Performance evaluation of the mouse version of the LabPET II PET scanner
Open accessAbstract
The LabPET II is a new positron emission tomography technology platform designed to achieve submillimetric spatial resolution imaging using fully pixelated avalanche photodiodes-based detectors and highly integrated parallel front-end processing electronics. The detector was designed as a generic building block to develop devices for preclinical imaging of small to mid-sized animals and for clinical imaging of the human brain. The aim of this work is to assess the physical characteristics and imaging performance of the mouse version of LabPET II scanner following the NEMA NU4-2008 standard and using high resolution phantoms and in vivo imaging applications. A reconstructed spatial resolution of 0.78 mm (0.5 μ l ) is measured close to the center of the radial field of view. With an energy window of 350 650 keV, the system absolute sensitivity is 1.2% and its maximum noise equivalent count rate reaches 61.1 kcps at 117 MBq. Submillimetric spatial resolution is achieved in a hot spot phantom and tiny bone structures were resolved with unprecedented contrast in the mouse. These results provide convincing evidence of the capabilities of the LabPET II technology for biomolecular imaging in preclinical research.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The best spatial resolution of small-animal PET scanners is 1-1.5 mm.
“the best resolution is still 1-1.5 mm [153].”
From Challenges and rewards of in vivo synaptic density imaging, and its application to the study of depression (2025) · cites this paper as [153]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Preclinical Alzheimer’s Disease: Definition, Natural History, and Diagnostic Criteria
Dubois, B., Hampel, H., Feldman, H. H., Scheltens, P., Aisen, P., Andrieu, S., Bakardjian, H., Benali, H., Bertram, L., Blennow, K., et al. (2016) · Alzheimer's & Dementia · doi:10.1016/j.jalz.2016.02.002 · Cited by 1,925 · see citing works
Bao, Q., Newport, D. F., Chen, M., Stout, D. B., & Chatziioannou, A. F. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.056374 · Cited by 306 · see citing works
Performance Evaluation of the LabPET APD-Based Digital PET Scanner
Bergeron, M., Cadorette, J., Beaudoin, J.-F., Lepage, M. D., Robert, G., Selivanov, V. V., Tétrault, M., Viscogliosi, N., Norenberg, J. P., Fontaine, R., et al. (2009) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.2010257 · Cited by 155 · see citing works
Self-citationPerformance measurements of small animal positron emission tomographs.
Weber, S., & Bruyndonckx, P. (2002) · JuSER (Forschungszentrum Jülich) · Cited by 154 · see citing works
Krishnamoorthy, S., Blankemeyer, E., Mollet, P., Surti, S., Van Holen, R., & Karp, J. S. (2018) · Physics in Medicine & Biology · doi:10.1088/1361-6560/aacec3 · Cited by 141 · see citing works
A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging
Yang, Y., Bec, J., Zhou, J., et al. (2016) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.165886 · Cited by 125 · see citing works
Inter-crystal scatter in a dual layer, high resolution LSO-APD positron emission tomograph
Rafecas, M., ning, G. B., Pichler, B. J., et al. (2003) · Physics in Medicine and Biology · doi:10.1088/0031-9155/48/7/302 · Cited by 120 · see citing works
A Digital Preclinical PET/MRI Insert and Initial Results
Weissler, B., Gebhardt, P., Dueppenbecker, P. M., Wehner, J., Schug, D., Lerche, C. W., Goldschmidt, B., Salomon, A., Verel, I., Heijman, E., et al. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2015.2427993 · Cited by 106 · see citing works
Tumor Texture Analysis in PET: Where Do We Stand?
Buvat, I., Orlhac, F., & Soussan, M. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.163469 · Cited by 104 · see citing works
Terry, G. E., Hirvonen, J., Liow, J., Zoghbi, S. S., Gladding, R. L., Tauscher, J. T., Schaus, J. M., Phebus, L. A., Felder, C. C., Morse, C. L., et al. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.067074 · Cited by 101 · see citing works
On the assessment of spatial resolution of PET systems with iterative image reconstruction
Gong, K., Cherry, S. R., & Qi, J. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/5/n193 · Cited by 94 · see citing works
Small animal PET: a review of what we have done and where we are going
Miyaoka, R. S., & Lehnert, A. L. (2020) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab8f71 · Cited by 87 · see citing works
Ko, G. B., Yoon, H. S., Kim, K. Y., et al. (2016) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.170019 · Cited by 80 · see citing works
Quantitative gated PET for the assessment of left ventricular function in small animals.
