CoChat

Evaluation of the Dinamap 8100 and Omron M1 blood pressure monitors for use in children

Barker, Mary E., Shiell, Alistair W., Law, Catherine M. · 2000 · Paediatric and Perinatal Epidemiology
Cited by 44

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

The aim of this study was to evaluate the Dinamap 8100 and the Omron M1 (the test devices) against standard criteria for use in children in a fieldwork setting. Device calibration and validation were carried out in accordance with the British Hypertension Society protocol for special groups such as children. A total of 102 children, recruited from infant and junior schools in Southampton, had six sequential measurements made of their blood pressure–four measurements with a mercury sphygmomanometer and two with one of two test devices, 55 children with the Dinamap 8100 and 47 with the Omron M1. Systolic and diastolic readings with the Dinamap 8100 were on average 11 mmHg higher (95% confidence interval [CI]+9, +12 mmHg) and 3 mmHg lower (95% CI −5, −1 mmHg), respectively, than measurements with the mercury sphygmomanometer, overestimating systolic pressures and underestimating diastolic pressures across the whole range observed. The Omron M1 gave readings lower by 1 mmHg on average for systolic pressures and 2 mmHg for diastolic pressures compared with the sphygmomanometer (95% CIs −4, +1 mmHg and −5, +1 mmHg respectively), specifically overestimating higher pressures and underestimating lower pressures. According to the criteria of the British Hypertension Society, neither the Dinamap 8100 nor the Omron M1 can be recommended for use in children in clinical situations in which accuracy of the absolute measurement is required. In epidemiological surveys, in which differences in blood pressure between groups of people are more important than absolute levels, it may be more appropriate to use these devices. Of the two, its more consistent performance supports the Dinamap 8100 as the instrument of choice in such studies of children.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1046/j.1365-3016.2000.00250.x

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 44

Metadata

Authors
Barker, M. E., Shiell, A. W., & Law, C. M.
Venue
Paediatric and Perinatal Epidemiology
Publisher
Wiley
Published
2000-04
Volume
14
Issue
2
Pages
179-186
ISSN
  • 0269-5022 (ISSN-L)
  • 1365-3016
OpenAlex
W2067434382
Cited by
44
References
1
FWCI
1.65
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.