Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children
Abstract
AIM: To validate Welch-Allyn Vital Sign Monitor, Dinamap Procare-120 and Datascope Accutorr Plus against auscultatory mercury sphygmomanometer in children aged 5-15 years old according to the International Protocol of European Society of Hypertension adapted for validation in children. METHOD: One hundred and thirty two children were studied (44 for each device; 67 boys, 65 girls). Each underwent seven sequential BP measurements on the right arm resting in the sitting position, alternately with the mercury sphygmomanometer read simultaneously by two independent, trained observers and the test device by a third observer. RESULTS: For the Welch-Allyn monitor, the mean+/-SD of differences (device minus auscultatory BP) were -4.39+/-4.82 mmHg for systolic blood pressure and -4.1+/-7.07 mmHg for diastolic blood pressure. The device failed phase 2.1 for both systolic blood pressure (55, 91 and 98% were within 5, 10 and 15 mmHg, respectively) and diastolic blood pressure measurements (46, 82 and 95% were within 5, 10 and 15 mmHg, respectively). For the Dinamap device, the mean+/-SD of differences were -3.08+/-5.21 mmHg for systolic blood pressure and -4.61+/-9.35 mmHg for diastolic blood pressure. The device passed phase 2.1 for systolic blood pressure (71, 96 and 98% were within 5, 10 and 15 mmHg, respectively) but failed for diastolic blood pressure (51, 72 and 91% were within 5, 10, and 15 mmHg, respectively). For the Datascope device, mean+/-SD of differences were -0.9+/-4.33 mmHg for systolic blood pressure and -1.20+/-6.48 mmHg for diastolic blood pressure. The device passed phase 2.1 in that 84, 97 and 99% of systolic blood pressure, and 61, 89 and 97% of diastolic blood pressure readings were within 5, 10 and 15 mmHg, respectively. It also passed phase 2.2 for both systolic blood pressure and diastolic blood pressure. CONCLUSION: We performed an independent validation of three oscillometric BP devices in children. Overall Datascope Accutorr Plus passed, whereas Welch-Allyn Vital Sign Monitor and Dinamap Procare-120 failed an adapted IP-ESH.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Jaddoe, V. W. V., de Jonge, L. L., Hofman, A., Franco, O. H., Steegers, E. A. P., & Gaillard, R. (2014) · BMJ · doi:10.1136/bmj.g14 · Cited by 442
Blood Pressure Percentiles by Age and Height From Nonoverweight Children and Adolescents in Germany
Neuhauser, H. K., Thamm, M., Ellert, U., Hense, H. W., & Rosario, A. S. (2011) · Pediatrics · doi:10.1542/peds.2010-1290 · Cited by 281
Childhood Cardiometabolic Outcomes of Maternal Obesity During Pregnancy: The Generation R Study
Gaillard, R., Steegers, E. A. P., Duijts, L., Felix, J. F., Hofman, A., Franco, O. H., & Jaddoe, V. W. V. (2014) · Hypertension · doi:10.1161/hypertensionaha.113.02671 · Cited by 269
Accuracy of Automated Blood Pressure Measurement in Children: Evidence, Issues, and Perspectives
Stergiou, G. S., Boubouchairopoulou, N., & Kollias, A. (2017) · Hypertension · doi:10.1161/hypertensionaha.116.08553 · Cited by 72
Metadata
- Authors
- Wong, S.-N., Tz Sung, R. Y., & Leung, L. C.-K.
- Venue
- Blood Pressure Monitoring
- Published
- 2006-10
- Volume
- 11
- Issue
- 5
- Pages
- 281-291
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W1979553170
- Cited by
- 166
- References
- 1
- FWCI
- 2.84
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children. https://app.cochat.ai/papers/10.1097/01.mbp.0000209082.09623.b4
- The paper
- Wong, S.-N., Tz Sung, R. Y., & Leung, L. C.-K. (2006). Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children. Blood Pressure Monitoring. https://doi.org/10.1097/01.mbp.0000209082.09623.b4
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.