CoChat

The Accuracy of Home Glucose Meters in Hypoglycemia

Sonmez, Alper, Yilmaz, Zeynep, Uckaya, Gokhan, Kilic, Selim, Tapan, Serkan, Taslipinar, Abdullah, Aydogdu, Aydogan, Yazici, Mahmut, Yilmaz, Mahmut Ilker, Serdar, Muhittin, Erbil, M. Kemal, Kutlu, Mustafa · 2010 · Diabetes Technology & Therapeutics
Cited by 23

The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.

Abstract

Background: Home glucose meters (HGMs) may not be accurate enough to sense hypoglycemia. We evaluated the accuracy and the capillary and venous comparability of five different HGMs (Optium Xceed™ [Abbott Diabetes Care, Alameda, CA, USA], Contour ® TS [Bayer Diabetes Care, Basel, Switzerland], Accu-Chek ® Go [Roche Ltd., Basel, Switzerland], OneTouch ® Select™ [Lifescan, Milpitas, CA, USA], and EZ Smart [Tyson Bioresearch Inc., Chu-Nan, Taiwan]) in an adult population. Methods: The insulin hypoglycemia test was performed to 59 subjects (56 males; 23.6 ± 3.2 years old). Glucose was measured from forearm venous blood and finger capillary samples both before and after regular insulin (0.1 U/kg) was injected. Venous samples were analyzed in the reference laboratory by the hexokinase method. In vitro tests for method comparison and precision analyses were also performed by spiking the glucose-depleted venous blood. Results: All HGMs failed to sense hypoglycemia to some extend. EZ Smart was significantly inferior in critical error Zone D, and OneTouch Select was significantly inferior in the clinically unimportant error Zone B. Accu-Chek Go, Optium Xceed, and Contour TS had similar performances and were significantly better than the other two HGMs according to error grid analysis or International Organization for Standardization criteria. The in vitro tests were consistent with the above clinical data. The capillary and venous consistencies of Accu-Chek Go and OneTouch Select were better than the other HGMs. Conclusions: The present results show that not all the HGMs are accurate enough in low blood glucose levels. The patients and the caregivers should be aware of these restrictions of the HGMs and give more credit to the symptoms of hypoglycemia than the values obtained by the HGMs. Finally, these results indicate that there is a need for the revision of the accuracy standards of HGMs in low blood glucose levels.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2009.0183

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 23

Metadata

Authors
Sonmez, A., Yilmaz, Z., Uckaya, G., Kilic, S., Tapan, S., Taslipinar, A., Aydogdu, A., Yazici, M., Yilmaz, M. I., Serdar, M., Erbil, M. K., & Kutlu, M.
Venue
Diabetes Technology & Therapeutics
Published
2010-08
Volume
12
Issue
8
Pages
619-626
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2069519004
Cited by
23
References
7
FWCI
1.76
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.