3D–2D image registration for target localization in spine surgery: investigation of similarity metrics providing robustness to content mismatch
Open accessAbstract
Abstract In image-guided spine surgery, robust three-dimensional to two-dimensional (3D–2D) registration of preoperative computed tomography (CT) and intraoperative radiographs can be challenged by the image content mismatch associated with the presence of surgical instrumentation and implants as well as soft-tissue resection or deformation. This work investigates image similarity metrics in 3D–2D registration offering improved robustness against mismatch, thereby improving performance and reducing or eliminating the need for manual masking. The performance of four gradient-based image similarity metrics (gradient information (GI), gradient correlation (GC), gradient information with linear scaling (GS), and gradient orientation (GO)) with a multi-start optimization strategy was evaluated in an institutional review board-approved retrospective clinical study using 51 preoperative CT images and 115 intraoperative mobile radiographs. Registrations were tested with and without polygonal masks as a function of the number of multistarts employed during optimization. Registration accuracy was evaluated in terms of the projection distance error (PDE) and assessment of failure modes (PDE > 30 mm) that could impede reliable vertebral level localization. With manual polygonal masking and 200 multistarts, the GC and GO metrics exhibited robust performance with 0% gross failures and median PDE < 6.4 mm (±4.4 mm interquartile range (IQR)) and a median runtime of 84 s (plus upwards of 1–2 min for manual masking). Excluding manual polygonal masks and decreasing the number of multistarts to 50 caused the GC-based registration to fail at a rate of >14%; however, GO maintained robustness with a 0% gross failure rate. Overall, the GI, GC, and GS metrics were susceptible to registration errors associated with content mismatch, but GO provided robust registration (median PDE = 5.5 mm, 2.6 mm IQR) without manual masking and with an improved runtime (29.3 s). The GO metric improved the registration accuracy and robustness in the presence of strong image content mismatch. This capability could offer valuable assistance and decision support in spine level localization in a manner consistent with clinical workflow.
AI snapshot
Written from the paper’s full text.
TL;DR
This study evaluated four gradient-based similarity metrics for 3D–2D CT-to-radiograph registration in spine surgery under image content mismatch. Gradient orientation (GO) was most robust, maintaining 0% gross failures without manual polygonal masking and with shorter runtime.
Key findings
- With manual polygonal masking and 200 multistarts, GC and GO achieved 0% gross failures and median projection distance error (PDE) < 6.4 mm (±4.4 mm IQR), with median runtime of 84 s plus 1–2 min for manual masking.
- Without manual polygonal masks and with 50 multistarts, GC failed at >14%, while GO maintained 0% gross failure, median PDE = 5.5 mm (2.6 mm IQR), and runtime = 29.3 s.
- GI and GS were susceptible to content mismatch; in the R(p, 200) configuration they had >16% failure, though GS improved gross failure rate over GI by approximately 50%. In R(r, 50), GC gross failure increased to 15%.
- GO had the largest median capture range in the superior–inferior direction: 375 mm ± 122 mm IQR, versus GI 192 mm ± 214 mm, GC 345 mm ± 185 mm, and GS 293 mm ± 180 mm.
Methods
- Institutional review board-approved retrospective clinical study using 51 preoperative CT images and 115 intraoperative mobile radiographs from thoracolumbar spine surgery; radiographs included 105 lateral and 10 anterior–posterior projections.
- Four gradient-based similarity metrics were tested: gradient information (GI), gradient correlation (GC), gradient information with linear scaling (GS), and gradient orientation (GO), using multistart optimization with CMA-ES.
- Registration performance was evaluated under three configurations—R(p, 200), R(p, 50), and R(r, 50)—using PDE, failure defined as PDE > 30 mm, runtime, objective-function search-space quality, and sensitivity to superior–inferior initialization.
Limitations
- The paper states that spine deformation is not specifically addressed by the rigid registration approach, and that ongoing clinical evaluation is needed.
- The paper states that manual polygonal masking is time-consuming, subjective, error-prone, and disruptive to workflow; GS and modified GC also require parameters that are non-trivial to determine automatically.
- Evident from the text: the study is retrospective and single-institution, uses mostly lateral radiographs, and relies on neuroradiologist-annotated vertebral centroids as reference truth.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
6 of 18 cited sources checked · 3 supported · 0 flagged · accuracy not yet measured
- Premisecitation
In previous work on vertebral level localization in spine surgery, the challenge due to content mismatch was mitigated by manually defined polygonal masks that excluded extraneous regions during similarity metric evaluation.
“However, the challenge due to content mismatch was mitigated by manually defined polygonal masks that excluded extraneous regions during similarity metric evaluation.”
Quote in the paperSubstantial claim1 of 2 sources checked- cites Lo et al 2015
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Otake et al 2015 source
The Patient study section (page unavailable) says a weighting mask can down-weight regions with deformation or instrumentation so they contribute less to the similarity metric. It says such masks could potentially be made manually, but does not describe polygonal masks or establish that manually defined masks were used to mitigate content mismatch in this study. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
GI and GC metrics have been shown previously to be useful when handling partial content mismatch.
“The GI and GC metrics have been shown previously (Penney et al 1998, Pluim et al 2000, Otake et al 2012b) to be useful when handling partial content mismatch.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites Penney et al 1998 source
- cites Pluim et al 2000 source
- cites Otake et al 2012b source
Methods §2.1 says GI is robust to image-content mismatches, including surgical tools visible in fluoroscopy but absent from the preoperative CT. The Introduction identifies GC as a local-intensity metric and describes that category as generally robust in 3D–2D registration, but the paper does not specifically establish GC’s usefulness for partial content mismatch. Page numbers were not provided in our copy. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
Gradient-based metrics have shown promise in 3D-2D image registration problems involving intraoperative radiographs and preoperative CT images.
“From the similarity metrics that have been previously investigated, gradient-based metrics have shown promise in 3D–2D image registration problems involving intraoperative radiographs and preoperative CT images (Penney et al 1998, Gendrin et al 2011, Otake et al 2012b, 2015).”
Quote in the paperSubstantial claim1 of 3 sources checked - Premisecitation
Previous work on vertebral level localization in spine surgery showed promising results with GC similarity.
“For example, previous work on vertebral level localization in spine surgery (Lo et al 2015, Otake et al 2015) showed promising results with GC similarity.”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites Lo et al 2015
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites Otake et al 2015 source
- Findingown data
GO exhibited the best overall performance in capture range, with a median capture range of 375 mm ± 122 mm IQR measured at the detector.
“GO exhibited the best overall performance with a median capture range of 375 mm ± 122 mm IQR measured at the detector.”
Quote in the paperSubstantial claim
- Findingown data
Overall, the GI, GC, and GS metrics were susceptible to registration errors associated with content mismatch.
“Overall, the GI, GC, and GS metrics were susceptible to registration errors associated with content mismatch,”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
- Findingown data
The clinical data exhibited a substantial degree of realistic variation in imaging protocols/techniques, image quality, and imaging systems used for preoperative and intraoperative imaging.
“These data exhibited a substantial degree of realistic variation in imaging protocols/techniques, image quality, and imaging systems used for preoperative and intraoperative imaging.”
Quote in the paperSubstantial claim - Findingown data
For all similarity metrics, the search space exhibits numerous false local optima associated with surgical instruments present in the radiograph but not the CT or the semiperiodic structure of the spine in the S-I direction.
“For all the metrics, the space exhibits numerous false local optima associated with surgical instruments present in the radiograph (but not the CT) or the semiperiodic structure of the spine itself in the S–I direction.”
Quote in the paperSubstantial claim - Findingown data
In the R(p, 50) configuration, the GC and GO metrics maintained superior performance compared to GI and GS.
“The GC and GO metrics maintained superior performance in comparison to GI and GS.”
Quote in the paperSubstantial claim - Findingown data
For the R(p, 200) configuration, both the GC and GO similarity metrics exhibited accurate and robust registration performance.
“For the R(p, 200) configuration (i.e. 2D polygonal masks and 200 multistarts), both the GC and GO metrics exhibited an accurate and robust performance.”
Quote in the paperSubstantial claim - Findingown data
For the GI, GS, and GC metrics compared to the GO metric, the PDE distribution was statistically significantly different (p < 0.05) in the cases marked in table 2.
“Comparing the GI, GS, and GC metrics to the GO metric, table 2 marks the cases (*) for which the PDE distribution was statistically significantly different (p < 0.05).”
Quote in the paperSubstantial claim - Findingown data
The GI and GS similarity metrics failed to achieve a clinically acceptable level of accuracy, with a failure rate greater than 16%.
“The GI and GS metrics failed to achieve a clinically acceptable level of accuracy, suffering a >16% failure rate”
Quote in the paperSubstantial claim - Findingown data
The GS metric improved the gross failure rate compared to the GI metric by approximately 50%.
“although GS improved the gross failure rate compared to GI by approximately 50%.”
Quote in the paperNot a substantial claim - Findingown data
The improvements in robustness observed with 200 multistarts come with approximately a three-fold increase in computation time.
“Thus, the improvements in robustness observed with 200 multistarts come with approximately three-fold increase in computation time.”
Quote in the paperNot a substantial claim - Findingown data
Reducing the number of multistarts to 50 in the R(p, 50) configuration improves runtime.
“Reducing to 50 multistarts in the R(p, 50) configuration (as shown in figure 4(b)) improves the runtime”
Quote in the paperNot a substantial claim - Findingown data
In the similarity metric maps, gradients produced by instrumentation appearing in the radiograph but not the CT tend to dominate the similarity metric value for the GI, GC, and GS metrics.
“the gradients produced by instrumentation appearing in the radiograph (but not the CT) tend to dominate the similarity metric value for the GI, GC, and GS metrics.”
