Broadband emissivity calibration of highly reflective samples at cryogenic temperatures
Abstract
We have developed a technique for measuring the broadband optical emissivity of high reflectivity samples at cryogenic temperatures. This measurement method employs a primary standard optical detector with high absorptance from visible wavelengths to beyond 200 µm. Background subtraction and quantification allow determination of the emissivity with total absolute uncertainty (k = 1) between approximately 0.0002 and 0.002 (for emissivities between 0.0035 and 0.092) over the temperature range from 80 K to 300 K. Using the irradiance data at the detector, precise measurements of the experimental geometry and diffraction calculations, the optical power emitted by the sample can be determined. Contact thermometry measurements for the sample can then be used to find its emissivity. The emissivity calibration technique is demonstrated for large plate samples made from polished stainless steel and is also extended to calculate the separate emissivities of multiple distinct regions on a more complex sample assembly.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Woods et al. (2012) at NIST developed a low-uncertainty, absolutely calibrated emissivity measurement system for the precise characterization of highly reflective materials.
“In 2012, Woods et al. [48] at the National Institute of Standards and Technology (NIST) developed a low-uncertainty, absolutely calibrated emissivity measurement system for the precise characterization of highly reflective materials.”
From Thermal Radiation Testing Methods at Cryogenic Temperatures: A Review (2026) · cites this paper as [48]
By incorporating shutter-partitioning technology, the Woods et al. system allowed independent emissivity measurements of multiple microregions on the same sample surface.
“By incorporating shutterpartitioning technology, it allowed independent emissivity measurements of multiple microregions on the same sample surface.”
From Thermal Radiation Testing Methods at Cryogenic Temperatures: A Review (2026) · cites this paper as [48]
The Woods et al. system enabled precise emissivity measurements of polished 304 stainless steel across the 80-300 K temperature range.
“Based on the radiometric method, this system enabled precise emissivity measurements of polished 304 stainless steel across the 80-300 K temperature range [38].”
From Thermal Radiation Testing Methods at Cryogenic Temperatures: A Review (2026) · cites this paper as [48]
The Woods et al. apparatus employed a helium-cooled, high-absorption, low-noise absolute cryogenic radiometer as the detector.
“The apparatus employed a heliumcooled, high-absorption, low-noise absolute cryogenic radiometer as the detector.”
From Thermal Radiation Testing Methods at Cryogenic Temperatures: A Review (2026) · cites this paper as [48]
Citations
Cites
No references could be resolved yet.
Research works printed but not linked: 14
- Ageladarakis P A Obert W DewHughes D, 1998, Cryogenic Engineering Conf. 1998, 715
- Lipa M Martinez A Lacroix B Chatain D, 2009, 23rd IEEE/NPSS Symp. on Fusion Engineering, 603
- 10.1016/j.cryogenics.2008.07.007 · DOI · Paper page
- 10.1117/12.624002 · DOI
- 10.1117/12.624020 · DOI
- 10.1016/j.cryogenics.2004.11.010 · DOI · Paper page
- 10.1364/JOSA.69.001384 · DOI
- 10.1364/AO.31.007219 · DOI
- Houston J M, 2006, Metrologia, 43, S31, 10.1088/0026-1394/43/2/S07 · DOI
- 10.1364/AO.44.000871 · DOI
- 10.1364/AO.40.004463 · DOI
- 10.1016/S1079-4042(05)41009-7 · DOI
- 10.1016/S1079-4042(05)41001-2 · DOI
- 10.1364/JOSA.68.001505 · DOI
Cited by
Sorted by citations
Thermal Radiation Testing Methods at Cryogenic Temperatures: A Review
Li, B., & Shen, F. (2026) · Cryo · doi:10.3390/cryo2010004 · Cited by 1
Metadata
- Authors
- Woods, S. I., Jung, T. M., Ly, G. T., & Yu, J.
- Venue
- Metrologia
- Publisher
- IOP Publishing
- Published
- 2012-10-15
- Volume
- 49
- Issue
- 6
- Pages
- 737-744
- ISSN
- 0026-1394
- 1681-7575
- Cited by
- 10
- References
- 0
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review