Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples
Abstract
A green mode of simultaneous homogeneous liquid–liquid microextraction combined with dispersive liquid–liquid microextraction assisted by sodium sulphate has been developed for the extraction and pre-concentration of some phthalate esters and di-(2-ethylhexyl) adipate from different aqueous samples prior to their determination by gas chromatography-mass spectrometry. In this method, an appropriate mixture of sodium sulphate solution (phase separation agent and disperser solvent) containing iso-propanol (extraction solvent) is hastily injected into an aqueous solution containing sodium sulphate and selected analytes. By doing so, the extractant is dispersed as tiny droplets into the aqueous phase. After centrifuging extractant containing the analytes is collected on the top of the aqueous phase. Under the optimum conditions, limits of detection and quantification were obtained in the ranges of 14–33, and 47–110 ng L–1, respectively. The enrichment factors and extraction recoveries of the phthalate esters ranged from 395 to 470 and 79–94%, respectively.
Overview
CoChat Summary
TL;DR
A simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction method using sodium sulphate, water as disperser, and iso-propanol as extractant was developed for selected phthalate esters and DEHA in aqueous samples before GC-MS. Under optimum conditions it gave LODs of 14–33 ng L⁻¹, LOQs of 47–110 ng L⁻¹, enrichment factors of 395–470, and extraction recoveries of 79–94%.
Key findings
- LODs and LOQs were 14–33 and 47–110 ng L⁻¹, respectively; enrichment factors ranged from 395 to 470 and extraction recoveries from 79 to 94%.
- Linearity was good for all analytes with r² ≥ 0.993; intraday and interday RSDs were 4–6% and 5–7%, respectively, at 250 ng L⁻¹.
- In real samples, DEP, DBP, and DEHP were found in mineral water, dextrose serum 5%, sodium chloride serum 0.9%, and medical injection water; DEHP was also found in tap water. For example, DEHP was 59 ± 4 µg L⁻¹ in mineral water, 74 ± 2 µg L⁻¹ in dextrose serum 5%, 65 ± 2 µg L⁻¹ in sodium chloride serum 0.9%, 28 ± 4 µg L⁻¹ in medical injection water, and 46 ± 4 µg L⁻¹ in tap water. DMP and DEHA were not detected in the reported samples.
- Matrix effects were considered negligible, with relative recoveries generally in the 88–99% range after spiking at 2 and 10 µg L⁻¹.
Methods
- The method couples HLLME and DLLME assisted by sodium sulphate: iso-propanol is the extraction solvent, distilled water is the disperser, and Na₂SO₄ acts as phase-separation/salting-out agent.
- For extraction, 5 mL aqueous sample containing 1.5 g Na₂SO₄ (30%, w/v) was mixed with 0.75 mL blank distilled water containing Na₂SO₄ (25%, w/v) and 115 µL iso-propanol containing DIBP internal standard; after centrifugation for 4 min at 1120 × g, about 10 µL organic phase was collected and 1 µL was injected into GC-MS.
- Analytes were separated on an HP-5 MS column and detected by mass spectrometry; samples included mineral water, dextrose serum 5%, sodium chloride serum 0.9%, medical injection water, tap water, well water, rain water, river water, and municipality wastewater as stated in the text.
Limitations
- The paper does not state a limitations section.
- Evident from the text: validation covered only five target analytes (DMP, DEP, DBP, DEHA, DEHP), and real-sample occurrence data were limited because DMP and DEHA were not detected in any reported sample and several analytes were not detected in rain, river, or tap water.
- Evident from the text: the procedure requires relatively high Na₂SO₄ concentrations (25–30%, w/v) and centrifugation, and it uses a homemade extraction vessel; these factors may affect practical throughput and standardization.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 1 Not valid · 9 Not checked · 39 Own results and interpretations
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The LODs and LOQs were in the ranges of 14-33, and 47-110 ng L -1 , respectively.”
- Also stated in: § ABSTRACT · plain
“Under the optimum conditions, limits of detection and quantification were obtained in the ranges of 14-33, and 47-110 ng L -1 , respectively.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § ABSTRACT · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“According to the chromatograms DEP, DBP, and DEHP were found in mineral water, dextrose serum 5%, sodium chloride serum 0.9%, and medical injection samples.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Repeatability of the introduced approach, stated as RSD%, was evaluated by employing the suggested approach on six (for intraday) and four solutions (for interday) spiked with analytes (250 ng L -1 , each analyte) and ranged from 4 to 6% and 5 to 7%, respectively.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The obtained results demonstrate a good linearity for all analytes with r 2 ≥ 0.993.”
boostedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The LODs of the introduced approach are comparable or lower than those of the previous approaches.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
25 more claims
Create a free account to see all 49 claims and every source verdict.
- It may state two separate claims as one. 8
- Probably not a substantial claim. 12
- The quote only partly states this claim. 2
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 3.8. Real samples analysis
“DIBP (IS) prepared in iso -propanol (10 mg L -1 ).”
§ 3.8. Real samples analysis
“(A) direct injection of standard solution of the target analytes at a concentration of 20 mg L -1 and IS at a concentration of 5 mg L -1”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 3.8. Real samples analysis
“Peaks identification: (1) DMP, (2) DEP, (3) DEP, (4) IS, (5) DEHA, and (6) DEHP.”
