Cascade Impactors for the Size Characterization of Aerosols from Medical Inhalers: Their Uses and Limitations
The DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
Cascade impactors, including the multi-stage liquid impinger, are by far the most widely encountered means for the in vitro determination of the particle size distribution of aerosols from medical inhalers, both in product development, batch release and in applications with add-on devices. This is because they directly measure aerodynamic size, which is the most relevant parameter to describe particle transport within the respiratory tract. At the same time, it is possible to quantify the mass of active pharmaceutical ingredient in different size ranges independent of other non-physiologically active components of the formulation. We begin by providing an overview of the operating principles of impactors and then highlight the various configurations and adaptations that have been adopted to characterize the various classes of inhaler. We continue by examining the limitations of the cascade impaction method, in particular looking at potential sources of measurement bias and discussing both appropriate and inappropriate uses of impactor-generated data. We also present a synopsis of current developments, including the Next Generation Pharmaceutical Impactor, and automation of cascade impactors for routine inhaler performance measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Respiratory Drug Delivery (1989)
Byron, P. R. (2018) · CRC Press eBooks · doi:10.4324/9780203710456 · Cited by 449
CLARK, A. R., & HOLLINGWORTH, A. M. (1993) · Journal of Aerosol Medicine · doi:10.1089/jam.1993.6.99 · Cited by 332 · cited 1×
“146,150 For the ideal inertial separator, it can be shown by application of equations 3 or 4, that the stage cut-CASCADE which is the basis of the calculation provided in the current compendial methods, where d 50,ref is fixed at 60 L/min.”
Canadian standards association
Husseini, A. (1993) · Journal of Cleaner Production · doi:10.1016/0959-6526(93)90012-z · Cited by 326
Characteristics of laminar jet impactors
Marple, V. A., & Liu, B. Y. H. (1974) · Environmental Science & Technology · doi:10.1021/es60092a003 · Cited by 293
Co-Chairmen of the Task Force:, Boe, J., Dennis, J. H., O'Driscoll, B. R., Members of Task Force:, Bauer, T. T., Carone, M., Dautzenberg, B., Diot, P., Heslop, K., et al. (2001) · European Respiratory Journal · doi:10.1183/09031936.01.00220001 · Cited by 280 · cited 1×
“80 The total flow rate of 15 L/min was chosen to match the mid-point flow of the sinusoidal pattern proposed for breathing simulator-based testing of these devices.”
Dispersion and Characterization of Pharmaceutical Dry Powder Aerosols
Dunber, C. A., Hickey, A. J., & Holzner, P. (1998) · KONA Powder and Particle Journal · doi:10.14356/kona.1998007 · Cited by 197
(2001) · CRC Press eBooks · doi:10.1201/b14022 · Cited by 153
May, K. R. (1966) · Bacteriological Reviews · doi:10.1128/br.30.3.559-570.1966 · Cited by 152
Marple, V. A., Olson, B. A., Santhanakrishnan, K., Mitchell, J. P., Murray, S. C., & Hudson-Curtis, B. L. (2003) · Journal of Aerosol Medicine · doi:10.1089/089426803769017668 · Cited by 151
Self-citationClark, A. R. (1995) · International Journal of Pharmaceutics · doi:10.1016/0378-5173(94)00255-4 · Cited by 145 · cited 1×
“70,71 Currently available light scattering methods, however, do not determine the mass of API and may therefore lead to misleading results, particularly with suspension formulations where droplets may be formed without containing any drug-bearing particles.”
Suman, J., Laube, B. L., & Dalby, R. N. (1999) · Pharmaceutical Research · doi:10.1023/a:1011933410898 · Cited by 140 · cited 1×
“9 They are therefore unsuitable to characterize aqueous droplet sprays produced by mechanical pump-driven nasal drug delivery devices, 10 other than to quantify the mass fraction, typically defined as being , 10 m m aerodynamic diameter, that is likely to penetrate beyond the r p C p } xr 0 nasal…”
Medical Aerosol Inhalers: Past, Present, and Future
Clark, A. R. (1995) · Aerosol Science and Technology · doi:10.1080/02786829408959755 · Cited by 135
A Personal Cascade Impactor Design, Evaluation and Calibration
Rubow, K. L., Marple, V. A., Olin, J. G., & McCawley, M. (1987) · American Industrial Hygiene Association Journal · doi:10.1080/15298668791385174 · Cited by 131 · cited 1×
“77 The impactor was developed for workplace personal exposure monitoring of particulates, and like the ACI, has subsequently been adapted for inhaler testing.”
