Determination of the weighted CT dose index in modern multi-detector CT scanners

Open access

Abstract

The aim of the present study was to (a) evaluate the underestimation in the value of the free-in-air (CTDI(air)) and the weighted CT dose index (CTDI(w)) determined with the standard 100 mm pencil chamber, i.e. the CTDI(100) concept, for the whole range of nominal radiation beam collimations selectable in a modern multi-slice CT scanner, (b) estimate the optimum length of the pencil-chamber and phantoms for accurate CTDI(w) measurements and (c) provide CTDI(w) values normalized to free-in-air CTDI for different tube-voltage, nominal radiation beam collimations and beam filtration values. The underestimation in the determination of CTDI(air) and CTDI(w) using the CTDI(100) concept was determined from measurements obtained with standard polymethyl-methacrylate (PMMA) phantoms and arrays of thermoluminescence dosimeters. The Monte Carlo N-Particle transport code was used to simulate standard CTDI measurements on a 16-slice CT scanner. The optimum pencil-chamber length for accurate determination of CTDI(w) was estimated as the minimum chamber length for which a further increase in length does not alter the value of the CTDI. CTDI(w)/CTDI(air) ratios were determined using Monte Carlo simulation and the optimum detector length for all selectable tube-voltage values and for three different values of beam filtration. To verify the Monte Carlo results, measured values of CTDI(w)/CTDI(air) ratios using the standard 100 mm pencil ionization chamber were compared with corresponding values calculated with Monte Carlo experiments. The underestimation in the determination of CTDI(air) using the 100 mm pencil chamber was less than 1% for all beam collimations. The underestimation in CTDI(w) was 15% and 27% for head and body phantoms, respectively. The optimum detector length for accurate CTDI(w) measurements was found to be 50 cm for the beam collimations commonly employed in modern multi-detector (MD) CT scanners. The ratio of CTDI(w)/CTDI(air) determined using the optimum detector length was found to be independent of beam collimation. Percentage differences between measured and calculated corresponding CTDI(w)/CTDI(air) ratios were always less than 8% for head and less than 5% for body PMMA phantoms. In conclusion, the CTDI(air) of MDCT scanners may be measured accurately with a 100 mm pencil chamber. However, the CTDI(100) concept was found to be inadequate for accurate CTDI(w) determination for the wide beam collimations commonly used in MDCT scanners. Accurate CTDI(w) determination presupposes the use of a pencil chamber and PMMA phantoms at least 50 cm long.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/0031-9155/52/21/010

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Research works printed but not linked: 1

  1. Jones D G, 1993, NRPB SR250

Cited by

Sorted by citations

See all 47

Metadata

Authors
Perisinakis, K., Damilakis, J., Tzedakis, A., Papadakis, A., Theocharopoulos, N., & Gourtsoyiannis, N.
Affiliations
University of Crete (GR)
Venue
Physics in Medicine and Biology
Published
2007-10-16
Volume
52
Issue
21
Pages
6485-6495
ISSN
  • 0031-9155 (ISSN-L)
  • 1361-6560
OpenAlex
W2056577594
Cited by
47
References
21
Citations by year
2026: 1 · 2025: 3 · 2024: 2 · 2023: 2 · 2022: 1 · 2021: 1 · 2020: 3 · 2019: 1 · 2018: 1 · 2017: 5 · 2016: 3 · 2015: 2 · 2014: 2 · 2013: 2 · 2012: 2
FWCI
2.49
Type
journal article
Language
en
Open access
Locations
  • Physics in Medicine and Biology (publishedVersion)
  • PubMed (publishedVersion)
Funders
Siemens
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
  • Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Digital Radiography and Breast Imaging
Keywords
pencil ionization chamber, multidetector computed tomography, radiation dosimetry, beam collimation, beam filtration, tube voltage, PMMA phantom, thermoluminescent dosimeters, Monte Carlo simulation, dose normalization, radiation protection
MeSH terms
Equipment Design, Humans, Monte Carlo Method, Radiation Dosage, Radiometry, Thermoluminescent Dosimetry, Tomography, X-Ray Computed, X-Rays, Reproducibility of Results, Models, Statistical, Tomography Scanners, X-Ray Computed, Phantoms, Imaging, Polymethyl Methacrylate

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review