Range uncertainty in proton therapy due to variable biological effectiveness

Open access
Carabe, Alejandro, Moteabbed, Maryam, Depauw, Nicolas, Schuemann, Jan, Paganetti, Harald·2012·Physics in Medicine and Biology

Abstract

Traditionally, dose in proton radiotherapy is prescribed as Gy(RBE) by scaling up the physical dose by 10%. The relative biological effectiveness (RBE) of protons is considered to vary with dose-averaged linear energy transfer (LET(d)), dose (d) and (α/β)(x). The increase of RBE with depth causes a shift of the falloff of the beam, i.e. a change of the beam range. The magnitude of this shift will depend on dose and (α/β)(x). The aim of this project was to quantify the dependence of the range shift on these parameters. Three double-scattered beams of different ranges incident on a computational phantom consisting of different regions of interest (ROIs) were used. Each ROI was assigned with (α/β)(x) values between 0.5 and 20 Gy. The distribution of LET(d) within each ROI was obtained from a Monte Carlo simulation. The LET(d) distribution depends on the beam energy and thus its nominal range. The RBE values within the ROIs were calculated for doses between 1 and 15 Gy using an in-house developed biophysical model. Dose-volume histograms of the RBE-weighted doses were extracted for each ROI for a 'fixed RBE' (RBE = 1.1) and a 'variable RBE' (RBE = f (d, α/β, LET(d))), and the percentage difference in range was obtained from the difference of the percentage volumes at the distal 80% of the dose. Range differences in normal tissue ((α/β)(x) = 3 Gy) of the order of 3-2 mm were obtained, respectively, for a shallow (physical range 4.8 cm) and a deep (physical range 12.8 cm) beam, when a dose of 1 Gy normalized to the mid-SOBP was delivered. As the dose increased to 15 Gy, the variable RBE decreases below 1.1 which induces ranges of about 1 mm shorter than those obtained with an RBE of 1.1. The shift in the range of an SOBP when comparing biological dose distributions obtained with a fixed or a variable RBE was quantified as a function of dose, (α/β)(x) and physical range (as a surrogate of the initial beam energy). The shift increases with the physical range but decreases with increasing dose or (α/β)(x). The results of our study allow a quantitative consideration of RBE-caused range uncertainties as a function of treatment site and dose in treatment planning.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/0031-9155/57/5/1159

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 246

Metadata

Authors
Carabe, A., Moteabbed, M., Depauw, N., Schuemann, J., & Paganetti, H.
Venue
Physics in Medicine and Biology
Published
2012-02-14
Volume
57
Issue
5
Pages
1159-1172
ISSN
  • 0031-9155 (ISSN-L)
  • 1361-6560
OpenAlex
W2163678892
Cited by
246
References
0
FWCI
12.09
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Range uncertainty in proton therapy due to variable biological effectiveness. https://app.cochat.ai/papers/10.1088/0031-9155/57/5/1159
The paper
Carabe, A., Moteabbed, M., Depauw, N., Schuemann, J., & Paganetti, H. (2012). Range uncertainty in proton therapy due to variable biological effectiveness. Physics in Medicine and Biology. https://doi.org/10.1088/0031-9155/57/5/1159

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.