Analytical Performance of Glucose Monitoring Systems at Different Blood Glucose Ranges and Analysis of Outliers in a Clinical Setting

Open access
Hasslacher, Christoph, Kulozik, Felix, Platten, Isabel · 2014 · Journal of Diabetes Science and Technology

The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.

Abstract

Background: We investigated the analytical accuracy of 27 glucose monitoring systems (GMS) in a clinical setting, using the new ISO accuracy limits. In addition to measuring accuracy at blood glucose (BG) levels < 100 mg/dl and > 100 mg/dl, we also analyzed devices performance with respect to these criteria at 5 specific BG level ranges, making it possible to further differentiate between devices with regard to overall performance. Methods: Carbohydrate meals and insulin injections were used to induce an increase or decrease in BG levels in 37 insulin-dependent patients. Capillary blood samples were collected at 10-minute intervals, and BG levels determined simultaneously using GMS and a laboratory-based method. Results obtained via both methods were analyzed according to the new ISO criteria. Results: Only 12 of 27 devices tested met overall requirements of the new ISO accuracy limits. When accuracy was assessed at BG levels < 100 mg/dl and > 100 mg/dl, criteria were met by 14 and 13 devices, respectively. A more detailed analysis involving 5 different BG level ranges revealed that 13 (48.1%) devices met the required criteria at BG levels between 50 and 150 mg/dl, whereas 19 (70.3%) met these criteria at BG levels above 250 mg/dl. The overall frequency of outliers was low. Conclusions: The assessment of analytical accuracy of GMS at a number of BG level ranges made it possible to further differentiate between devices with regard to overall performance, a process that is of particular importance given the user-centered nature of the devices’ intended use.

Overview

Integrity signals

Registry

Crossref · 10.1177/1932296814522804

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Verification

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 16 references

Research works printed but not linked: 2

  1. International Organization for Standardization, 2003, DIN EN ISO 15197
  2. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen, 2008, Dtsch Arztebl, 105, A341

Cited by

See all 24

Metadata

Authors
Hasslacher, C., Kulozik, F., & Platten, I.
Affiliations
Diabetesinstitut Heidelberg (DE)
Venue
Journal of Diabetes Science and Technology
Published
2014-02-18
Volume
8
Issue
3
Pages
466-472
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2163658918
Cited by
22
References
16
Citations by year
2024: 1 · 2023: 1 · 2022: 2 · 2021: 1 · 2020: 1 · 2019: 1 · 2018: 2 · 2017: 4 · 2016: 3 · 2015: 6
FWCI
3.49
Type
journal article
Language
en
Open access
Licence
journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
Keywords
blood glucose, analytical accuracy, type 1 diabetes, capillary blood glucose, outlier analysis, device performance comparison, hyperglycemia
MeSH terms
Blood Glucose, Diabetes Mellitus, Dietary Carbohydrates, Equipment Design, Female, Humans, Hypoglycemic Agents, Insulin, Male, Materials Testing, Middle Aged, Predictive Value of Tests, Time Factors, Blood Glucose Self-Monitoring, Reproducibility of Results, Biomarkers, Treatment Outcome

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Analytical Performance of Glucose Monitoring Systems at Different Blood Glucose Ranges and Analysis of Outliers in a Clinical Setting. https://app.cochat.ai/papers/10.1177/1932296814522804
The paper
Hasslacher, C., Kulozik, F., & Platten, I. (2014). Analytical Performance of Glucose Monitoring Systems at Different Blood Glucose Ranges and Analysis of Outliers in a Clinical Setting. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296814522804

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.