In vivo impedance spectroscopy of deep brain stimulation electrodes
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Deep brain stimulation (DBS) represents a powerful clinical technology, but a systematic characterization of the electrical interactions between the electrode and the brain is lacking. The goal of this study was to examine the in vivo changes in the DBS electrode impedance that occur after implantation and during clinically relevant stimulation. Clinical DBS devices typically apply high-frequency voltage-controlled stimulation, and as a result, the injected current is directly regulated by the impedance of the electrode-tissue interface. We monitored the impedance of scaled-down clinical DBS electrodes implanted in the thalamus and subthalamic nucleus of a rhesus macaque using electrode impedance spectroscopy (EIS) measurements ranging from 0.5 Hz to 10 kHz. To further characterize our measurements, equivalent circuit models of the electrode-tissue interface were used to quantify the role of various interface components in producing the observed electrode impedance. Following implantation, the DBS electrode impedance increased and a semicircular arc was observed in the high-frequency range of the EIS measurements, commonly referred to as the tissue component of the impedance. Clinically relevant stimulation produced a rapid decrease in electrode impedance with extensive changes in the tissue component. These post-operative and stimulation-induced changes in impedance could play an important role in the observed functional effects of voltage-controlled DBS and should be considered during clinical stimulation parameter selection and chronic animal research studies.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Foreign body reaction to biomaterials
Anderson, J. M., Rodriguez, A., & Chang, D. T. (2008) · Seminars in Immunology · doi:10.1016/j.smim.2007.11.004 · Cited by 5,138 · cited 1×
“…was further supported by the time course of these impedance changes that closely paralleled the time course for establishing the chronic foreign body reaction that is characterized by the formation of foreign body giant cells and granulation tissue around the implanted device (Anderson et al 2008).”
BackgroundBiran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893 · cited 4×
“Direct adhesion of cells and the development of an encapsulation layer around electrodes implanted in the central nervous system has been documented for both cortical microelectrode (Szarowski et al., 2003; Biran et al., 2005) and DBS applications (Haberler et al.”
BackgroundMethodInfluentialUnterberg, A. W., Stover, J., Kress, B., & Kiening, K. L. (2004) · Neuroscience · doi:10.1016/j.neuroscience.2004.06.046 · Cited by 881 · cited 2×
“These changes in the electrode-electrolyte interface impedance could also be due to changes in the ionic composition of the cerebrospinal fluid near the electrode during vasogenic and cellular edema (Unterberg et al., 2004).”
Background
Cited by
Glial responses to implanted electrodes in the brain
Salatino, J. W., Ludwig, K. A., Kozai, T. D. Y., & Purcell, E. K. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0154-1 · Cited by 642
“For constant-voltage stimulation paradigms, the day-to-day changes in impedance caused by consolidation of the glial scar during the first three to six months post-implantation may dramatically alter the effective current reaching neural tissue 39 .”
Bilateral adaptive deep brain stimulation is effective in Parkinson's disease
Little, S., Beudel, M., Zrinzo, L., Foltynie, T., Limousin, P., Hariz, M., Neal, S., Cheeran, B., Cagnan, H., Gratwicke, J., et al. (2016) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp-2015-310972 · Cited by 385
Lozano, A. M., Giacobbe, P., Hamani, C., Rizvi, S. J., Kennedy, S. H., Kolivakis, T. T., Debonnel, G., Sadikot, A. F., Lam, R. W., Howard, A. K., et al. (2012) · Journal of Neurosurgery · doi:10.3171/2011.10.jns102122 · Cited by 356
Koeglsperger, T., Palleis, C., Hell, F., Mehrkens, J. H., & Bötzel, K. (2019) · Frontiers in Neurology · doi:10.3389/fneur.2019.00410 · Cited by 235
“In addition to the micro-lesion effect, the fluctuation of impedances may bias the determination of the therapeutic window in the early post-operative period (230) which might become more relevant hen using constant-voltage stimulation (CVS) where the current delivered is inversely proportional to the”
MethodAdvances in closed-loop deep brain stimulation devices
Parastarfeizabadi, M., & Kouzani, A. Z. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0295-1 · Cited by 232
“during the first 3 months following DBS surgery [95, 96] as well as post-implantation [26, 97, 98].”
Background
Metadata
- Authors
- Lempka, S. F., Miocinovic, S., Johnson, M. D., Vitek, J. L., & McIntyre, C. C.
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2009-06-03
- Volume
- 6
- Issue
- 4
- Pages
- 046001
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2161248222
- Cited by
- 229
- References
- 3
- FWCI
- 4.50
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: In vivo impedance spectroscopy of deep brain stimulation electrodes. https://app.cochat.ai/papers/10.1088/1741-2560/6/4/046001
- The paper
- Lempka, S. F., Miocinovic, S., Johnson, M. D., Vitek, J. L., & McIntyre, C. C. (2009). In vivo impedance spectroscopy of deep brain stimulation electrodes. Journal of Neural Engineering. https://doi.org/10.1088/1741-2560/6/4/046001
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.