A Comparison of the Lactate Pro, Accusport, Analox GM7 and Kodak Ektachem Lactate Analysers in Normal, Hot and Humid Conditions
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
This study aimed to compare the performance of a new portable lactate analyser against other standard laboratory methods in three conditions, normal (20 +/- 1.3 degrees C; 40 +/- 5 % RH), hot (40 +/- 2.5 degrees C; 40 +/- 5 % RH), and humid (20 +/- 1.1 degrees C; 82 +/- 6 % RH) conditions. Seven healthy males, ([Mean +/- SE]: age, 26.3 +/- 1.3 yr; height, 177.7 +/- 1.6 cm; weight, 77.4 +/- 0.9 kg, .VO(2)max, 56.1 +/- 1.9 ml x kg x min(-1)) undertook a maximal cycle ergometry test to exhaustion in the three conditions. Blood was taken every 3 min at the end of each stage and was analysed using the Lactate Pro LT-1710, the Accusport, the Analox GM7 and the Kodak Ektachem systems. The MANOVA (Analyser Type x Condition x Workload) indicated no interaction effect (F(42,660), = 0.45, p > 0.99, Power = 0.53). The data across all workloads indicated that the machines measured significantly differently to each other (F(4,743) = 14.652, p < 0.0001, Power = 1.00). The data were moderately to highly correlated. We conclude that the Lactate Pro is a simple and effective measurement device for taking blood lactate in a field or laboratory setting. However, we would caution against using this machine to compare data from other machines.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Measures of Reliability in Sports Medicine and Science
Hopkins, W. G. (2000) · Sports Medicine · doi:10.2165/00007256-200030010-00001 · Cited by 4,635
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Integration of the Physiological Factors Determining Endurance Performance Ability
COYLE, E. F. (1995) · Exercise and Sport Sciences Reviews · doi:10.1249/00003677-199500230-00004 · Cited by 424
Cited by
Lactate Threshold Concepts: How Valid are They?
Faude, O., Kindermann, W., & Meyer, T. (2009) · Sports Medicine · doi:10.2165/00007256-200939060-00003 · Cited by 882
Blood Lactate Measurements and Analysis during Exercise: A Guide for Clinicians
Goodwin, M. L., Harris, J. E., Hernández, A., & Gladden, L. B. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100414 · Cited by 554
Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus
Tanner, R. K., Fuller, K. L., & Ross, M. L. R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1379-9 · Cited by 274
Matthew, D., & Delextrat, A. (2009) · Journal of Sports Sciences · doi:10.1080/02640410902926420 · Cited by 244
“986; McNaughton, Thompson, Philips, Backx, & Crickmore, 2002 ).”
Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer
Baldari, C., Bonavolontà, V., Emerenziani, G. P., Gallotta, M. C., Silva, A. J., & Guidetti, L. (2009) · European Journal of Applied Physiology · doi:10.1007/s00421-009-1107-5 · Cited by 171
“Other works, instead, have compared AL and LP to other Wxed analyzers without taking into consideration the three validation parameters we have assessed (i.e. accuracy, reliability and linearity) at the same time (Mc Naughton et al. 2002; Nordström et al. 1998).”
Background
Metadata
- Authors
- Mc Naughton, L. R., Thompson, D., Philips, G., Backx, K., & Crickmore, L.
- Venue
- International Journal of Sports Medicine
- Publisher
- Georg Thieme Verlag KG
- Published
- 2002-02
- Volume
- 23
- Issue
- 2
- Pages
- 130-135
- ISSN
- 0172-4622 (ISSN-L)
- 1439-3964
- OpenAlex
- W2079880346
- Cited by
- 86
- References
- 4
- FWCI
- 3.42
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.