In vitro optoacoustic flow cytometry with light scattering referencing

Open access
Seeger, Markus, Stiel, Andre C., Ntziachristos, Vasilis · 2021 · Scientific Reports

The DOI matches its Crossref record; 26 of the 26 references it cites were found in a registry or an index.

Abstract

Morphological and functional optoacoustic imaging is enhanced by dedicated transgene reporters, in analogy to fluorescence methods. The development of optoacoustic reporters using protein engineering and directed evolution would be accelerated by high-throughput in-flow screening for intracellular, genetically encoded, optoacoustic contrast. However, accurate characterization of such contrast is impeded because the optoacoustic signals depend on the cell’s size and position in the flow chamber. We report herein an optoacoustic flow cytometer (OA-FCM) capable of precise measurement of intracellular optoacoustic signals of genetically-encoded chromoproteins in flow. The novel system records light-scattering as a reference for the detected optoacoustic signals in order to account for cell size and position, as well as excitation light flux in the focal volume, which we use to reference the detected optoacoustic signals to enhance the system’s precision. The OA-FCM was calibrated using micrometer-sized particles to showcase the ability to assess in-flow objects in the size range of single-cells. We demonstrate the capabilities of our OA-FCM to identify sub-populations in a mixture of two E. coli stocks expressing different reporter-proteins with a precision of over 90%. High-throughput screening of optoacoustic labels could pave the way for identifying genetically encoded optoacoustic reporters by transferring working concepts of the fluorescence field such as directed evolution and activated cell sorting.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1038/s41598-021-81584-y

Cites

26 of 26 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 Schematic depiction and validation of the OA-FCM system. (a) The OA-FCM system is based on focusing a 532 nm laser into a microfluidic channel, which is connected to a syringe pump for controlling the flow of suspensions of particles or cells. OA and LS signals are recording simultaneously. (b) Close-up depiction of the microfluidic interrogation area. Raw (c) OA and (d) LS signal detected from a single bead indicated by an arrow flowing through the interrogation area flowing at 100 μL/min. (e) OA and (f) LS maximum amplitude projection of the record window shown in (c) and (d). (g) OA- and LS-trajectories recorded over 5 s show transits of individual beads. (h) Auto-correlation curves of the OA and LS trajectories in (e) reveal interrogation time of ~ 300 μs. Abbreviations: AE: active element; AcL: acoustic lens; AL: achromatic doublet lens; AMP: low noise amplifier; AWG: arbitrary waveform function generator; DAQ: data acquisition card; IM: inverted microscope; L: planoconvex lens; M: dielectric mirror; ND: neutral density filter; OL: microscope objective lens; PH: pinhole; S: high-precision motorized stage; SP: syringe pump; UT: Ultrasound transducer; W: Waste.
Figure 1 Schematic depiction and validation of the OA-FCM system. (a) The OA-FCM system is based on focusing a 532 nm laser into a microfluidic channel, which is connected to a syringe pump for controlling the flow of suspensions of particles or cells. OA and LS signals are recording simultaneously. (b) Close-up depiction of the microfluidic interrogation area. Raw (c) OA and (d) LS signal detected from a single bead indicated by an arrow flowing through the interrogation area flowing at 100 μL/min. (e) OA and (f) LS maximum amplitude projection of the record window shown in (c) and (d). (g) OA- and LS-trajectories recorded over 5 s show transits of individual beads. (h) Auto-correlation curves of the OA and LS trajectories in (e) reveal interrogation time of ~ 300 μs. Abbreviations: AE: active element; AcL: acoustic lens; AL: achromatic doublet lens; AMP: low noise amplifier; AWG: arbitrary waveform function generator; DAQ: data acquisition card; IM: inverted microscope; L: planoconvex lens; M: dielectric mirror; ND: neutral density filter; OL: microscope objective lens; PH: pinhole; S: high-precision motorized stage; SP: syringe pump; UT: Ultrasound transducer; W: Waste.CC BY · PubMed Central
Figure 2 Characterization of in-flow signals at 100 μL/min. Normalized OA and LS trajectories of (a) water as the solvent for μB and CNPs as well as of (b) PBS as the buffer for the cell suspension exhibit no detected transit event over 10 s record time. Water-suspended (c) μB and (d) reveal OA and LS signals of transits of individual particles. Unnormalized histograms are shown Gaussian-shaped distribution over the entire 1 min measurement. Normalized OA-FCM trajectories of E. coli expressing (e) mKO or (f) mCherry in PBS reveal detectable in-flow signals following a Gaussian-shaped distribution in the associated histograms.
Figure 2 Characterization of in-flow signals at 100 μL/min. Normalized OA and LS trajectories of (a) water as the solvent for μB and CNPs as well as of (b) PBS as the buffer for the cell suspension exhibit no detected transit event over 10 s record time. Water-suspended (c) μB and (d) reveal OA and LS signals of transits of individual particles. Unnormalized histograms are shown Gaussian-shaped distribution over the entire 1 min measurement. Normalized OA-FCM trajectories of E. coli expressing (e) mKO or (f) mCherry in PBS reveal detectable in-flow signals following a Gaussian-shaped distribution in the associated histograms.CC BY · PubMed Central
Figure 3 Signal distribution of in-flow single transit events of E. coli cells expressing optoacoustic labels enabled separation by applying linear discrimination. (a) Norm. OA signals against norm. LS signals of µB and (b) CNP, and (c) E. coli cells expressing mKO. Inset depicts separation by linear discrimination of 96.0%. (d) Analogues analysis of E. coli cells expressing mCherry yield a discrimination of 92.6%. (e) Data merge of the separate measurements and (f) measurement of cell stock mixture (mixing ratio 1:1) of mKO and mCherry E. coli cells yield a linear discrimination of ~ 2:1 resembling the ~ 3:1 concentration difference in stock solutions. (g) Varying the slope of the linear discrimination and Gaussian fitting peaked in ~ 0.0209 a.u. with an average separation of ~ 92.5%. (h) Flow speed analysis shows consistent linear discrimination by constant true-positive (TP) and false-negative (FN) assignments of signals. (i) ROC analysis reveals an AUC of ~ 0.8.
Figure 3 Signal distribution of in-flow single transit events of E. coli cells expressing optoacoustic labels enabled separation by applying linear discrimination. (a) Norm. OA signals against norm. LS signals of µB and (b) CNP, and (c) E. coli cells expressing mKO. Inset depicts separation by linear discrimination of 96.0%. (d) Analogues analysis of E. coli cells expressing mCherry yield a discrimination of 92.6%. (e) Data merge of the separate measurements and (f) measurement of cell stock mixture (mixing ratio 1:1) of mKO and mCherry E. coli cells yield a linear discrimination of ~ 2:1 resembling the ~ 3:1 concentration difference in stock solutions. (g) Varying the slope of the linear discrimination and Gaussian fitting peaked in ~ 0.0209 a.u. with an average separation of ~ 92.5%. (h) Flow speed analysis shows consistent linear discrimination by constant true-positive (TP) and false-negative (FN) assignments of signals. (i) ROC analysis reveals an AUC of ~ 0.8.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 26 references

