Advantages of soft subdural implants for the delivery of electrochemical neuromodulation therapies to the spinal cord
Open accessThe DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: We recently developed soft neural interfaces enabling the delivery of electrical and chemical stimulation to the spinal cord. These stimulations restored locomotion in animal models of paralysis. Soft interfaces can be placed either below or above the dura mater. Theoretically, the subdural location combines many advantages, including increased selectivity of electrical stimulation, lower stimulation thresholds, and targeted chemical stimulation through local drug delivery. However, these advantages have not been documented, nor have their functional impact been studied in silico or in a relevant animal model of neurological disorders using a multimodal neural interface. APPROACH: We characterized the recruitment properties of subdural interfaces using a realistic computational model of the rat spinal cord that included explicit representation of the spinal roots. We then validated and complemented computer simulations with electrophysiological experiments in rats. We additionally performed behavioral experiments in rats that received a lateral spinal cord hemisection and were implanted with a soft interface. MAIN RESULTS: In silico and in vivo experiments showed that the subdural location decreased stimulation thresholds compared to the epidural location while retaining high specificity. This feature reduces power consumption and risks of long-term damage in the tissues, thus increasing the clinical safety profile of this approach. The hemisection induced a transient paralysis of the leg ipsilateral to the injury. During this period, the delivery of electrical stimulation restricted to the injured side combined with local chemical modulation enabled coordinated locomotor movements of the paralyzed leg without affecting the non-impaired leg in all tested rats. Electrode properties remained stable over time, while anatomical examinations revealed excellent bio-integration properties. SIGNIFICANCE: Soft neural interfaces inserted subdurally provide the opportunity to deliver electrical and chemical neuromodulation therapies using a single, bio-compatible and mechanically compliant device that effectively alleviates locomotor deficits after spinal cord injury.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078
Electronic dura mater for long-term multimodal neural interfaces
Minev, I. R., Musienko, P., Hirsch, A., Barraud, Q., Wenger, N., Moraud, E. M., Gandar, J., Capogrosso, M., Milekovic, T., Asboth, L., et al. (2015) · Science · doi:10.1126/science.1260318 · Cited by 1,082
Self-citationModels of recruitment and rate coding organization in motor-unit pools.
Fuglevand, A. J., Winter, D. A., & Patla, A. E. (1993) · Journal of Neurophysiology · doi:10.1152/jn.1993.70.6.2470 · Cited by 988
Restoring Voluntary Control of Locomotion after Paralyzing Spinal Cord Injury
van den Brand, R., Heutschi, J., Barraud, Q., DiGiovanna, J., Bartholdi, K., Huerlimann, M., Friedli, L., Vollenweider, I., Moraud, E. M., Duis, S., et al. (2012) · Science · doi:10.1126/science.1217416 · Cited by 767 · cited 1×
“Kinematics, kinetics, and muscle activity were recorded following protocols that have been described in details previously (Van den Brand 2012, Wenger 2016).”
MethodSelf-citationAngeli, C. A., Edgerton, V. R., Gerasimenko, Y. P., & Harkema, S. J. (2014) · Brain · doi:10.1093/brain/awu038 · Cited by 750
Transformation of nonfunctional spinal circuits into functional states after the loss of brain input
Courtine, G., Gerasimenko, Y., van den Brand, R., Yew, A., Musienko, P., Zhong, H., Song, B., Ao, Y., Ichiyama, R. M., Lavrov, I., et al. (2009) · Nature Neuroscience · doi:10.1038/nn.2401 · Cited by 706 · cited 1×
“While these results offer encouraging perspectives for clinical applications, rodents studies also emphasized that the recovery of functional locomotor movements after more severe forms of spinal cord injury was contingent on the addition of a serotonergic replacement therapy (Courtine et al 2009, van den Brand et al…”
Self-citationA brain–spine interface alleviating gait deficits after spinal cord injury in primates
Capogrosso, M., Milekovic, T., Borton, D., Wagner, F., Moraud, E. M., Mignardot, J.-B., Buse, N., Gandar, J., Barraud, Q., Xing, D., et al. (2016) · Nature · doi:10.1038/nature20118 · Cited by 635
Self-citationGosselin, M.-C., Neufeld, E., Moser, H., Huber, E., Farcito, S., Gerber, L., Jedensjö, M., Hilber, I., Gennaro, F. D., Lloyd, B., et al. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/18/5287 · Cited by 520
A Computational Model for Epidural Electrical Stimulation of Spinal Sensorimotor Circuits
Capogrosso, M., Wenger, N., Raspopovic, S., Musienko, P., Beauparlant, J., Bassi Luciani, L., Courtine, G., & Micera, S. (2013) · The Journal of Neuroscience · doi:10.1523/jneurosci.1688-13.2013 · Cited by 435 · cited 3×
“…to compute the electric potential distribution over the volume conductor during electrical stimulation (Bossetti et al 2008, Raspopovic et al 2011, Capogrosso et al 2013, Lempka et al 2015), reducing the Maxwell’s equations to the Laplace equation ∇ · j(x) = 0, ∀x ∈ Ω where j(x) = σ(x) E(x) is…”
BackgroundResultInfluentialSelf-citationWenger, N., Moraud, E. M., Gandar, J., Musienko, P., Capogrosso, M., Baud, L., Le Goff, C. G., Barraud, Q., Pavlova, N., Dominici, N., et al. (2016) · Nature Medicine · doi:10.1038/nm.4025 · Cited by 369
Self-citationMoraud, E. M., Capogrosso, M., Formento, E., Wenger, N., DiGiovanna, J., Courtine, G., & Micera, S. (2016) · Neuron · doi:10.1016/j.neuron.2016.01.009 · Cited by 220
Self-citationWenger, N., Moraud, E. M., Raspopovic, S., Bonizzato, M., DiGiovanna, J., Musienko, P., Morari, M., Micera, S., & Courtine, G. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008325 · Cited by 209 · cited 2×
“Bipedal locomotion was recorded on a treadmill using a robotic bodyweight support system (Robomedica) that was adjusted to provide optimal assistance during bipedal stepping (Wenger et al 2014).”
Method
Cited by
Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454
“Particularly in neuroprostheses implanted in nervous tissue—some of the softest tissue in the body—implants implementing flexible or stretchable materials can achieve a much lower degree of FBR and tissue damage when chronically implanted (Nguyen et al., 2014; Minev et al., 2015; Lacour et al., 2016; Capogrosso et al.…”
BackgroundGreiner, N., Barra, B., Schiavone, G., Lorach, H., James, N., Conti, S., Kaeser, M., Fallegger, F., Borgognon, S., Lacour, S., et al. (2021) · Nature Communications · doi:10.1038/s41467-020-20703-1 · Cited by 199
Self-citationCarnicer-Lombarte (2019) · doi:10.1101/829648 · Cited by 46
Metadata
- Authors
- Capogrosso, M., Gandar, J., Greiner, N., Moraud, E. M., Wenger, N., Shkorbatova, P., Musienko, P., Minev, I., Lacour, S., & Courtine, G.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2018-02-16
- Volume
- 15
- Issue
- 2
- Pages
- 026024
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2791986753
- Cited by
- 48
- References
- 12
- FWCI
- 3.20
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.