A study on electrode gels for skin conductance measurements
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Low-frequency skin conductance is used within several clinical applications and is mainly sensitive to sweating and the moisture content of the stratum corneum, but also how electrodes introduce changes in the electrical properties. Four electrode gels were investigated with regard to sorption characteristics and electrical properties. Skin conductance time series were collected from 18 test subjects during relaxation, exercise and recovery, wearing different pairs of electrodes contralaterally on the hypothenar and the T9 dermatome. Pressure test was applied on the T9 electrodes. Impedance frequency sweeps were taken on the T9 electrodes the same day and the next, parameterized to the Cole model. ANOVA on the initial skin conductance level change, exercise response amplitude, recovery offset and pressure-induced changes revealed significant differences among gel types. The wetter gels caused a higher positive level change, a greater response amplitude, larger recovery offset and greater pressure-induced artifacts compared to the solid gels. Sweating on the T9 site led to negative skin conductance responses for the wetter gels. Correlations were found between the desorption measurements and the initial skin conductance level change (hypothenar: R = 0.988 T9: R = 0.901) RM-ANOVA on the Cole parameters revealed a significant decrease in R(s) of the most resistive gel. Clinical implications are discussed.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Detecting Stress During Real-World Driving Tasks Using Physiological Sensors
Healey, J. A., & Picard, R. W. (2005) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2005.848368 · Cited by 2,207
Boucsein, W. (1992) · doi:10.1007/978-1-4757-5093-5 · Cited by 818
Discriminating Stress From Cognitive Load Using a Wearable EDA Device
Setz, C., Arnrich, B., Schumm, J., La Marca, R., Troster, G., & Ehlert, U. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2036164 · Cited by 742 · cited 2×
“Keywords: sweat, skin conductance, electrodes, water sorption, stratum corneum, electrodermal activity, sweat glands”
BackgroundMethod
Cited by
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434
Publication recommendations for electrodermal measurements
Society for Psychophysiological Research Ad Hoc Committee on Electrodermal Measures (2012) · Psychophysiology · doi:10.1111/j.1469-8986.2012.01384.x · Cited by 1,288
“Hydrogel is an ionic material that does not wet the skin, and measurements have shown (Tronstad et al., 2010) that a solid gel may have little free water to exchange with the ducts and corneum.”
BackgroundKleckner, I. R., Jones, R. M., Wilder-Smith, O., Wormwood, J. B., Akcakaya, M., Quigley, K. S., Lord, C., & Goodwin, M. S. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2758643 · Cited by 121
“Recording artifacts can arise from a variety of sources, such as: (1) movement artifacts from pressure or movement of the electrodes relative to the skin [20]; (2) participants intentionally or unintentionally touching the sensors (especially for individuals with sensory sensitivity conditions such as autism); or (3)…”
Daily Electrode Change and Effect on Cardiac Monitor Alarms: An Evidence-Based Practice Approach
Cvach, M. M., Biggs, M., Rothwell, K. J., & Charles-Hudson, C. (2013) · Journal of Nursing Care Quality · doi:10.1097/ncq.0b013e31827993bc · Cited by 75
“Level II 3 A/B • Electrodes used beyond first day exhibit poor electrode-skin contact; inaccurate impedance measurement (10) • Factors causing electrodes to become disconnected (12): ◦ Patient movement; perspiration; manipulation; defective electrodes • Best performing electrodes (14): ◦ Good adhesive properties; low…”
Background
Metadata
- Authors
- Tronstad, C., Johnsen, G. K., Grimnes, S., & Martinsen, Ø. G.
- Venue
- Physiological Measurement
- Publisher
- IOP Publishing
- Published
- 2010-09-01
- Volume
- 31
- Issue
- 10
- Pages
- 1395-1410
- ISSN
- 0967-3334 (ISSN-L)
- 1361-6579
- OpenAlex
- W2020816939
- Cited by
- 93
- References
- 3
- FWCI
- 4.73
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A study on electrode gels for skin conductance measurements. https://app.cochat.ai/papers/10.1088/0967-3334/31/10/008
- The paper
- Tronstad, C., Johnsen, G. K., Grimnes, S., & Martinsen, Ø. G. (2010). A study on electrode gels for skin conductance measurements. Physiological Measurement. https://doi.org/10.1088/0967-3334/31/10/008
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.