BabySQUID: A mobile, high-resolution multichannel magnetoencephalography system for neonatal brain assessment
The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
We developed a prototype of a mobile, high-resolution, multichannel magnetoencephalography (MEG) system, called babySQUID, for assessing brain functions in newborns and infants. Unlike electroencephalography, MEG signals are not distorted by the scalp or the fontanels and sutures in the skull. Thus, brain activity can be measured and localized with MEG as if the sensors were above an exposed brain. The babySQUID is housed in a moveable cart small enough to be transported from one room to another. To assess brain functions, one places the baby on the bed of the cart and the head on its headrest with MEG sensors just below. The sensor array consists of 76 first-order axial gradiometers, each with a pickup coil diameter of 6mm and a baseline of 30mm, in a high-density array with a spacing of 12–14mm center-to-center. The pickup coils are 6±1mm below the outer surface of the headrest. The short gap provides unprecedented sensitivity since the scalp and skull are thin (as little as 3–4mm altogether) in babies. In an electromagnetically unshielded room in a hospital, the field sensitivity at 1kHz was ∼17fT∕√Hz. The noise was reduced from ∼400to200fT∕√Hz at 1Hz using a reference cancellation technique and further to ∼40fT∕√Hz using a gradient common mode rejection technique. Although the residual environmental magnetic noise interfered with the operation of the babySQUID, the instrument functioned sufficiently well to detect spontaneous brain signals from babies with a signal to noise ratio (SNR) of as much as 7.6:1. In a magnetically shielded room, the field sensitivity was 17fT∕√Hz at 20Hz and 30fT∕√Hz at 1Hz without implementation of reference or gradient cancellation. The sensitivity was sufficiently high to detect spontaneous brain activity from a 7month old baby with a SNR as much as 40:1 and evoked somatosensory responses with a 50Hz bandwidth after as little as four averages. We expect that both the noise and the sensor gap can be reduced further by approximately half with a gain in SNR of about four. Thus, we conclude from the performance of the prototype that it should be feasible to improve the babySQUID to detect cortical activity in infants in real time with high spatial resolution.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Even systems suitable for foetal MEG are used in clinical research [16,26].”
From Biomagnetism using SQUIDs: status and perspectives (2006) · cites this paper as [26]
Citations
Cites
Formulas for Stress and Strain
Roark, R. J., Young, W. C., & Plunkett, R. (1976) · Journal of Applied Mechanics · doi:10.1115/1.3423917 · Cited by 1,927
Positron emission tomography study of human brain functional development
Chugani, H. T., Phelps, M. E., & Mazziotta, J. C. (1987) · Annals of Neurology · doi:10.1002/ana.410220408 · Cited by 1,548
Signal-space projection method for separating MEG or EEG into components
Uusitalo, M. A., & Ilmoniemi, R. J. (1997) · Medical & Biological Engineering & Computing · doi:10.1007/bf02534144 · Cited by 913
Magnetic Sensors and Magnetometers
(ed), P. R. (2002) · Measurement Science and Technology · doi:10.1088/0957-0233/13/4/707 · Cited by 663
A Critical Period of Brain Development: Studies of Cerebral Glucose Utilization with PET
Chugani, H. T. (1998) · Preventive Medicine · doi:10.1006/pmed.1998.0274 · Cited by 589
Suppression of Interference and Artifacts by the Signal Space Separation Method
Taulu, S., Kajola, M., & Simola, J. (2004) · Brain Topography · doi:10.1023/b:brat.0000032864.93890.f9 · Cited by 541
Diffusion tensor imaging of normal and injured developing human brain-a technical review
Neil, J., Miller, J., Mukherjee, P., & Hüppi, P. S. (2002) · NMR in Biomedicine · doi:10.1002/nbm.784 · Cited by 402
Validation of SOBI components from high-density EEG
Tang, A. C., Sutherland, M. T., & McKinney, C. J. (2005) · NeuroImage · doi:10.1016/j.neuroimage.2004.11.027 · Cited by 173
Atlas of Roentgenographic Measurement
(1959) · JAMA · doi:10.1001/jama.1959.03010270108031 · Cited by 144
Magnetic localization of a dipolar current source implanted in a sphere and a human cranium
Barth, D. S., Sutherling, W. W., Broffman, J., & Beatty, J. (1986) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(86)90094-5 · Cited by 134
Research works printed but not linked: 5
- 2000, Neuorology of the Newborn
- 2005, Clin. Neurophysiol., 116, 1142
- 1999, Electroencephalogr. Clin. Neurophysiol., 110, 230
- 1997, J. Japanese Biomag. Bioelectromag. Soc., 10, 180
- 2004, Brain Topogr, 16, 269
Cited by
Superconducting quantum interference device instruments and applications
Fagaly, R. L. (2006) · Review of Scientific Instruments · doi:10.1063/1.2354545 · Cited by 594
Portable Magnetometry for Detection of Biomagnetism in Ambient Environments
Limes, M. E., Foley, E. L., Kornack, T. W., Caliga, S., McBride, S., Braun, A., Lee, W., Lucivero, V. G., & Romalis, M. V. (2020) · Physical Review Applied · doi:10.1103/physrevapplied.14.011002 · Cited by 257
Denoising based on time-shift PCA
de Cheveigné, A., & Simon, J. Z. (2007) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2007.06.003 · Cited by 214
Biomagnetism using SQUIDs: status and perspectives
Sternickel, K., & Braginski, A. I. (2006) · Superconductor Science and Technology · doi:10.1088/0953-2048/19/3/024 · Cited by 180
Exploring early human brain development with structural and physiological neuroimaging
Vasung, L., Abaci Turk, E., Ferradal, S. L., Sutin, J., Stout, J. N., Ahtam, B., Lin, P.-Y., & Grant, P. E. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2018.07.041 · Cited by 177
Metadata
- Authors
- Okada, Y., Pratt, K., Atwood, C., Mascarenas, A., Reineman, R., Nurminen, J., & Paulson, D.
- Affiliations
- University of New Mexico (US)
- Tristan Technologies (United States) (US)
- Helsinki University Hospital (FI)
- Venue
- Review of Scientific Instruments
- Publisher
- AIP Publishing
- Published
- 2006-02-01
- Volume
- 77
- Issue
- 2
- ISSN
- 0034-6748 (ISSN-L)
- 1089-7623
- OpenAlex
- W2084082143
- Cited by
- 53
- References
- 14
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 2 · 2023: 1 · 2022: 1 · 2021: 2 · 2020: 3 · 2019: 6 · 2018: 3 · 2017: 5 · 2016: 4 · 2015: 5 · 2014: 6 · 2013: 2
- FWCI
- 1.29
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Review of Scientific Instruments (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › EEG and Brain-Computer Interfaces
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Atomic and Subatomic Physics Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Magnetic Field Sensors Techniques
- Keywords
- magnetoencephalography, infant brain activity, axial gradiometer, high-density sensor arrays, magnetic noise, field sensitivity, somatosensory evoked potentials, spontaneous brain activity, magnetic shielding, SQUID magnetometer, infant neuroimaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: BabySQUID: A mobile, high-resolution multichannel magnetoencephalography system for neonatal brain assessment. https://app.cochat.ai/papers/10.1063/1.2168672
- The paper
- Okada, Y., Pratt, K., Atwood, C., Mascarenas, A., Reineman, R., Nurminen, J., & Paulson, D. (2006). BabySQUID: A mobile, high-resolution multichannel magnetoencephalography system for neonatal brain assessment. Review of Scientific Instruments. https://doi.org/10.1063/1.2168672
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.