Croteau, É., Bénard, F., Cadorette, J., et al. (2003) · PubMed · Cited by 77 · see citing works
Self-citationPerformance of a PET Insert for High-Resolution Small-Animal PET/MRI at 7 Tesla
Stortz, G., Thiessen, J. D., Bishop, D., et al. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.116.187666 · Cited by 64 · see citing works
Bayesian reconstruction of photon interaction sequences for high-resolution PET detectors
Pratx, G., & Levin, C. S. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/17/001 · Cited by 62 · see citing works
Imaging performance of LabPET APD-based digital PET scanners for pre-clinical research
Bergeron, M., Cadorette, J., Tétrault, M., et al. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/3/661 · Cited by 58 · see citing works
Self-citationPET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner
Omidvari, N., Cabello, J., Topping, G. J., Schneider, F. R., Paul, S. M., Schwaiger, M., & Ziegler, S. I. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa910d · Cited by 56 · see citing works
Hallen, P., Schug, D., & Schulz, V. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-0279-2 · Cited by 56 · see citing works
A new dual threshold time-over-threshold circuit for fast timing in PET
Grant, A. M., & Levin, C. S. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/13/3421 · Cited by 54 · see citing works
NEMA NU-4 Performance Evaluation of the IRIS PET/CT Preclinical Scanner
Belcari, N., Camarlinghi, N., Ferretti, S., et al. (2017) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2017.2707300 · Cited by 50 · see citing works
Wang, L., Zhu, J. J., Liang, X., Niu, M., Wu, X., Kao, C.-M., Kim, H., & Xie, Q. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/1/137 · Cited by 45 · see citing works
Hallen, P., Schug, D., Weissler, B., et al. (2018) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/aae6c2 · Cited by 36 · see citing works
SimPET: a Preclinical PET Insert for Simultaneous PET/MR Imaging
Son, J., Kim, K. Y., Park, J. Y., Kim, K., Lee, Y., Ko, G. B., & Lee, J. S. (2020) · Molecular Imaging and Biology · doi:10.1007/s11307-020-01491-y · Cited by 36 · see citing works
Michaud, J.-B., Tétrault, M., Beaudoin, J.-F., et al. (2014) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2014.2372788 · Cited by 32 · see citing works
Self-citationNEMA NU‐4 2008 performance evaluation of Xtrim‐PET: A prototype SiPM‐based preclinical scanner
Amirrashedi, M., Sarkar, S., Ghafarian, P., Shahraki, R. H., Geramifar, P., Zaidi, H., & Ay, M. R. (2019) · Medical Physics · doi:10.1002/mp.13785 · Cited by 31 · see citing works
Vrigneaud, J., McGrath, J., Courteau, A., et al. (2018) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aac4f7 · Cited by 26 · see citing works
Development of an Ultra High Resolution PET Scanner for Imaging Rodent Paws: PawPET
Godinez, F., Gong, K., Zhou, J., et al. (2017) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2017.2765486 · Cited by 18 · see citing works
A preclinical PET detector constructed with a monolithic scintillator ring
Xu, J., Xie, S., Zhang, X., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab2ca4 · Cited by 18 · see citing works
Dual-threshold Time-over-Threshold nonlinearity correction for PET detectors
Gaudin, É., Thibaudeau, C., Arpin, L., Bouziri, H., Fontaine, R., & Lecomte, R. (2020) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2020.164100 · Cited by 17 · see citing works
Self-citationEmbedded real time digital signal processing unit for a 64-channel PET detector module
Arpin, L., Koua, K., Panier, S., et al. (2011) · 2006 IEEE Nuclear Science Symposium Conference Record · doi:10.1109/nssmic.2011.6154367 · Cited by 16 · see citing works
Self-citationA new pulse width signal processing with delay-line and non-linear circuit (for ToT)
Orita, T., Takahashi, H., Shimazoe, K., et al. (2011) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2011.01.023 · Cited by 14 · see citing works
NEMA NU 4-2008 and in vivo imaging performance of RAYCAN trans-PET/CT X5 small animal imaging system
Teuho, J., Han, C. Y., Riehakainen, L., Honkaniemi, A., Tirri, M. E., Liljenbäck, H., Virta, J., Gu, S., Liu, S., Wan, L., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab1856 · Cited by 14 · see citing works
