Automatic localization of vertebral levels in x-ray fluoroscopy using 3D–2D registration: a tool to reduce wrong-site surgery
Open accessAbstract
Surgical targeting of the incorrect vertebral level (wrong-level surgery) is among the more common wrong-site surgical errors, attributed primarily to the lack of uniquely identifiable radiographic landmarks in the mid-thoracic spine. The conventional localization method involves manual counting of vertebral bodies under fluoroscopy, is prone to human error and carries additional time and dose. We propose an image registration and visualization system (referred to as LevelCheck), for decision support in spine surgery by automatically labeling vertebral levels in fluoroscopy using a GPU-accelerated, intensity-based 3D-2D (namely CT-to-fluoroscopy) registration. A gradient information (GI) similarity metric and a CMA-ES optimizer were chosen due to their robustness and inherent suitability for parallelization. Simulation studies involved ten patient CT datasets from which 50 000 simulated fluoroscopic images were generated from C-arm poses selected to approximate the C-arm operator and positioning variability. Physical experiments used an anthropomorphic chest phantom imaged under real fluoroscopy. The registration accuracy was evaluated as the mean projection distance (mPD) between the estimated and true center of vertebral levels. Trials were defined as successful if the estimated position was within the projection of the vertebral body (namely mPD <5 mm). Simulation studies showed a success rate of 99.998% (1 failure in 50 000 trials) and computation time of 4.7 s on a midrange GPU. Analysis of failure modes identified cases of false local optima in the search space arising from longitudinal periodicity in vertebral structures. Physical experiments demonstrated the robustness of the algorithm against quantum noise and x-ray scatter. The ability to automatically localize target anatomy in fluoroscopy in near-real-time could be valuable in reducing the occurrence of wrong-site surgery while helping to reduce radiation exposure. The method is applicable beyond the specific case of vertebral labeling, since any structure defined in pre-operative (or intra-operative) CT or cone-beam CT can be automatically registered to the fluoroscopic scene.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
GI and GC metrics have been shown previously to be useful when handling partial content mismatch.
“The GI and GC metrics have been shown previously (Penney et al 1998, Pluim et al 2000, Otake et al 2012b) to be useful when handling partial content mismatch.”
From 3D–2D image registration for target localization in spine surgery: investigation of similarity metrics providing robustness to content mismatch (2016) · cites this paper as Otake et al 2012b
Methods §2.1 says GI is robust to image-content mismatches, including surgical tools visible in fluoroscopy but absent from the preoperative CT. The Introduction identifies GC as a local-intensity metric and describes that category as generally robust in 3D–2D registration, but the paper does not specifically establish GC’s usefulness for partial content mismatch. Page numbers were not provided in our copy. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
The citation lists haven’t been fetched yet
References 53 · Cited by 79
Research works printed but not linked: 40
- Armato SG McLennan G Bidaut L McNitt-Gray MF Meyer CR Reeves AP 2011 The lung image database consortium (LIDC) and image database resource initiative (IDRI): A completed reference database of lung nodules on CT scans AAPM PMC3041807 21452728
- Birkfellner W Stock M Figl M Gendrin C Hummel J Dong S 2009 Stochastic rank correlation: A robust merit function for 2D/3D registration of image data obtained at different energies Medical Physics 36 8 3420 3428 19746775 PMC2855948
- Cabral B Cam N Foran J 1994 Accelerated volume rendering and tomographic reconstruction using texture mapping hardware VVS ‘94: Proceedings of the 1994 Symposium on Volume Visualization 91 98
- Canale ST 2005 Wrong-site surgery: A preventable complication Clinical Orthopaedics and Related Research 433 26 29 15805933
