The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Thermal characterization and analysis of microliter liquid volumes using the three-omega method
Abstract
Thermal phenomena in many biological systems offer an alternative detection opportunity for quantifying relevant sample properties. While there is substantial prior work on thermal characterization methods for fluids, the push in the biology and biomedical research communities towards analysis of reduced sample volumes drives a need to extend and scale these techniques to these volumes of interest, which can be below 100 pl. This work applies the 3ω technique to measure the temperature-dependent thermal conductivity and heat capacity of de-ionized water, silicone oil, and salt buffer solution droplets from 24 to 80 °C. Heater geometries range in length from 200 to 700 μm and in width from 2 to 5 μm to accommodate the size restrictions imposed by small volume droplets. We use these devices to measure droplet volumes of 2 μl and demonstrate the potential to extend this technique down to pl droplet volumes based on an analysis of the thermally probed volume. Sensitivity and uncertainty analyses provide guidance for relevant design variables for characterizing properties of interest by investigating the tradeoffs between measurement frequency regime, device geometry, and substrate material. Experimental results show that we can extract thermal conductivity and heat capacity with these sample volumes to within less than 1% of thermal properties reported in the literature.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The calculated error of 3ω measurement due to radiation is less than 2% even at 1000 K.
“The calculated error of 3ω measurement due to radiation is less than 2% even at a high temperature of 1000 K [85].”
From Measurement Techniques for Thermal Conductivity and Interfacial Thermal Conductance of Bulk and Thin Film Materials (2016) · cites this paper as [85]
Citations
Cites
Thermal conductivity measurement from 30 to 750 K: the 3ω method
Cahill, D. G. (1990) · Review of Scientific Instruments · doi:10.1063/1.1141498 · Cited by 1,928
3 omega method for specific heat and thermal conductivity measurements
Lu, L., Yi, W., & Zhang, D. L. (2002) · arXiv · doi:10.48550/arxiv.quant-ph/0202038 · Cited by 447
1 ω , 2 ω , and 3ω methods for measurements of thermal properties
Dames, C., & Chen, G. (2005) · Review of Scientific Instruments · doi:10.1063/1.2130718 · Cited by 275
A Transient Hot Wire Thermal Conductivity Apparatus for Fluids
Roder, H. M. (1981) · Journal of Research of the National Bureau of Standards · doi:10.6028/jres.086.020 · Cited by 154
Flagged first, then by citations
Research works printed but not linked: 23
- 2005, Clin. Chem., 51, 1770, 10.1373/clinchem.2005.054924 · DOI
- 1999, Proc. Natl. Acad. Sci. U. S. A., 96, 7853, 10.1073/pnas.96.14.7853 · DOI
- 1981, Rev. Sci. Instrum., 52, 229, 10.1063/1.1136577 · DOI
- 2004, Appl. Phys. Lett., 84, 3597, 10.1063/1.1737794 · DOI
- 2009, Proc. Natl. Acad. Sci. U. S. A., 106, 15225, 10.1073/pnas.0901447106 · DOI
- 2005, Appl. Phys. Lett., 86, 034101, 10.1063/1.1850186 · DOI
- 2008, Thermochim. Acta, 477, 48, 10.1016/j.tca.2008.08.007 · DOI
- 2002, Appl. Phys. Lett., 80, 2029, 10.1063/1.1457532 · DOI
- 2007, Rev. Sci. Instrum., 78, 084902, 10.1063/1.2777162 · DOI
- 2008, Rev. Sci. Instrum., 79, 064901, 10.1063/1.2937180 · DOI
- 1996, Rev. Sci. Instrum., 67, 29, 10.1063/1.1146545 · DOI
- 2012, J. Eng. Thermophys., 21, 60, 10.1134/S1810232812010067 · DOI
- 2009, Rev. Sci. Instrum., 80, 024901, 10.1063/1.3082036 · DOI
- 2012, Rev. Sci. Instrum., 83, 106102, 10.1063/1.4757400 · DOI
- 2007, Int. J. Thermophys., 28, 1255, 10.1007/s10765-007-0254-3 · DOI
- 2011, Rev. Sci. Instrum., 82, 064903, 10.1063/1.3593372 · DOI
- 2011, Lab Chip, 11, 2378, 10.1039/c0lc00728e · DOI
- 1996, NISTIR 5928
- Feldman, 1999, HighTemp. High-Pressures, 31, 293, 10.1068/htrt171 · DOI
- 1999, J. Appl. Phys., 86, 3959, 10.1063/1.371314 · DOI
- 2014, Rev. Sci. Instrum., 85, 064903, 10.1063/1.4884638 · DOI
- 1959, Conduction of Heat in Solids, 193
- 2001, Rev. Sci. Instrum., 72, 2996, 10.1063/1.1378340 · DOI · Paper page
Cited by
Sorted by citations
Zhao, D., Qian, X., Gu, X., Jajja, S. A., & Yang, R. (2016) · Journal of Electronic Packaging · doi:10.1115/1.4034605 · Cited by 559
Metadata
- Authors
- Roy-Panzer, S., Kodama, T., Lingamneni, S., Panzer, M. A., Asheghi, M., & Goodson, K. E.
- Affiliations
- Stanford University 1 Department of Electrical Engineering, , Stanford, California 94305, USA
- Stanford University 2 Department of Mechanical Engineering, , Stanford, California 94305, USA
- ORCID
- Panzer, Matthew A.
- Venue
- Review of Scientific Instruments
- Publisher
- AIP Publishing
- Published
- 2015-02-01
- Volume
- 86
- Issue
- 2
- ISSN
- 0034-6748
- 1089-7623
- Cited by
- 20
- References
- 4
- Language
- en
- Funders
- National Science Foundation (Graduate Research Fellowship)
- Semiconductor Research Corporation (2252.001)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review