The theory and fundamentals of bioimpedance analysis in clinical status monitoring and diagnosis of diseases.
Open accessAbstract
Bioimpedance analysis is a noninvasive, low cost and a commonly used approach for body composition measurements and assessment of clinical condition. There are a variety of methods applied for interpretation of measured bioimpedance data and a wide range of utilizations of bioimpedance in body composition estimation and evaluation of clinical status. This paper reviews the main concepts of bioimpedance measurement techniques including the frequency based, the allocation based, bioimpedance vector analysis and the real time bioimpedance analysis systems. Commonly used prediction equations for body composition assessment and influence of anthropometric measurements, gender, ethnic groups, postures, measurements protocols and electrode artifacts in estimated values are also discussed. In addition, this paper also contributes to the deliberations of bioimpedance analysis assessment of abnormal loss in lean body mass and unbalanced shift in body fluids and to the summary of diagnostic usage in different kinds of conditions such as cardiac, pulmonary, renal, and neural and infection diseases.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Resistance of a conductor is expressed by R = ρL2/V, so V = ρL2/R, where ρ is the conductor resistivity, L is the length, and V is volume.
“R of the conductor is expressed by R = ρL2/V, so V = ρL2/R, wherein ρ is the conductor resistivity, L is the length, and V is volume (48, 49).”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [49]
The electrical properties of the human body may explain the use of BIA to estimate FFM by the resistance index.
“The electrical properties of the human body may explain the use of BIA to estimate FFM by the resistance index.”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [49]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Dispersion and Absorption in Dielectrics I. Alternating Current Characteristics
Cole, K. S., & Cole, R. H. (1941) · The Journal of Chemical Physics · doi:10.1063/1.1750906 · Cited by 10,152 · see citing works
Bioelectrical impedance analysis?part I: review of principles and methods
KYLE, U. (2004) · Clinical Nutrition · doi:10.1016/j.clnu.2004.06.004 · Cited by 3,200 · see citing works
ESPEN Guidelines for Nutrition Screening 2002
Kondrup, J. (2003) · Clinical Nutrition · doi:10.1016/s0261-5614(03)00098-0 · Cited by 3,159 · see citing works
Bioelectrical impedance analysis—part II: utilization in clinical practice
Kyle, U. G., Bosaeus, I., De Lorenzo, A. D., Deurenberg, P., Elia, M., Manuel Gómez, J., Lilienthal Heitmann, B., Kent-Smith, L., Melchior, J.-C., Pirlich, M., et al. (2004) · Clinical Nutrition · doi:10.1016/j.clnu.2004.09.012 · Cited by 2,350 · see citing works
Assessment of fat-free mass using bioelectrical impedance measurements of the human body
Lukaski, H., Johnson, P., Bolonchuk, W., & Lykken, G. (1985) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/41.4.810 · Cited by 1,896 · see citing works
Body mass index as a measure of body fatness: age- and sex-specific prediction formulas
Deurenberg, P., Weststrate, J., & Seidell, J. C. (1991) · British Journal of Nutrition · doi:10.1079/bjn19910073 · Cited by 1,397 · see citing works
Validation of tetrapolar bioelectrical impedance method to assess human body composition
Lukaski, H. C., Bolonchuk, W. W., Hall, C. B., & Siders, W. A. (1986) · Journal of Applied Physiology · doi:10.1152/jappl.1986.60.4.1327 · Cited by 1,376 · cited 4× · see citing works
“It was introduced by Hoofer [4] and later revised by Nyboer [3] to decrease the contact impedance between skin and electrodes, and validated by Lukaski [17] in 140 normal adults.”
MethodInfluentialBody composition of reference children from birth to age 10 years
Fomon, S. J., Haschke, F., Ziegler, E., & Nelson, S. (1982) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/35.5.1169 · Cited by 1,264 · see citing works
Bioimpedance and Bioelectricity Basics
(2015) · Elsevier eBooks · doi:10.1016/c2012-0-06951-7 · Cited by 1,175 · see citing works
Norman, K., Stobäus, N., Pirlich, M., & Bosy-Westphal, A. (2012) · Clinical Nutrition · doi:10.1016/j.clnu.2012.05.008 · Cited by 1,049 · see citing works
A new method for monitoring body fluid variation by bioimpedance analysis: The RXc graph
Piccoli, A., Rossi, B., Pillon, L., & Bucciante, G. (1994) · Kidney International · doi:10.1038/ki.1994.305 · Cited by 711 · see citing works
Sun, S. S., Chumlea, W. C., Heymsfield, S. B., Lukaski, H. C., Schoeller, D., Friedl, K., Kuczmarski, R. J., Flegal, K. M., Johnson, C. L., & Hubbard, V. S. (2003) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/77.2.331 · Cited by 701 · cited 5× · see citing works
“[86] developed prediction Equations (19) and (20) of the TBW reference to dilution method using SF-BIA from a multi-ethnic pool of 1830 people aged from 12 to 94 years old:”
BackgroundMethodInfluentialPredicting body cell mass with bioimpedance by using theoretical methods: a technological review
LORENZO, A. D., Andréoli, A., Matthie, J. R., & Withers, P. O. (1997) · Journal of Applied Physiology · doi:10.1152/jappl.1997.82.5.1542 · Cited by 654 · see citing works
Body fluid volume determination via body composition spectroscopy in health and disease
Moissl, U. M., Wabel, P., Chamney, P. W., Bosaeus, I. G., Levin, N. W., Bosy‐Westphal, A., Korth, O., Müller, M. J., Ellegård, L., Malmros, V., et al. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/9/012 · Cited by 605 · see citing works
Single prediction equation for bioelectrical impedance analysis in adults aged 20–94 years
Kyle, U. G., Genton, L., Karsegard, L., Slosman, D. O., & Pichard, C. (2001) · Nutrition · doi:10.1016/s0899-9007(00)00553-0 · Cited by 573 · cited 5× · see citing works
“Several studies were conducted using the BIA method on children [68,105] adults [13], and elders [106,107].”
BackgroundMethodInfluential(1996) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.524s · Cited by 549 · see citing works
Correlation of whole-body impedance with total body water volume.
Hoffer, E. C., Meador, C. K., & Simpson, D. C. (1969) · Journal of Applied Physiology · doi:10.1152/jappl.1969.27.4.531 · Cited by 527 · see citing works
Electrical Impedance Plethysmography
Nyboer, J., Mm, K., & Hannapel, L. (1950) · Circulation · doi:10.1161/01.cir.2.6.811 · Cited by 493 · see citing works
Jaffrin, M. Y., & Morel, H. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2008.06.009 · Cited by 490 · see citing works
Bioelectric impedance phase angle and body composition
Baumgartner, R., Chumlea, W., & Roche, A. (1988) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/48.1.16 · Cited by 479 · see citing works
Kotler, D., Burastero, S., Wang, J., & Pierson, R. (1996) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.489s · Cited by 472 · see citing works
Body composition interpretation
Kyle, U. G., Schutz, Y., Dupertuis, Y. M., & Pichard, C. (2003) · Nutrition · doi:10.1016/s0899-9007(03)00061-3 · Cited by 456 · see citing works
Fat-free and fat mass percentiles in 5225 healthy subjects aged 15 to 98 years
Kyle, U. G., Genton, L., Slosman, D. O., & Pichard, C. (2001) · Nutrition · doi:10.1016/s0899-9007(01)00555-x · Cited by 440 · see citing works
THE MEASUREMENT OF TOTAL BODY WATER IN THE HUMAN SUBJECT BY DEUTERIUM OXIDE DILUTION
Schloerb, P. R., Friis-Hansen, B., Edelman, I. S., Solomon, A. K., & Moore, F. D. (1950) · Journal of Clinical Investigation · doi:10.1172/jci102366 · Cited by 411 · see citing works
Jebb, S. A., Cole, T. J., Doman, D., Murgatroyd, P. R., & Prentice, A. M. (2000) · British Journal Of Nutrition · doi:10.1017/s0007114500000155 · Cited by 378 · see citing works
Deurenberg, P., Kooy, K. V. D., Leenen, R. C. A., Weststrate, J. A., & Seidell, J. C. (1991) · PubMed · Cited by 347 · see citing works
Clinical characteristics influencing bioelectrical impedance analysis measurements
Kushner, R., Gudivaka, R., & Schoeller, D. A. (1996) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.423s · Cited by 337 · see citing works
Lukaski, H. C. (2013) · European Journal of Clinical Nutrition · doi:10.1038/ejcn.2012.149 · Cited by 306 · see citing works
Bioimpedance analysis: evaluation of leg-to-leg system based on pressure contact foot-pad electrodes
EZ, C. N., Gallagher, D., Visser, M., Pi‐Sunyer, F. X., Wang, Z., & Heymsfield, S. B. (1997) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199704000-00015 · Cited by 281 · see citing works
Single- and multifrequency models for bioelectrical impedance analysis of body water compartments
Gudivaka, R., Schoeller, D. A., Kushner, R. F., & Bolt, M. J. G. (1999) · Journal of Applied Physiology · doi:10.1152/jappl.1999.87.3.1087 · Cited by 261 · see citing works
Body composition: Why, when and for who?
