Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model

Open access
Jebb, Susan A., Cole, Timothy J., Doman, Deanne, Murgatroyd, Peter R., Prentice, Andrew M.·2000·British Journal of Nutrition

Abstract

The Tanita body-fat analyser is a novel device to estimate body fat, based on the principles of bioelectrical impedance. It differs from other impedance systems which use surface electrodes in that the subjects stand bare-footed on a metal sole-plate which incorporates the electrodes, hence impedance is measured through the legs and lower trunk. In 104 men and 101 women (16–78 years and BMI 16–41 kg/m 2) the mean bias in body-fat mass measured using the Tanita body-fat analyser was 0·8 (2SD 7·9) KG RELATIVE TO A FOUR-COMPARTMENT MODEL. THIS IS COMPARABLE TO THE OTHER PREDICTION TECHNIQUES TESTED (CONVENTIONAL TETRAPOLAR IMPEDANCE -1·3 (2sd 6·9) kg, skinfold thicknesses 0·3 (2sd 7·4) kg, and BMI-based formulas -0·2 (2sd 9·0) kg and -0·6 (2sd 8·5) kg), but the agreement was poorer than for ‘reference’ methods to measure body fat (density 0·2 (2sd 3·7) kg, total body water -0·9 (2sd 3·4) kg and dual-energy X-ray absorptiometry 0·1 (2sd 5·0) kg). The present paper also describes the derivation of a new prediction equation for the calculation of body composition from the Tanita body-fat analyser. The equation incorporates sex, age, and a log-transformation of height, weight and the measured impedance to predict body fat measured by a four-compartment model. This approach is recommended in the derivation of other prediction equations in body composition analysis. Using this novel prediction equation the residual standard deviations were 4·8 % for men and 3·3 % for women. A similar analysis using data collected with a conventional tetrapolar system yielded residual standard deviations of 4·3 % for men and 3·1 % for women. This demonstrates that the practical simplicity of the novel Tanita method is not associated with a clinically significant decrement in performance relative to a traditional impedance device.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1017/s0007114500000155

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 378

Metadata

Authors
Jebb, S. A., Cole, T. J., Doman, D., Murgatroyd, P. R., & Prentice, A. M.
Venue
British Journal of Nutrition
Published
2000-02
Volume
83
Issue
2
Pages
115-122
ISSN
  • 0007-1145 (ISSN-L)
  • 1475-2662
OpenAlex
W2148969315
Cited by
378
References
0
FWCI
7.37
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model. https://app.cochat.ai/papers/10.1017/s0007114500000155
The paper
Jebb, S. A., Cole, T. J., Doman, D., Murgatroyd, P. R., & Prentice, A. M. (2000). Evaluation of the novel Tanita body-fat analyser to measure body composition by comparison with a four-compartment model. British Journal of Nutrition. https://doi.org/10.1017/s0007114500000155

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.