LaBr3:Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time
Open accessThe DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
The use of time-of-flight (TOF) information in positron emission tomography (PET) enables significant improvement in image noise properties and, therefore, lesion detection. Silicon photomultipliers (SiPMs) are solid-state photosensors that have several advantages over photomultiplier tubes (PMTs). SiPMs are small, essentially transparent to 511 keV gamma rays and insensitive to magnetic fields. This enables novel detector designs aimed at e.g. compactness, high resolution, depth-of-interaction (DOI) correction and MRI compatibility. The goal of the present work is to study the timing performance of SiPMs in combination with LaBr(3):Ce(5%), a relatively new scintillator with promising characteristics for TOF-PET. Measurements were performed with two, bare, 3 mm x 3 mm x 5 mm LaBr(3):Ce(5%) crystals, each coupled to a 3 mm x 3 mm SiPM. Using a (22)Na point source placed at various positions in between the two detectors, a coincidence resolving time (CRT) of approximately 100 ps FWHM for 511 keV annihilation photon pairs was achieved, corresponding to a TOF positioning resolution of approximately 15 mm FWHM. At the same time, pulse height spectra with well-resolved full-energy peaks were obtained. To our knowledge this is the best CRT reported for SiPM-based scintillation detectors to date. It is concluded that SiPM-based scintillation detectors can provide timing resolutions at least as good as detectors based on PMTs.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Simultaneous PET-MRI: a new approach for functional and morphological imaging
Judenhofer, M. S., Wehrl, H. F., Newport, D. F., Catana, C., Siegel, S. B., Becker, M., Thielscher, A., Kneilling, M., Lichy, M. P., Eichner, M., et al. (2008) · Nature Medicine · doi:10.1038/nm1700 · Cited by 981 · cited 7×
“The above results were achieved using detectors based on 3 mm × 3 mm × 5 mm LaBr3:Ce(5%) crystals and 3 mm × 3 mm SiPMs. To obtain sufficient system sensitivity, a clinical TOF-PET scanner might, for example, be based on several stacked layers of such detector elements. Alternatively, the detector design might be base…”
MethodResultInfluentialSurti, S., Kuhn, A., Werner, M. E., Perkins, A. E., Kolthammer, J. A., & Karp, J. S. (2007) · PubMed · Cited by 639
Catana, C., Wu, Y., Judenhofer, M. S., Qi, J., Pichler, B. J., & Cherry, S. R. (2006) · PubMed · Cited by 350
Recent developments in PET detector technology
Lewellen, T. K. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/17/r01 · Cited by 336
Muehllehner, G., & Karp, J. S. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/13/r08 · Cited by 288
Multimodality imaging of structure and function
Townsend, D. W. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/4/r01 · Cited by 286
A novel, SiPM-array-based, monolithic scintillator detector for PET
Schaart, D. R., van Dam, H. T., Seifert, S., Vinke, R., Dendooven, P., Löhner, H., & Beekman, F. J. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/11/015 · Cited by 286
Self-citationImpact of Time-of-Flight on PET Tumor Detection
Kadrmas, D. J., Casey, M. E., Conti, M., Jakoby, B. W., Lois, C., & Townsend, D. W. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.063016 · Cited by 277
An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging
Lois, C., Jakoby, B. W., Long, M. J., Hubner, K. F., Barker, D. W., Casey, M. E., Conti, M., Panin, V. Y., Kadrmas, D. J., & Townsend, D. W. (2010) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.068098 · Cited by 265 · cited 1×
“…in positron emission tomography (PET) has recently been demonstrated to enable significant improvement in image noise properties and, therefore, lesion detection, especially in heavier patients (Kadrmas et al 2009, Lois et al 2010, Moses 2007, Muehllehner and Karp 2006, Surti et al 2007).”