Thibault, R., Genton, L., & Pichard, C. (2012) · Clinical Nutrition · doi:10.1016/j.clnu.2011.12.011 · Cited by 249 · see citing works
Piccoli, A., Pillon, L., & Dumler, F. (2002) · Nutrition · doi:10.1016/s0899-9007(01)00665-7 · Cited by 247 · see citing works
Membranes, ions, and impulses : a chapter of classical biophysics
Cole, K. S. (1968) · University of California Press eBooks · Cited by 245 · see citing works
Electrical Properties of Tissue
Duck, F. A. (1990) · Elsevier eBooks · doi:10.1016/b978-0-12-222800-1.50010-3 · Cited by 242 · see citing works
Moon, J. R. (2013) · European Journal of Clinical Nutrition · doi:10.1038/ejcn.2012.165 · Cited by 237 · see citing works
Principles and Biophysics of Dialysis
SARGENT, J. A., & Gotch, F. A. (1983) · Springer eBooks · doi:10.1007/978-94-009-6768-7_3 · Cited by 231 · see citing works
Assessing Body Composition among 3‐ to 8‐Year‐Old Children: Anthropometry, BIA, and DXA
Eisenmann, J. C., Heelan, K. A., & Welk, G. J. (2004) · Obesity Research · doi:10.1038/oby.2004.203 · Cited by 228 · see citing works
The Validity of Bioelectrical Impedance Models in Clinical Populations
Buchholz, A. C., Bartok, C., & Schoeller, D. A. (2004) · Nutrition in Clinical Practice · doi:10.1177/0115426504019005433 · Cited by 228 · see citing works
Toso, S., Piccoli, A., Gusella, M., Menon, D., Bononi, A., Crepaldi, G., & Ferrazzí, E. (2000) · Nutrition · doi:10.1016/s0899-9007(99)00230-0 · Cited by 218 · see citing works
Bioimpedance Spectroscopy for Clinical Assessment of Fluid Distribution and Body Cell Mass
Earthman, C. P., Traughber, D., Dobratz, J. R., & Howell, W. H. (2007) · Nutrition in Clinical Practice · doi:10.1177/0115426507022004389 · Cited by 210 · see citing works
Techniques of Body Composition Assessment: A Review of Laboratory and Field Methods
Wagner, D. R., & Heyward, V. H. (1999) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.1999.10608031 · Cited by 209 · see citing works
Analysis of Body Composition: A Critical Review of the Use of Bioelectrical Impedance Analysis
Mialich, M. S., Sicchieri, J. M. F., & Jordão, A. A. (2014) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) · doi:10.12691/ijcn-2-1-1 · Cited by 205 · see citing works
A new technique for establishing dry weight in hemodialysis patients via whole body bioimpedance
Chamney, P. W., Krämer, M., Rode, C., Kleinekofort, W., & Wizemann, V. (2002) · Kidney International · doi:10.1046/j.1523-1755.2002.00377.x · Cited by 203 · see citing works
Limitations and validation of bioelectrical impedance analysis in morbidly obese patients
Coppini, L. Z., Waitzberg, D. L., & Campos, A. C. L. (2005) · Current Opinion in Clinical Nutrition & Metabolic Care · doi:10.1097/01.mco.0000165013.54696.64 · Cited by 199 · see citing works
Genton, L., Hans, D. B., Kyle, U. G., & Pichard, C. (2002) · Nutrition · doi:10.1016/s0899-9007(01)00700-6 · Cited by 174 · see citing works
Predicting body fatness: the body mass index vs estimation by bioelectrical impedance.
Roubenoff, R., Dallal, G. E., & Wilson, P. W. F. (1995) · American Journal of Public Health · doi:10.2105/ajph.85.5.726 · Cited by 169 · see citing works
Body composition determination by bioimpedance: an update
Jaffrin, M. Y. (2009) · Current Opinion in Clinical Nutrition & Metabolic Care · doi:10.1097/mco.0b013e32832da22c · Cited by 167 · see citing works
Validity of body composition methods across ethnic population groups
Deurenberg, P., & Deurenberg‐Yap, M. (2003) · Acta Diabetologica · doi:10.1007/s00592-003-0077-z · Cited by 166 · see citing works
Segmental body composition assessed by bioelectrical impedance analysis and DEXA in humans
Bracco, D., Thiébaud, D., Chioléro, R. L., Landry, M., Burckhardt, P., & Schütz, Y. (1996) · Journal of Applied Physiology · doi:10.1152/jappl.1996.81.6.2580 · Cited by 159 · see citing works
Localized bioimpedance analysis in the evaluation of neuromuscular disease
Rutkove, S. B., Aaron, R., & Shiffman, C. A. (2002) · Muscle & Nerve · doi:10.1002/mus.10048 · Cited by 150 · see citing works
Comparison of body composition methods: a literature analysis
Fogelholm, M., & Lichtenbelt, W. V. M. (1997) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1600448 · Cited by 147 · see citing works
Multi-frequency impedance for the prediction of extracellular water and total body water
Deurenberg, P., Tagliabue, A., & Schouten, F. J. M. (1995) · British Journal Of Nutrition · doi:10.1079/bjn19950038 · Cited by 147 · see citing works
UTTER, A. C., Nieman, D. C., Ward, A., & Butterworth, D. E. (1999) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/69.4.603 · Cited by 146 · see citing works
Pichard, C., Kyle, U. G., Bracco, D., Slosman, D. O., Morabia, A., & Schütz, Y. (2000) · Nutrition · doi:10.1016/s0899-9007(00)00256-2 · Cited by 144 · see citing works
Cornish, B., Ward, L. C., Thomas, B. J., Jebb, S. A., & Elia, M. (1996) · PubMed · Cited by 136 · see citing works
Gender Differences in Body Composition from Childhood to Old Age: An Evolutionary Point of View
Kirchengast, S. (2010) · Journal of Life Sciences · doi:10.1080/09751270.2010.11885146 · Cited by 135 · see citing works
Analytic assessment of the various bioimpedance methods used to estimate body water
Matthie, J. R., Zarowitz, B. J. M., LORENZO, A. D., Andréoli, A., Katzarski, K. S., Pan, G., & Withers, P. O. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.84.5.1801 · Cited by 134 · see citing works
Chertow, G. M., Lazarus, J. M., Lew, N. L., Ma, L., & Lowrie, E. G. (1997) · Kidney International · doi:10.1038/ki.1997.216 · Cited by 131 · see citing works
Human hydrometry: comparison of multifrequency bioelectrical impedance with2H2O and bromine dilution
Ellis, K. J., & Wong, W. W. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.85.3.1056 · Cited by 130 · see citing works
Principles of electrical engineering materials and devices
Kasap, S. O. (1996) · Medical Entomology and Zoology · Cited by 128 · see citing works
Evaluation of body fat estimated from body mass index, skinfolds and impedance. A comparative study.
