Disinfection of Needleless Catheter Connectors and Access Ports With Alcohol May Not Prevent Microbial Entry: The Promise of a Novel Antiseptic-Barrier Cap

Menyhay, Steve Z., Maki, Dennis G. · 2006 · Infection Control & Hospital Epidemiology
Cited by 127

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background.: Needleless valve connectors for vascular catheters are widely used throughout the United States because they reduce the risk of biohazardous injuries from needlesticks and exposure to bloodborne pathogens, such as human immunodeficiency virus and hepatitis C virus. Patients with long-term central venous catheters are at significant risk of acquiring catheter-related bloodstream infections caused by microbes that gain access through the connection between the administration set and the catheter or an injection port. Most healthcare practitioners wipe the membranous septum of the needleless connector or the injection port with 70% alcohol before accessing it. We report a simulation study of the efficacy of conventional alcohol disinfection before access, compared with that of a novel antiseptic-barrier cap that, when threaded onto a needleless luer-activated valved connector, allows a chlorhexidine-impregnated sponge to come into continuous contact with the membranous surface; after removal of the cap, there is no need to disinfect the surface with alcohol before accessing it. Methods.: One hundred five commercial, needleless luer-activated valved connectors, each accessible by a blunt male-connector luer-lock attachment, were purchased from 3 manufacturers and were tested. The membranous septum of each test device was first heavily contaminated with ~10 5 colony-forming units of Enterococcus faecalis and then was allowed to dry for 24 hours. Fifteen of the contaminated devices were not disinfected (positive controls), 30 were conventionally disinfected with a commercial 70% alcohol pledget, and 60 had the antiseptic cap threaded onto the connector and then removed after 10 minutes. The test connectors were then accessed with a sterile syringe containing nutrient broth media, which was injected, captured on the downstream side of the intraluminal fluid pathway, and cultured quantitatively. Results.: All 15 control connectors (100%) showed massive transmission of microorganisms across the membranous septum (4,500-10,000 colony-forming units). Of the 30 connectors accessed after conventional disinfection with 70% alcohol, 20 (67%) showed transmission of microorganisms (442-25,000 colony-forming units). In contrast, of the 60 connectors cultured after application of the novel antiseptic cap, only 1 (1.6%) showed any transmission of microorganisms (P<.001). Conclusions.: The findings of this study show that, if the membranous septum of a needleless luer-activated connector is heavily contaminated, conventional disinfection with 70% alcohol does not reliably prevent entry of microorganisms. In contrast, the antiseptic-barrier cap provided a high level of protection, even in the presence of very heavy contamination. This novel technology deserves to be studied in a clinical trial.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1086/500280

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

Metadata

Authors
Menyhay, S. Z., & Maki, D. G.
Venue
Infection Control & Hospital Epidemiology
Published
2006-01
Volume
27
Issue
1
Pages
23-27
ISSN
  • 0899-823X (ISSN-L)
  • 1559-6834
OpenAlex
W2162004715
Cited by
127
References
2
FWCI
8.98
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.