Quote in the paperSubstantial claim - Findingown data
Under the same conditions of no manual polygonal masks and 50 multistarts, GO maintained robustness with a 0% gross failure rate.
“however, GO maintained robustness with a 0% gross failure rate.”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
- Findingown data
Excluding manual polygonal masks and decreasing the number of multistarts to 50 caused the GC-based registration to fail at a rate of >14%.
“Excluding manual polygonal masks and decreasing the number of multistarts to 50 caused the GC-based registration to fail at a rate of >14%;”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
- Findingown data
A modified GC metric that truncates β percentiles from the top-most and bottom-most values prior to GC summation showed improved overall performance compared to conventional untruncated GC.
“While the modified metric showed an improved overall performance compared to conventional (untruncated) GC”
Quote in the paperSubstantial claim - Findingreasoning
Due to its ability to handle mismatch, GO yielded comparable performance when the number of multistarts was decreased from 200 to 50.
“Due to its ability to handle mismatch, GO yielded a comparable performance when the number of multistarts was decreased from 200 to 50.”
Quote in the paperSubstantial claim - Findingnone given
The paper extended the LevelCheck 3D-2D registration algorithm to incorporate an improved objective function that exhibits stronger robustness to the image content mismatch observed in real clinical images.
“In conclusion, we have extended the LevelCheck 3D–2D registration algorithm to incorporate an improved objective function that exhibits stronger robustness to the image content mismatch observed in real clinical images.”
Quote in the paperSubstantial claim - Findingown data
Increasing the number of multistarts from 50 to 200 reduced the failure rate (PDE > 30 mm).
“The benefit of increasing the number of multistarts is evident in table 2 (column R(p, 200) relative to R(p, 50)), showing a reduction in failure rate (i.e. PDE > 30 mm).”
Quote in the paperSubstantial claim - Findingown data
Increasing the number of multistarts did not necessarily improve the precision of registration, as observed in the median ± IQR PDE.
“However, increasing the number of multistarts did not necessarily improve the precision of the registration, as observed in the median ± IQR PDE”
Quote in the paperSubstantial claim - Interpretationreasoning
The min operator applied in the GI and GS metrics imparts a flattening effect in the search space and can challenge the optimizer to reach the solution.
“The min operator applied in these metrics imparts a flattening effect in the search space and can challenge the optimizer to reach the solution.”
Quote in the paperSubstantial claim - Interpretationreasoning
The RGO(r, 50) configuration presents an implementation advantageous with respect to clinical workflow via a simple, potentially automatic definition of a rectangular mask.
“and presents an implementation that is advantageous with respect to clinical workflow (via a simple, potentially automatic definition of a rectangular mask).”
Quote in the paperNot a substantial claim - Interpretationown data
The challenging configuration RGO(r, 50) produces a similar PDE distribution to the more manually intensive RGO(p, 200) and RGO(p, 50) configurations.
“This suggests that the challenging configuration RGO(r, 50) produces a similar PDE distribution to the more manually intensive RGO(p, 200) and RGO(p, 50) configurations.”
Quote in the paperSubstantial claim - Interpretationreasoning
Unless extraneous gradients are masked from the image, the GC registration is susceptible to failure.
“Therefore, unless these extraneous gradients are masked from the image (as with the manual p mask as in previous work (Lo et al 2015), the registration is susceptible to failure.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites Lo et al 2015
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
The main advantage of GO is that both instrumentation and vertebrae gradients contribute similar weights to the overall metric value.
“The main advantage of GO is that both the instrumentation and vertebrae gradients contribute similar weights to the overall metric value.”
Quote in the paperSubstantial claim - Interpretationreasoning
As long as vertebral bodies produce a larger spatial distribution of gradients than extraneous objects, the desired solution of accurately aligning vertebrae is more favorable within the GO search space.
“As long as vertebral bodies produce a larger spatial distribution of gradients in comparison to those produced by the extraneous objects, the desired solution of accurately aligning vertebrae is more favorable within the GO search space.”
Quote in the paperSubstantial claim - Interpretationreasoning
The slight improvement in GO registration for R(r, 50) compared with R(p, 50) is attributed to additional relevant anatomical information included with rectangle masking but masked out with polygonal masking.
“A slight improvement in the GO registration when compared with R(p, 50) is attributed to the additional benefit of relevant anatomical information included with rectangle masking but masked out with polygonal masking.”
Quote in the paperNot a substantial claim - Interpretationreasoning
The GO capture range spans approximately 12 vertebrae in the S-I direction.
“This capture range spans approximately 12 vertebra in the S–I direction.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Image content mismatch caused by surgical instruments introduced between the preoperative CT and intraoperative radiographs can be problematic in achieving robust registration.
“Image content mismatch caused by surgical instruments introduced between the preoperative CT and intraoperative radiographs can be problematic in achieving robust registration.”
Quote in the paperSubstantial claim - Premisenone given
GI and GC metrics have been developed with the inherent capacity to handle registrations with partial mismatch.
“while GI and GC metrics have been developed with the inherent capacity to handle registrations with partial mismatch.”
Quote in the paperSubstantial claim - Premisecitation
Gradient information (GI), gradient correlation (GC), and gradient difference (GD) have shown superior performance in 3D-2D registration applications in previous studies.
“Gradient information (GI), gradient correlation (GC), and gradient difference (GD) have shown superior performance in 3D–2D registration applications in previous studies”
Quote in the paperSubstantial claim - Premisereasoning
The assumption that anatomical structures produce a constant gradient magnitude in both DRRs and radiographs is often not valid, since extraneous objects and differences in real/simulated projections may contribute to the similarity metric value.
“However, this assumption is often not valid, since extraneous objects, differences in real/simulated projections may contribute to the similarity metric value.”
Quote in the paperSubstantial claim - Premisenone given
In image-guided spine surgery, robust 3D-2D registration of preoperative CT and intraoperative radiographs can be challenged by image content mismatch associated with surgical instrumentation, implants, and soft-tissue resection or deformation.
“In image-guided spine surgery, robust three-dimensional to two-dimensional (3D–2D) registration of preoperative computed tomography (CT) and intraoperative radiographs can be challenged by the image content mismatch associated with the presence of surgical instrumentation and implants as well as soft-tissue resection or deformation.”
Quote in the paperSubstantial claim - Interpretationcitation
The performance of the GC metric is consistent with that reported in Lo et al 2015.
“The performance exhibited by the GC metric is consistent with that in (Lo et al 2015).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Lo et al 2015
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisenone given
The image similarity metric is an important component within the registration framework that plays an important role in the robustness of registration in the presence of content mismatch.
“The image similarity metric is an important component within the registration framework that plays an important role in the robustness of registration in the presence of content mismatch.”
Quote in the paperSubstantial claim - Interpretationown data
The capture range results suggest a high level of robustness to S-I initialization error, particularly considering the ability to move through multiple false local maxima associated with one-level error, two-level error, and so on.
“The result is encouraging in that it suggests a high level of robustness to S–I initialization error, particularly considering the ability to move through multiple false local maxima associated with one-level error, two-level error, and so on.”
Quote in the paperSubstantial claim - Interpretationreasoning
For the GI and GC metrics, the dominance of instrumentation gradients tends to direct registration to a false erroneous solution.
“For the GI and GC metrics, this tends to direct the registration to a false (erroneous) solution.”
Quote in the paperSubstantial claim - Interpretationreasoning
Our results indicate that similarity metrics that rely upon gradient magnitudes tend to be more susceptible to registration failure under image content mismatch.
“Our results indicate that similarity metrics that rely upon gradient magnitudes tend to be more susceptible to registration failure under such conditions.”
Quote in the paperSubstantial claim - Interpretationreasoning
The susceptibility of gradient-magnitude metrics may be attributed to extraneous objects producing stronger gradient magnitudes than vertebrae, which can dominate the search space and drive the solution to false local maxima.
“This may be attributed to the fact that the extraneous objects produce stronger gradient magnitudes than those of the vertebrae and therefore can dominate the search space and drive the solution to false local maxima.”
Quote in the paperSubstantial claim - Interpretationreasoning
The modified GC metric suffered from the same limitation of automatically determining the parameter β that is dependent on the degree of image content mismatch.
“it suffered from the same limitation of automatically determining the parameter β that is dependent on the degree of image content mismatch.”
Quote in the paperSubstantial claim - Interpretationreasoning
The slight improvement in registration accuracy for RGO(r, 50) compared to RGO(p, 50) could be attributable to additional image content available with the simple rectangular mask.
“The slight improvement in registration accuracy for RGO(r, 50) compared to RGO(p, 50) could be attributable to this additional image content available with the simple rectangular mask.”
Quote in the paperNot a substantial claim - Interpretationnone given
Complex polygonal masks about surgical implants within vertebrae can partially occlude vertebral boundaries (cortical margins) that are salient features for registration.
“complex polygonal masks about surgical implants within verte-brae can partially occlude vertebral boundaries (cortical margins) that are salient features for registration.”
Quote in the paperSubstantial claim - Interpretationreasoning
One challenge in the design of the GS metric is the selection of the parameter that attempts to match the gradient magnitudes of similar anatomical objects in the DRR and radiograph.
“One challenge in the design of the GS metric is the selection of the parameter that attempts to match the gradient magnitudes of similar anatomical objects in the DRR and radiograph.”
Quote in the paperSubstantial claim - Interpretationreasoning
The rigid registration framework may be sufficient if the labels can be projected in close proximity to the accurate vertebral levels when deformation occurs.
“Given the clinical need to identify the correct vertebral level within approximately 30 mm PDE (i.e. within the correct vertebral body), the rigid registration framework may be sufficient if the labels can be projected in close proximity to the accurate vertebral levels when deformation occurs.”