Heitmann, B. L. (1990) · PubMed · Cited by 120 · see citing works
Hannan, W. J., Cowen, S. J., Plester, C. E., Fearon, K. C. H., & Debeau, A. (1995) · Clinical Science · doi:10.1042/cs0890651 · Cited by 119 · see citing works
Basics of broadband impedance spectroscopy measurements using periodic excitations
Sanchez, B., Vandersteen, G., Bragós, R., & Schoukens, J. (2012) · Measurement Science and Technology · doi:10.1088/0957-0233/23/10/105501 · Cited by 115 · see citing works
Bio-electrical properties of tissue impedance measurements
Thommaset, A. (1962) · Lyon médical/Lyon médical · Cited by 115 · see citing works
Body composition variations in ageing.
Resić, J., Cikes-Culić, V., Zemunik, T., & Markotić, A. (2011) · PubMed · Cited by 113 · see citing works
Dynamics of segmental extracellular volumes during changes in body position by bioimpedance analysis
Zhu, F., Schneditz, D., Wang, E., & Levin, N. W. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.85.2.497 · Cited by 113 · see citing works
Impedance: a valid method in assessment of body composition?
Heitmann, B. L. (1994) · PubMed · Cited by 107 · see citing works
Thomas, B. J., Ward, L. C., & Cornish, B. (1998) · Applied Radiation and Isotopes · doi:10.1016/s0969-8043(97)00052-3 · Cited by 106 · see citing works
Bioelectrical Impedance Analysis For Measurement of Body Fluid Volumes
Thomas, B. J., Cornish, B., & Ward, L. C. (1992) · Journal of Clinical Engineering · doi:10.1097/00004669-199211000-00016 · Cited by 105 · see citing works
Scharfetter, H., Hartinger, P., Hinghofer-Szalkay, H. G., & Hutten, H. (1998) · Physiological Measurement · doi:10.1088/0967-3334/19/2/012 · Cited by 95 · see citing works
Pierson, R. N., Wang, J., Colt, E. W. D., & Neumann, P. H. (1982) · Journal of Chronic Diseases · doi:10.1016/0021-9681(82)90056-x · Cited by 94 · see citing works
Scharfetter, H., Monif, M., László, Z., Lambauer, T., Hutten, H., & Hinghofer-Szalkay, H. G. (1997) · Kidney International · doi:10.1038/ki.1997.150 · Cited by 94 · see citing works
Matthie, J. R. (2005) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00145.2005 · Cited by 94 · see citing works
A comparison of bioimpedance methods for detection of body cell mass change in HIV infection
Earthman, C. P., Matthie, J. R., Reid, P. M., Harper, I. T., Ravussin, É., & Howell, W. H. (2000) · Journal of Applied Physiology · doi:10.1152/jappl.2000.88.3.944 · Cited by 90 · see citing works
Optimal multisine excitation design for broadband electrical impedance spectroscopy
Sanchez, B., Vandersteen, G., Bragós, R., & Schoukens, J. (2011) · Measurement Science and Technology · doi:10.1088/0957-0233/22/11/115601 · Cited by 89 · see citing works
Tengvall, M., Ellegård, L., Malmros, V., Bosaeus, N., Lissner, L., & Bosaeus, I. (2009) · Clinical Nutrition · doi:10.1016/j.clnu.2008.10.005 · Cited by 87 · see citing works
Deurenberg, P., Deurenberg-Yap, M., & Schouten, F. (2002) · European Journal of Clinical Nutrition · doi:10.1038/sj.ejcn.1601303 · Cited by 87 · see citing works
Zhu, F., Kuhlmann, M. K., Kotanko, P., Seibert, E., Leonard, E. F., & Levin, N. W. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/6/s42 · Cited by 80 · see citing works
THOMASSET, M. A. (1962) · PubMed · Cited by 79 · see citing works
Sergi, G., Bussolotto, M., Perini, P., Calliari, I., Giantin, V., Ceccon, A., Scanferla, F., Bressan, M., Moschini, G., & Enzi, G. (1994) · Annals of Nutrition and Metabolism · doi:10.1159/000177806 · Cited by 76 · see citing works
Buffa, R., Mereu, R. M., Putzu, P. F., Floris, G., & Marini, E. (2010) · The journal of nutrition health & aging · doi:10.1007/s12603-010-0115-9 · Cited by 76 · see citing works
Cox-Reijven, P. L. M., & Soeters, P. B. (2000) · International Journal of Obesity · doi:10.1038/sj.ijo.0801123 · Cited by 76 · see citing works
Simpson, J. A., Lobo, D. N., ANDERSON, J. A., Macdonald, I. A., Perkins, A. C., Neal, K. R., Allison, S. P., & Rowlands, B. J. (2001) · Clinical Nutrition · doi:10.1054/clnu.2001.0398 · Cited by 76 · see citing works
Deurenberg, P., & Deurenberg‐Yap, M. (2002) · Asia Pacific Journal of Clinical Nutrition · doi:10.1046/j.1440-6047.2002.00258.x · Cited by 76 · see citing works
Rikkert, M. G. M. O., Deurenberg, P., Jansen, R. W. M. M., van't Hof, M. A., & Hoefnagels, W. H. L. (1997) · Journal of the American Geriatrics Society · doi:10.1111/j.1532-5415.1997.tb02934.x · Cited by 75 · see citing works
Tetrapolar body impedance is influenced by body posture and plasma sodium concentration.
Roos, A. N., Westendorp, R. G. J., Frölich, M., & Meinders, A. E. (1992) · PubMed · Cited by 75 · see citing works
Sanchez, B., Louarroudi, E., Jorge, E., Juan Cinca, A., Bragós, R., & Pintelon, R. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/3/339 · Cited by 74 · see citing works
Ayllón, D., Seoane, F., & Gil‐Pita, R. (2009) · doi:10.1109/iembs.2009.5334494 · Cited by 73 · see citing works
Bioimpedance or anthropometry?
Diaz, E. O., Villar, J., Immink, M., & Gonzales, T. (1989) · PubMed · Cited by 71 · see citing works
Xie, X., Kolthoff, N., Bärenholt, O., & Sp, N. (1999) · International Journal of Obesity · doi:10.1038/sj.ijo.0801034 · Cited by 69 · see citing works
Kim, J. H., Choi, S. H., Lim, S., Kim, K. W., Lim, J. Y., Cho, N. H., Park, K. S., & Jang, H. C. (2014) · Archives of Gerontology and Geriatrics · doi:10.1016/j.archger.2013.11.002 · Cited by 67 · see citing works
Patel, R. V., Peterson, E. L., Silverman, N., & Zarowitz, B. J. M. (1996) · Critical Care Medicine · doi:10.1097/00003246-199611000-00011 · Cited by 67 · see citing works
Saragat, B., Buffa, R., Mereu, E., RUI, M. D., Coin, A., Sergi, G., & Marini, E. (2013) · Experimental Gerontology · doi:10.1016/j.exger.2013.11.016 · Cited by 66 · see citing works
Marini, E., Sergi, G., Succa, V., Saragat, B., Sarti, S., Coin, A., Manzato, E., & Buffa, R. (2012) · The journal of nutrition health & aging · doi:10.1007/s12603-012-0411-7 · Cited by 65 · see citing works
Tanaka, N. I., Miyatani, M., Masuo, Y., Fukunaga, T., & Kanehisa, H. (2007) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00255.2007 · Cited by 65 · see citing works
Ward, L. C., Essex, T., & Cornish, B. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/9/007 · Cited by 64 · see citing works
A portable bio-impedance system for monitoring lung resistivity
Zlochiver, S., Arad, M., Radai, M. M., Barak‐Shinar, D., Krief, H., Engelman, T., Ben-Yehuda, R., Adunsky, A., & Abboud, S. (2006) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2006.02.005 · Cited by 62 · see citing works
Sanchez, B., Schoukens, J., Bragós, R., & Vandersteen, G. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2166116 · Cited by 58 · see citing works
Measurement of total body water by bioelectrical impedance in chronic renal failure.
