Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF)
Open accessAbstract
Background: Bayesian penalized likelihood reconstruction for PET (e.g., GE Q.Clear) aims at improving convergence of lesion activity while ensuring sufficient signal-to-noise ratio (SNR). This study evaluated reconstructed spatial resolution, maximum/peak contrast recovery (CRmax/CRpeak) and SNR of Q.Clear compared to time-of-flight (TOF) OSEM with and without point spread function (PSF) modeling. Methods: The NEMA IEC Body phantom was scanned five times (3 min scan duration, 30 min between scans, background, 1.5–3.9 kBq/ml F18) with a GE Discovery MI PET/CT (3-ring detector) with spheres filled with 8-, 4-, or 2-fold the background activity concentration (SBR 8:1, 4:1, 2:1). Reconstruction included Q.Clear (beta, 150/300/450), “PSF+TOF 4/16” (iterations, 4; subsets, 16; in-plane filter, 2.0 mm), “OSEM+TOF 4/16” (identical parameters), “PSF+TOF 2/17” (2 it, 17 ss, 2.0 mm filter), “OSEM+TOF 2/17” (identical), “PSF+TOF 4/8” (4 it, 8 ss, 6.4 mm), and “OSEM+TOF 2/8” (2 it, 8 ss, 6.4 mm). Spatial resolution was derived from 3D sphere activity profiles. RC as (sphere activity concentration [AC]/true AC). SNR as (background mean AC/background AC standard deviation). Results: Spatial resolution of Q.Clear 150 was significantly better than all conventional algorithms at SBR 8:1 and 4:1 (Wilcoxon, each p < 0.05). At SBR 4:1 and 2:1, the spatial resolution of Q.Clear 300/450 was similar or inferior to PSF+TOF 4/16 and OSEM+TOF 4/16. Small sphere CRpeak generally underestimated true AC, and it was similar for Q.Clear 150/300/450 as with PSF+TOF 4/16 or PSF+TOF 2/17 (i.e., relative differences < 10%). Q.Clear provided similar or higher CRpeak as OSEM+TOF 4/16 and OSEM+TOF 2/17 resulting in a consistently better tradeoff between CRpeak and SNR with Q.Clear. Compared to PSF+TOF 4/8/OSEM+TOF 2/8, Q.Clear 150/300/450 showed lower SNR but higher CRpeak. Conclusions: Q.Clear consistently improved reconstructed spatial resolution at high and medium SBR compared to PSF+TOF and OSEM+TOF, but only with beta = 150. However, this is at the cost of inferior SNR with Q.Clear 150 compared to Q.Clear 300/450 and PSF+TOF 4/16/PSF+TOF 2/17 while CRpeak for the small spheres did not improve considerably. This suggests that Q.Clear 300/450 may be advantageous for the 3-ring detector configuration because the tradeoff between CR and SNR with Q.Clear 300/450 was superior to PSF+TOF 4/16, OSEM+TOF 4/16, and OSEM+TOF 2/17. However, it requires validation by systematic evaluation in patients at different activity and acquisition protocols.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Even though improved scanner technology and reconstruction algorithms lead to a noticeably improved reconstructed spatial resolution compared to older systems (about 4 mm vs. 6 mm) [41], it is still much worse than the resolution of computed tomography (CT) or magnetic resonance imaging (MRI).”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [41]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Several studies on subjective image quality in whole-body [18F]FDG-PET with PL reconstruction found that image quality was highest at beta values of 450 (to 600) despite lower lesion CR or “image sharpness” at these beta values compared to lower values [41,108,112,175,179].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [41]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“However, this difference is directly dependent on the user-defined penalization factor beta during PL reconstruction, and at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers [107,112,113], inter-method SUV differences may no longer be significant [41,108,111].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [41]
The cited source says what the claim attributes to it.