Quote in the paperSubstantial claim- Rests on an assumption: The clinical need is to identify the correct vertebral level within approximately 30 mm PDE.
- May combine two claims. 34
- Classified as probably not a substantial claim. 13
May combine two claims.
PremisecitationGradient orientation on its own has been widely used for feature matching, object detection, and image registration, and shown promise in robust matching under partial occlusion and view point changes.
“While all of the above metrics rely upon gradient magnitudes, gradient orientation on its own has been widely used for feature matching (Lowe 2004), object detection (Dalal and Triggs 2005), and image registration (De Nigris et al 2012), and shown promise in robust matching under partial occlusion and view point changes.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites Lowe 2004 source
- cites Dalal and Triggs 2005 source
The Abstract describes histograms of oriented gradients used for human detection, supporting the object-detection example attributed to Dalal and Triggs. It does not establish that gradient orientation was used on its own, independent of other gradient information. (LLM, conf=0.82)
Experimental — accuracy not yet measured - cites De Nigris et al 2012 source
Classified as probably not a substantial claim.
PremisecitationThe average thoracolumbar vertebral height in the superior-inferior direction is 22 mm.
“Considering an average thoracolumbar vertebral height (superior–inferior direction) of 22 mm (Busscher et al 2010)”
Quote in the paperSubstantial claim1 of 1 sources checked- cites Busscher et al 2010 source
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenFor the acquired radiographs, the magnification factor at the detector typically ranged from 1.4 to 1.6.
“For the acquired radiographs, the magnification factor at the detector typically ranged from 1.4–1.6.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe GI, GC, and GS metrics demonstrated broad capture ranges: GI 192 mm ± 214 mm IQR, GC 345 mm ± 185 mm IQR, and GS 293 mm ± 180 mm IQR.
“The other metrics demonstrated a fairly broad capture range as well: GI capture range = 192 mm ± 214 mm IQR; GC capture range = 345 mm ± 185 mm IQR; and GS capture range = 293 mm ± 180 mm IQR.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe PDE following initialization ranged from 11.3 to 111.1 mm (median PDE = 43.1 mm, interquartile range 20.9 mm).
“The PDE following initialization ranged from 11.3 to 111.1 mm (median PDE = 43.1 mm, interquartile range 20.9 mm).”
Quote in the paperNot a substantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
May combine two claims.
Findingown dataIn the R(r, 50) configuration, the GO metric maintained geometric accuracy and robustness.
“The GO metric, on the other hand, maintained geometric accuracy and robustness.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe GC metric shows a more distinct optimum at the solution but also exhibits sharp local maxima attributed to strong gradient magnitudes far from the solution from surgical instruments.
“On the other hand, GC shows a more distinct optimum at the solution, but it also exhibits sharp local maxima attributed to strong gradient magnitudes far from the solution (surgical instruments).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the figure 3 simulation, while all metrics align corresponding square and circle elements at the desired solution, the mismatching triangle at the degenerate solution causes the GI and GC metrics to yield a larger relative contribution to the overall metric value.
“While all metrics align corresponding square and circle elements at the desired solution, the mismatching triangle at the degenerate solution causes the GI, GC metrics to yield a larger relative contribution to the overall metric value.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataGO provided robust registration (median PDE = 5.5 mm, 2.6 mm IQR) without manual masking and with an improved runtime (29.3 s).
“but GO provided robust registration (median PDE = 5.5 mm, 2.6 mm IQR) without manual masking and with an improved runtime (29.3 s).”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
May combine two claims.
Findingown dataIn the R(r, 50) configuration, the performance of the GI, GC, and GS metrics degraded significantly, with the gross failure rate for GC increasing to 15%.
“The performance of the GI, GC, and GS metrics degraded significantly, e.g. the gross failure rate for GC increasing to 15%.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor GI and GS, the peak at the solution is less distinct from surrounding local maxima, although GS appears improved over GI.
“For GI and GS, the peak at the solution is less distinct from surrounding local maxima, although GS appears improved over GI (consistent with improved robustness noted above).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe average computation time ranged from 23-29 s for the S = 50 configurations, compared to 65-85 s for the S = 200 configuration.
“the average computation time ranged from 23–29 s for the S = 50 configurations, compared to 65–85 s for the S = 200 configuration.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAmong the RGO(p, 200), RGO(p, 50), and RGO(r, 50) distributions, the Wilcoxon signed rank test found p > 0.94, failing to reject the null hypothesis that the PDE distributions have the same median.
“Among distributions RGO(p, 200), RGO(p, 50), and RGO(r, 50), we found p-value >0.94 failing to reject the null hypothesis that the PDE distributions have the same median.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataFor the GO metric, gradient magnitude values do not dominate the overall metric value, and the desired solution is reached with corresponding vertebrae gradients aligned robustly against extraneous gradients and occlusions from instrumentation.
“For the GO metric, the gradient magnitude values do not dominate the overall metric value, and the desired solution is reached wherein corresponding vertebrae gradients align in a manner that is robust against extraneous gradients and occlusions from instrumentation.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe GO metric exhibits the most distinct global optimum at the solution and suppresses false maxima far from the solution.
“The GO metric exhibits the most distinct global optimum at the solution and suppresses false maxima far from the solution.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataWith manual polygonal masking and 200 multistarts, the GC and GO metrics exhibited robust performance with 0% gross failures and median PDE < 6.4 mm (±4.4 mm interquartile range) and a median runtime of 84 s (plus upwards of 1-2 min for manual masking).
“With manual polygonal masking and 200 multistarts, the GC and GO metrics exhibited robust performance with 0% gross failures and median PDE < 6.4 mm (±4.4 mm interquartile range (IQR)) and a median runtime of 84 s (plus upwards of 1–2 min for manual masking).”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
May combine two claims.
Findingown dataThe RGO(r, 50) configuration outperformed all of the GI, GS, and GC configurations in terms of geometric accuracy, robustness, and run time.
“Overall, the RGO(r, 50) configuration outperformed all of the GI, GS, and GC configurations in terms of geometric accuracy, robustness, and run time”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataReducing the number of multistarts to 50 in the R(p, 50) configuration compromises registration robustness and increases the failure rate for all similarity metrics.
“but compromises the robustness of the registration and results in an increasing failure rate for all of the similarity metrics.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe mm-scale variations in PDE between multistart configurations were not statistically significant (p = 0.52, 0.72, 0.10, and 0.07 for GI, GC, GS, and GO, respectively) and are likely not clinically significant with respect to localizing a particular vertebral body.
“but such mm-scale variations in PDE in table 2 were not statistically significant (p = 0.52, 0.72, 0.10, and 0.07 for GI, GC, GS, and GO, respectively) and are likely not clinically significant with respect to localizing a particular vertebral body.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAlthough the GS metric somewhat improved registration performance compared to GI, it did not exhibit particularly strong robustness to failure (>16%) over all tested configurations.
“Although the GS metric somewhat improved the registration performance compared to GI, it did not exhibit particularly strong robustness to failure (>16%) over all the tested configurations.”
Quote in the paperSubstantial claimMay combine two claims.
Premisenone givenIn spine surgery, sources of mismatch such as surgical instrumentation in the intraoperative radiograph and non-rigid anatomical deformation or positioning differences are common occurrences and present particular challenges to 3D-2D registration.
“In spine surgery, for example, the following sources of mismatch are common occurrences (as shown in figure 1) and present particular challenges to 3D–2D registration: (a) The presence of surgical instrumentation in the intraoperative radiograph (e.g. retractors, screws, implants, clips, etc); and (b) non-rigid anatomical deformation”
Quote in the paperSubstantial claimMay combine two claims.
Premisenone givenAutomatic or semi-automatic segmentation as a pre-processing step to mask extraneous objects increases computational burden and makes the registration algorithm susceptible to segmentation errors.
“However, such an approach increases the computational burden and makes the registration algorithm susceptible to segmentation errors.”
Quote in the paperSubstantial claimMay combine two claims.
Premisenone givenGC can accommodate differences in gradient images up to a linear factor and therefore offers some degree of robustness against mismatch.
“GC can accommodate such differences up to a linear factor and therefore offers some degree of robustness against mismatch:”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe broad capture ranges are partly attributable to the similarity metric and perhaps more so to the robust optimization strategy involving multiple parallel searches from multiple initial seed points.
“Such broad capture ranges are partly attributable to the similarity metric and perhaps more so to the robust optimization strategy involving multiple parallel searches from multiple initial seed points.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisereasoningIncreasing the number of multistarts boosts the density of seed points positioned within the multidimensional space and therefore improves the robustness of finding the global optima at the cost of additional computation.
“Increasing the number of multistarts (denoted S in figure 2) boosts the density of seed points positioned within the multidimensional space and therefore improves the robustness of finding the global optima at the cost of additional computation in performing parallel searches (roughly proportional to the number of multi-starts).”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Premisenone givenThe min operator in GI assumes that anatomical structures produce a constant gradient magnitude in both DRRs and radiographs.
“The min operator in GI assumes that anatomical structures produce a constant gradient magnitude in both DRRs and radiographs.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationImage gradients can filter out low spatial frequency differences such as soft tissue and act as a high-pass filter to focus registration on bony anatomy visible in x-ray images.
“Image gradients can be used to filter out low spatial frequency differences such as soft tissue (Penney et al 1998, Pluim et al 2000) and act as a high-pass filter to focus the registration on the bony anatomy clearly visible in x-ray images.”
Quote in the paperSubstantial claim0 of 2 sources checkedMay combine two claims.
InterpretationreasoningThe GO similarity metric exhibits robust performance against content mismatch and enables the definition of a simple rectangular mask of the radiograph collimators to reduce or eliminate the need for polygonal masking.
