Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrations
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
Background: While point‐of‐care testing (POCT) is being used increasingly as a basis for deciding on perioperative erythrocyte transfusion, no valid standards currently exist concerning the accuracy of Hb concentration measurements. For clinical employment, however, the confidence limits (±2 SD) of these measurements should lie close to 5 g/l. The aim of the present study was to evaluate the accuracy and precision of point‐of‐care testing for blood hemoglobin concentration (cHb in g/l) measurements in critically ill patients. Methods: Fifty blood samples from 50 postoperative patients requiring intensive care treatment were withdrawn from a cannula in the radial artery into a 2‐ml heparinized syringe (containing wet sodium heparinate in the conus), in a 2‐ml Monovette with 50 IE lithium heparinate, and into a 2.7‐ml cuvette with 1.6 mg potassium EDTA/ml blood. The POCT battery consisted of two blood gas analyzers (ABL TM 625 and 725, Radiometer, Copenhagen), the HemoCue® system (Mallinckrodt Medical, Germany), and an automated hematology analyzer (M‐2000®, Sysmex, Germany). The cyanmethemoglobin method served as the reference ‘gold standard’ procedure. The blood gas analyzer and HemoCue® systems were tested using dry and wet heparinized blood samples. Results: Hemoglobin concentrations of the reference measurements ranged from 73.9 to 159.4 g/l. The automated hematology analyzer method did reveal a small but systematic deviation for higher cHb values. For the blood gas analyzer and HemoCue® system procedures there was no systematic deviation of bias for either the first measurement or the averaged data. Bland & Altman analysis revealed a larger scattering for the wet heparinized samples. Conclusions: The above‐stated requirement for POCT systems, i.e. that the confidence limits should lie close to 5 g/l cHb, held true for the dry heparinized samples of the blood gas analyzer (1st measurement and mean of 2), the HemoCue® system (mean of 3) and the automated hematology analyzer.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A Multicenter, Randomized, Controlled Clinical Trial of Transfusion Requirements in Critical Care
Hébert, P. C., Wells, G., Blajchman, M. A., Marshall, J., Martin, C., Pagliarello, G., Tweeddale, M., Schweitzer, I., & Yetisir, E. (1999) · New England Journal of Medicine · doi:10.1056/nejm199902113400601 · Cited by 5,187
Cited by
Detection of anaemia from retinal fundus images via deep learning
Mitani, A., Huang, A., Venugopalan, S., Corrado, G. S., Peng, L., Webster, D. R., Hammel, N., Liu, Y., & Varadarajan, A. V. (2019) · Nature Biomedical Engineering · doi:10.1038/s41551-019-0487-z · Cited by 230
Point-of-care testing: where is the evidence? A systematic survey
Pecoraro, V., Germagnoli, L., & Banfi, G. (2014) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-0386 · Cited by 56
Interpreting arterial blood gas results
Cowley, N. J., Owen, A., & Bion, J. F. (2013) · BMJ · doi:10.1136/bmj.f16 · Cited by 46
Metadata
- Authors
- Gehring, H., Hornberger, C., Dibbelt, L., Roth‐Isigkeit, A., Gerlach, K., Schumacher, J., & Schmucker, P.
- Venue
- Acta Anaesthesiologica Scandinavica
- Publisher
- Wiley
- Published
- 2002-08-21
- Volume
- 46
- Issue
- 8
- Pages
- 980-986
- ISSN
- 0001-5172 (ISSN-L)
- 1399-6576
- OpenAlex
- W2051615659
- Cited by
- 98
- References
- 2
- FWCI
- 0.55
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.