§ 3.8. Real samples analysis
“According to the chromatograms DEP, DBP, and DEHP were found in mineral water”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 3.9. Comparison of the suggested approach with other ones
“| SHLLME-DLLME-GC-MS i | Aqueous sample | 14-33 | 110-2,000,000 | 4-6 | 395-470 | This method |”
§ 3.7. Method validation
“| DMP | 33 | 110 | 110-200,000 | 0.998 | 6 | 6 | 430 ± 25 | 86 ± 5 |”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ 3.1. Selection of extraction solvent
“collected on the top of the solution as an organic layer after centrifuging for 5 min at 5000 × g.”
§ 3.2. Selection of extraction solvent volume
“centrifugation rate, 1120 × g; and centrifugation time, 5 min.”
Citations
Cites
null (2016) · Philosophy Study · doi:10.17265/2159-5313/2016.09.003 · Cited by 28,317
Solid phase microextraction with thermal desorption using fused silica optical fibers
Arthur, C. L., & Pawliszyn, J. (1990) · Analytical Chemistry · doi:10.1021/ac00218a019 · Cited by 4,997
Determination of organic compounds in water using dispersive liquid–liquid microextraction
Rezaee, M., Assadi, Y., Milani Hosseini, M.-R., Aghaee, E., Ahmadi, F., & Berijani, S. (2006) · Journal of Chromatography A · doi:10.1016/j.chroma.2006.03.007 · Cited by 3,402
Gray, L. E. (2000) · Toxicological Sciences · doi:10.1093/toxsci/58.2.350 · Cited by 1,084
Monitoring phthalate exposure in humans
Latini, G. (2005) · Clinica Chimica Acta · doi:10.1016/j.cccn.2005.05.003 · Cited by 454
Liang, P., Xu, J., & Li, Q. (2008) · Analytica Chimica Acta · doi:10.1016/j.aca.2007.12.025 · Cited by 276
Calafat, A. M., Slakman, A. R., Silva, M. J., Herbert, A. R., & Needham, L. L. (2004) · Journal of Chromatography B · doi:10.1016/j.jchromb.2004.02.006 · Cited by 206
Polo, M., Llompart, M., Garcia-Jares, C., & Cela, R. (2005) · Journal of Chromatography A · doi:10.1016/j.chroma.2004.12.040 · Cited by 164
Graphene-dispersive solid-phase extraction of phthalate acid esters from environmental water
Wu, X., Hong, H., Liu, X., Guan, W., Meng, L., Ye, Y., & Ma, Y. (2012) · The Science of The Total Environment · doi:10.1016/j.scitotenv.2012.11.060 · Cited by 151
Phthalate and adipate esters in Australian packaging materials
Balafas, D. (1999) · Food Chemistry · doi:10.1016/s0308-8146(98)00240-4 · Cited by 145
Research works printed but not linked: 2
- Rezaeepour R., 2015, Methods, 7, 3253
- Heydari R., 2018, J. Anal. Chem., 73, 110
Cited by
Chen, Z., Zhao, L., Zhang, Z., Wu, J., Zhang, L., Jing, X., & Wang, X. (2023) · Talanta · doi:10.1016/j.talanta.2023.124802 · Cited by 24
AZIMI, M., MANZOORI, J. L., AMJADI, M., et al. (2021) · Analytical Sciences · doi:10.2116/analsci.21p108 · Cited by 5
Fard, S. M. B., Ahmadi, S. H., Hajimahmodi, M., Fazaeli, R., & Amini, M. (2021) · International Journal of Environmental & Analytical Chemistry · doi:10.1080/03067319.2021.1912335 · Cited by 2
Metadata
- Authors
- Mogaddam, M. R. A., Farajzadeh, M. A., Nemati, M., Mohebbi, A., & Khodadadeian, F.
- Affiliations
- Tabriz University of Medical Sciences (IR)
- University of Tabriz (IR)
- Azarbaijan Shahid Madani University (IR)
- Corresponding author
- Mohebbi, Ali
- Khodadadeian, Fariba
- Venue
- International Journal of Environmental Analytical Chemistry
- Publisher
- Informa UK Limited
- Published
- 2020-06-23
- Volume
- 102
- Issue
- 16
- Pages
- 4466-4477
- ISSN
- 0306-7319 (ISSN-L)
- 1029-0397
- OpenAlex
- W3037711337
- Semantic Scholar
- 383cc3f35c385397049890183937be9c2f591a73
- Cited by
- 3
- References
- 36
- Citations by year
- 2023: 1 · 2021: 2
- FWCI
- 0.39
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- International Journal of Environmental & Analytical Chemistry (publishedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical chemistry methods development
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Chromatography
- Physical Sciences › Environmental Science › Health, Toxicology and Mutagenesis › Effects and risks of endocrine disrupting chemicals
- Keywords
- homogeneous liquid-liquid microextraction, dispersive liquid-liquid microextraction, plasticizers, phthalates, di(2-ethylhexyl) adipate, sodium sulfate, isopropanol, preconcentration, aqueous sample, green analytical chemistry
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples. https://app.cochat.ai/papers/10.1080/03067319.2020.1784886
- The paper
- Mogaddam, M. R. A., Farajzadeh, M. A., Nemati, M., Mohebbi, A., & Khodadadeian, F. (2022). Simultaneous homogeneous liquid–liquid microextraction and dispersive liquid–liquid microextraction for extraction of some plasticizers from polymeric containers and aqueous samples. International Journal of Environmental Analytical Chemistry. https://doi.org/10.1080/03067319.2020.1784886
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.