- Fahrig R Holdsworth DW 2000 Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Image-based correction of gantry motion nonidealities Medical Physics 27 1 30 38 10659735
- Fu D Kuduvalli G 2008 A fast, accurate, and automatic 2D-3D image registration for image-guided cranial radiosurgery Medical Physics 35 5 2180 2194 18561693
- Gendrin C Markelj P Pawiro SA Spoerk J Bloch C Weber C 2011 Validation for 2D/3D registration II: The comparison of intensity- and gradient-based merit functions using a new gold standard data set AAPM PMC3089767 21520861
- Gueziec A Kazanzides P Williamson B Taylor RH 1998 Anatomy-based registration of CT-scan and intraoperative X-ray images for guiding a surgical robot Medical Imaging, IEEE Transactions on 17 5 715 728 9874295
- Hansen N Lozano JA Larranaga P Inza I Bengoetxea E 2006 The CMA evolution strategy: A comparing review Towards a new evolutionary computation. advances on estimation of distribution algorithms 75 102 Springer
- Hansen N Niederberger ASP Guzzella L Koumoutsakos P 2009 A method for handling uncertainty in evolutionary optimization with an application to feedback control of combustion Evolutionary Computation, IEEE Transactions on 13 1 180 197
- Hsiang J 2011 Wrong-level surgery: A unique problem in spine surgery Surgical Neurology International 2 47 21660270 PMC3108446
- Hsu W Sciubba DM Sasson AD Khavkin Y Wolinsky JP Gailloud P 2008 Intraoperative localization of thoracic spine level with preoperative percutaneous placement of intravertebral polymethylmethacrylate Journal of Spinal Disorders & Techniques 21 1 72 75 18418141
- Hurvitz A Joskowicz L 2008 Registration of a CT-like atlas to fluoroscopic X-ray images using intensity correspondences International Journal of Computer Assisted Radiology and Surgery 3 6 493 504
- Jin J Ryu S Faber K Mikkelsen T Chen Q Li S 2006 2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions Medical Physics 33 12 4557 4566 17278807
- Khamene A Bloch P Wein W Svatos M Sauer F 2006 Automatic registration of portal images and volumetric CT for patient positioning in radiation therapy Medical Image Analysis 10 1 96 112 16150629
- Kwaan MR Studdert DM Zinner MJ Gawande AA 2006 Incidence, patterns, and prevention of wrong-site surgery Archives of Surgery 141 4 353 358 16618892
- Markelj P Tomaževič D Likar B Pernuš F 2012 A review of 3D/2D registration methods for image-guided interventions Medical Image Analysis 16 3 642 661 20452269
- McLaughlin RA Hipwell J Hawkes DJ Noble JA Byrne JV Cox TC 2005 A comparison of a similarity-based and a feature-based 2-D-3-D registration method for neurointerventional use Medical Imaging, IEEE Transactions on 24 8 1058 1066 16092337
- Mody MG Nourbakhsh A Stahl DL Gibbs M Alfawareh M Garges KJ 2008 The prevalence of wrong level surgery among spine surgeons Spine 33 2 194 198 18197106
- Munbodh R Tagare HD Chen Z Jaffray DA Moseley DJ Knisely JPS 2009 2D-3D registration for prostate radiation therapy based on a statistical model of transmission images Medical Physics 36 10 4555 4568 19928087
- Navab N Bani-Hashemi A Nadar M Wiesent K Durlak P Brunner T Wells W Colchester A Delp S 1998 3D reconstruction from projection matrices in a C-arm based 3D-angiography system Medical image computing and computer-assisted interventation — MICCAI’98 119 129 Springer Berlin/Heidelberg
- Nelder JA Mead R 1965 A simplex method for function minimization The Computer Journal 7 4 308 313
- Otake Y Armand M Armiger RS Kutzer MD Basafa E Kazanzides P 2012 Intraoperative image-based multiview 2D/3D registration for image-guided orthopaedic surgery: Incorporation of fiducial-based C-arm tracking and GPU-acceleration Medical Imaging, IEEE Transactions on 31 4 948 962 PMC4451116 22113773
- Patil S Lindley EM Burger EL Yoshihara H Patel VV 2012 Pedicle screw placement with O-arm and stealth navigation Orthopedics 35 1 e61 5 22229616
- Penney GP Weese J Little JA Desmedt P Hill DLG Hawkes DJ 1998 A comparison of similarity measures for use in 2-D-3-D medical image registration Medical Imaging, IEEE Transactions on 17 4 586 595 9845314
- Pluim JP Maintz JB Viergever MA 2000 Image registration by maximization of combined mutual information and gradient information IEEE Transactions on Medical Imaging 19 8 809 814 11055805
