3D–2D registration in mobile radiographs: algorithm development and preliminary clinical evaluation

Open access
Otake, Yoshito, Wang, Adam S, Uneri, Ali, Kleinszig, Gerhard, Vogt, Sebastian, Aygun, Nafi, Lo, Sheng-fu L, Wolinsky, Jean-Paul +2 more·2015·Physics in Medicine & Biology

Abstract

An image-based 3D–2D registration method is presented using radiographs acquired in the uncalibrated, unconstrained geometry of mobile radiography. The approach extends a previous method for six degree-of-freedom (DOF) registration in C-arm fluoroscopy (namely ‘LevelCheck’) to solve the 9-DOF estimate of geometry in which the position of the source and detector are unconstrained. The method was implemented using a gradient correlation similarity metric and stochastic derivative-free optimization on a GPU. Development and evaluation were conducted in three steps. First, simulation studies were performed that involved a CT scan of an anthropomorphic body phantom and 1000 randomly generated digitally reconstructed radiographs in posterior–anterior and lateral views. A median projection distance error (PDE) of 0.007 mm was achieved with 9-DOF registration compared to 0.767 mm for 6-DOF. Second, cadaver studies were conducted using mobile radiographs acquired in three anatomical regions (thorax, abdomen and pelvis) and three levels of source-detector distance (~800, ~1000 and ~1200 mm). The 9-DOF method achieved a median PDE of 0.49 mm (compared to 2.53 mm for the 6-DOF method) and demonstrated robustness in the unconstrained imaging geometry. Finally, a retrospective clinical study was conducted with intraoperative radiographs of the spine exhibiting real anatomical deformation and image content mismatch (e.g. interventional devices in the radiograph that were not in the CT), demonstrating a PDE = 1.1 mm for the 9-DOF approach. Average computation time was 48.5 s, involving 687 701 function evaluations on average, compared to 18.2 s for the 6-DOF method. Despite the greater computational load, the 9-DOF method may offer a valuable tool for target localization (e.g. decision support in level counting) as well as safety and quality assurance checks at the conclusion of a procedure (e.g. overlay of planning data on the radiograph for verification of the surgical product) in a manner consistent with natural surgical workflow.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/0031-9155/60/5/2075

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • In previous work on vertebral level localization in spine surgery, the challenge due to content mismatch was mitigated by manually defined polygonal masks that excluded extraneous regions during similarity metric evaluation.

    “However, the challenge due to content mismatch was mitigated by manually defined polygonal masks that excluded extraneous regions during similarity metric evaluation.”

    From 3D–2D image registration for target localization in spine surgery: investigation of similarity metrics providing robustness to content mismatch (2016) · cites this paper as Otake et al 2015

    The Patient study section (page unavailable) says a weighting mask can down-weight regions with deformation or instrumentation so they contribute less to the similarity metric. It says such masks could potentially be made manually, but does not describe polygonal masks or establish that manually defined masks were used to mitigate content mismatch in this study. (LLM, conf=0.85)

    Experimental — accuracy not yet measured
  • Previous work on vertebral level localization in spine surgery showed promising results with GC similarity.

    “For example, previous work on vertebral level localization in spine surgery (Lo et al 2015, Otake et al 2015) showed promising results with GC similarity.”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

