Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold
The DOI matches its Crossref record.
Abstract
We studied quantitatively the effects of the discontinuity introduced in an otherwise homogeneous background by the cold walls of the standard spherical glass inserts commonly used in phantom measurements for calibration of threshold-based approaches to volumetric evaluation of PET investigations. We concentrated especially on the question of threshold-based volume determination. We computed analytically the convolution of an isotropic Gaussian point-spread function with the insert geometry (hot sphere + cold wall + warm background) and derived the theoretical background dependence of the volume reproducing threshold. This analysis shows a clear wall-related reduction of the optimal threshold with increasing background. The predictions of our theoretical analysis were verified in phantom measurements at background fractions between 0 and 0.29. Defining the background-corrected relative threshold [formula: see text] (T(abs): absolute volume reproducing threshold, A: measured activity at centre, B: background), we find that for a wall-less sphere T is independent of the background level. In the presence of cold walls, T drops (for not too small spheres, where recovery at the centre approaches 100%) from about 43% at B/A = 0 to about 25% at B/A = 0.5. Applying these thresholds to wall-less spheres leads to sizeable overestimates of the true volumes (43% at B/A = 0.5 for a sphere of 6 ml volume). We conclude that phantom measurements with standard sphere inserts for calibration of optimal thresholding algorithms introduce a systematic bias if performed at finite background levels. The observed background dependence is an artefact of the measurement procedure and does not reflect the conditions present in actual patient investigations.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
A review on segmentation of positron emission tomography images
Foster, B., Bagci, U., Mansoor, A., Xu, Z., & Mollura, D. J. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.04.014 · Cited by 370
Lee, J. A. (2010) · Radiotherapy and Oncology · doi:10.1016/j.radonc.2010.07.003 · Cited by 159
“Though thin, these walls have a non-negligible impact on the images and should be accounted for [35,46,55,62].”
BackgroundRogasch, J. M., Suleiman, S., Hofheinz, F., Bluemel, S., Lukas, M., Amthauer, H., & Furth, C. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-0270-y · Cited by 63
Self-citationA method for model-free partial volume correction in oncological PET
Hofheinz, F., Langner, J., Petr, J., Beuthien-Baumann, B., Oehme, L., Steinbach, J., Kotzerke, J., & van den Hoff, J. (2012) · EJNMMI Research · doi:10.1186/2191-219x-2-16 · Cited by 55
“Moreover, the standard spherical phantom inserts are hollow glass spheres whose cold walls can have a strong influence on the measured partial volume effects [31,32] which are, therefore, not representative for the conditions found in real data.”
BackgroundSelf-citationRogasch, J. M., Hofheinz, F., Lougovski, A., Furth, C., Ruf, J., Großer, O. S., Mohnike, K., Hass, P., Walke, M., Amthauer, H., et al. (2014) · EJNMMI Physics · doi:10.1186/2197-7364-1-12 · Cited by 53
Self-citation
Metadata
- Authors
- Hofheinz, F., Dittrich, S., Pötzsch, C., & Hoff, J. V. D.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2010-01-28
- Volume
- 55
- Issue
- 4
- Pages
- 1099-1113
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W1980291386
- Cited by
- 111
- References
- 0
- FWCI
- 6.99
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. https://app.cochat.ai/papers/10.1088/0031-9155/55/4/013
- The paper
- Hofheinz, F., Dittrich, S., Pötzsch, C., & Hoff, J. V. D. (2010). Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/55/4/013
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.