Interference with Cardiac Pacemakers by Cellular Telephones
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
BACKGROUND: A growing body of evidence suggests that electromagnetic interference may occur between cardiac pacemakers and wireless hand-held (cellular) telephones, posing a potential public health problem. Electromagnetic interference may occur when the pacemaker is exposed to an electromagnetic field generated by the cellular telephone. METHODS: In this multicenter, prospective, crossover study, we tested 980 patients with cardiac pacemakers with five types of telephones (one analogue and four digital) to assess the potential for interference. Telephones were tested in a test mode and were programmed to transmit at the maximal power, simulating the worst-case scenario; in addition, one telephone was tested during actual transmission to simulate actual use. Patients were electrocardiographically monitored while the telephones were tested at the ipsilateral ear and in a series of maneuvers directly over the pacemaker. Interference was classified according to the type and clinical significance of the effect. RESULTS: The incidence of any type of interference was 20 percent in the 5533 tests, and the incidence of symptoms was 7.2 percent. The incidence of clinically significant interference was 6.6 percent. There was no clinically significant interference when the telephone was placed in the normal position over the ear. Interference that was definitely clinically significant occurred in only 1.7 percent of tests, and only when the telephone was held over the pacemaker. Interference was more frequent with dual-chamber pacemakers (25.3 percent) than with single-chamber pacemakers (6.8 percent, P<0.001) and more frequent with pacemakers without feed-through filters (28.9 to 55.8 percent) than with those with such filters (0.4 to 0.8 percent, P=0.01). CONCLUSIONS: Cellular telephones can interfere with the function of implanted cardiac pacemakers. However, when telephones are placed over the ear, the normal position, this interference does not pose a health risk.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Multicenter Experience With a Bipolar Tined Polyurethane Ventricular Lead
Hayes, D. L., Graham, K. J., IRWIN, M. E., Vidaillet, H. J., Jr., DISLER, G., SWEESY, M. W., KINCAID, D. K., Osborn, M. J., Suman, V. J., Neubauer, S. A., et al. (1995) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1995.tb04740.x · Cited by 45
Self-citation
Cited by
Authors/Task Force Members, Vardas, P. E., Auricchio, A., Blanc, J.-J., Daubert, J.-C., Drexler, H., Ector, H., Gasparini, M., Linde, C., Morgado, F. B., et al. (2007) · European Heart Journal · doi:10.1093/eurheartj/ehm305 · Cited by 902
Crossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., Hoyt, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.12.023 · Cited by 340
Crossley, G. H., Poole, J. E., Rozner, M. A., Asirvatham, S. J., Cheng, A., Chung, M. K., Ferguson, T. B., Gallagher, J. D., Gold, M. R., Hoyt, R. H., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.05.010 · Cited by 300
Authors/Task Force Members, Vardas, P. E., Auricchio, A., Blanc, J.-J., Daubert, J.-C., Drexler, H., Ector, H., Gasparini, M., Linde, C., Morgado, F. B., et al. (2007) · Europace · doi:10.1093/europace/eum189 · Cited by 291
WALNUT: Waging Doubt on the Integrity of MEMS Accelerometers with Acoustic Injection Attacks
Trippel, T., Weisse, O., Xu, W., Honeyman, P., & Fu, K. (2017) · 2017 IEEE European Symposium on Security and Privacy (EuroS&P) · doi:10.1109/eurosp.2017.42 · Cited by 279
Metadata
- Authors
- Hayes, D. L., Wang, P. J., Reynolds, D. W., Estes, N. A. M., Griffith, J. L., Steffens, R. A., Carlo, G. L., Findlay, G. K., & Johnson, C. M.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 1997-05-22
- Volume
- 336
- Issue
- 21
- Pages
- 1473-1479
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2317624582
- Cited by
- 282
- References
- 1
- FWCI
- 8.14
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.