Cited by

See all 8

Metadata

Authors
Seeger, M., Stiel, A. C., & Ntziachristos, V.
Affiliations
Helmholtz Munich (DE)
Corresponding author
Ntziachristos, Vasilis
Venue
Scientific Reports
Published
2021-01-26
Volume
11
Issue
1
Pages
2181-2181
ISSN
2045-2322 (ISSN-L)
OpenAlex
W3124549009
Cited by
8
References
26
Citations by year
2025: 2 · 2023: 1 · 2022: 3 · 2021: 2
FWCI
0.45
Type
journal article
Language
en
Open access
Licence
Locations
  • Scientific Reports (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • mediaTUM – the media and publications repository of the Technical University Munich (Technical University Munich) (submittedVersion, cc-by)
  • Site cant be reached (submittedVersion, cc-by)
Funders
  • Deutsche Forschungsgemeinschaft (NT 3/10-1, STI656/1-1)
  • Projekt DEAL
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Photoacoustic and Ultrasonic Imaging
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Advanced Fluorescence Microscopy Techniques
  • Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Bio-sensing Technologies
Keywords
photoacoustic imaging, chromoproteins, high-throughput screening, Escherichia coli, subpopulation identification, directed evolution
MeSH terms
Escherichia coli, Flow Cytometry, Scattering, Radiation, Signal Processing, Computer-Assisted, Reproducibility of Results, Photoacoustic Techniques

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: In vitro optoacoustic flow cytometry with light scattering referencing. https://app.cochat.ai/papers/10.1038/s41598-021-81584-y
The paper
Seeger, M., Stiel, A. C., & Ntziachristos, V. (2021). In vitro optoacoustic flow cytometry with light scattering referencing. Scientific Reports. https://doi.org/10.1038/s41598-021-81584-y

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.