Performance evaluation of the MOLECUBES β -CUBE—a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors
Open accessAbstract
The MOLECUBES β -CUBE scanner is the newest amongst commercially available preclinical PET scanners for dedicated small animal imaging. The scanner is compact, lightweight and utilizes a small footprint to facilitate bench-top imaging. It can be used individually, or in combination with the X-CUBE CT scanner, which provides the ability to perform all necessary PET data corrections and provide fully quantitative PET images. The PET detector comprises of an 8 mm thick monolithic LYSO scintillator read-out by an array of 3 mm × 3 mm Hamamatsu silicon photomultipliers. The monolithic scintillator provides the ability to measure depth-of-interaction which aids in the development of such a compact scanner. With a scanner diameter of 7.6 cm and axial length of 13 cm it is suitable for imaging both whole-body mice and rats. This paper presents the design and imaging performance of the β -CUBE scanner. NEMA NU4-2008 characterization and a variety of phantom and animal imaging studies to demonstrate the quantitative imaging performance of the PET scanner are presented. Spatial resolution of 1 mm is measured with a filtered-back projection reconstruction algorithm at the center of the scanner and DOI measurement helps maintain the excellent spatial resolution over the entire imaging FOV. An absolute peak sensitivity of 12.4% is measured with a 255–765 keV energy window. The scanner demonstrates good count-rate performance, with a peak NEC of 300 kcps and 160 kcps measured with ~900 µ Ci in the NEMA mouse and rat phantoms, respectively. Imaging data with the NEMA image quality phantom and Micro Derenzo phantoms demonstrate the ability to achieve good image quality and accurate quantitative data. Image uniformity of 7.4% and spill-over ratio of 8% were measured. The superior spatial resolution, excellent energy resolution and sensitivity also provide superior contrast recovery, with ~70% recovery for the 2 mm rods. While current commercial preclinical PET scanners have spatial resolution in the 1–2 mm range, the 1 mm 3 volumetric resolution presents significant improvement over current commercially available preclinical PET scanners. In combination with the X-CUBE scanner it provides the ability to perform fully quantitative imaging with spatially co-registered high-resolution 3D PET-CT images.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In all cases, images were corrected for radionuclide decay, tissue attenuation, detector dead-time loss, and photon scatter according to the manufacturer’s methods (Krishnamoorthy et al., 2018).”
From Plcg2M28L Interacts With High Fat/High Sugar Diet to Accelerate Alzheimer’s Disease-Relevant Phenotypes in Mice (2022) · cites this paper as Krishnamoorthy et al., 2018
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
MicroPET: a high resolution PET scanner for imaging small animals
Cherry, S. R., Shao, Y., Silverman, R. W., Meadors, K., Siegel, S. B., Chatziioannou, A. F., Young, J. W., Jones, W., Moyers, J. C., Newport, D. F., et al. (1997) · IEEE Transactions on Nuclear Science · doi:10.1109/23.596981 · Cited by 611
Strother, S. C., Casey, M., & Hoffman, E. J. (1990) · IEEE Transactions on Nuclear Science · doi:10.1109/23.106715 · Cited by 549
Burger, C., Goerres, G. W., Schoenes, S., Buck, A., Lonn, A. H. R., & von Schulthess, G. K. (2002) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-002-0796-3 · Cited by 429
Treatment of axial data in three-dimensional PET.
Daube-Witherspoon, M. E., & Muehllehner, G. (1987) · PubMed · Cited by 357
Chatziioannou, A. F., Cherry, S. R., Shao, Y., Silverman, R. W., Meadors, K., Farquhar, T. H., Pedarsani, M., & Phelps, M. E. (1999) · PubMed · Cited by 347
A Single Scatter Simulation Technique for Scatter Correction in 3D PET
Watson, C. C., Newport, D. F., & Casey, M. (1996) · Computational imaging and vision · doi:10.1007/978-94-015-8749-5_18 · Cited by 342
Bao, Q., Newport, D. F., Chen, M., Stout, D. B., & Chatziioannou, A. F. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.056374 · Cited by 306
Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.
