Droplet-Based Microfluidic Flow Injection System with Large-Scale Concentration Gradient by a Single Nanoliter-Scale Injection for Enzyme Inhibition Assay
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
We described a microfluidic chip-based system capable of generating droplet array with a large scale concentration gradient by coupling flow injection gradient technique with droplet-based microfluidics. Multiple modules including sample injection, sample dispersion, gradient generation, droplet formation, mixing of sample and reagents, and online reaction within the droplets were integrated into the microchip. In the system, nanoliter-scale sample solution was automatically injected into the chip under valveless flow injection analysis mode. The sample zone was first dispersed in the microchannel to form a concentration gradient along the axial direction of the microchannel and then segmented into a linear array of droplets by immiscible oil phase. With the segmentation and protection of the oil phase, the concentration gradient profile of the sample was preserved in the droplet array with high fidelity. With a single injection of 16 nL of sample solution, an array of droplets with concentration gradient spanning 3-4 orders of magnitude could be generated. The present system was applied in the enzyme inhibition assay of β-galactosidase to preliminarily demonstrate its potential in high throughput drug screening. With a single injection of 16 nL of inhibitor solution, more than 240 in-droplet enzyme inhibition reactions with different inhibitor concentrations could be performed with an analysis time of 2.5 min. Compared with multiwell plate-based screening systems, the inhibitor consumption was reduced 1000-fold.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Millisecond Kinetics on a Microfluidic Chip Using Nanoliters of Reagents
Song, H., & Ismagilov, R. F. (2003) · Journal of the American Chemical Society · doi:10.1021/ja0354566 · Cited by 693
Cited by
Emerging Droplet Microfluidics
Shang, L., Cheng, Y., & Zhao, Y. (2017) · Chemical Reviews · doi:10.1021/acs.chemrev.6b00848 · Cited by 1,584
Analytical detection techniques for droplet microfluidics—A review
Zhu, Y., & Fang, Q. (2013) · Analytica Chimica Acta · doi:10.1016/j.aca.2013.04.064 · Cited by 365
Self-citationConcentration gradient generation methods based on microfluidic systems
Wang, X., Liu, Z., & Pang, Y. (2017) · RSC Advances · doi:10.1039/c7ra04494a · Cited by 219
A review of chemical gradient systems for cell analysis
Somaweera, H., Ibraguimov, A., & Pappas, D. (2016) · Analytica Chimica Acta · doi:10.1016/j.aca.2015.12.008 · Cited by 98
Sun, M., & Vanapalli, S. A. (2013) · Analytical Chemistry · doi:10.1021/ac303526y · Cited by 26
Metadata
- Authors
- Cai, L.-F., Zhu, Y., Du, G.-S., & Fang, Q.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2011-12-19
- Volume
- 84
- Issue
- 1
- Pages
- 446-452
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2321432786
- Cited by
- 108
- References
- 1
- FWCI
- 6.20
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›