Microfabricated Electrophoresis Chips for Simultaneous Bioassays of Glucose, Uric Acid, Ascorbic Acid, and Acetaminophen
Abstract
A micromachined capillary electrophoresis chip is described for simultaneous measurements of glucose, ascorbic acid, acetaminophen, and uric acid. Fluid control is used to mix the sample and enzyme glucose oxidase (GOx). The enzymatic reaction, a catalyzed aerobic oxidation of glucose to gluconic acid and hydrogen peroxide, occurs along the separation channel. The enzymatically liberated neutral peroxide species is separated electrophoretically from the anionic uric and ascorbic acids in the separation/reaction channel. The three oxidizable species are detected at the downstream gold-coated thick-film amperometric detector at different migration times. Glucose can be detected within less than 100 s, and detection of all electroactive constituents is carried out within 4 min. Measurements of glucose in the presence of acetaminophen, a neutral compound, are accomplished by comparing the responses in the presence and absence of GOx in the running buffer. The reproducibility of the on-chip glucose measurements is improved greatly by using uric acid as an internal standard. Factors influencing the performance, including the GOx concentration, field strength, and detection potential, are optimized. Such coupling of enzymatic assays with electrophoretic separations on a microchip platform holds great promise for rapid testing of metabolites (such as glucose or lactate), as well as for the introduction of high-speed clinical microanalyzers based on multichannel chips.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Microfluidic chips for clinical and forensic analysis
Verpoorte, E. (2002) · ELECTROPHORESIS · doi:10.1002/1522-2683(200203)23:5<677::aid-elps677>3.0.co;2-8 · Cited by 526
Biosensor Recognition Elements
(2008) · Current Issues in Molecular Biology · doi:10.21775/cimb.010.001 · Cited by 264
Gazdik, Z., Zitka, O., Petrlova, J., Adam, V., Zehnalek, J., Horna, A., Reznicek, V., Beklova, M., & Kizek, R. (2008) · Sensors · doi:10.3390/s8117097 · Cited by 170
Metadata
- Authors
- Wang, J., Chatrathi, M. P., Tian, B., & Polsky, R.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2000-04-20
- Volume
- 72
- Issue
- 11
- Pages
- 2514-2518
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2080070718
- Cited by
- 178
- References
- 0
- FWCI
- 16.17
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Microfabricated Electrophoresis Chips for Simultaneous Bioassays of Glucose, Uric Acid, Ascorbic Acid, and Acetaminophen. https://app.cochat.ai/papers/10.1021/ac991489l
- The paper
- Wang, J., Chatrathi, M. P., Tian, B., & Polsky, R. (2000). Microfabricated Electrophoresis Chips for Simultaneous Bioassays of Glucose, Uric Acid, Ascorbic Acid, and Acetaminophen. Analytical Chemistry. https://doi.org/10.1021/ac991489l
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.