BackgroundProspects for time-of-flight PET using LSO scintillator
Moses, W. W., & Derenzo, S. E. (1999) · IEEE Transactions on Nuclear Science · doi:10.1109/23.775565 · Cited by 240
Novel type of avalanche photodetector with Geiger mode operation
Golovin, V., & Saveliev, V. (2004) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2003.11.085 · Cited by 233
Limited Geiger-mode microcell silicon photodiode: new results
Bondarenko, G., Buzhan, P., Dolgoshein, B., Golovin, V, Guschin, E., Ilyin, A., Kaplin, V, Karakash, A., Klanner, R., Pokachalov, V, et al. (2000) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/s0168-9002(99)01219-x · Cited by 220
Recent advances and future advances in time-of-flight PET
Moses, W. W. (2007) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2007.06.038 · Cited by 205
The imaging performance of a LaBr3-based PET scanner
Daube-Witherspoon, M. E., Surti, S., Perkins, A., Kyba, C. C. M., Wiener, R., Werner, M. E., Kulp, R., & Karp, J. S. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/1/004 · Cited by 122
Renker, D. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.10.016 · Cited by 117 · cited 1×
“…photomultipliers (SiPMs) are turning into a promising alternative for PMTs, as of recently (Antich et al 1997, Bondarenko et al 2000, Britvitch et al 2007, Golovin and Saveliev 2004, Herbert et al 2007, Lewellen 2008, McElroy et al 2007, Musienko et al 2007, Renker 2007, Yamamoto et al 2007).”
BackgroundDevelopment of Multi-Pixel Photon Counter (MPPC)
Yamamoto, K., Yamamura, K., Sato, K., Kamakura, S., Ota, T., Suzuki, H. S., & Ohsuka, S. (2007) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2007.4437286 · Cited by 91 · cited 2×
“…photomultipliers (SiPMs) are turning into a promising alternative for PMTs, as of recently (Antich et al 1997, Bondarenko et al 2000, Britvitch et al 2007, Golovin and Saveliev 2004, Herbert et al 2007, Lewellen 2008, McElroy et al 2007, Musienko et al 2007, Renker 2007, Yamamoto et al 2007).”
BackgroundEvaluation of silicon photomultipliers: A promising new detector for MR compatible PET
McElroy, D. P., Saveliev, V. I., Reznik, A., & Rowlands, J. A. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.10.040 · Cited by 63
The Silicon Photomultiplier for application to high-resolution Positron Emission Tomography
Herbert, D., Moehrs, S., D’Ascenzo, N., Belcari, N., Guerra, A. D., Morsani, F., & Saveliev, V. I. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.11.020 · Cited by 53
Characterisation of Geiger-mode avalanche photodiodes for medical imaging applications
Britvitch, I., Johnson, I., Renker, D., Stoykov, A., & Eckart Lorenz, G. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.10.089 · Cited by 37 · cited 1×
“…silicon photomultipliers (SiPMs) are turning into a promising alternative for PMTs, as of recently (Antich et al 1997, Bondarenko et al 2000, Britvitch et al 2007, Golovin and Saveliev 2004, Herbert et al 2007, Lewellen 2008, McElroy et al 2007, Musienko et al 2007, Renker 2007, Yamamoto et…”
BackgroundStudy of multi-pixel Geiger-mode avalanche photodiodes as a read-out for PET
Musienko, Y., Auffray, E., Lecoq, P. R., Reucroft, S., Swain, J. D., & Trummer, J. (2006) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2006.10.109 · Cited by 19 · cited 1×
“…photomultipliers (SiPMs) are turning into a promising alternative for PMTs, as of recently (Antich et al 1997, Bondarenko et al 2000, Britvitch et al 2007, Golovin and Saveliev 2004, Herbert et al 2007, Lewellen 2008, McElroy et al 2007, Musienko et al 2007, Renker 2007, Yamamoto et al 2007).”
BackgroundEnergy and Timing Response of Six Prototype Scintillators for TOF-PET
Kyba, C. C. M., Glodo, J., van Loef, E. V. D., Karp, J. S., & Shah, K. S. (2008) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.921939 · Cited by 18
An investigation of waveform sampling for improved signal processing in TOF PET
Wiener, R. I., Surti, S., Kyba, C. C. M., Newcomer, F. M., Berg, R. V., & Karp, J. S. (2008) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2008.4774183 · Cited by 12
Cited by
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335
PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging
Vandenberghe, S., & Marsden, P. K. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/4/r115 · Cited by 258
Metadata
- Authors
- Schaart, D. R., Seifert, S., Vinke, R., van Dam, H. T., Dendooven, P., Löhner, H., & Beekman, F. J.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2010-03-19
- Volume
- 55
- Issue
- 7
- Pages
- N179-N189
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2151969366
- Cited by
- 225
- References
- 22
- FWCI
- 20.11
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.