Woodrow, G., Oldroyd, B., Turney, J. H., Davies, P., Day, J., & Smith, M. A. (1996) · PubMed · Cited by 56 · see citing works
A review of body water status and the effects of age and body fatness in children and adults.
Chumlea, W. C., Schubert, C. M., Sun, S. S., Demerath, E. W., Towne, B., & Siervogel, R. M. (2007) · PubMed · Cited by 54 · see citing works
Adjusting Dry Weight by Extracellular Volume and Body Composition in Hemodialysis Patients
Chen, Y., Chen, H.-H., Yeh, J.-C., & Chen, S. (2002) · The Nephron journals/Nephron journals · doi:10.1159/000064471 · Cited by 53 · see citing works
A new method for monitoring hydration at high altitude by bioimpedance analysis
Piccoli, A., Piazza, P. A., Noventa, D., Pillon, L., & Zaccaria, M. (1996) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-199612000-00012 · Cited by 53 · see citing works
Effects of skin blood flow and temperature on bioelectric impedance after exercise
Liang, M. T. C., NORRIS, S., EXERCISE, MASS, F.-F., VOLUME, S. B., VELOCITY, F., & EQUATIONS, P. (1993) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-199311000-00005 · Cited by 52 · see citing works
Ward, L. C., Elia, M., & Cornish, B. (1998) · Physiological Measurement · doi:10.1088/0967-3334/19/1/005 · Cited by 48 · see citing works
Thomasset, A. (1963) · PubMed · Cited by 48 · see citing works
de Azevedo, Z. M. A., Moore, D. C. B. C., de Matos, F. A. A., Fonseca, V. M., Peixoto, M. V. M., Elsas, M., Santinoni, E., dos Anjos, L. A., & Ramos, E. G. (2013) · Clinical Nutrition · doi:10.1016/j.clnu.2013.01.011 · Cited by 47 · see citing works
Paton, N. I., Elia, M., Jennings, G., Ward, L. C., & Griffin, G. E. (1998) · Nutrition · doi:10.1016/s0899-9007(98)00105-1 · Cited by 47 · see citing works
Total body water measurement by a modification of the bioimpedance spectroscopy method
Jaffrin, M. Y., Fenech, M., Moreno, M. V., & Kieffer, R. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0099-0 · Cited by 44 · see citing works
Bioelectrical impedance analysis measurements as part of a national nutrition survey
Kuczmarski, R. J. (1996) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/64.3.453s · Cited by 43 · see citing works
Body composition at the bedside
Ghosh, S. K., Meister, D., Cowen, S. J., Hannan, J., & Ferguson, A. (1997) · European Journal of Gastroenterology & Hepatology · doi:10.1097/00042737-199708000-00009 · Cited by 43 · see citing works
Baarends, E. M., Lichtenbelt, W. D. V. M., Wouters, E. F. �., & Schols, A. M. W. J. (1998) · Clinical Nutrition · doi:10.1016/s0261-5614(98)80038-1 · Cited by 42 · see citing works
Validation of a three-frequency bioimpedance spectroscopic method for body composition analysis
Ward, L. C., Dyer, J. M., Byrne, N. M., Sharpe, K., & Hills, A. P. (2007) · Nutrition · doi:10.1016/j.nut.2007.06.009 · Cited by 41 · see citing works
Cox-Reijven, P. L. M., van Kreel, B., & Soeters, P. B. (2003) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/78.6.1111 · Cited by 40 · see citing works
Bracco, D., Revelly, J., Berger, M. M., & Chioléro, R. L. (1998) · Critical Care Medicine · doi:10.1097/00003246-199806000-00029 · Cited by 39 · see citing works
Sanchez, B., Rojas, C. R., Vandersteen, G., Bragós, R., & Schoukens, J. (2012) · Measurement Science and Technology · doi:10.1088/0957-0233/23/8/085702 · Cited by 38 · see citing works
Hoyle, G. E., Chua, M., & Soiza, R. L. (2010) · QJM · doi:10.1093/qjmed/hcq157 · Cited by 38 · see citing works
Body composition changes by DXA, BIA and skinfolds during exercise training in women
Sillanpää, E., HÄKKINEN, A. H., & Häkkinen, K. (2013) · European Journal of Applied Physiology · doi:10.1007/s00421-013-2669-9 · Cited by 37 · see citing works
Dielectric properties of emulsions
Hanai, T. (1961) · Colloid & Polymer Science · doi:10.1007/bf01521332 · Cited by 36 · see citing works
Ibrahim, F., Taib, M. N., Abas, W. A. B. W., Guan, C. C., & Sulaiman, S. (2005) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2004.840237 · Cited by 36 · see citing works
Novel approach of processing electrical bioimpedance data using differential impedance analysis
Sanchez, B., Bandarenka, A. S., Vandersteen, G., Schoukens, J., & Bragós, R. (2013) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2013.03.006 · Cited by 35 · see citing works
Deurenberg, P., Wolde-Gebriel, Z., & Schouten, F. J. M. (1995) · Annals of Nutrition and Metabolism · doi:10.1159/000177868 · Cited by 33 · see citing works
Assessment of body composition in adolescent subjects with anorexia nervosa by bioimpedance
Moreno, M. V., Djeddi, D., & Jaffrin, M. Y. (2007) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.09.005 · Cited by 33 · see citing works
Prediction of body composition in elderly men over 75 years of age
Fuller, N. J., Sawyer, M. B., Laskey, M. A., Paxton, P., & Elia, M. (1996) · Annals of Human Biology · doi:10.1080/03014469600004352 · Cited by 32 · see citing works
Schols, A. M. W. J., Dingemans, A. C., Soeters, P. B., & Wouters, E. F. �. (1990) · Clinical Nutrition · doi:10.1016/0261-5614(90)90035-q · Cited by 31 · see citing works
Bioimpedance and Bioimpedance Vector Analysis in patients with Anorexia Nervosa
Haas, V. K., Riedl, A., Hofmann, T., Nischan, A., Burghardt, R., Boschmann, M., & Klapp, B. F. (2012) · European Eating Disorders Review · doi:10.1002/erv.1166 · Cited by 30 · see citing works
Nigam, P., Misra, A., & Colles, S. L. (2013) · Diabetes & Metabolic Syndrome Clinical Research & Reviews · doi:10.1016/j.dsx.2013.02.031 · Cited by 29 · see citing works
A comparison of segmental and wrist-to-ankle methodologies of bioimpedance analysis
Thomas, B. J., Cornish, B., Ward, L. C., & Patterson, M. A. (1998) · Applied Radiation and Isotopes · doi:10.1016/s0969-8043(97)00058-4 · Cited by 27 · see citing works
Total Body Water Measured by Bioelectrical Impedance in Patients After Hemodialysis
Pastan, S. O., & Gassensmith, C. M. (1992) · ASAIO Journal · doi:10.1097/00002480-199207000-00016 · Cited by 26 · see citing works
Deurenberg, P., Andréoli, A., & LORENZO, A. D. (1996) · Annals of Human Biology · doi:10.1080/03014469600004252 · Cited by 25 · see citing works
Seibert, E., Müller, S. G., Fries, P., Pattmöller, J., Kuss, O., Heine, G., Girndt, M., Schneider, G., Kotanko, P., Zhu, F., et al. (2013) · Kidney & Blood Pressure Research · doi:10.1159/000343400 · Cited by 24 · see citing works
Bioelectrical impedance analysis to assess changes intotal body water in patients with cancer
Simons, J., Schols, A. M. W. J., Westerterp, K. R., ten Velde, G. P. M., & Woliters, E. F. M. (1999) · Clinical Nutrition · doi:10.1016/s0261-5614(99)80047-8 · Cited by 24 · see citing works