Real Time Artificial Neural Network FPGA Implementation for Triple Coincidences Recovery in PET
Geoffroy, C., Michaud, J.-B., Tétrault, M., et al. (2015) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2015.2432754 · Cited by 13 · see citing works
NEMA NU-4 performance evaluation of a non-human primate animal PET
Chai, P., Baotong, F., Zhang, Z., Haohui, T., Shuangquan, L., Xiaoli, S., Peilin, W., Xiaoming, W., Xu-dong, Z., & Long, W. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab1614 · Cited by 9 · see citing works
Lee, S., Choi, Y., Jung, J. H., et al. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aafeff · Cited by 8 · see citing works
Development and preliminary results of Xtrim-PET, a modular cost-effective preclinical scanner
Sajedi, S., Zeraatkar, N., Taheri, M., Kaviani, S. I., Khanmohammadi, H., Sarkar, S., Sabet, H., & Ay, M. R. (2019) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2019.06.009 · Cited by 8 · see citing works
Performance evaluation of the PET component of a sequential APD-based micro-PET/MR imaging system
Vrigneaud, J., Walker, P., Barbier, B., et al. (2017) · Biomedical Physics & Engineering Express · doi:10.1088/2057-1976/aa686d · Cited by 7 · see citing works
Rucher, G., Camelière, L., Fendri, J., et al. (2018) · Molecular Imaging and Biology · doi:10.1007/s11307-018-1202-2 · Cited by 5 · see citing works
A Monte Carlo detector response model for the IRIS PET preclinical scanner
Pilleri, A., Camarlinghi, N., Guerra, A. D., Sportelli, G., & Belcari, N. (2019) · Physica Medica · doi:10.1016/j.ejmp.2018.12.035 · Cited by 2 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Fontaine R, Arpin L, Bouchard J, Paille M, Bergeron M, Pepin C-M, Cadorette J & Lecomte R (2016). The hardware architecture of the LabPET II-mouse, a highly integrated APD-based PET scanner, 2016 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Vol. M06–2, pp. 1– 3. Strasbourg, France, Oct. 29-Nov. 6, 2016.
- Leroux J-D (2014). Contribution aux méthodes de reconstruction d’images appliquées à la tomographie d’émission par positrons par l’exploitation des symétries du système, PhD thesis, Université de Sherbrooke, Sherbrooke (Québec) Canada.
- McKechnie TS (2016). Telescope Resolution and Optical Tolerance Specifications, Springer International Publishing, Cham, pp. 405– 464.
Cited by
Sorted by citations
Toyonaga, T., Fesharaki-Zadeh, A., Strittmatter, S. M., Carson, R. E., & Cai, Z. (2022) · Frontiers in Neuroscience · doi:10.3389/fnins.2022.787404 · Cited by 20 · see citing works
Asch, R. H., Abdallah, C. G., Carson, R. E., & Esterlis, I. (2025) · Neuropsychopharmacology · doi:10.1038/s41386-024-01913-3 · Cited by 13 · see citing works
- Authors
- Gaudin, É., Thibaudeau, C., Arpin, L., Leroux, J.-D., Toussaint, M., Beaudoin, J.-F., Cadorette, J., Paillé, M., Pepin, C. M., Koua, K., Bouchard, J., Viscogliosi, N., Paulin, C., Fontaine, R., & Lecomte, R.
- Affiliations
- Université de Sherbrooke (CA)
- IR&T Inc., Sherbrooke, QC,
- Novalgo Inc., Sherbrooke, QC,
- Sherbrooke Molecular Imaging Center and Department of Nuclear Medicine and Radiobiology, Université de Sherbrooke, Sherbrooke, QC,
- ORCID
- Venue
- Physics in Medicine & Biology
- Publisher
- IOP Publishing
- Published
- 2021-03-08
- Volume
- 66
- Issue
- 6
- Pages
- 065019
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- PMC
- PMC7979522
- OpenAlex
- W3118791832
- Cited by
- 16
- References
- 40
- Citations by year
- 2025: 3 · 2024: 5 · 2023: 3 · 2022: 3 · 2021: 2
- FWCI
- 1.72
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Physics in Medicine and Biology (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Natural Sciences and Engineering Research Council of Canada
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Keywords
- positron emission tomography, small animal imaging, mouse imaging, noise equivalent count rate, absolute sensitivity, bone imaging, biomolecular imaging
- MeSH terms
- Animals, Brain, Calibration, Electronics, Equipment Design, Humans, Image Processing, Computer-Assisted, Sensitivity and Specificity, Reproducibility of Results, Phantoms, Imaging, Fluorodeoxyglucose F18, Imaging, Three-Dimensional, Positron-Emission Tomography, Mice, Whole Body Imaging
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review