- Qi W Gu L Xu J Dohi T Sakuma I Liao H 2008 Non-rigid 2D-3D registration based on support vector regression estimated similarity metric Medical imaging and augmented reality 339 348 Springer Berlin/Heidelberg
- Russakoff DB Rohlfing T Mori K Rueckert D Ho A Adler JR Jr 2005 Fast generation of digitally reconstructed radiographs using attenuation fields with application to 2D-3D image registration Medical Imaging, IEEE Transactions on 24 11 1441 1454 16279081
- Russakoff DA Rohlfing T Maurer CR Jr 2003 Fast intensity-based 2D-3D image registration of clinical data using light fields Proc. 9th IEEE Int. Conf. Computer Vision (ICCV) 416 423
- Russakoff D Rohlfing T Ho A Kim D Shahidi R Adler J Gee J Maintz J Vannier M 2003 Evaluation of intensity-based 2D-3D spine image registration using clinical gold-standard data Biomedical image registration 151 160 Springer Berlin/Heidelberg
- Sadowsky O 2008 Image registration and hybrid volume reconstruction of bone anatomy using a statistical shape atlas PhD Dissertation the Johns Hopkins University
- Sadowsky O Chintalapani G Taylor RH 2007 Deformable 2D-3D registration of the pelvis with a limited field of view, using shape statistics Medical Image Computing and Computer-Assisted Intervention: MICCAI…International Conference on Medical Image Computing and Computer-Assisted Intervention 10 Pt 2 519 526 18044608
- Shams R Barnes N 2007 Speeding up mutual information computation using NVIDIA CUDA hardware Digital Image Computing Techniques and Applications, 9th Biennial Conference of the Australian Pattern Recognition Society on 555 560
- Siddon RL 1985 Fast calculation of the exact radiological path for a three-dimensional CT array Medical Physics 12 2 252 255 4000088
- Spoerk J Bergmann H Wanschitz F Dong S Birkfellner W 2007 Fast DRR splat rendering using common consumer graphics hardware Medical Physics 34 11 4302 4308 18072495
- Studholme C Hill DLG Hawkes DJ 1997 Automated three-dimensional registration of magnetic resonance and positron emission tomography brain images by multiresolution optimization of voxel similarity measures Medical Physics 24 1 25 35 9029539
- Tomazevic D Likar B Slivnik T Pernus F 2003 3-D/2-D registration of CT and MR to X-ray images Medical Imaging, IEEE Transactions on 22 11 1407 1416 14606674
- Tonn JC Schichor C Schnell O Zausinger S Uhl E Morhard D 2011 Intraoperative computed tomography Acta Neurochirurgica Supplement 109 163 167 20960337
- Upadhyaya CD Wu J Chin CT Balamurali G Mummaneni PV 2012 Avoidance of wrong-level thoracic spine surgery: Intraoperative localization with preoperative percutaneous fiducial screw placement Journal of Neurosurgery: Spine 16 3 280 284 22054638
- van de Kraats EB Penney GP Tomazevic D van Walsum T Niessen WJ 2005 Standardized evaluation methodology for 2-D-3-D registration IEEE Transactions on Medical Imaging 24 9 1177 1189 16156355
- Authors
- Otake, Y., Schafer, S., Stayman, J. W., Zbijewski, W., Kleinszig, G., Graumann, R., Khanna, A. J., & Siewerdsen, J. H.
- Affiliations
- Johns Hopkins University (US)
- Siemens Healthineers (Germany) (DE)
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2012-08-03
- Volume
- 57
- Issue
- 17
- Pages
- 5485-5508
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- PMC
- PMC3429949
- OpenAlex
- W2070914374
- Cited by
- 79
- References
- 53
- Citations by year
- 2026: 2 · 2025: 1 · 2024: 4 · 2023: 8 · 2022: 7 · 2021: 6 · 2020: 4 · 2019: 4 · 2018: 8 · 2017: 10 · 2016: 5 · 2015: 11 · 2014: 6 · 2013: 2 · 2012: 1
- FWCI
- 3.65
- Type
- journal article
- Language
- en
- Open access
- green
- europepmc.org
- Locations
- Physics in Medicine and Biology (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Johns Hopkins University
- National Institutes of Health
- National Cancer Institute (R01CA127444)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Health Sciences › Medicine › Surgery › Spinal Fractures and Fixation Techniques
- Keywords
- wrong-site surgery, X-ray fluoroscopy, 2D-3D registration, spinal surgery, surgical decision support, anthropomorphic chest phantom, radiation dose reduction, cone-beam computed tomography
- MeSH terms
- Automation, Fluoroscopy, Humans, Spine, Tomography, X-Ray Computed, Phantoms, Imaging, Imaging, Three-Dimensional, Surgery, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review