“The GO similarity metric thereby exhibits robust performance against content mismatch and enables the definition of a simple rectangular mask of the radiograph collimators to reduce or eliminate the need for polygonal masking.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe GO metric improved the registration accuracy and robustness in the presence of strong image content mismatch.
“The GO metric improved the registration accuracy and robustness in the presence of strong image content mismatch.”
Quote in the paperSubstantial claim- The authors state this as an assumption: The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
May combine two claims.
InterpretationreasoningThe definition of manual polygonal masks is subject to user variability and tends to exclude regions that may be disadvantageous to registration.
“the definition of manual polygonal masks is subject to user variability and tends to exclude regions that may be disadvantageous to registration.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenManual or semi-automatic methods to mask out extraneous or missing content are time consuming, user dependent, error prone, and disruptive to workflow.
“Manual/semi-automatic methods to mask out such extraneous (or missing) content are time consuming, user dependent, error prone, and disruptive to workflow.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
Interpretationown dataThe GO metric capability could offer valuable assistance and decision support in spine level localization in a manner consistent with clinical workflow.
“This capability could offer valuable assistance and decision support in spine level localization in a manner consistent with clinical workflow.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenThe definition of polygons to mask extraneous objects is a slow, subjective step that is error prone and limits the practical utility of image-based registration within streamlined workflow.
“The definition of polygons to mask extraneous objects is a slow, subjective step that is error prone and limits the practical utility of image-based registration within streamlined workflow.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Premisenone givenContent mismatch caused by extraneous objects and tissue deformation can create a challenging search space with multiple local optima.
“Content mismatch caused by extraneous objects and tissue deformation can create a challenging search space with multiple local optima.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningFor the GS metric, the effect of instrumentation gradients is mitigated partly by the scaling parameter α, but excluding extraneous gradients requires estimating α to accurately match corresponding gradients in the two images.
“For the GS metric, this effect is mitigated partly by the scaling parameter α; however, the exclusion of extraneous gradients requires estimating α to accurately match the corresponding gradients in the two images.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe challenging search space shows multiple local maxima resulting from extraneous objects and stands to benefit from the multistart optimization strategy to improve robustness.
“The challenging search space illustrated in figure 6 shows multiple local maxima resulting from extraneous objects and stands to benefit from the multistart optimization strategy to improve robustness.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataDeformation of the spine is not specifically addressed in the current rigid registration approach, but the large clinical dataset presented realistic levels of anatomical deformation.
“Deformation of the spine is not specifically addressed in the current (rigid) registration approach, but the large clinical dataset did present realistic levels of anatomical deformation.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe corresponding increase in robustness from more multistarts has diminishing return due to the curse of dimensionality in the 6D search space, whereas computation time increases approximately linearly with the number of multistarts.
“Moreover, the corresponding increase in robustness has diminishing return due to the ‘curse of dimensionality’ in the 6D search space, whereas the computation time increases approximately linearly with the number of multistarts.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningThe GS parameter was evaluated by computing the ratio of maximum gradient magnitudes observed between the two images, but the parameter should be adjusted according to mismatch caused by strong gradient magnitudes, and automatic determination for any given pair of images is non-trivial.
“We evaluated this parameter by computing the ratio of maximum gradient magnitudes observed between the two images; however, the parameter should be adjusted according to mismatch caused by strong gradient magnitudes. Automatic determination of the parameter for any given pair of images is non-trivial.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Interpretationown dataThe registration framework should rely less upon the number of multistarts, an attribute exhibited with the incorporation of the GO similarity metric.
“Therefore, it is desirable for the registration framework to rely less upon the number of multistarts, an attribute exhibited with the incorporation of the GO similarity metric.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningThese improvements better enable clinical translation of the registration algorithm.
“These improvements better enable clinical translation of the registration algorithm”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThe polygonal masking approach requires the user to anticipate and identify problematic gradients in a radiograph with many subtleties.
“Another drawback of the polygonal masking approach is that it requires the user to anticipate and identify problematic gradients in a radiograph with many subtleties.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningThe small variations in PDE could be attributed to errors in localization of the vertebrae centroid (expert reader truth definition), the stochasticity of the CMA-ES optimizer, and/or fundamental limitations arising from the use of similarity metrics as surrogate measures of registration accuracy during the optimization process.
“The small variations in PDE could be attributed to errors in the localization of the vertebrae centroid (expert reader truth definition), the stochasticity of the CMA-ES optimizer, and/or fundamental limitations arising from the use of similarity metrics as surrogate measures of registration accuracy during the optimization process.”
Quote in the paperNot a substantial claim
The authors state this as an assumption.
The transformation between preoperative CT and intraoperative radiograph can be modeled as a rigid 6D transformation.
“The registration is performed by optimizing the image similarity between the preoperative CT image and the intraoperative radiograph (I2:R2→R) in a rigid 6D transformation space.”
The paper relies on this without support.
Disparity in the superior–inferior (S–I) position is the most susceptible factor for level misidentification in spine registration.
“disparity in the S–I position is believed to be the most susceptible for level misidentification (e.g. registration errors resulting in label positions away from the true level by one or more levels in the S–I direction).”
The paper relies on this without support.
The clinical need is to identify the correct vertebral level within approximately 30 mm PDE.
“Given the clinical need to identify the correct vertebral level within approximately 30 mm PDE (i.e. within the correct vertebral body)”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Elliptical mask's minor axis is longer than its major axis
MinorThe stated minor axis is 50 mm, while the major axis is 25 mm. A minor axis cannot be longer than the major axis.
“minor axis 50 mm”
“major axis 25 mm”
Cites
Distinctive Image Features from Scale-Invariant Keypoints
Lowe, D. G. (2004) · International Journal of Computer Vision · doi:10.1023/b:visi.0000029664.99615.94 · Cited by 55,589 · cited 1× · see citing works
“…(GO): While all of the above metrics rely upon gradient magnitudes, gradient orientation on its own has been widely used for feature matching (Lowe 2004), object detection (Dalal and Triggs 2005), and image registration (De Nigris et al 2012), and shown promise in robust matching under…”
BackgroundHistograms of Oriented Gradients for Human Detection
Dalal, N., & Triggs, B. (2005) · 2005 IEEE Computer Society Conference on Computer Vision and Pattern Recognition (CVPR'05) · doi:10.1109/cvpr.2005.177 · Cited by 33,584 · cited 2× · see citing works
“Gradient orientation (GO): While all of the above metrics rely upon gradient magnitudes, gradient orientation on its own has been widely used for feature matching (Lowe 2004), object detection (Dalal and Triggs 2005), and image registration (De Nigris et al 2012), and shown promise in robust matching under partial occ…”
BackgroundMethodThe CMA Evolution Strategy: A Comparing Review
Hansen, N. (2007) · Studies in fuzziness and soft computing · doi:10.1007/3-540-32494-1_4 · Cited by 1,907 · cited 1× · see citing works
“To achieve robust performance under these conditions, the registration process incorporates a multistart optimization with the covariance matrix adaptation; evolution strategy (CMA-ES) reported by (Hansen 2006).”
MethodA comparison of similarity measures for use in 2-D-3-D medical image registration
Penney, G. P., Weese, J., Little, J. A., Desmedt, P., Hill, D. L. G., & hawkes, D. J. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.730403 · Cited by 782 · see citing works
Image registration by maximization of combined mutual information and gradient information
Pluim, J. P. W., Maintz, J. B. A., & Viergever, M. A. (2000) · IEEE Transactions on Medical Imaging · doi:10.1109/42.876307 · Cited by 670 · see citing works
Towards a New Evolutionary Computation
Lozano, J. A., Larrañaga, P., Inza, I., & Bengoetxea, E. (2006) · Springer eBooks · doi:10.1007/3-540-32494-1 · Cited by 519 · see citing works
Automated model-based vertebra detection, identification, and segmentation in CT images
Klinder, T., Östermann, J., Ehm, M., et al. (2009) · Medical Image Analysis · doi:10.1016/j.media.2009.02.004 · Cited by 331 · cited 1× · see citing works
“Labeling in CT can potentially be done automatically (Klinder et al 2009, Ma and Lu 2013), but in this study, a boardcertified neuroradiologist annotated the vertebral labels at the approximate centroid of each vertebral body.”
Comparative anatomical dimensions of the complete human and porcine spine
Busscher, I., Ploegmakers, J. J. W., Verkerke, G. J., & Veldhuizen, A. G. (2010) · European Spine Journal · doi:10.1007/s00586-010-1326-9 · Cited by 295 · see citing works
The Prevalence of Wrong Level Surgery Among Spine Surgeons
Mody, M. G., Nourbakhsh, A., Stahl, D. L., Gibbs, M., Alfawareh, M. D., & Garges, K. J. (2008) · Spine · doi:10.1097/brs.0b013e31816043d1 · Cited by 221 · cited 1× · see citing works
“This could potentially decrease wrong-level errors, reported to occur in approximately 1 in 3110 spine surgery procedures (Mody et al 2008), and reduce the time and stress associated with resolving anatomical uncertainty.”
BackgroundIntensity-based 2-D-3-D registration of cerebral angiograms
Hipwell, J. H., Penney, G., McLaughlin, R. A., Rhode, K., Summers, P. E., Cox, T. C. S., Byrne, J. V., Noble, J. A., & Hawkes, D. J. (2003) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2003.819283 · Cited by 146 · cited 2× · see citing works
“…Metrics Validation Penney et al (1998) CT to projection x-ray in spine surgery NCC, entropy, MI, GC, PI, GD Phantom + simulated clinical images Hipwell et al (2003) Magnetic resonance angiography (MRA) to digital subtraction angiography (DSA) in neurointerventions NCC, GC, entropy, MI, PI, GD…”
BackgroundKhamene, A., Bloch, P., Wein, W., Svatos, M. M., & Sauer, F. (2005) · Medical Image Analysis · doi:10.1016/j.media.2005.06.002 · Cited by 130 · cited 2× · see citing works
“…for clinical applications ranging from image guidance in the spine (Penney et al 1998) and intracranial interventions (Hipwell et al 2003, McLaughlin et al 2005) to patient positioning/target alignment in radiotherapy (Khamene et al 2006, Kim et al 2007, Wu et al 2009, Gendrin et al 2011).”