Scanferla, F., Landini, S., Fracasso, A., Morachiello, P., Righetto, F., Toffoletto, P., & Bazzato, G. (1990) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/5.suppl_1.167 · Cited by 24 · see citing works
Buendía, R., Bogónez-Franco, P., Nescolarde, L., & Seoane, F. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2012.05.011 · Cited by 23 · see citing works
Libraty, D. H., Endy, T. P., Kalayanarooj, S., Chansiriwongs, W., Nisalak, A., Green, S., Ennis, F. A., & Rothman, A. L. (2002) · Transactions of the Royal Society of Tropical Medicine and Hygiene · doi:10.1016/s0035-9203(02)90104-5 · Cited by 22 · see citing works
Dynamic assessment of fat-free mass during catabolism and recovery
Kyle, U. G., & Pichard, C. (2000) · Current Opinion in Clinical Nutrition & Metabolic Care · doi:10.1097/00075197-200007000-00013 · Cited by 21 · see citing works
Pichler, G. P., Amouzadeh-Ghadikolai, O., Leis, A., & Skrabal, F. (2012) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2012.07.006 · Cited by 21 · see citing works
Adverse effects of near current-electrode placement in non-invasive bio-impedance measurements
Shiffman, C. A. (2013) · Physiological Measurement · doi:10.1088/0967-3334/34/11/1513 · Cited by 20 · see citing works
Piccoli, A., Pillon, L., Pisanello, L., & Zacchello, G. (1999) · Nutrition · doi:10.1016/s0899-9007(98)00139-7 · Cited by 18 · see citing works
Ward, L. C., & Heitmann, B. L. (2001) · Clinical Nutrition · doi:10.1054/clnu.2000.0362 · Cited by 16 · see citing works
Cumming, K., Hoyle, G. E., Hutchison, J. D., & Soiza, R. L. (2014) · The journal of nutrition health & aging · doi:10.1007/s12603-014-0446-z · Cited by 14 · see citing works
A comparison of the whole-body and segmental methodologies of bioimpedance analysis
Thomas, B. J., Cornish, B., Pattemore, M., Jacobs, M., & Ward, L. C. (2003) · Acta Diabetologica · doi:10.1007/s00592-003-0074-2 · Cited by 13 · see citing works
Ward, L. C., & Heitmann, B. L. (2000) · Nutrition · doi:10.1016/s0899-9007(00)00225-2 · Cited by 13 · see citing works
Siváková, D., Vondrová, D., Valkovič, P., Cvíčelová, M., Daňková, Z., & Luptáková, L. (2013) · Open Life Sciences · doi:10.2478/s11535-013-0216-7 · Cited by 8 · see citing works
A Device for Monitoring Hydration State in Hemodialysis Patients Using a Calf Bioimpedance Technique
Zhu, F., Kuhlman, M., Kotanko, P., Handelman, G. J., Leonard, E. F., & Levin, N. W. (2007) · Springer eBooks · doi:10.1007/978-3-540-73841-1_200 · Cited by 6 · see citing works
A comparison between two methods for measuring limb resistances with wrist and ankle electrodes
Jaffrin, M. Y., Bousbiat, S., & Dongmo, E. (2011) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2011.03.003 · Cited by 6 · see citing works
Schulz, H., Teske, D., Penven, D., & Tomczak, J. (2006) · Nutrition · doi:10.1016/j.nut.2006.04.007 · Cited by 6 · see citing works
Relationship Between Physical Measures of Anthropometry and Bioimpedance Measures
Serrano, M. D. M., de Espinosa, M. G. M., & Zamorano, E. M. (2012) · Springer eBooks · doi:10.1007/978-1-4419-1788-1_25 · Cited by 5 · see citing works
Mridha, S. (2010) · British Journal of Sports Medicine · doi:10.1136/bjsm.2010.078725.123 · Cited by 4 · see citing works
Atilano-Carsi, X., Miguel, J. L. S., Martínez, J., Sánchez, R. J., & Selgas, R. (2012) · Kidney Research and Clinical Practice · doi:10.1016/j.krcp.2012.04.331 · Cited by 4 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Pitfalls in the measurement of muscle mass: A need for a reference standard
Buckinx, F., Landi, F., Cesari, M., Fielding, R. A., Visser, M., Engelke, K., Maggi, S., Dennison, E., Al‐Daghri, N. M., Allepaerts, S., et al. (2018) · Journal of Cachexia, Sarcopenia and Muscle · doi:10.1002/jcsm.12268 · Cited by 765 · see citing works
Advanced Body Composition Assessment: From Body Mass Index to Body Composition Profiling
Borga, M., West, J., Bell, J. D., Harvey, N. C., Romu, T., Heymsfield, S. B., & Leinhard, O. D. (2018) · Journal of Investigative Medicine · doi:10.1136/jim-2018-000722 · Cited by 626 · see citing works
Strengths and Limitations of BMI in the Diagnosis of Obesity: What is the Path Forward?
Sweatt, K., Garvey, W. T., & Martins, C. (2024) · Current Obesity Reports · doi:10.1007/s13679-024-00580-1 · Cited by 398 · see citing works
Electrical impedance spectroscopy (EIS) for biological analysis and food characterization: a review
Grossi, M., & Riccò, B. (2017) · Journal of sensors and sensor systems · doi:10.5194/jsss-6-303-2017 · Cited by 395 · see citing works
Kuriyan, R. (2018) · Indian Journal of Medical Research · doi:10.4103/ijmr.ijmr_1777_18 · Cited by 384 · see citing works
Campa, F., Toselli, S., Mazzilli, M., Gobbo, L. A., & Coratella, G. (2021) · Nutrients · doi:10.3390/nu13051620 · Cited by 315 · see citing works
Chang, K. C., Hsu, T.-H., Wu, W., Huang, K., & Han, D. (2017) · Age and Ageing · doi:10.1093/ageing/afx094 · Cited by 278 · see citing works
Low muscle mass and risk of type 2 diabetes in middle-aged and older adults: findings from the KoGES
Son, J. W., Lee, S. S., Kim, S. R., Yoo, S. J., Cha, B. Y., Son, H., & Cho, N. H. (2017) · Diabetologia · doi:10.1007/s00125-016-4196-9 · Cited by 223 · see citing works
Sarcopenia in cancer: Risking more than muscle loss
Anjanappa, M., Corden, M., Green, A., Roberts, D. L., Hoskin, P., McWilliam, A., & Choudhury, A. (2020) · Technical Innovations & Patient Support in Radiation Oncology · doi:10.1016/j.tipsro.2020.10.001 · Cited by 203 · see citing works
Naranjo-Hernández, D., Reina‐Tosina, J., & Min, M. (2019) · Journal of Sensors · doi:10.1155/2019/9210258 · Cited by 195 · see citing works
Abasi, S., Aggas, J. R., Garayar-Leyva, G. G., Walther, B. K., & Guiseppi‐Elie, A. (2022) · ACS Measurement Science Au · doi:10.1021/acsmeasuresciau.2c00033 · Cited by 183 · see citing works
Clinical and Genetic Determinants of Varicose Veins
Fukaya, E., Flores, A. M., Lindholm, D. P., Gustafsson, S., Zanetti, D., Ingelsson, E., & Leeper, N. J. (2018) · Circulation · doi:10.1161/circulationaha.118.035584 · Cited by 165 · see citing works
Electrical Impedance Tomography for Cardio-Pulmonary Monitoring
Putensen, C., Hentze, B., Muenster, S., & Muders, T. (2019) · Journal of Clinical Medicine · doi:10.3390/jcm8081176 · Cited by 160 · see citing works
Santos, H. O., Price, J. C., & Bueno, A. A. (2020) · Nutrients · doi:10.3390/nu12103159 · Cited by 157 · see citing works
Beaudart, C., Bruyère, O., Geerinck, A., Hajaoui, M., Scafoglieri, A., Perkisas, S., Bautmans, I., Gielen, E., Reginster, J.-Y., & Buckinx, F. (2020) · Clinical Nutrition ESPEN · doi:10.1016/j.clnesp.2019.09.012 · Cited by 113 · see citing works