The citation lists haven’t been fetched yet

References 31 · Cited by 41

Research works printed but not linked: 24

  1. Bentley JL 1975 Multidimensional binary search trees used for associative searching Commun ACM 18 509 17
  2. Birkfellner W Stock M Figl M Gendrin C Hummel J Dong S Bergmann H 2009 Stochastic rank correlation: a robust merit function for 2D/3D registration of image data obtained at different energies Med Phys 36 3420 8 19746775 PMC2855948
  3. Boisvert J Cheriet F Pennec X Labelle H Ayache N 2008 Geometric variability of the scoliotic spine using statistics on articulated shape models IEEE Trans Med Imaging 27 557 68 18390352
  4. Cho Y Moseley DJ Siewerdsen JH Jaffray DA 2005 Accurate technique for complete geometric calibration of cone-beam computed tomography systems Med Phys 32 968 83 15895580
  5. Dang H Otake Y Schafer S Stayman JW Kleinszig G Siewerdsen JH 2012 Robust methods for automatic image-to-world registration in cone-beam CT interventional guidance Med Phys 39 6484 98 23039683 PMC3477200
  6. Hansen N 2006 The CMA evolution strategy: a comparing review Towards a New Evolutionary Computation 192 Lozano J 75 102 Berlin Springer
  7. Kirkpatrick S Gelatt CD Vecchi MP 1983 Optimization by simulated annealing Science 220 671 80 17813860
  8. Moura DC Boisvert J Barbosa JG Labelle H Tavares JM 2011 Fast 3D reconstruction of the spine from biplanar radiographs using a deformable articulated model Med Eng Phys 33 924 33 21481628
  9. Munbodh R Tagare HD Chen Z Jaffray DA Moseley DJ Knisely JP Duncan JS 2009 2D–3D registration for prostate radiation therapy based on a statistical model of transmission images Med Phys 36 4555 68 19928087
  10. Otake Y Armand M Armiger RS Kutzer MD Basafa E Kazanzides P Taylor RH 2012 Intraoperative image-based multiview 2D/3D registration for image-guided orthopaedic surgery: incorporation of fiducial-based C-arm tracking and GPU-acceleration IEEE Trans Med Imaging 31 948 62 22113773 PMC4451116
  11. Otake Y Armand M Sadowsky O Armiger R Kazanzides P Taylor R 2010 An iterative framework for improving the accuracy of intraoperative intensity-based 2D/3D registration for image-guided orthopedic surgery Information Processing in Computer-Assisted Interventions 6135 Navab N Jannin P 23 33 Berlin Springer
  12. Palumbo MA Bianco AJ Esmende S Daniels AH 2013 Wrong-site spine surgery J Am Acad Orthop Surg 21 312 20 23637150
  13. Penney GP Weese J Little JA Desmedt P Hill DL Hawkes DJ 1998 A comparison of similarity measures for use in 2D–3D medical image registration IEEE Trans Med Imaging 17 586 95 9845314
  14. Russakoff D Rohlfing T Ho A Kim D Shahidi R Adler J Jr Maurer C Jr 2003 Evaluation of intensity-based 2D–3D spine image registration using clinical gold-standard data Biomedical Image Registration 2717 Gee J 151 60 Berlin Springer
  15. Schumann S Liu L Tannast M Bergmann M Nolte L-P Zheng G 2013 An integrated system for 3D hip joint reconstruction from 2D x-rays: a preliminary validation study Ann Biomed Eng 41 2077 87 23670657
  16. Siddon RL 1985 Fast calculation of the exact radiological path for a 3D CT array Med Phys 12 252 5 4000088
  17. Tornai GJ Cserey G Pappas I 2012 Fast DRR generation for 2D to 3D registration on GPUs Med Phys 39 4795 9 22894404
  18. Uneri A Otake Y Wang AS Kleinszig G Vogt S Khanna AJ Siewerdsen JH 2013 3D–2D registration for surgical guidance: effect of projection view angles on registration accuracy Phys Med Biol 59 271 87 24351769 PMC4927006
  19. Uneri A Wang S Otake Y Kleinszig G Vogt S Khanna A Siewerdsen J 2014 Evaluation of low-dose limits in 3D–2D registration for surgical guidance Phys Med Biol 59 5329 45 25146673
  20. van de Kraats EB Penney GP Tomazevic D van Walsum T Niessen WJ 2005 Standardized evaluation methodology for 2D–3D registration IEEE Trans Med Imaging 24 1177 89 16156355
  21. Varnavas A Carrell T Penney G 2013 Increasing the automation of a 2D–3D registration system IEEE Trans Med Imaging 32 387 99 23362246
  22. Zhang J Lv L Shi X Wang Y Guo F Zhang Y Li H 2013 3D reconstruction of the spine from biplanar radiographs based on contour matching using the Hough transform IEEE Trans Biomed Eng 60 1954 64 23412567
  23. Zheng G 2010 Statistical shape model-based reconstruction of a scaled, patient-specific surface model of the pelvis from a single standard AP x-ray radiograph Med Phys 37 1424 39 20443464
  24. Zheng G Zhang X Steppacher SD Murphy SB Siebenrock KA Tannast M 2009 HipMatch: an object-oriented cross-platform program for accurate determination of cup orientation using 2D–3D registration of single standard x-ray radiograph and a CT volume Comput Methods Programs Biomed 95 236 48 19328585

Authors
Otake, Y., Wang, A. S., Uneri, A., Kleinszig, G., Vogt, S., Aygun, N., Lo, S.-F. L., Wolinsky, J.-P., Gokaslan, Z. L., & Siewerdsen, J. H.
Affiliations
  • Johns Hopkins University (US)
  • Siemens Healthineers (Germany) (DE)
  • Johns Hopkins University
Corresponding author
Siewerdsen, Jeffrey H
Venue
Physics in Medicine & Biology
Published
2015-02-12
Volume
60
Issue
5
Pages
2075-2090
ISSN
  • 0031-9155 (ISSN-L)
  • 1361-6560
OpenAlex
W2031954532
Cited by
41
References
31
Citations by year
2026: 2 · 2025: 2 · 2024: 1 · 2023: 6 · 2022: 3 · 2021: 4 · 2020: 2 · 2019: 3 · 2018: 4 · 2017: 8 · 2016: 4 · 2015: 1
FWCI
4.92
Type
journal article
Language
en
Open access
Locations
  • Physics in Medicine and Biology (publishedVersion)
  • PubMed (publishedVersion)
  • Europe PMC (PubMed Central) (acceptedVersion)
  • PubMed Central (acceptedVersion)
Funders
National Institute of Biomedical Imaging and Bioengineering (R01EB017226)
Topics
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Medical Image Segmentation Techniques
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
Keywords
2D-3D registration, mobile radiography, derivative-free optimization, GPU acceleration, spine, surgical target localization, cadaveric study, quality assurance
MeSH terms
Abdomen, Algorithms, Cadaver, Computer Simulation, Fluoroscopy, Humans, Pelvis, Retrospective Studies, Spine, Radiography, Thoracic, Tomography, X-Ray Computed, Phantoms, Imaging, Imaging, Three-Dimensional

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review