Modelling and algebraic formulation of regional aerosol deposition in man
Rudolf, G., Köbrich, R., & STAHLHOFEN, W. (1990) · Journal of Aerosol Science · doi:10.1016/0021-8502(90)90266-z · Cited by 112
Lung Deposition and Respirable Mass During Wet Nebulization
Sangwan, S., Condos, R., & Smaldone, G. C. (2003) · Journal of Aerosol Medicine · doi:10.1089/089426803772455631 · Cited by 53 · cited 1×
“Nevertheless, indirect evidence of its suitability is given in an article in this issue, 104 which demonstrates a correlation between lung deposition based on g -scintigraphy and CI-measured FPF , 2.5 m m .”
InfluentialChristopher, D., Curry, P., Doub, B., Furnkranz, K., Lavery, M., Lin, K. K., Lyapustina, S., Mitchell, J. P., Rogers, B., Strickland, H., et al. (2003) · Journal of Aerosol Medicine · doi:10.1089/089426803769017604 · Cited by 50 · cited 4×
“39 Since the pressure differential within the CI and the surrounding ambient atmosphere is greatest at the stages nearest to the vacuum source, it follows that air leakage due to sealing defects are most likely to occur at these locations.”
InfluentialSelf-citationIn Vitro Testing of Two Formoterol Dry Powder Inhalers at Different Flow Rates
Weuthen, T., Roeder, S., Brand, P., Müllinger, B., & Scheuch, G. (2002) · Journal of Aerosol Medicine · doi:10.1089/089426802760292636 · Cited by 49 · cited 4×
“Although most investigators utilize the com-pendial method for evaluating DPIs, the study by Weuthen et al. 152 is of significance, in that it employed an alternative strategy making it possible to characterize inhaler performance at a range of different flow rates from 28.3 to 80 L/min, whilst operating an ACI at a f…”
InfluentialBasic Principles of Particle Behavior in the Human Respiratory Tract
Heyder, J., & Svartengren, M. (2001) · Lung biology in health and disease · doi:10.1201/b14022-3 · Cited by 48 · cited 2×
“7 In response to this concern, the reporting of performance data based on two sub-fractions (fine component having d ae , 4.7 m m [FPF 4.7 m m ] and extra-fine component with d ae , 1.1 m m [EPF 1.1 m m ]) has been proposed in a recently developed Canadian Standard for spacers and holding chambers.”
InfluentialPhysical and chemical properties of aerosols
Colbeck, I. (1998) · Blackie Academic & Professional eBooks · Cited by 42
Thiel, C. G. (1998) · Journal of Aerosol Medicine · doi:10.1089/jam.1998.11.suppl_1.s-43 · Cited by 30 · cited 3×
“23 In an effort to minimize the amount of data required for the description of inhaler-based aerosols, Thiel, 29 like Berg et al., 23 proposed sep-arating the dose captured in the induction port and pre-separator (if used), from the “non-ballis-tic fraction” (NBF) entering the impactor.”
Mitchell, J. P., NAGEL, M. W., & Rau, J. L. (1999) · Respiratory Care · doi:10.1177/002013249904400106 · Cited by 13 · cited 1×
“The additional information is particularly useful for the assessment of add-on devices with pMDIs, 40,41 where it may be important to have an indication of the magnitude of reduction of the coarse component brought about by the spacer or holding chamber.”
Self-citationDetermination of Large Aerosol Particle Size by Elutriation
TILLERY, M. I., & Buchan, R. M. (2002) · Applied Occupational and Environmental Hygiene · doi:10.1080/10473220290106695 · Cited by 6 · cited 1×
“11 Elutriators that operate on the principle of sedimentation have been developed for size separating larger particles, 12 but these devices have not so far been widely applied to the testing of either nasal or oral inhalers.”
Cited by
Influence of particle size on regional lung deposition – What evidence is there?
Carvalho, T. C., Peters, J. I., & Williams III, R. O. (2011) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2010.12.040 · Cited by 614
Dry Powder Inhaler Formulation
Telko, M. J., & Hickey, A. J. (2005) · Respiratory Care · doi:10.4187/respcare.05501209 · Cited by 498
Fromen, C. A., Rahhal, T. B., Robbins, G. R., Kai, M. P., Shen, T. W., Luft, J. C., & DeSimone, J. M. (2016) · Nanomedicine: Nanotechnology, Biology and Medicine · doi:10.1016/j.nano.2015.11.002 · Cited by 182
Metadata
- Authors
- Mitchell, J. P., & Nagel, M. W.
- Venue
- Journal of Aerosol Medicine
- Publisher
- SAGE Publications
- Published
- 2003-12
- Volume
- 16
- Issue
- 4
- Pages
- 341-377
- ISSN
- 0894-2684 (ISSN-L)
- 1557-9026
- OpenAlex
- W2103012009
- Cited by
- 189
- References
- 22
- FWCI
- 4.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.