Electrical Impedance Tomography for Biomedical Applications: Circuits and Systems Review
Wu, Y., Hanzaee, F. F., Jiang, D., Bayford, R. H., & Demosthenous, A. (2021) · IEEE Open Journal of Circuits and Systems · doi:10.1109/ojcas.2021.3075302 · Cited by 107 · see citing works
Yamada, Y., Yoshida, T., Yokoyama, K., Watanabe, Y., Miyake, M., Yamagata, E., Yamada, M., Kimura, M., & Study, K.-K. (2016) · The Journals of Gerontology Series A · doi:10.1093/gerona/glw125 · Cited by 104 · see citing works
Latent Factors Limiting the Performance of sEMG-Interfaces
Lobov, S. A., Krilova, N. P., Kastalskiy, I. A., Kazantsev, V., & Makarov, V. A. (2018) · Sensors · doi:10.3390/s18041122 · Cited by 102 · see citing works
Applications of bioimpedance measurement techniques in tissue engineering
Amini, M., Hisdal, J., & Kalvøy, H. (2018) · Journal of Electrical Bioimpedance · doi:10.2478/joeb-2018-0019 · Cited by 99 · see citing works
Sheppard, V. B., Hicks, J., Makambi, K. H., Hurtado‐de‐Mendoza, A., Demark‐Wahnefried, W., & Adams-Campbell, L. L. (2015) · Contemporary Clinical Trials · doi:10.1016/j.cct.2015.12.005 · Cited by 92 · see citing works
Electrical Impedance Tomography: From the Traditional Design to the Novel Frontier of Wearables
Pennati, F., Angelucci, A., Morelli, L., Bardini, S., Barzanti, E., Cavallini, F., Conelli, A., Federico, G. D., Paganelli, C., & Aliverti, A. (2023) · Sensors · doi:10.3390/s23031182 · Cited by 88 · see citing works
Phase angle as a marker of muscle quality: A systematic review and meta-analysis
Pereira, J. P. D. C., Rebouças, A. D. S., Prado, C. M. M., González, M. C., Cabral, P. C., Diniz, A. D. S., Fayh, A. P. T., & Silva, F. M. (2024) · Clinical Nutrition · doi:10.1016/j.clnu.2024.11.008 · Cited by 87 · see citing works
Lee, L., Liao, Y.-S., Lu, H.-K., Hsiao, P.-L., Chen, Y.-Y., Ching-Chi, C., & Hsieh, K.-C. (2017) · PLoS ONE · doi:10.1371/journal.pone.0171568 · Cited by 84 · see citing works
Campa, F., Matias, C. N., Gatterer, H., Toselli, S., Koury, J. C., Andréoli, A., Melchiorri, G., Sardinha, L. B., & Silva, A. M. (2019) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph16245066 · Cited by 76 · see citing works
Park, J. H., Jo, Y., & Lee, J. (2018) · The Korean Journal of Internal Medicine · doi:10.3904/kjim.2018.197 · Cited by 72 · see citing works
Son, J. W., Han, B., Bennett, J. P., Heymsfield, S., & Lim, S. (2025) · Obesity Reviews · doi:10.1111/obr.13844 · Cited by 74 · see citing works
Bera, T. K., Nagaraju, J., & Gilles Lubineau, A. (2016) · Journal of Visualization · doi:10.1007/s12650-016-0351-0 · Cited by 69 · see citing works
Uğraş, S. (2020) · Libyan Journal of Medicine · doi:10.1080/19932820.2020.1741904 · Cited by 66 · see citing works
Costa‐Urrutia, P., Vizuet-Gámez, A., Ramirez-Alcántara, M., Guillén-González, M. Á., Contreras, O. M., Valdés-Moreno, M. I., Musalem-Younes, C., Solares-Tlapechco, J., Granados, J., Franco‐Trecu, V., et al. (2019) · PLoS ONE · doi:10.1371/journal.pone.0212792 · Cited by 63 · see citing works
Developments in the human machine interface technologies and their applications: a review
Singh, H. P., & Kumar, P. (2021) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2021.1936237 · Cited by 65 · see citing works
Non-Invasive Electrochemical Biosensors Operating in Human Physiological Fluids
Falk, M., Psotta, C., Cirovic, S., & Shleev, S. (2020) · Sensors · doi:10.3390/s20216352 · Cited by 62 · see citing works
The role of non-invasive devices for the telemonitoring of heart failure patients
Faragli, A., Abawi, D., Quinn, C., Cvetković, M., Schlabs, T., Tahirović, E., Düngen, H., Pieske, B. M., Kelle, S., Edelmann, F., et al. (2020) · Heart Failure Reviews · doi:10.1007/s10741-020-09963-7 · Cited by 60 · see citing works
Psoriatic disease and body composition: A systematic review and narrative synthesis
Blake, T., Gullick, N., Hutchinson, C. E., & Barber, T. M. (2020) · PLoS ONE · doi:10.1371/journal.pone.0237598 · Cited by 59 · see citing works
Feng, Q., Bešević, J., Conroy, M. C., Omiyale, W., Lacey, B., & Allen, N. E. (2024) · Clinical Nutrition ESPEN · doi:10.1016/j.clnesp.2024.06.040 · Cited by 59 · see citing works
Krzesiński, P., Jankowska, E. A., Siebert, J., Galas, A. A., Piotrowicz, K., Stańczyk, A., Siwołowski, P., Gutknecht, P. G., Chrom, P., Murawski, P., et al. (2021) · European Journal of Heart Failure · doi:10.1002/ejhf.2358 · Cited by 51 · see citing works
Techapiroontong, S., Limpuangthip, N., Tumrasvin, W., & Sirotamarat, J. (2022) · BMC Oral Health · doi:10.1186/s12903-022-02103-7 · Cited by 51 · see citing works
Equivalent circuit models using CPE for impedance spectroscopy of electronic ceramics
Pandey, S., Kumar, D., Parkash, O., & Pandey, L. (2017) · Integrated ferroelectrics · doi:10.1080/10584587.2017.1376984 · Cited by 50 · see citing works
Piuzzi, E., Pisa, S., Pittella, E., Podestà, L., & Sangiovanni, S. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2887303 · Cited by 49 · see citing works
MALNUTRITION AND CLINICAL OUTCOMES IN SURGICAL PATIENTS WITH COLORECTAL DISEASE
Nishiyama, V. K. G., Albertini, S. M., de Moraes, C. M. Z. G., de Godoy, M. F., & Netinho, J. G. (2018) · Arquivos de Gastroenterologia · doi:10.1590/s0004-2803.201800000-85 · Cited by 49 · see citing works
Wearable Belt With Built-In Textile Electrodes for Cardio—Respiratory Monitoring
Piuzzi, E., Pisa, S., Pittella, E., Podestà, L., & Sangiovanni, S. (2020) · Sensors · doi:10.3390/s20164500 · Cited by 49 · see citing works
Stratton, M. T., Smith, R. W., Harty, P. S., Rodriguez, C., Johnson, B. A., Dellinger, J. R., Williams, A. D., White, S. J., Benavides, M. L., & Tinsley, G. M. (2021) · European Journal of Clinical Nutrition · doi:10.1038/s41430-020-00811-3 · Cited by 46 · see citing works
Ward, L. C., Isenring, E. A., Dyer, J. M., Kagawa, M., & Essex, T. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/7/1529 · Cited by 47 · see citing works
Pathiraja, A. A., Weerakkody, R. A., von Roon, A. C., Ziprin, P., & Bayford, R. H. (2020) · Journal of Translational Medicine · doi:10.1186/s12967-020-02395-9 · Cited by 47 · see citing works
Analytical Model for Blood Glucose Detection Using Electrical Impedance Spectroscopy
Pedro, B. G., Marcôndes, D. W. C., & Bertemes-Filho, P. (2020) · Sensors · doi:10.3390/s20236928 · Cited by 46 · see citing works
Bioimpedance analysis is safe in patients with implanted cardiac electronic devices
Chabin, X., Taghli-Lamallem, O., Mulliez, A., Bordachar, P., Jean, F., Futier, E., Massoullié, G., Andonache, M., Souteyrand, G., Ploux, S., et al. (2018) · Clinical Nutrition · doi:10.1016/j.clnu.2018.02.029 · Cited by 44 · see citing works
