239-x

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

  • EEG-based neonatal seizure detection with Support Vector Machines

    Temko, A., Thomas, E., Marnane, W., Lightbody, G., & Boylan, G. (2011) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.06.034 · Cited by 339 · cited 1×

    “31,40 Newly developed In conclusion, this clinical investigation was the first to assess the performance of a machine-learning algorithm for neonatal seizure detection in real time and in the real-world setting of busy NICUs throughout Europe.”
    BackgroundSelf-citation

Show all 1 references

Cited by

See all 151

Metadata

Authors
Pavel, A. M., Rennie, J. M., de Vries, L. S., Blennow, M., Foran, A., Shah, D. K., Pressler, R. M., Kapellou, O., Dempsey, E. M., Mathieson, S. R., Pavlidis, E., van Huffelen, A. C., Livingstone, V., Toet, M. C., Weeke, L. C., Finder, M., Mitra, S., Murray, D. M., Marnane, W. P., & Boylan, G. B.
Venue
The Lancet Child & Adolescent Health
Publisher
Elsevier BV
Published
2020-10
Volume
4
Issue
10
Pages
740-749
ISSN
2352-4642 (ISSN-L)
OpenAlex
W3080952300
Cited by
151
References
1
FWCI
17.17
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A machine-learning algorithm for neonatal seizure recognition: a multicentre, randomised, controlled trial. https://app.cochat.ai/papers/10.1016/s2352-4642(20)30239-x
The paper
Pavel, A. M., Rennie, J. M., de Vries, L. S., Blennow, M., Foran, A., Shah, D. K., Pressler, R. M., Kapellou, O., Dempsey, E. M., Mathieson, S. R., Pavlidis, E., van Huffelen, A. C., Livingstone, V., Toet, M. C., Weeke, L. C., Finder, M., Mitra, S., Murray, D. M., Marnane, W. P., & Boylan, G. B. (2020). A machine-learning algorithm for neonatal seizure recognition: a multicentre, randomised, controlled trial. The Lancet Child & Adolescent Health. https://doi.org/10.1016/s2352-4642(20)30239-x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.