BackgroundA protocol for evaluation of similarity measures for rigid registration
Škerl, D., Likar, B., & Pernuš, F. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.874963 · Cited by 93 · cited 1× · see citing works
“The profiles were normalized such that a unit change in all the transform directions (i.e. both rotations and translations) cause the same mean shift in physical dimensions in the image (Škerl et al 2006).”
BackgroundDemons deformable registration of CT and cone‐beam CT using an iterative intensity matching approach
Nithiananthan, S., Schäfer, S., Uneri, A., et al. (2011) · Medical Physics · doi:10.1118/1.3555037 · Cited by 87 · see citing works
Self-citationOtake, Y., Wang, A. S., Webster Stayman, J., Uneri, A., Kleinszig, G., Vogt, S., Khanna, A. J., Gokaslan, Z. L., & Siewerdsen, J. H. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/23/8535 · Cited by 85 · cited 4× · see citing works
“For the multistart strategy, the six-dimensional (6D) search space is partitioned to multiple, equal-size subspaces following a k-d tree partitioning algorithm, and a separate CMA-ES search was performed for each subspace initialized at their respective centers (Otake et al 2013).”
BackgroundMethodInfluentialSelf-citationJun, M., Lü, L., Zhan, Y., et al. (2010) · Lecture notes in computer science · doi:10.1007/978-3-642-15705-9_3 · Cited by 84 · cited 1× · see citing works
“Labeling in CT can potentially be done automatically (Klinder et al 2009, Ma and Lu 2013), but in this study, a boardcertified neuroradiologist annotated the vertebral labels at the approximate centroid of each vertebral body.”
Otake, Y., Schafer, S., Stayman, J. W., Zbijewski, W., Kleinszig, G., Graumann, R., Khanna, A. J., & Siewerdsen, J. H. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/17/5485 · Cited by 79 · cited 10× · see citing works
“From the similarity metrics that have been previously investigated, gradient-based metrics have shown promise in 3D–2D image registration problems involving intraoperative radiographs and preoperative CT images (Penney et al 1998, Gendrin et al 2011, Otake et al 2012b, 2015).”
BackgroundMethodInfluentialSelf-citationMulti-modal image registration based on gradient orientations of minimal uncertainty
De Nigris, D., Collins, D. L., & Arbel, T. (2012) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2012.2218116 · Cited by 73 · cited 1× · see citing works
“…metrics rely upon gradient magnitudes, gradient orientation on its own has been widely used for feature matching (Lowe 2004), object detection (Dalal and Triggs 2005), and image registration (De Nigris et al 2012), and shown promise in robust matching under partial occlusion and view point changes.”
MethodJun, M., & Lü, L. (2013) · Computer Vision and Image Understanding · doi:10.1016/j.cviu.2012.11.016 · Cited by 66 · cited 1× · see citing works
“Labeling in CT can potentially be done automatically (Klinder et al 2009, Ma and Lu 2013), but in this study, a boardcertified neuroradiologist annotated the vertebral labels at the approximate centroid of each vertebral body.”
Wu, J., Kim, M., Peters, J., et al. (2009) · Medical Physics · doi:10.1118/1.3250843 · Cited by 64 · cited 3× · see citing works
“…for clinical applications ranging from image guidance in the spine (Penney et al 1998) and intracranial interventions (Hipwell et al 2003, McLaughlin et al 2005) to patient positioning/target alignment in radiotherapy (Khamene et al 2006, Kim et al 2007, Wu et al 2009, Gendrin et al 2011).”
BackgroundMethodMcLaughlin, R. A., Hipwell, J., Hawkes, D. J., Noble, J. A., Byrne, J. V., & Cox, T. C. (2005) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2005.852067 · Cited by 63 · cited 3× · see citing works
“While the amount of clinical data used for validation in the studies listed in table 1 is limited, clinical images often pose additional challenges for similarity metrics during registration (Penney et al 1998, McLaughlin et al 2005, Wu et al 2009).”
BackgroundMethodUneri, A., Nithiananthan, S., Schäfer, S., et al. (2012) · Medical Physics · doi:10.1118/1.4767757 · Cited by 43 · cited 1× · see citing works
“The development and validation of image registration methods (Nithiananthan et al 2011, Uneri et al 2013) that are robust in the presence of content mismatch are important to achieving reliable registration in a manner consistent with the needs of intraoperative workflow and clinical integration.”
MethodSelf-citation3D–2D registration in mobile radiographs: algorithm development and preliminary clinical evaluation
Otake, Y., Wang, A. S., Uneri, A., Kleinszig, G., Vogt, S., Aygun, N., Lo, S.-F. L., Wolinsky, J.-P., Gokaslan, Z. L., & Siewerdsen, J. H. (2015) · Physics in Medicine & Biology · doi:10.1088/0031-9155/60/5/2075 · Cited by 41 · cited 7× · see citing works
“From the similarity metrics that have been previously investigated, gradient-based metrics have shown promise in 3D–2D image registration problems involving intraoperative radiographs and preoperative CT images (Penney et al 1998, Gendrin et al 2011, Otake et al 2012b, 2015).”
BackgroundResultInfluentialSelf-citationGendrin, C., Markelj, P., Pawiro, S. A., Spoerk, J., Bloch, C., Weber, C., Figl, M., Bergmann, H., Birkfellner, W., Likar, B., et al. (2011) · Medical Physics · doi:10.1118/1.3553403 · Cited by 38 · see citing works
Automatic Localization of Target Vertebrae in Spine Surgery
Lo, S.-F. L., Otake, Y., Puvanesarajah, V., et al. (2015) · Spine · doi:10.1097/brs.0000000000000814 · Cited by 32 · see citing works
Self-citationComparison of Similarity Measures for Rigid-body CT/Dual X-ray Image Registrations
Kim, J., Kim, J., Li, S., et al. (2007) · Technology in Cancer Research & Treatment · doi:10.1177/153303460700600411 · Cited by 24 · cited 3× · see citing works
“…for clinical applications ranging from image guidance in the spine (Penney et al 1998) and intracranial interventions (Hipwell et al 2003, McLaughlin et al 2005) to patient positioning/target alignment in radiotherapy (Khamene et al 2006, Kim et al 2007, Wu et al 2009, Gendrin et al 2011).”
BackgroundStatistical Segmentation of Surgical Instruments in 3-D Ultrasound Images
Linguraru, M. G., Vasilyev, N. V., del Nido, P. J., et al. (2007) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2007.03.003 · Cited by 22 · cited 1× · see citing works
“Alternatively, automatic/semi-automatic segmentation methods (Linguraru et al 2007) could be implemented as a pre-processing step to mask extraneous objects within the field-of-view of the radiograph.”
MethodOtake, Y., Schafer, S., Stayman, J. W., Zbijewski, W. B., Kleinszig, G., Graumann, R., Khanna, A. J., & Siewerdsen, J. H. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.911308 · Cited by 11 · cited 10× · see citing works
“From the similarity metrics that have been previously investigated, gradient-based metrics have shown promise in 3D–2D image registration problems involving intraoperative radiographs and preoperative CT images (Penney et al 1998, Gendrin et al 2011, Otake et al 2012b, 2015).”
BackgroundMethodInfluentialSelf-citation
Flagged first, then by citations
Cited by
Sorted by citations
Deep learning in medical image registration: a survey
Haskins, G., Kruger, U., & Yan, P. (2020) · Machine Vision and Applications · doi:10.1007/s00138-020-01060-x · Cited by 704 · see citing works
“They were able to significantly outperform registration approaches that used a state-of-the-art similarity metric given by [24].”
Deep learning in medical image registration: a review
Fu, Y., Lei, Y., Wang, T., Curran, W. J., Liu, T., & Yang, X. (2020) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ab843e · Cited by 661 · see citing works
Dilated FCN for Multi-Agent 2D/3D Medical Image Registration
Miao, S., Piat, S., Fischer, P., Tuysuzoglu, A., Mewes, P., Mansi, T., & Liao, R. (2018) · Proceedings of the AAAI Conference on Artificial Intelligence · doi:10.1609/aaai.v32i1.11576 · Cited by 113 · see citing works
Colonoscopy 3D video dataset with paired depth from 2D-3D registration
Bobrow, T. L., Golhar, M., Vijayan, R. C., et al. (2023) · Medical Image Analysis · doi:10.1016/j.media.2023.102956 · Cited by 113 · see citing works
“Depth-based registrations were optimized using L 1 , L 2 , Normalized Cross Correlation (NCC), and Gradient Correlation (GC) loss functions (De Silva et al., 2016).”
MethodMultiview 2D/3D Rigid Registration via a Point-Of-Interest Network for Tracking and Triangulation
Liao, H., Lin, W.-A., Zhang, J., Zhang, J., Luo, J., & Zhou, S. K. (2019) · 2019 IEEE/CVF Conference on Computer Vision and Pattern Recognition (CVPR) · doi:10.1109/cvpr.2019.01292 · Cited by 66 · see citing works
Han, R., Uneri, A., Vijayan, R., et al. (2020) · Medical Image Analysis · doi:10.1016/j.media.2020.101917 · Cited by 64 · see citing works
“For example, image registration between preoperative CT and 2D fluoroscopy have been proposed for target localization and guidance in spine surgery (De Silva et al., 2016; Uneri et al., 2014).”
