Physical characteristics of the ECAT EXACT3D positron tomograph
The DOI matches its Crossref record.
Abstract
The 'EXACT3D' positron tomograph, which is now in routine clinical research use, was developed with the aim of achieving unprecedented sensitivity, high spatial and temporal resolution and simplicity of design using proven detector technology. It consists of six rings of standard detector blocks (CTI/Siemens EXACT HR+) with 4.39 mm x 4.05 mm x 30 mm elements, giving an axial field of view (FOV) of 23.4 cm. This extended FOV and the absence of interplane septa and retractable transmission rod sources has allowed greatly simplified gantry and detector cassette design. Operation in exclusive 3D mode requires an alternative to the conventional coincidence method for transmission scanning, and a single photon approach using a hydraulically driven 137Cs point source has been implemented. The tomograph has no other moving parts. A single time frame of data without any compression is very large (> 300 Mbyte) and two approaches are employed to overcome this difficulty: (a) adjacent sinograms can be summed automatically into different combinations and (b) listmode (event-by-event) acquisition has been instituted, which is both storage efficient (particularly for acquisition of sparse data sets) and maximizes temporal resolution. The high-speed I/O and computing hardware can maintain a sustained acquisition rate of about 4 million coincidence events per second. A disadvantage of the large axial FOV in 3D is the increased sensitivity to activity outside the coincidence FOV. However, this can be minimized by additional side shielding. The mean spatial resolution is 4.8 +/- 0.2 mm FWHM (transaxial, 1 cm off-axis) and 5.6 +/- 0.5 mm (axial, on-axis). Its absolute efficiency is 5.8% for a line source in air (just spanning the axial FOV) and 10% for a central point source (with thresholds of 350-650 keV). For a uniform 20 cm diameter cylinder, the efficiency is 69 kcps kBq(-1) ml(-1) (after subtraction of a scatter fraction of 42%). Sensitivity relative to the EXACT HR+ (with four rings of blocks) is 2.5 (3D) and 12 (2D) times respectively. The rate of random events in blood flow studies in the brain and body, using 15O-labelled water, can be controlled by limiting the administered dose and inserting additional side shielding.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
In vivo imaging of microglial activation with [11C](R)-PK11195 PET in idiopathic Parkinson's disease
Gerhard, A., Pavese, N., Hotton, G., Turkheimer, F., Es, M., Hammers, A., Eggert, K., Oertel, W., Banati, R. B., & Brooks, D. J. (2006) · Neurobiology of Disease · doi:10.1016/j.nbd.2005.08.002 · Cited by 1,200
Edison, P., Archer, H. A., Gerhard, A., Hinz, R., Pavese, N., Turkheimer, F. E., Hammers, A., Tai, Y. F., Fox, N., Kennedy, A., et al. (2008) · Neurobiology of Disease · doi:10.1016/j.nbd.2008.08.001 · Cited by 513
Microglial activation correlates with severity in Huntington disease: A clinical and PET study
Pavese, N., Gerhard, A., Tai, Y. F., Ho, A. K., Turkheimer, F., Barker, R. A., Brooks, D. J., & Piccini, P. (2006) · Neurology · doi:10.1212/01.wnl.0000222734.56412.17 · Cited by 462
Microglial activation in presymptomatic Huntington's disease gene carriers
Tai, Y. F., Pavese, N., Gerhard, A., Tabrizi, S. J., Barker, R. A., Brooks, D. J., & Piccini, P. (2007) · Brain · doi:10.1093/brain/awm044 · Cited by 440
“The camera has a transaxial spatial resolution of 4.8 0.2mm and axial resolution of 5.6 0.5mm after image reconstruction (Spinks et al., 2000).”
BackgroundMethodInfluentialRajappan, K., Rimoldi, O. E., Dutka, D. P., Ariff, B., Pennell, D. J., Sheridan, D. J., & Camici, P. G. (2002) · Circulation · doi:10.1161/hc0402.102931 · Cited by 358
Metadata
- Authors
- Spinks, T. J., Jones, T., Bloomfield, P. M., Bailey, D. L., Miller, M., Hogg, D., Jones, W. F., Vaigneur, K., Reed, J., Young, J., Newport, D., Moyers, C., Casey, M. E., & Nutt, R.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2000-08-25
- Volume
- 45
- Issue
- 9
- Pages
- 2601-2618
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W1977582341
- Cited by
- 155
- References
- 0
- FWCI
- 9.78
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Physical characteristics of the ECAT EXACT3D positron tomograph. https://app.cochat.ai/papers/10.1088/0031-9155/45/9/313
- The paper
- Spinks, T. J., Jones, T., Bloomfield, P. M., Bailey, D. L., Miller, M., Hogg, D., Jones, W. F., Vaigneur, K., Reed, J., Young, J., Newport, D., Moyers, C., Casey, M. E., & Nutt, R. (2000). Physical characteristics of the ECAT EXACT3D positron tomograph. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/45/9/313
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.