14C Background Levels in An Accelerator Mass Spectrometry System

Open access
Vogel, J S, Nelson, D E, Southon, J R·1987·Radiocarbon

Abstract

The levels and sources of the measurement background in an AMS 14C dating system have been studied in detail. The relative contributions to the total background from combustion, graphitization, storage, handling, and from the accelerator were determined by measuring the C concentrations in samples of anthracite coal ranging in size from 15μg to 20mg. The results show that, for the present system, the uncertainty in the background is greater than that due to measurement precision alone for very old or for very small samples. While samples containing 100μg of carbon can yield useful 14C dates throughout the Holocene, 200 to 500μg are required for dating late Pleistocene materials. With the identification of the procedures that introduce contamination, the level and uncertainty of the total system background should both be reducible to the point that 100μg of carbon would be sufficient for dating most materials.

Overview

CoChat Summary

TL;DR

The paper quantifies sources of ¹⁴C measurement background in an AMS dating system using ¹⁴C-free anthracite coal samples from 15 μg to 20 mg. It finds combustion is the largest contamination source and that background uncertainty dominates for very small or very old samples, with 100 μg useful for Holocene dating and 200–500 μg needed for late Pleistocene materials.

Key findings

  • For anthracite samples >500 μg, total system background averaged 0.48 ± 0.16% modern (equivalent to 43 kyr ¹⁴C age); for samples <500 μg, contamination behaved as a constant amount of modern material, best fit as 2.2 ± 1.1 μg modern carbon equivalent.
  • Combustion contributed 1.5 ± 1.0 μg of the 2.2 ± 1.1 μg total background; graphitization contributed 0.36 ± 0.19 μg; storage and handling had upper limits of ca. 0.2 μg and <0.1 μg modern carbon equivalent, respectively.
  • Machine background from scattered ¹³C was equivalent to ca. 60–65 kyr (ca. 0.04% modern), and the authors assumed a machine background of ca. 55 kyr.
  • Water-vapor cleansing reduced large anthracite backgrounds from 0.59 ± 0.24% modern to 0.44 ± 0.13% modern; Pyrex combustion tubes gave a total background of 1.1 μg versus 2.2 μg modern carbon for Vycor.
  • 100 μg carbon yielded useful ¹⁴C dates throughout the Holocene; 200–500 μg were required for late Pleistocene materials, and 500 μg was sufficient through 40,000 years. A 200 μg sample of 35 kyr carbon would yield an age range of 32.5–39.5 kyr.

Methods

  • ¹⁴C-free anthracite coal was prepared and measured in the same manner as unknown samples, with carbon amounts ranging from 15 μg to 20 mg; combustion used evacuated Vycor tubes with copper oxide, followed by catalytic graphitization onto iron powder and AMS detection in continuous dual-beam mode.
  • Contamination contributions were determined by varying carbon amounts in individual preparation steps and measuring the resulting graphite ¹⁴C concentration, assuming most extraneous carbon was modern; machine background was estimated using spectroscopic graphite and graphite produced from old materials of known ages.
  • Measurements were made relative to identically prepared NBS oxalic acid standards, with typical ¹⁴C count rates of 50–75 per second; most anthracite measurements consisted of three 20-minute runs over 8–12 hours.

Limitations

  • The paper states it could not distinguish whether the machine-background ¹⁴C was inherent to the graphite, added during handling, or came from ion-source parts other than the sample, and no graphite known to be free of ¹⁴C was available.
  • In-situ contamination before sample selection is sample- and purification-dependent and was not discussed further.
  • The source of the coal contaminant was still under investigation, and the authors speculate that standard pretreatment procedures may not be adequate for coal used as a background material for AMS ¹⁴C determinations.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1017/s0033822200043733