BackgroundSelf-citationMedical image registration using unsupervised deep neural network: A scoping literature review
Abbasi, S., Tavakoli, M., Boveiri, H. R., Mosleh‐Shirazi, M. A., Khayami, R., Khorasani, H., Javidan, R., & Mehdizadeh, A. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.103444 · Cited by 63 · see citing works
“4 is vital procedure in different clinical application such as image-guided [5, 6], motion tracking [7], segmentation [8, 9], dose accumulation [10, 11, 12], image reconstruction [13, 14].”
BackgroundAtlas-based automatic planning and 3D–2D fluoroscopic guidance in pelvic trauma surgery
Han, R., Uneri, A., Silva, T. D., Ketcha, M. D., Goerres, J., Vogt, S. K., Kleinszig, G., Osgood, G. M., & Siewerdsen, J. H. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab1456 · Cited by 57 · see citing works
Self-citationAMNet: Adaptive multi-level network for deformable registration of 3D brain MR images
Che, T., Wang, X., Zhao, K., Zhao, Y., Zeng, D., Li, Q., Zheng, Y., Yang, N., Wang, J., & Li, S. (2023) · Medical Image Analysis · doi:10.1016/j.media.2023.102740 · Cited by 52 · see citing works
Hajek, J., Unberath, M., Fotouhi, J., Bier, B., Lee, S. C., Osgood, G. M., Maier, A., Armand, M., & Navab, N. (2018) · Lecture notes in computer science · doi:10.1007/978-3-030-00937-3_35 · Cited by 52 · see citing works
“the VT C describing the transformation from 3D image volumes to an intra-operatively acquired X-ray image. For pre-operative data, VT C can be estimated via image-based 3D/2D registration, e.g. as in [9,10]. If the Carm is CBCT capable and the 3D volume is acquired intra-procedurally, VT C is known and can be dened as…”
Method3D/2D model-to-image registration by imitation learning for cardiac procedures
Toth, D. F., Miao, S., Kurzendorfer, T., Rinaldi, C. A., Liao, R., Mansi, T., Rhode, K., & Mountney, P. (2018) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-018-1774-y · Cited by 51 · see citing works
Uneri, A., Silva, T. D., Goerres, J., Jacobson, M. W., Ketcha, M. D., Reaungamornrat, S., Kleinszig, G., Vogt, S. K., Khanna, A. J., Osgood, G. M., et al. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa62c5 · Cited by 42 · see citing works
Self-citationSilva, T. D., Chew, E. Y., Hotaling, N. A., & Cukras, C. A. (2020) · Biomedical Optics Express · doi:10.1364/boe.408573 · Cited by 42 · see citing works
Self-citationAn application independent review of multimodal 3D registration methods
Saiti, E., & Theoharis, T. (2020) · Computers & Graphics · doi:10.1016/j.cag.2020.07.012 · Cited by 40 · see citing works
“While the method achieves a high robustness and out-performs approaches that use the state-of-the-art similarity metric of [218] , registration accuracy remains challenging.”
MethodCorrection of patient motion in cone-beam CT using 3D–2D registration
Ouadah, S., Jacobson, M. W., Stayman, J. W., Ehtiati, T., Weiss, C. R., & Siewerdsen, J. H. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa9254 · Cited by 39 · see citing works
Self-citationKetcha, M. D., Silva, T. D., Uneri, A., Jacobson, M. W., Goerres, J., Kleinszig, G., Vogt, S. K., Wolinsky, J.-P., & Siewerdsen, J. H. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa6b3e · Cited by 37 · see citing works
“…provides reasonable registration accuracy overall, even in the presence of deformation (Otake et al 2013); however, recent clinical studies (De Silva et al 2016b) found a reduction in surgical confidence arising even from fairly small errors for which the label was still within cortical…”
BackgroundMethodResultInfluentialSelf-citationImage-to-Graph Convolutional Network for 2D/3D Deformable Model Registration of Low-Contrast Organs
Nakao, M., Nakamura, M., & Matsuda, T. (2022) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2022.3194517 · Cited by 37 · see citing works
“Optimization-based 2D/3D registration has been extensively researched over the past 20 years [6], [34]–[37].”
BackgroundRegistration of MRI to intraoperative radiographs for target localization in spinal interventions
Silva, T. D., Uneri, A., Ketcha, M. D., Reaungamornrat, S., Goerres, J., Jacobson, M. W., Vogt, S. K., Kleinszig, G., Khanna, A. J., Wolinsky, J.-P., et al. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/62/2/684 · Cited by 29 · see citing works
Self-citationEsfandiari, H., Anglin, C., Guy, P., Street, J., Weidert, S., & Hodgson, A. J. (2019) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-019-02024-x · Cited by 27 · see citing works
Han, R., Uneri, A., Ketcha, M. D., et al. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab843c · Cited by 28 · see citing works
Self-citationA Review of Keypoints’ Detection and Feature Description in Image Registration
Liu, C., Xu, J., & Wang, F. (2021) · Scientific Programming · doi:10.1155/2021/8509164 · Cited by 47 · see citing works
“To FAST, AOS (additive operator splitting) is exploited to accelerate calculation [54].”
MethodSimilarity attention-based CNN for robust 3D medical image registration
Zhu, F., Wang, S., Li, D., & Li, Q. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.104403 · Cited by 26 · see citing works
Utility of the LevelCheck Algorithm for Decision Support in Vertebral Localization
Silva, T. D., Lo, S.-F. L., Aygün, N., et al. (2016) · Spine · doi:10.1097/brs.0000000000001589 · Cited by 25 · see citing works
Self-citationGoerres, J., Uneri, A., Jacobson, M. W., Ramsay, B., Silva, T. D., Ketcha, M. D., Han, R., Manbachi, A., Vogt, S. K., Kleinszig, G., et al. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa954f · Cited by 24 · see citing works
Self-citationVirtual fluoroscopy for intraoperative C-arm positioning and radiation dose reduction
Silva, T. D., Punnoose, J., Uneri, A., et al. (2018) · Journal of medical imaging · doi:10.1117/1.jmi.5.1.015005 · Cited by 24 · see citing works
“First, patient position is estimated by registering a single radiograph to the CT image via a 3D–2D image registration algorithm that has been shown in previous work to be robust over a wide variety of clinical imaging scenarios, including the presence of surgical instrumentation and soft-tissue deformation.(8,9,13,14…”
MethodSelf-citationSpineCloud: image analytics for predictive modeling of spine surgery outcomes
Silva, T. D., Vedula, S. S., Perdomo‐Pantoja, A., Vijayan, R. C., Doerr, S. A., Uneri, A., Han, R., Ketcha, M. D., Skolasky, R. L., Witham, T. F., et al. (2020) · Journal of medical imaging · doi:10.1117/1.jmi.7.3.031502 · Cited by 23 · see citing works
“26,27 31 Example image-based features include automatic labeling of vertebral levels, global spinal alignment, endplate angles, intervertebral space, number of levels treated, and measurements of surgical instrumentation.”
BackgroundSelf-citationDong, G., Dai, J., Li, N., Zhang, C., He, W., Liu, L., Chan, Y., Li, Y., Xie, Y., & Liang, X. (2023) · Bioengineering · doi:10.3390/bioengineering10020144 · Cited by 24 · see citing works
C-arm positioning using virtual fluoroscopy for image-guided surgery
de Silva, T., Punnoose, J., Uneri, A., et al. (2017) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2256028 · Cited by 23 · see citing works
“We utilize robust 3D-2D registration approaches[4]–[7] previously developed and validated in clinical images [8], [9].”
Self-citationCo-localized augmented human and X-ray observers in collaborative surgical ecosystem
Fotouhi, J., Unberath, M., Song, T., Hajek, J., Lee, S. C., Bier, B., Maier, A., Osgood, G. M., Armand, M., & Navab, N. (2019) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-019-02035-8 · Cited by 19 · see citing works
Mi, J., Yin, W., Zhao, L., Chen, Y., Zhou, Y., & Feng, Q. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acd29d · Cited by 21 · see citing works
Robotic drill guide positioning using known-component 3D–2D image registration
Yi, T. Y., Ramchandran, V., Siewerdsen, J. H., et al. (2018) · Journal of medical imaging · doi:10.1117/1.jmi.5.2.021212 · Cited by 19 · see citing works
Self-citationUnsupervised Cross-Modality Domain Adaptation Network for X-Ray to CT Registration
Zheng, S., Yang, X. N., Wang, Y., Ding, M., & Hou, W. W. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3135890 · Cited by 18 · see citing works
Metrics for Assessing the Similarity of Microwave Breast Imaging Scans of Healthy Volunteers
Lavoie, B. R., Bourqui, J., Fear, E. C., et al. (2018) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2018.2806878 · Cited by 18 · see citing works
“A number of metrics have been devised for quantifying the similarity of images [12]–[15].”
BackgroundZhang, X., Uneri, A., Wu, P., Ketcha, M. D., Jones, C. K., Huang, Y., Lo, S.-F. L., Helm, P. A., & Siewerdsen, J. H. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abde96 · Cited by 17 · see citing works
Self-citationGeng, H., Xiao, D., Yang, S., Fan, J., Fu, T., Lin, Y., Bai, Y., Ai, D., Song, H., Wang, Y., et al. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acede5 · Cited by 17 · see citing works
Morales‐Rodríguez, C., Vannini, A., Scanu, B., et al. (2025) · Current Forestry Reports · doi:10.1007/s40725-024-00237-1 · Cited by 17 · see citing works
Reynolds, T., Ma, Y., Kanawati, A. J., et al. (2022) · Investigative Radiology · doi:10.1097/rli.0000000000000885 · Cited by 16 · see citing works
Self-citationSiewerdsen, J. H., Levine, M., Silva, T. D., Vijayan, R. C., Doerr, S. A., Uneri, A., Vedula, S. S., Theodore, N., & Ketcha, M. D. (2019) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2513915 · Cited by 15 · see citing works
“, pedicle screws).(4,5) Robust, accurate, and automatic vertebrae centroid annotation thus poses a challenging problem for learning algorithms, especially in large data sets presenting a broad spectrum of realistic spinal pathology.”