Wang, Y., Seidel, J., Tsui, B. M. W., Vaquero, J. J., & Pomper, M. G. (2006) · PubMed · Cited by 289
A novel, SiPM-array-based, monolithic scintillator detector for PET
Schaart, D. R., van Dam, H. T., Seifert, S., Vinke, R., Dendooven, P. G., Löhner, H., & Beekman, F. J. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/11/015 · Cited by 286
Tai, Y., Chatziioannou, A. F., Yang, Y., Silverman, R. W., Meadors, K., Siegel, S. B., Newport, D. F., Stickel, J. R., & Cherry, S. R. (2003) · Physics in Medicine and Biology · doi:10.1088/0031-9155/48/11/303 · Cited by 275
Research works printed but not linked: 2
- Casey, 2017, A next generation SiPM based PET/CT system with improved time and spatial resolution, J. Nucl. Med., 58, 1332S
- Mollet, 2017, The β-CUBE, a high-end compact preclinical benchtop PET for total body imaging, J. Nucl. Med., 58, 393S
Cited by
Sorted by citations
Items 101–141 of 141
Proton Beam Range Verification with Secondary Radiation from Titanium Implants
Bäcker, C. M., Bäumer, C., Bley, A., Costa, P. F., Gerhardt, M., Herrmann, K., Kauer, S., Kroeninger, K. A., Nitsch, C., Siregar, H. M., et al. (2019) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic42101.2019.9059997 · Cited by 4
Performance evaluation of S-PET, a compact, LYSO/BGO phoswich detector small animal PET prototype
Yang, H., He, W., Zhao, X., Li, G. D., Zhao, Y., Huang, W., Huang, Y., Niu, M., Liang, Y., & Gu, Z. (2026) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ae4b02 · Cited by 3
Giacomo, S. D., Ronchi, M., Amadori, M., Borghi, G., Carminati, M., & Fiorini, C. E. (2025) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2025.3530774 · Cited by 3
Effect of depth of interaction resolution on the spatial resolution of SIAT aPET
Kuang, Z., Zhang, L., Ren, N., Kinyanjui, S. M., Liu, Z., Sun, T., Hu, Z., & Yang, Y. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ad078b · Cited by 3
Ribeiro, F. M., Encarnação, P. M. C. C., Silva, A. L. M., Correia, P. M. M., Pinto, A., Castro, I. F., SANTOS, A. C. D. S., & Veloso, J. F. C. A. (2025) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2025.3546120 · Cited by 3
Investigation of a Model-Based Time-Over-Threshold Technique for Phoswich Crystal Discrimination
Bouziri, H., Pepin, C. M., Koua, K., Benhouria, M., Paulin, C., Ouyang, J., Normandin, M. D., Pratte, J. ‐., Fakhri, G. E., Lecomte, R., et al. (2021) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2021.3077412 · Cited by 3
Single‐Photon Emission Computed Tomographic Imaging in Live Animals
Yagi, Y., Kawashima, H., Arimitsu, K., Hasegawa, K., & Kimura, H. (2019) · doi:10.1002/9783527344406.ch6 · Cited by 3
Improvement of depth estimation for crosshair light sharing TOF-DOI PET detector
Yoshida, E., Obata, F., & Yamaya, T. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ae128d · Cited by 2
β -eye: A benchtop system for in vivo molecular screening of labeled compounds
Fysikopoulos, E., Rouchota, M. G., Γεωργίου, Μ., Sfyris, C., Cheimarios, N., Sarpaki, S., Kostopoulos, S. A., Glotsos, D. T., Larimer, B. M., Hunter, C. L., et al. (2021) · Applied Radiation and Isotopes · doi:10.1016/j.apradiso.2021.110034 · Cited by 2
Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008
Wei, Q., Wang, Y., Huang, X., Li, D., Zheng, Y., Wang, P., Sun, X.-L., Chai, P., Han, X., Liu, S., et al. (2024) · Medical Physics · doi:10.1002/mp.17469 · Cited by 2
Chen, K., Qin, J., Wang, S., Zuo, Y., Zhao, L., Yang, Y., & Du, J. (2025) · Journal of Instrumentation · doi:10.1088/1748-0221/20/05/p05036 · Cited by 2
