The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Cell lysis and DNA extraction of gram-positive and gram-negative bacteria from whole blood in a disposable microfluidic chip
Abstract
Sepsis caused by gram positive and gram negative bacteria is the leading cause of death in noncoronary ICUs and the tenth leading cause of death in the United States. We have developed a microfluidic sample preparation platform for rapid on-chip detection of infectious organisms for point-of-care diagnostics. The microfluidic chips are made of a robust thermoplastic and can be easily multiplexed for high throughput applications. Bacteria are lysed on-chip via hybrid chemical/mechanical method. Once lysed, the bacterial DNA is isolated using a microscale silica bead/polymer composite solid-phase-extraction (SPE) column. Lysis was confirmed using off-chip real time PCR. We isolated and detected both gram-negative (Escherichia coli) and gram-positive (Bacillussubtilis and Enterococcus faecalis) bacterial genomic DNA from microliter scale spiked whole human blood samples. The system performs better for gram-negative bacteria than it does for gram-positive bacteria, with limits of detection at 10(2) CFU/ml and 10(3)-10(4) CFU/ml, respectively. Total extraction times are less than one hour and can be further decreased by altering the channel geometry and pumping configuration.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Epidemiology of sepsis: an update
Angus, D. C., & Wax, R. S. (2001) · Critical Care Medicine · doi:10.1097/00003246-200107001-00035 · Cited by 813
Vince, A., Lepej, S. Ž., Baršić, B., Dušek, D., Mitrović, Z., Serventi-Seiwerth, R., & Labar, B. (2008) · Journal of Medical Microbiology · doi:10.1099/jmm.0.47797-0 · Cited by 59
Cited by
Point of Care Diagnostics: Status and Future
Gubala, V., Harris, L. F., Ricco, A. J., Tan, M. X., & Williams, D. E. (2012) · Analytical Chemistry · doi:10.1021/ac2030199 · Cited by 1,194
A Review on Macroscale and Microscale Cell Lysis Methods
Shehadul Islam, M., Aryasomayajula, A., & Selvaganapathy, P. (2017) · Micromachines · doi:10.3390/mi8030083 · Cited by 541
Advances and challenges in biosensor-based diagnosis of infectious diseases
Sin, M. L., Mach, K. E., Wong, P. K., & Liao, J. C. (2014) · Expert Review of Molecular Diagnostics · doi:10.1586/14737159.2014.888313 · Cited by 435
Microfluidics for medical diagnostics and biosensors
Rivet, C., Lee, H., Hirsch, A., Hamilton, S., & Lu, H. (2011) · Chemical Engineering Science · doi:10.1016/j.ces.2010.08.015 · Cited by 356
Advances in microfluidic PCR for point-of-care infectious disease diagnostics
Park, S., Zhang, Y., Lin, S., Wang, T.-H., & Yang, S. (2011) · Biotechnology Advances · doi:10.1016/j.biotechadv.2011.06.017 · Cited by 299
Metadata
- Authors
- Mahalanabis, M., Al-Muayad, H., Kulinski, M. D., Altman, D., & Klapperich, C. M.
- Venue
- Lab on a Chip
- Publisher
- Royal Society of Chemistry (RSC)
- Published
- 2009
- Volume
- 9
- Issue
- 19
- Pages
- 2811
- ISSN
- 1473-0189 (ISSN-L)
- 1473-0197
- OpenAlex
- W2147268130
- Cited by
- 178
- References
- 2
- FWCI
- 6.60
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Cell lysis and DNA extraction of gram-positive and gram-negative bacteria from whole blood in a disposable microfluidic chip. https://app.cochat.ai/papers/10.1039/b905065p
- The paper
- Mahalanabis, M., Al-Muayad, H., Kulinski, M. D., Altman, D., & Klapperich, C. M. (2009). Cell lysis and DNA extraction of gram-positive and gram-negative bacteria from whole blood in a disposable microfluidic chip. Lab on a Chip. https://doi.org/10.1039/b905065p
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.