The Accuracy of the CGMS ™ in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy Study
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
The accuracy of the Continuous Glucose Monitoring System, CGMS (Medtronic MiniMed, Northridge, CA) was assessed in children and adolescents with type 1 diabetes mellitus (T1DM) when compared with reference serum glucose levels during spontaneous fluctuations in glucose levels over 24 h and during acute hyper- and hypoglycemia. Ninety-one subjects with type 1 diabetes (3.5-17.7 years old) wore one or two CGMSs while blood samples were obtained for serum glucose determinations (made at a central laboratory) hourly during the day, every 30 min overnight, and every 5 min during meal-induced hyperglycemia and insulin-induced hypoglycemia tests, resulting in 6778 CGMS-reference glucose pairs. CGMS function was assessed on each of the 3 days of sensor life. The median relative absolute difference (RAD) between the CGMS and reference values was 18% (25th, 75th percentiles = 8%, 34%). Similar results were obtained on each of the 3 days of sensor life. Accuracy was worse during hypoglycemia than during hyperglycemia. Modified sensors that first became available in November 2002 were more accurate than were the original sensors (median RAD = 11% vs. 19%) and had better precision (r = 0.92 vs. r = 0.77) during time periods in which two CGMSs were simultaneously used. The CGMS sensors that have been in clinical use until recently are often inaccurate in quantifying glucose values in children with T1DM. However, recent modifications to the sensor have resulted in substantially better accuracy and reliability. This improved function, if confirmed by additional data, may enhance the clinical utility of the CGMS.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Skeie, S., Thue, G., Nerhus, K., & Sandberg, S. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.7.994 · Cited by 131 · cited 1×
“10 These criteria were met 96% of the time by our 1,741 Ultra glucose values but were met only 56% of the time by the CGMS values.”
Amin, R., Ross, K., Acerini, C. L., Edge, J. A., Warner, J., & Dunger, D. B. (2003) · Diabetes Care · doi:10.2337/diacare.26.3.662 · Cited by 106
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Care of Children and Adolescents With Type 1 Diabetes
Silverstein, J., Klingensmith, G., Copeland, K., Plotnick, L., Kaufman, F., Laffel, L., Deeb, L., Grey, M., Anderson, B., Holzmeister, L. A., et al. (2005) · Diabetes Care · doi:10.2337/diacare.28.1.186 · Cited by 1,312
Klonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Murphy, H. R., Rayman, G., Lewis, K., Kelly, S., Johal, B., Duffield, K., Fowler, D., Campbell, P. J., & Temple, R. C. (2008) · BMJ · doi:10.1136/bmj.a1680 · Cited by 358
Metadata
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2003-10
- Volume
- 5
- Issue
- 5
- Pages
- 781-789
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W4253301460
- Cited by
- 113
- References
- 3
- FWCI
- 9.32
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The Accuracy of the CGMS ™ in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy Study. https://app.cochat.ai/papers/10.1089/152091503322526987
- The paper
- The Accuracy of the CGMS ™ in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy Study. (2003). Diabetes Technology & Therapeutics. https://doi.org/10.1089/152091503322526987
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.