Wearable Bioimpedance Monitoring: Viewpoint for Application in Chronic Conditions
Groenendaal, W., Lee, S., & Hoof, C. A. V. (2021) · JMIR Biomedical Engineering · doi:10.2196/22911 · Cited by 43 · see citing works
Bioimpedance analysis of vascular tissue and fluid flow in human and plant body: A review
Prasad, A., & Roy, M. (2020) · Biosystems Engineering · doi:10.1016/j.biosystemseng.2020.06.006 · Cited by 43 · see citing works
Bioimpedance Sensor and Methodology for Acute Pain Monitoring
Ghita, M., Neckebroek, M. M., Juchem, J., Copoţ, D., Muresan, C. I., & Ionescu, C. M. (2020) · Sensors · doi:10.3390/s20236765 · Cited by 44 · see citing works
Sahathevan, S., Khor, B., Singh, B. K. S., Sabatino, A., Fiaccadori, E., Daud, Z. A. M., Ali, M., Narayanan, S. S., Tallman, D. A., Chinna, K., et al. (2020) · Nutrients · doi:10.3390/nu12113597 · Cited by 43 · see citing works
Shah, U. A., Ballinger, T. J., Bhandari, R., Dieli-Conwright, C. M., Guertin, K. A., Hibler, E. A., Kalam, F., Lohmann, A. E., & Ippolito, J. E. (2023) · JNCI Monographs · doi:10.1093/jncimonographs/lgad001 · Cited by 43 · see citing works
Structured Design Methodology to Achieve a High SNR Electrical Impedance Tomography
Takhti, M., & Odame, K. M. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2894157 · Cited by 41 · see citing works
Label-free impedance flow cytometry for nanotoxicity screening
Ostermann, M., Sauter, A., Xue, Y., Birkeland, E. S., Schoelermann, J., Holst, B., & Cimpan, M. R. (2020) · Scientific Reports · doi:10.1038/s41598-019-56705-3 · Cited by 40 · see citing works
He, Q., Xia, B., Liu, A., Li, M., Zhou, Z., Cheung, E. C., Kuo, Z. C., Wang, B., Li, F., Tang, Y., et al. (2021) · International Journal of Cancer · doi:10.1002/ijc.33697 · Cited by 39 · see citing works
Peinado, A. B., Alfaro-Magallanes, V. M., Romero‐Parra, N., Barba‐Moreno, L., Rael, B., Maestre-Cascales, C., Rojo‐Tirado, M. Á., de Castro, E. A., Benito, P. J., Ortega‐Santos, C. P., et al. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18020735 · Cited by 38 · see citing works
Martínez-Rodríguez, A., Martínez-Olcina, M., Hernández, M., Rubio‐Arias, J. Á., Sánchez‐Sánchez, J., Cobos, D. L., Vicente-Martínez, M., Carvalho, M. J., & Sánchez-Sáez, J. A. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18062837 · Cited by 38 · see citing works
Ghio, S., Acquaro, M., Agostoni, P., Ambrosio, G., Carluccio, E., Castiglione, V., Colombo, D., D’Alto, M., Dellegrottaglie, S., Dini, F. L., et al. (2022) · Heart Failure Reviews · doi:10.1007/s10741-022-10276-0 · Cited by 37 · see citing works
Yang, L. J., Zhang, G., Song, J., Dai, M., Xu, C., Dong, X., & Fu, F. (2016) · Sensors · doi:10.3390/s16111942 · Cited by 37 · see citing works
Impedance Analysis to Evaluate Nutritional Status in Physiological and Pathological Conditions
Catapano, A., Trinchese, G., Cimmino, F., Petrella, L., D’Angelo, M., Maio, G. D., Crispino, M., Cavaliere, G., Monda, M., & Mollica, M. P. (2023) · Nutrients · doi:10.3390/nu15102264 · Cited by 40 · see citing works
Esen, İ., Arslan, H., Esen, S. A., Gülşen, M., Kültekin, N., & Özdemir, O. (2024) · Medicine · doi:10.1097/md.0000000000037258 · Cited by 35 · see citing works
Phase Angle Is a Marker of Muscle Quantity and Strength in Overweight/Obese Former Athletes
Matias, C. N., Campa, F., Nunes, C. L., Francisco, R., Jesus, F., Cardoso, M., Valamatos, M. J., Mil-Homens, P. V., Sardinha, L. B., Martins, P., et al. (2021) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph18126649 · Cited by 35 · see citing works
Remote Monitoring to Reduce Heart Failure Readmissions
Emani, S. (2017) · Current Heart Failure Reports · doi:10.1007/s11897-017-0315-2 · Cited by 35 · see citing works
Type 1 diabetes and body composition in youth: A systematic review
Calella, P., Gallé, F., Fornelli, G., Liguori, G., & Valerio, G. (2019) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3211 · Cited by 34 · see citing works
Shajith Anoop, S., Krakauer, J. C., Krakauer, N. Y., & Misra, A. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2020-001324 · Cited by 34 · see citing works
Nutritional assessment and support during continuous renal replacement therapy
Ostermann, M., Lumlertgul, N., & Mehta, R. L. (2021) · Seminars in Dialysis · doi:10.1111/sdi.12973 · Cited by 34 · see citing works
A review of bio-impedance devices
Showkat, I., Khanday, F. A., & Beigh, M. R. (2023) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02763-1 · Cited by 37 · see citing works
Broers, N. J. H., Canaud, B., Dekker, M. J. E., SANDE, F. M. V. D., Stuard, S., Wabel, P., & Kooman, J. P. (2020) · Hemodialysis International · doi:10.1111/hdi.12812 · Cited by 33 · see citing works
Vaquero-Cristóbal, R., Albaladejo-Saura, M., Luna-Badachi, A. E., & Esparza-Ros, F. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17217777 · Cited by 32 · see citing works
Rosas‐Carrasco, Ó., Valenzuela, R. E. R., & López-Teros, M. T. (2021) · Frontiers in Medicine · doi:10.3389/fmed.2021.617126 · Cited by 32 · see citing works
Thanapholsart, J., Khan, E., & Lee, G. (2022) · Biological Research For Nursing · doi:10.1177/10998004221132838 · Cited by 32 · see citing works
Mahdavi, R., Hosseinpour, P., Abbasvandi, F., Mehrvarz, S., Yousefpour, N., Ataee, H., Parniani, M., Mamdouh, A., Ghafari, H., & Abdolahad, M. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112421 · Cited by 31 · see citing works
Simple Wireless Impedance Pneumography System for Unobtrusive Sensing of Respiration
Aqueveque, P., Gómez, B., Monsalve, E., Germany, E., Ortega-Bastidas, P., Dubó, S., & Pino, E. J. (2020) · Sensors · doi:10.3390/s20185228 · Cited by 31 · see citing works
Body Composition Over the Menstrual and Oral Contraceptive Cycle in Trained Females
Rael, B., Romero‐Parra, N., Alfaro-Magallanes, V. M., Barba‐Moreno, L., Cupeiro, R., de Jonge, X. J., Peinado, A. B., & _, _. (2020) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2020-0038 · Cited by 31 · see citing works
A New Venous Entry Detection Method Based on Electrical Bio-impedance Sensing
Cheng, Z., Davies, B. L., Caldwell, D. G., & Mattos, L. S. (2018) · Annals of Biomedical Engineering · doi:10.1007/s10439-018-2025-7 · Cited by 31 · see citing works
Effectiveness of impedance parameters for muscle quality evaluation in healthy men
Sato, H., Nakamura, T., Kusuhara, T., Kobara, K., Kuniyasu, K., Kawashima, T., & Hanayama, K. (2020) · The Journal of Physiological Sciences · doi:10.1186/s12576-020-00780-z · Cited by 30 · see citing works
Sanai, F., Sahid, A. S., Huvanandana, J., Spoa, S., Boyle, L. H., Hribar, J., Wang, D. T.-Y., Kwan, B., Colagiuri, S., Cox, S. J., et al. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211054110 · Cited by 30 · see citing works