BackgroundSelf-citationAutomatic masking for robust 3D-2D image registration in image-guided spine surgery
Ketcha, M. D., Silva, T. D., Uneri, A., et al. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2216913 · Cited by 14 · see citing works
Self-citationRealistic C-arm to pCT registration for vertebral localization in spine surgery
Naik, R. R., Anitha, H., Bhat, S. N., Ampar, N., & Kundangar, R. S. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02600-5 · Cited by 14 · see citing works
“Specifically, to improve robustness against image content mismatch, various gradient-based similarity metrics have been applied in vertebral centroid registration between CT and lateral plane X-ray images [19].”
BackgroundMulti-View Global 2D-3D Registration Based on Branch and Bound Algorithm
Pan, J., Min, Z., Zhang, A., Ma, H., & Meng, M. Q. ‐. (2019) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio49542.2019.8961379 · Cited by 13 · see citing works
Image-Guided Tethering Spine Surgery With Outcome Prediction Using Spatio-Temporal Dynamic Networks
Mandel, W., Oulbacha, R., Roy‐Beaudry, M., et al. (2020) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2020.3030741 · Cited by 13 · see citing works
“Alternatively, datadriven methods used global inter-connected representations of the spine [40] or unsupervised vertebral level identification techniques [41] in order to fuse intra-operative cone beam CT images with pre-operative plans.”
Method3D vertebrae labeling in spine CT: an accurate, memory-efficient (Ortho2D) framework
Huang, Y., Uneri, A., Jones, C. K., et al. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac07c7 · Cited by 13 · see citing works
Self-citationSilva, T. D., Cool, D. W., Yuan, J., et al. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2703150 · Cited by 12 · see citing works
“and less computationally intensive than global methods with multiple initializations as in [26] and [27].”
MethodSelf-citationIntensity‐based 2D‐3D registration for an ACL reconstruction navigation system
Guo, N., Yang, B., Ji, X., et al. (2019) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2008 · Cited by 11 · see citing works
Kang, H., Jiang, H., Zhou, X., Yu, H., Hara, T., Fujita, H., & Yao, Y. (2019) · IEEE Access · doi:10.1109/access.2019.2961268 · Cited by 11 · see citing works
“Commonly used cost functions include mean square error (MSE) [6], mutual information (MI) [7], normalized mutual information (NMI) [8], normalized cross-correlation (NCC) [9] and gradient correlation (GC) [10].”
BackgroundHCS-Net: Multi-level deformation strategy combined with quadruple attention for image registration
Ou, Z., Lu, X., & Gu, Y. (2023) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2023.107832 · Cited by 10 · see citing works
Lü, S., Jiang, P., Liang, Y., et al. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3485629 · Cited by 9 · see citing works
“…GO represents the Gradient Orientation used as the image similarity measure function, which has been proven to be more robust in clinical images [37]; P ) denotes the projection of the patient’s preoperative 3D pelvic image X ( 3 d ) P using ray projection; and CAiCA 1 T represents the…”
Chen, M., Zhang, Z., Gu, S., & Kong, Y. (2024) · IEEE International Conference on Acoustics Speech and Signal Processing · doi:10.1109/icassp48485.2024.10446096 · Cited by 8 · see citing works
“In these traditional non-learning-based method [4, 5, 6], the 3D X-ray attenuation map is used to create the Digital Reconstructed Radiograph (DRR), which simulates the attenuation of virtual X-rays.”
MethodManbachi, A., Silva, T. D., Uneri, A., et al. (2018) · Annals of Biomedical Engineering · doi:10.1007/s10439-018-2099-2 · Cited by 8 · see citing works
Self-citationAgent-Based Methods for Medical Image Registration
Miao, S., & Liao, R. (2019) · Advances in computer vision and pattern recognition · doi:10.1007/978-3-030-13969-8_16 · Cited by 9 · see citing works
Robust Orthogonal-View 2-D/3-D Rigid Registration for Minimally Invasive Surgery
An, Z., Ma, H., Liu, L., Wang, Y., Lu, H., Zhou, C., Xiong, R., & Hu, J. (2021) · Micromachines · doi:10.3390/mi12070844 · Cited by 7 · see citing works
“For comparison, we implement the state-of-the-art 2-D/3-D registration approach POINT2 [15], which can also be used as an initial pose estimator for optimization-based methods [9,31] called POINT2+opt.”
BackgroundMethodAnatomy-Aware Inference of the 3D Standing Spine Posture from 2D Radiographs
Bayat, A., Pace, D. F., Sekuboyina, A., Payer, C., Štern, D., Urschler, M., Kirschke, J. S., & Menze, B. H. (2022) · Tomography · doi:10.3390/tomography8010039 · Cited by 7 · see citing works
“Other approaches for this purpose introduce an automated 3D–2D spine registration algorithm [7,8], wherein authors suggest a multi-stage registration method by introducing a comparison metric for a CT projection and a radiograph.”
MethodVijayan, R. C., Sheth, N., Mekki, L., et al. (2022) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ac9e3c · Cited by 7 · see citing works
Self-citationChi, Y., Shen, C., Li, B., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab4c0d · Cited by 6 · see citing works
Bindels, B. J. J., Weijers, R. A. M., van Mourik, M. S., et al. (2022) · International Journal of Computer Assisted Radiology and Surgery · doi:10.1007/s11548-022-02733-w · Cited by 6 · see citing works
Ushakiran, U., Naik, R. R., Anitha, H., & Bhat, S. N. (2023) · IEEE Access · doi:10.1109/access.2023.3286531 · Cited by 6 · see citing works
Abderezaei, J., Pionteck, A., Chopra, A., et al. (2023) · Lecture notes in computer science · doi:10.1007/978-3-031-44153-0_4 · Cited by 6 · see citing works
Ketcha, M. D., Silva, T. D., Han, R., et al. (2019) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2019.2907868 · Cited by 6 · see citing works
“examine 3D-2D registration scenarios, which previous studies [11], [12] showed to have better performance when using”
BackgroundSelf-citationJobidon-Lavergne, H., Kadoury, S., Knez, D., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab1bfa · Cited by 6 · see citing works
Deformable registration of lateral cephalogram and cone‐beam computed tomography image
Zhang, Y., Qin, H., Li, P., Pei, Y., Guo, Y., Xu, T., & Zha, H. (2021) · Medical Physics · doi:10.1002/mp.15214 · Cited by 6 · see citing works
Contrastive learning based method for X-ray and CT registration under surgical equipment occlusion
Wang, X., Zhang, Z., Xu, S., Luo, X., Zhang, B., & Wu, X.-J. (2024) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2024.108946 · Cited by 5 · see citing works
Fluoroscopic guidance of a surgical robot: pre-clinical evaluation in pelvic guidewire placement
Vijayan, R. C., Han, R., Wu, P., et al. (2021) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2582188 · Cited by 5 · see citing works
Self-citationChen, L., Zhang, X., He, Y., et al. (2021) · International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2283 · Cited by 5 · see citing works
“Silva et al.(29) studied the index of image similarity in 3D‐2D registration.”
BackgroundInfluentialFeature-based multimodal registration framework for vertebral pose estimation
Kiran, U., Bhat, S. N., Anitha, H., et al. (2023) · European Spine Journal · doi:10.1007/s00586-023-08054-z · Cited by 4 · see citing works
A new method to reconstruct intra-fractional prostate motion in volumetric modulated arc therapy
Chi, Y., Rezaeian, N. H., Shen, C., Zhou, Y., Lu, W., Yang, M., Hannan, R., & Jia, X. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa6e37 · Cited by 4 · see citing works
Drill-mounted video guidance for orthopaedic trauma surgery
Vagdargi, P., Sheth, N., Sisniega, A., et al. (2021) · Journal of medical imaging · doi:10.1117/1.jmi.8.1.015002 · Cited by 4 · see citing works
“Real-time guidance is achieved by virtual overlay of the registered drill axis on fluoroscopy or in CT. Performance was evaluated in terms of target registration error (TRE), conformance within clinically relevant pelvic bone corridors, and runtime.”