Monitoring the biodistribution of radiolabeled therapeutics in mice
Fung, E. K., & ZANZONICO, P. B. (2023) · Methods in cell biology · doi:10.1016/bs.mcb.2023.02.012 · Cited by 2
Recent advances in digital Positron Emission Tomography
D’Ascenzo, N., Antonecchia, E., Bender, V., Gao, M., Jacob, C., Li, X., Ling, Y., Kattan, M., Li, J., Tian, C., et al. (2020) · Journal of Instrumentation · doi:10.1088/1748-0221/15/10/c10029 · Cited by 2
Du, J., & Du, S. (2024) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2024.3361891 · Cited by 1
Raylman, R. R., Stolin, A., Jaliparthi, G., & Martone, P. F. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/adc537 · Cited by 1
An open-source irradiation and data-handling framework for pre-clinical ion-beam research
Abdarahmane, I., Wolf, L., Kuess, P., Heilemann, G., Stocchiero, S., Knäusl, B., Feinerer, I., Zeilinger, M., & Georg, D. (2025) · Zeitschrift für Medizinische Physik · doi:10.1016/j.zemedi.2025.08.002 · Cited by 1
Azimi, M.-S., Karakatsanis, N. A., Sabet, H., Rafecas, M., Nehmeh, S. A., Abass, A. S., King, M. A., & Zaidi, H. (2026) · Computational and Structural Biotechnology Journal · doi:10.34133/csbj.0046 · Cited by 1
Performance of high-resolution PET detectors based on long semi-monolithic scintillator slabs
Kinyanjui, S. M., Kuang, Z., Liu, Z., Ren, N., & Yang, Y. (2026) · Nuclear Science and Techniques · doi:10.1007/s41365-026-01920-7 · Cited by 1
Zammit, M. D., Kao, C.-M., Zhang, H. J., Tsai, H., Holderman, N., Mitchell, S. J., Tanios, E., Bhuiyan, M. P. I., Freifelder, R. H., Kucharski, A., et al. (2023) · International Journal of Molecular Sciences · doi:10.3390/ijms242115510 · Cited by 1
Banka, V. K., Sainas, S., Martino, E., Wang, J., Lolli, M. L., & Ding, Y. (2024) · RSC Medicinal Chemistry · doi:10.1039/d4md00433g · Cited by 1
Wang, Z., Huang, X., Li, D., Zhou, W., Wang, Y., Zeng, X., Zhang, Z., Zheng, Y., Pan, W., Zhu, M., et al. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/adebd9 · Cited by 1
3D Spatial resolution proprieties of Molecubes β-Cube: characterization with different isotopes
Presotto, L., Spangler‐Bickell, M. G., Belloli, S., Moresco, R. M., Picchio, M., Gilardi, M. C., Gianolli, L., & Bettinardi, V. (2019) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic42101.2019.9059746 · Cited by 1
Performance of Dual-Ended Readout PET Detectors Based on SiPMs with Different Microcell Sizes
Yang, Q., Sang, Z., Zhang, C., & Du, J. (2020) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic42677.2020.9507942 · Cited by 1
Instrumentierung in der präklinischen PET- und SPECT-Bildgebung
Moraitis, A., & Costa, P. F. (2022) · Angewandte Nuklearmedizin · doi:10.1055/a-1715-5254 · Cited by 1
Buonanno, L., Pedretti, B., D’Adda, I., Alaimo, C., Carminati, M., & Fiorini, C. E. (2021) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic44867.2021.9875942 · Cited by 1
Iterative Algorithm for Optimal Super Resolution Sampling
Galve, P., López-Montes, A., Udías, J. M., & Herraiz, J. L. (2019) · Springer proceedings in physics · doi:10.1007/978-3-030-22204-8_10
Preclinical PET and SPECT Instrumentation
Dexter, K., Foster, J., Sosabowski, J. K., & Petřík, M. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822960-6.00055-7
Zammit, M. D., Kao, C.-M., Zhang, H. J., Holderman, N., Mitchell, S. J., Tanios, E., Zhang, V., Bhuiyan, M. P. I., Freifelder, R. H., Green, W. N., et al. (2022) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2022.08.05.502975
Sadremomtaz, A., Taherparvar, P., & saber, M. (2023) · Research Square · doi:10.21203/rs.3.rs-3599114/v1