In the citing paper
“Using a commercially available PL reconstruction (General Electric [GE] Q.Clear), several phantom studies have shown that PL reconstruction can increase the CR of standardized sphere inserts compared to conventional TOF and PSF reconstruction [41,46,101,107,108].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [41]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335
Teoh, E. J., McGowan, D. R., Macpherson, R. E., Bradley, K. M., & Gleeson, F. V. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.159301 · Cited by 265
EM algorithm system modeling by image-space techniques for PET reconstruction
Reader, A. J., Julyan, P. J., Williams, H., Hastings, D. L., & Zweit, J. (2003) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2003.817327 · Cited by 191
Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET
Rapisarda, E., Bettinardi, V, Thielemans, K., & Gilardi, M. C. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/14/012 · Cited by 142
Pan, T., Einstein, S. A., Kappadath, S. C., Grogg, K. S., Lois Gomez, C., Alessio, A. M., Hunter, W. C., El Fakhri, G., Kinahan, P. E., & Mawlawi, O. R. (2019) · Medical Physics · doi:10.1002/mp.13576 · Cited by 122
Effects of cold sphere walls in PET phantom measurements on the volume reproducing threshold
Hofheinz, F., Dittrich, S., Pötzsch, C., & van den Hoff, J. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/4/013 · Cited by 111
Self-citationLindström, E., Sundin, A., Trampal, C., Lindsjö, L., Ilan, E., Danfors, T., Antoni, G., Sörensen, J., & Lubberink, M. (2018) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.200790 · Cited by 99
Akamatsu, G., Mitsumoto, K., Taniguchi, T., Tsutsui, Y., Baba, S., & Sasaki, M. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2013.09.030 · Cited by 96
Armstrong, I. S., Kelly, M. D., Williams, H. A., & Matthews, J. C. (2014) · EJNMMI Physics · doi:10.1186/s40658-014-0099-3 · Cited by 85
Cited by
Language Analysis of Speakers with Dementia of the Alzheimer’s Type
Guinn, C. I., & Habash, A. (2012) · National Conference on Artificial Intelligence · Cited by 71
Miwa, K., Wagatsuma, K., Nemoto, R., Masubuchi, M., Kamitaka, Y., Yamao, T., Hiratsuka, S., Yamaguchi, M., Yoshii, T., Kobayashi, R., et al. (2020) · Annals of Nuclear Medicine · doi:10.1007/s12149-020-01500-8 · Cited by 58
Influences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 57
“Clear), several phantom studies have shown that PL reconstruction can increase the CR of standardized sphere inserts compared to conventional TOF and PSF reconstruction [41,46,101,107,108].”
BackgroundInfluentialSelf-citationMiwa, K., Yoshii, T., Wagatsuma, K., Nezu, S., Kamitaka, Y., Yamao, T., Kobayashi, R., Fukuda, S., Yakushiji, Y., Miyaji, N., et al. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00527-w · Cited by 41
Yoshii, T., Miwa, K., Yamaguchi, M., Shimada, K., Wagatsuma, K., Yamao, T., Kamitaka, Y., Hiratsuka, S., Kobayashi, R., Ichikawa, H., et al. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00325-8 · Cited by 32
Metadata
- Authors
- Rogasch, J. M., Suleiman, S., Hofheinz, F., Bluemel, S., Lukas, M., Amthauer, H., & Furth, C.
- Affiliations
- Humboldt-Universität zu Berlin (DE)
- Helmholtz-Zentrum Dresden-Rossendorf (DE)
- Corresponding author
- Rogasch, Julian M.
- Venue
- EJNMMI Physics
- Published
- 2020-01-10
- Volume
- 7
- Issue
- 1
- Pages
- 2-2
- ISSN
- 2197-7364 (ISSN-L)
- PMC
- PMC6954158
- OpenAlex
- W2999826695
- Cited by
- 63
- References
- 18
- Citations by year
- 2026: 3 · 2025: 6 · 2024: 5 · 2023: 15 · 2022: 17 · 2021: 10 · 2020: 6
- FWCI
- 5.84
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- EJNMMI Physics (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Refubium (Universitätsbibliothek der Freien Universität Berlin) (submittedVersion, cc-by)
- Universitätsbibliothek der FU Berlin Hochschulschriftenstelle u. Dokumentenserver (cc-by)
- Funders
- Deutsche Forschungsgemeinschaft
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Keywords
- Bayesian penalized likelihood reconstruction, Q.Clear, FDG-PET, point spread function, spatial resolution, contrast recovery, signal-to-noise ratio, lesion-to-background ratio, PET image quality, iterative reconstruction, 18F-FDG
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF). https://app.cochat.ai/papers/10.1186/s40658-020-0270-y
- The paper
- Rogasch, J. M., Suleiman, S., Hofheinz, F., Bluemel, S., Lukas, M., Amthauer, H., & Furth, C. (2020). Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF). EJNMMI Physics. https://doi.org/10.1186/s40658-020-0270-y
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.