Uncertainty and bias in contrast concentration measurements using spoiled gradient echo pulse sequences
Open accessAbstract
Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is a widely used technique for assessing tissue physiology. Spoiled gradient echo (SPGR) pulse sequences are one of the most common methods for acquisition of DCE-MRI data, providing high temporal and spatial resolution with strong T(1)-weighting. Conversion of SPGR signal to concentration is briefly reviewed, and a new closed-form expression for concentration measurement uncertainty for finite signal-to-noise ratio (SNR) and baseline scan time is derived. This result is applicable to arbitrary concentration-dependent relaxation rate and is valid over the same domain as the theoretical SPGR signal equation. Expressions for the lower and upper bounds on measurable concentration are also derived. The existence of a concentration- and tissue-dependent optimal flip angle that minimizes concentration uncertainty is demonstrated and it is shown that, for clinically relevant pulse sequence parameters, this optimal flip angle is significantly larger than the corresponding Ernst angle. Analysis of three pulse sequences from the DCE-MRI literature shows that optimization of flip angle using the methods discussed here leads to potential improvements of 10-1166% in effective SNR over the 0.5-5.0 mM concentration range with minimal or no loss of measurement accuracy down to 0.1 mM. In vivo data from three study patients provide further support for our theoretical expression for concentration measurement uncertainty, with predicted and experimental estimates agreeing to within +/- 30%. Equations for concentration bias resulting from biases in flip angle and from pre-contrast relaxation time and contrast relaxivity (both longitudinal and transverse) are also derived in closed-form. The resulting equations show the potential for significant contributions to bias in concentration measurement arising from even relatively small mis-specification of flip angle and/or pre-contrast longitudinal relaxation time, particularly at high contrast concentrations.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
T1, T2 relaxation and magnetization transfer in tissue at 3T
Stanisz, G. J., Odrobina, E. E., Pun, J., Escaravage, M., Graham, S. J., Bronskill, M. J., & Henkelman, R. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20605 · Cited by 1,266
Cited by
Quantitative susceptibility mapping (QSM): Decoding MRI data for a tissue magnetic biomarker
Wang, Y., & Liu, T. (2015) · Magnetic Resonance in Medicine · doi:10.1002/mrm.25358 · Cited by 878
Liu, J., Liu, T., de Rochefort, L., Ledoux, J., Khalidov, I., Chen, W., Tsiouris, A. J., Wisnieff, C., Spincemaille, P., Prince, M. R., et al. (2012) · NeuroImage · doi:10.1016/j.neuroimage.2011.08.082 · Cited by 498
Classic models for dynamic contrast‐enhanced MRI
Sourbron, S. P., & Buckley, D. L. (2013) · NMR in Biomedicine · doi:10.1002/nbm.2940 · Cited by 434
Heye, A. K., Culling, R. D., Valdés Hernández, M. D. C., Thrippleton, M. J., & Wardlaw, J. M. (2014) · NeuroImage: Clinical · doi:10.1016/j.nicl.2014.09.002 · Cited by 406
Liu, T., Wisnieff, C., Lou, M., Chen, W., Spincemaille, P., & Wang, Y. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24272 · Cited by 360
Metadata
- Authors
- Schabel, M. C., & Parker, D. L.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2008-04-17
- Volume
- 53
- Issue
- 9
- Pages
- 2345-2373
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2021539460
- Cited by
- 221
- References
- 1
- FWCI
- 10.99
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Uncertainty and bias in contrast concentration measurements using spoiled gradient echo pulse sequences. https://app.cochat.ai/papers/10.1088/0031-9155/53/9/010
- The paper
- Schabel, M. C., & Parker, D. L. (2008). Uncertainty and bias in contrast concentration measurements using spoiled gradient echo pulse sequences. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/53/9/010
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.