Effects of Random Coincidences on Image Quality for a Preclinical CZT PET System
Groll, A. N., Stanford-Hill, R., & Levin, C. S. (2022) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic44845.2022.10399083
Positron emission tomography instrumentation and image reconstruction
Hu, Z., Sun, T., Wang, H., Kuang, Z., Liu, Z., & Yang, Y. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-95223-1.00010-2
Chyzh, R., Tauzin, G., Yvon, D., Sung, C.-H., Breton, D., Maalmi, J., Schäfers, K. P., Weinheimer, C., & Shary, V. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/07/p07018
Encarnação, P. M. C. C., Correia, P. M. M., Ribeiro, F. M., & Veloso, J. F. C. A. (2025) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/adc5f5
Performance evaluation of the nanoScan® P123S total-body PET
Réti, D., Corral, C.-A., Cranston, I., Reid, V. J. M., O’Rourke, K. M., Morgan, T. E. F., Montagne, A., Jansen, M. A., Burianova, V. K., Sutherland, A. J., et al. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00817-5
Wang, H., Wang, K., Yuan, Z., Li, C., Liao, R., Hou, Y., Hua, J., Tang, Y. P., Ruan, Q. Z., Han, D., et al. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00813-9
Teuho, J., Riehakainen, L., Honkaniemi, A., Moisio, O., Han, C., Tirri, M. E., Liu, S., Grönroos, T. J., Liu, J., Wan, L., et al. (2020) · Figshare · doi:10.6084/m9.figshare.13364390
Wang, S., Chen, K., Liu, J., Zhao, L., Ye, B. J., Qin, J., Du, J., Zuo, Y., & Yang, Y. (2025) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2025.171255
Mincke, J., Kloth, K. J., Verbeke, S., Kersemans, K., Vandenberghe, S., Vanhove, C., & Steppe, K. (2026) · Plant Methods · doi:10.1186/s13007-026-01525-6
Zhu, Z., Ren, D., Chen, W., Shen, W., Huang, X., Ren, N., Kuang, Z., Hu, Z., Yang, Y., & Liu, Z. (2026) · EJNMMI Physics · doi:10.1186/s40658-026-00908-x
An optically coupled hybrid semi-monolithic PET detector for simplified DOI and position decoding
Zhao, Y., Yang, H., Huang, W., Zhang, X., Jiang, X., Niu, M., Zeng, X., He, W. J., Wei, Q., & Gu, Z. (2026) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ae97d3
Preclinical PET Systems: Comparative Analysis and Technical Specification Guide
Pereira, L. R. M., & Gontijo, R. M. G. (2026) · Brazilian Journal of Radiation Technology Research. · doi:10.70745/bjrtr.v1.232
Metadata
- Authors
- Krishnamoorthy, S., Blankemeyer, E., Mollet, P., Surti, S., Van Holen, R., & Karp, J. S.
- Affiliations
- University of Pennsylvania (US)
- Corresponding author
- Krishnamoorthy, Srilalan
- Venue
- Physics in Medicine & Biology
- Publisher
- IOP Publishing
- Published
- 2018-07-27
- Volume
- 63
- Issue
- 15
- Pages
- 155013
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- PMC
- PMC6145835
- OpenAlex
- W2810955814
- Cited by
- 141
- References
- 44
- Citations by year
- 2026: 8 · 2025: 14 · 2024: 19 · 2023: 13 · 2022: 19 · 2021: 28 · 2020: 22 · 2019: 14 · 2018: 4
- FWCI
- 11.75
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Physics in Medicine and Biology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institutes of Health (R01-CA113941)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Keywords
- small animal imaging, silicon photomultipliers, spatial resolution, count rate performance, PET quantification, PET/CT, image quality, contrast recovery, rats and mice
- MeSH terms
- Animals, Scintillation Counting, Sensitivity and Specificity, Phantoms, Imaging, Positron-Emission Tomography, Mice, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review