Extremity compartment syndrome: A review with a focus on non-invasive methods of diagnosis
Novák, M., Penhaker, M., Raska, P., Pleva, L., & Schmidt, M. (2022) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2022.801586 · Cited by 30 · see citing works
Impedance-Readout Integrated Circuits for Electrical Impedance Spectroscopy: Methodological Review
Cheon, S.-I., Choi, H., Kang, H., Suh, J.-H., Park, S., Kweon, S.-J., Je, M., & Ha, S. (2023) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2023.3319212 · Cited by 29 · see citing works
Samouda, H., & Langlet, J. (2022) · BMC Endocrine Disorders · doi:10.1186/s12902-022-01111-6 · Cited by 26 · see citing works
Törpel, A., Peter, B., & Schega, L. (2020) · Frontiers in Physiology · doi:10.3389/fphys.2020.00335 · Cited by 29 · see citing works
Cammarota, C., & Pinto, A. (2020) · Journal of Applied Statistics · doi:10.1080/02664763.2020.1763930 · Cited by 27 · see citing works
Riezzo, G., Prospero, L., D’Attoma, B., Ignazzi, A., Bianco, A., Franco, I., Curci, R., Campanella, A., Bonfiglio, C., Osella, A. R., et al. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12165359 · Cited by 27 · see citing works
The Exercise, Arterial Modulation and Nutrition in Youth South Africa Study (ExAMIN Youth SA)
Kruger, R., Monyeki, M. A., Schutte, A. E., Smith, W. T., Mels, C. M. C., Kruger, H. S., Pienaar, A. E., Gafane-Matemane, L. F., Breet, Y., Lammertyn, L., et al. (2020) · Frontiers in Pediatrics · doi:10.3389/fped.2020.00212 · Cited by 27 · see citing works
Estimation of Muscle Mass in the Integrated Assessment of Patients on Hemodialysis
Sabatino, A., Broers, N. J. H., SANDE, F. M. V. D., Hemmelder, M. H., Fiaccadori, E., & Kooman, J. P. (2021) · Frontiers in Nutrition · doi:10.3389/fnut.2021.697523 · Cited by 27 · see citing works
AlKhatib, A. H., & Obita, G. (2024) · Nutrients · doi:10.3390/nu16111730 · Cited by 27 · see citing works
Buckinx, F., Reginster, J., Brunois, T., Lenaerts, C., Beaudart, C., Croisier, J., Pétermans, J., & Bruyère, O. (2017) · Europe PMC (PubMed Central) · Cited by 27 · see citing works
Normal Values of Hemoglobin Mass and Blood Volume in Young, Active Women and Men
Falz, R., Fikenzer, S., Hoppe, S., & Busse, M. W. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0826-9235 · Cited by 26 · see citing works
Logothetis, I., Fernández, R., Troynikov, O., Dabnichki, P., Pirogova, E., & Gil, I. (2019) · Measurement Science and Technology · doi:10.1088/1361-6501/ab3201 · Cited by 26 · see citing works
de Oliveira, A. L. M., O’Connor, T., Patton, D., Strapp, H., & Moore, Z. E. H. (2022) · Journal of Wound Care · doi:10.12968/jowc.2022.31.3.254 · Cited by 26 · see citing works
Górnicka, M., Szewczyk, K., Bialkowska, A. B., Jančichová, K., Habánová, M., Górnicki, K., & Hamułka, J. (2022) · Applied Sciences · doi:10.3390/app12126165 · Cited by 26 · see citing works
Time Stamp – A Novel Time-to-Digital Demodulation Method for Bioimpedance Implant Applications
Wu, Y., Jiang, D., Habibollahi, M., Almarri, N., & Demosthenous, A. (2020) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2020.3012057 · Cited by 25 · see citing works
Choi, H., Kim, K., Kim, Y. S., & Yim, J. (2020) · Clinical Nutrition Research · doi:10.7762/cnr.2020.9.3.205 · Cited by 25 · see citing works
Bioelectrical Impedance Analysis for Evaluation of Body Composition: A Review
Aldobali, M., & Pal, K. (2021) · International Congress of Advanced Technology and Engineering (ICOTEN) · doi:10.1109/icoten52080.2021.9493494 · Cited by 25 · see citing works
Merrigan, J. J., Stute, N. L., Eckerle, J. J., Mackowski, N. S., Walters, J., O’Connor, M., Barrett, K. N., Robert, R., Strang, A. J., & Hagen, J. A. (2022) · Journal of Exercise and Nutrition · doi:10.53520/jen2022.103133 · Cited by 25 · see citing works
Analysis and classification of human skin diseases
Manoorkar, P. B., Kamat, D. K., & Patil, P. M. (2016) · International Conference on Automatic Control and Dynamic Optimization Techniques · doi:10.1109/icacdot.2016.7877750 · Cited by 25 · see citing works
Wearable cuffless blood pressure monitoring: From flexible electronics to machine learning
Hua, J., Su, M., Wu, J., Zhou, Y., Guo, Y., Shi, Y., & Pan, L. (2024) · Wearable electronics. · doi:10.1016/j.wees.2024.05.004 · Cited by 24 · see citing works
Krzesiński, P., Siebert, J., Jankowska, E. A., Banasiak, W., Piotrowicz, K., Stańczyk, A., Galas, A. A., Walczak, A., Murawski, P., Chrom, P., et al. (2021) · ESC Heart Failure · doi:10.1002/ehf2.13330 · Cited by 24 · see citing works
Nejadgholi, I., & Bolić, M. (2015) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2015.05.004 · Cited by 24 · see citing works
Dry Textile Electrodes for Wearable Bio-Impedance Analyzers
Corchia, L., Monti, G., Raheli, F., Candelieri, G., & Tarricone, L. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2972603 · Cited by 24 · see citing works
Calella, P., Valerio, G., Brodlie, M., Taylor, J., Donini, L. M., & Siervo, M. (2019) · Nutrition in Clinical Practice · doi:10.1002/ncp.10247 · Cited by 23 · see citing works
Electrical and Hormonal Biomarkers in Cachectic Elderly Women with Chronic Heart Failure
Sobieszek, G., Powrózek, T., Mazurek, M., Skwarek-Dziekanowska, A., & Malecka-Massalska, T. J. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9041021 · Cited by 23 · see citing works
- Authors
- Khalil, S., Mohktar, M., & Ibrahim, F.
- Affiliations
- University of Malaya (MY)
- Corresponding author
- Ibrahim, Fatimah
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2014-06-19
- Volume
- 14
- Issue
- 6
- Pages
- 10895-10928
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC4118362
- OpenAlex
- W2082211146
- Cited by
- 712
- References
- 149
- Citations by year
- 2026: 64 · 2025: 83 · 2024: 91 · 2023: 86 · 2022: 94 · 2021: 72 · 2020: 74 · 2019: 51 · 2018: 48 · 2017: 25 · 2016: 16 · 2015: 7 · 2014: 1
- FWCI
- 1.92
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Physiology › Body Composition Measurement Techniques
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrical and Bioimpedance Tomography
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- bioelectrical impedance analysis, body composition, bioimpedance vector analysis, prediction equations, anthropometry, lean mass loss, fluid shifts, heart disease, lung disease, kidney disease, neurological disorders, infectious diseases
- MeSH terms
- Body Composition, Diagnosis, Computer-Assisted, Equipment Design, Humans, Models, Biological, Monitoring, Physiologic, Plethysmography, Impedance, Biosensing Techniques, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review