MethodInfluentialSelf-citationImage-guided robotic k-wire placement for orthopaedic trauma surgery
Vijayan, R. C., Han, R., Wu, P., et al. (2020) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2549713 · Cited by 4 · see citing works
Self-citationSiewerdsen, J. H., Uneri, A., Zhang, X., et al. (2018) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2293874 · Cited by 3 · see citing works
“The objective function is solved using covariance matrix adaptation evolution strategy (CMA-ES), with registration of anatomy employing gradient orientation (GO) similarity for robustness to content mismatch, and component registration using gradient correlation (GC) to exploit high contrast edges.(11) A CBCT image ob…”
Self-citationSisniega, A., Hernandez, A. M., Aminololama‐Shakeri, S., et al. (2024) · Medical Physics · doi:10.1002/mp.17452 · Cited by 3 · see citing works
Self-citationUshakiran, U., Anitha, H., Bhat, S. N., et al. (2022) · International Conference on Smart Technologies in Computing, Electrical and Electronics (ICSTCEE) · doi:10.1109/icstcee56972.2022.10099932 · Cited by 3 · see citing works
“, [11] used gradient orientation similarity metric with CMA-ES”
MethodCollaboration calibration and three-dimensional localization in multi-view system
Feng, S., Wu, C., Zhang, Y., et al. (2018) · International Journal of Advanced Robotic Systems · doi:10.1177/1729881418813778 · Cited by 3 · see citing works
Vijayan, R. C., Han, R., Wu, P., et al. (2021) · Journal of medical imaging · doi:10.1117/1.jmi.8.3.035001 · Cited by 3 · see citing works
Self-citationJiang, Y., Pei, Y., Xu, T., Yuan, X., & Zha, H. (2024) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2024.3456251 · Cited by 2 · see citing works
Preventing Wrong-Level Spine Surgery
Agolia, J. P., Robertson, S., Turel, K. E., & Kasper, E. M. (2024) · Acta neurochirurgica. Supplementum · doi:10.1007/978-3-031-61601-3_1 · Cited by 3 · see citing works
Kong, D., Hu, S., Zhang, W., Zhao, G., Bai, X., Wang, X., Gurure, D. M., Li, X., Li, X., & Wang, Y. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.107751 · Cited by 2 · see citing works
Siewerdsen, J. H., Ramsay, B., Silva, T. D., et al. (2018) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2293857 · Cited by 2 · see citing works
Self-citationPan, J., Min, Z., Zhang, A., Ma, H., & Meng, M. Q. ‐. (2021) · IEEE International Conference on Robotics and Biomimetics (ROBIO) · doi:10.1109/robio54168.2021.9739622 · Cited by 2 · see citing works
“In Image-guided Intervention of Hip Joint [5], pre-operative 3D CT is implemented to provide the patient’s information to plan the surgery [6], [7].”
BackgroundEvaluation of Similarity Measure for Feature based 3D-2D Registration in Vertebral Pose Estimation
Ushakiran, U., Bhat, S. N., Naik, R. R., & Anitha, H. (2023) · International Conference for Emerging Technology (INCET) · doi:10.1109/incet57972.2023.10169995 · Cited by 2 · see citing works
“, [6] evaluated the accuracy of the registration method in the presence of surgical instruments by comparing various gradient similarity measures.”
MethodRigid Shape Registration Based on Extended Hamiltonian Learning
Yi, J., Zhang, S., Cao, Y., et al. (2020) · Entropy · doi:10.3390/e22050539 · Cited by 2 · see citing works
Towards Automated Spine Mobility Quantification: A Locally Rigid CT to X-ray Registration Framework
Drobny, D., Ranzini, M. B. M., Isaac, A., Vercauteren, T. K., Sebastien Ourselin, A., Choi, D., & Marc Modat, A. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-50120-4_7 · Cited by 2 · see citing works
Representing Ambiguity in Registration Problems with Conditional Invertible Neural NetworksPreprint
Trofimova, D., Adler, T. J., Kausch, L., Ardizzone, L., Maier-Hein, K. H., Köthe, U., Rother, C., & Maier‐Hein, L. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2012.08195 · Cited by 5 · see citing works
Collision Constraints and Diffeomorphisms: Dealing With Physics in Deformable Image Registration
Alscher, T., Petersen, J., Lauze, F., Erleben, K., & Darkner, S. (2025) · IEEE Access · doi:10.1109/access.2025.3588760 · Cited by 1 · see citing works
“Practical applications for image registration in the medical field incorporate atlas creation [5], surgery image guidance [6], and image fusion [7], among others.”
Real-time image-based 3D-2D registration for ultrasound-guided spinal interventions
Siewerdsen, J. H., Silva, T. D., Uneri, A., et al. (2018) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2293710 · Cited by 1 · see citing works
Self-citationSparse Branch-and-Bound 2D-3D Registration Based on Prior Correspondences of Dense Points
Pan, J., Min, Z., Shen, Y., Liu, T., & Meng, M. Q. ‐. (2022) · International Conference on Advanced Robotics and Mechatronics (ICARM) · doi:10.1109/icarm54641.2022.9959548 · Cited by 1 · see citing works
Huang, Y., Zhang, X., Johnston, A., et al. (2023) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2654329 · Cited by 1 · see citing works
Self-citationVijayan, R. C., Venkataraman, K., Wei, J., et al. (2023) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2654481 · Cited by 1 · see citing works
Self-citationEvaluating similarity measure for multimodal 3D to 2D registration
Kiran, U., Naik, R. R., Bhat, S. N., & Anitha, H. (2023) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/ace9e1 · Cited by 1 · see citing works
“To overcome the intensity difference between the multi-modal images, De Silva et al [10] employed a segmentation algorithm in the pre-processing step.”
MethodCompensation for Patient Movements in CBCT Imaging for Dental Applications
Ali, A. S. R. A., Sarti, C., & Landi, C. (2023) · Lecture notes in computer science · doi:10.1007/978-3-031-43153-1_45 · Cited by 1 · see citing works
Vijayan, R. C., Sheth, N., Mekki, L., et al. (2022) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2612986 · Cited by 1 · see citing works
“A comparison of the objective functions (RMSD, NCC, GC, NGI, and GO) as shown in Figure 3a indicates that RMSD (PDE = 13.9 6.9 mm) and NCC (PDE = 13.0 5.9 mm) perform poorly for conventional local 3D- 2D lung registration, suffering from local minima presented by strong rib gradients.”
BackgroundInfluentialSelf-citationRegistration made easy -- standalone orthopedic navigation with HoloLensPreprint
Liebmann, F., Roner, S., von Atzigen, M., Wanivenhaus, F., Neuhaus, C., Spirig, J. M., Scaramuzza, D., Sutter, R., Snedeker, J. G., Farshad, M., et al. (2020) · arXiv (Cornell University) · doi:10.48550/arxiv.2001.06209 · Cited by 3 · see citing works
“measure which can be optimized in an iterative fashion. However, surgical tools, implants and resected soft tissue which are usually not present in preoperative images make this approach challenging [6]. DRR generation is computationally expensive and a balance between robustness and computational cost has to be found…”
BackgroundTechnical note: known-component registration for robotic drill guide positioning
Yi, T. Y., Ramchandran, V., Siewerdsen, J. H., et al. (2018) · SPIE Medical Imaging: Image-Guided Procedures, Robotic Interventions, and Modeling · doi:10.1117/12.2322408
Self-citationChaitanya, B. H. R., Anitha, H., Shyamasunder, N. B., et al. (2018) · International Conference on Advances in Computing, Communications and Informatics (ICACCI) · doi:10.1109/icacci.2018.8554401
“But among the similarity metrics, GO has an advantage that both the instrumentation and vertebrae gradients contribute similar weights to the overall metric value thereby exhibiting stronger robustness to mismatch.[13].”
InfluentialLiao, H., Lin, W.-A., Zhang, J., Zhang, J., Luo, J., & Zhou, S. (2019) · arXiv (Cornell University) · doi:10.48550/arxiv.1903.03896
Ketcha, M. D. (2020)
Self-citationLung CT image registration based on end-to-end unsupervised learning
Zhu, Z., Ji, Y., & Wei, Y. (2021) · International Conference on Communication, Image and Signal Processing (CCISP) · doi:10.1109/ccisp52774.2021.9639353
An improved multi-resolution 2D/3D registration method
Cao, Y., He, F., Qu, F., Wang, T., Chen, Y., Shi, W., & Jiang, Z. (2022) · 2022 3rd International Conference on Computer Vision, Image and Deep Learning & International Conference on Computer Engineering and Applications (CVIDL & ICCEA) · doi:10.1109/cvidliccea56201.2022.9824315
A Review of Honey Application in Marinades Towards Hetero-Cyclic Amines (HCA) Formation
Mat, N. N. S., & Sarbon, N. M. (2022) · Advances in environmental engineering and green technologies book series · doi:10.4018/978-1-6684-6265-2.ch001
Topologically preserved registration of 3D CT images with deep networks
Liu, H., Gong, G., Zou, W., Hu, N., & Wang, J. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acb197
Registration: 2D/3D rigid registration
Liao, H., Zhou, S., & Luo, J. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-824383-1.00014-9
- Authors
- De Silva, T., Uneri, A., Ketcha, M. D., Reaungamornrat, S., Kleinszig, G., Vogt, S., Aygun, N., Lo, S.-F., Wolinsky, J.-P., & Siewerdsen, J. H.
- Affiliations
- Johns Hopkins University
- Johns Hopkins University (US)
- Siemens (Germany) (DE)
- Johns Hopkins Medical Institute
- Corresponding author
- Siewerdsen, J H
- Venue
- Physics in Medicine & Biology
- Publisher
- IOP Publishing
- Published
- 2016-03-18
- Volume
- 61
- Issue
- 8
- Pages
- 3009-3025
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- PMC
- PMC4915376
- OpenAlex
- W2329673985
- Semantic Scholar
- b79816585c14219cea6401f1f6b70af222d8b96a
- Cited by
- 107
- References
- 27
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 9 · 2023: 16 · 2022: 12 · 2021: 13 · 2020: 14 · 2019: 13 · 2018: 13 · 2017: 8 · 2016: 2
- FWCI
- 5.49
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Physics in Medicine and Biology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion)
- Funders
- National Institutes of Health (R01-EB-017226)
- Siemens Healthcare
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
- Health Sciences › Medicine › Surgery › Spinal Fractures and Fixation Techniques
- Keywords
- image-guided spine surgery, image similarity measures, gradient direction, gradient correlation, gradient information, preoperative CT, multi-start optimization, surgical instrumentation, soft tissue deformation
- MeSH terms
- Humans, Retrospective Studies, Spinal Diseases, Tomography, X-Ray Computed, Imaging, Three-Dimensional, Surgery, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-14.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-14.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review