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

u e r sess oaer o wu o sae er o e eo ozarr r oo suaeder oer n nr using different sized cold traps. Reaction volumes as small as 3ml were possible.
u e r sess oaer o wu o sae er o e eo ozarr r oo suaeder oer n nr using different sized cold traps. Reaction volumes as small as 3ml were possible.From the PDF
Fig 2 The total system background expressed as a 14C concentration vs anthracite sample weight from 10μg to 20mg. Samples above the 500μg in size showed a relatively constant concentration. Smaller samples had a concentration inversely proportional to their weight and were best fit by concentration (% mod) = 2.2 ± 1.1µg dívided by the sample size in micrograms.
Fig 2 The total system background expressed as a 14C concentration vs anthracite sample weight from 10μg to 20mg. Samples above the 500μg in size showed a relatively constant concentration. Smaller samples had a concentration inversely proportional to their weight and were best fit by concentration (% mod) = 2.2 ± 1.1µg dívided by the sample size in micrograms.From the PDF
Fig 3 Total system background vs the inverse of the graphitized sample size. The anthracite was combusted as large (>1mg) samples and the CÕ2 was divided into smaller samples before graphitization.
Fig 3 Total system background vs the inverse of the graphitized sample size. The anthracite was combusted as large (>1mg) samples and the CÕ2 was divided into smaller samples before graphitization.From the PDF
Fig 4 The total measurement uncertainty vs Holocene material age for different sample sizes with an assumed measurement precision of 2%. The background was assumed to have the form shown in Fig 2.
Fig 4 The total measurement uncertainty vs Holocene material age for different sample sizes with an assumed measurement precision of 2%. The background was assumed to have the form shown in Fig 2.From the PDF
Fig 5 The final measured age range for samples of various sizes assuming a concentration measurement precision of 3% with the background shown in Figure 2. The background uncertainty dominates throughout the late Pleistocene portion of the plot. For example, 200μg of 35 kyr carbon would yield an age range of 32.5 to 39.5 kyr.
Fig 5 The final measured age range for samples of various sizes assuming a concentration measurement precision of 3% with the background shown in Figure 2. The background uncertainty dominates throughout the late Pleistocene portion of the plot. For example, 200μg of 35 kyr carbon would yield an age range of 32.5 to 39.5 kyr.From the PDF

Verification

Claims the paper makes

1 Valid · 0 Not valid · 6 Not checked · 40 Own results and interpretations

31 more claims

Create a free account to see all 47 claims and every source verdict.

See all claims — free account
  • It may state two separate claims as one. 11
  • Probably not a substantial claim. 1

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

1 of 1

Citations

Cites

Show all 12 references

Research works printed but not linked: 1

  1. Eitel, 1966, Silicate science, 4

Cited by

See all 286

Metadata

Authors
Vogel, J. S., Nelson, D. E., & Southon, J. R.
Affiliations
Simon Fraser University (CA)
Venue
Radiocarbon
Published
1987
Volume
29
Issue
3
Pages
323-333
ISSN
  • 0033-8222 (ISSN-L)
  • 1945-5755
OpenAlex
W2505528999
Cited by
286
References
12
Citations by year
2026: 3 · 2025: 5 · 2024: 2 · 2023: 3 · 2022: 1 · 2021: 4 · 2020: 2 · 2019: 10 · 2018: 5 · 2017: 7 · 2016: 5 · 2015: 7 · 2014: 1 · 2013: 7 · 2012: 8
FWCI
12.34
Type
journal article
Language
en
Open access
Locations
  • Radiocarbon (publishedVersion)
  • UA Campus Repository (The University of Arizona) (acceptedVersion)
Topics
  • Social Sciences › Archeology › Archaeology and ancient environmental studies
  • Physical Sciences › Physics and Astronomy › Radiation › Radioactive Decay and Measurement Techniques
  • Physical Sciences › Earth and Planetary Sciences › Atmospheric Science › Geology and Paleoclimatology Research
Keywords
accelerator mass spectrometry, background contamination, sample preparation, graphitization, sample storage, sample handling, anthracite coal, sample size requirements, measurement uncertainty

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

A language model drew the claims from the paper on 2026-09-27, under version 3 of the claim definition (prompt extract-v3).

A second model labelled each claim on 2026-09-27 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: 14C Background Levels in An Accelerator Mass Spectrometry System. https://app.cochat.ai/papers/10.1017/s0033822200043733
The paper
Vogel, J. S., Nelson, D. E., & Southon, J. R. (1987). 14C Background Levels in An Accelerator Mass Spectrometry System. Radiocarbon. https://doi.org/10.1017/s0033822200043733

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.