Accuracy and Reliability of Continuous Glucose Monitoring Systems: A Head-to-Head Comparison
Open accessThe DOI matches its Crossref record.
Abstract
OBJECTIVE: This study assessed the accuracy and reliability of three continuous glucose monitoring (CGM) systems. RESEARCH DESIGN AND METHODS: We studied the Animas® (West Chester, PA) Vibe™ with Dexcom® (San Diego, CA) G4™ version A sensor (G4A), the Abbott Diabetes Care (Alameda, CA) Freestyle® Navigator I (NAV), and the Medtronic (Northridge, CA) Paradigm® with Enlite™ sensor (ENL) in 20 patients with type 1 diabetes mellitus. All systems were investigated both in a clinical research center (CRC) and at home. In the CRC, patients received a meal with a delayed and increased insulin dose to induce a postprandial glucose peak and nadir. Hereafter, randomization determined which two of the three systems would be worn at home until the end of functioning, attempting use beyond manufacturer-specified lifetime. Patients performed at least five reference finger sticks per day. An analysis of variance was performed on all data points ≥15 min apart. RESULTS: Overall average mean absolute relative difference (MARD) (SD) measured at the CRC was 16.5% (14.3%) for NAV and 16.4% (15.6%) for ENL, outperforming G4A at 20.5% (18.2%) (P<0.001). Overall MARD when assessed at home was 14.5% (16.7%) for NAV and 16.5 (18.8%) for G4A, outperforming ENL at 18.9% (23.6%) (P=0.006). Median time until end of functioning was similar: 10.0 (1.0) days for G4A, 8.0 (3.5) days for NAV, and 8.0 (1.5) days for ENL (P=0.119). CONCLUSIONS: In the CRC, G4A was less accurate than NAV and ENL sensors, which seemed comparable. However, at home, ENL was less accurate than NAV and G4A. Moreover, CGM systems often show sufficient accuracy to be used beyond manufacturer-specified lifetime.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Continuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 737
Mechanisms of sampling interstitial fluid from skin using a microneedle patch
Samant, P. P., & Prausnitz, M. R. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1716772115 · Cited by 363
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“73,92,112 However, to the best of the authors’ knowledge, a more thorough statistical examination of those concerns is currently missing.”
BackgroundLimits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
“Accuracy measured in clinical settings may be quite different from accuracy observed under at-home conditions [14,22].”
BackgroundSelf-citation
Metadata
- Authors
- Luijf, Y. M., Mader, J. K., Doll, W., Pieber, T., Farret, A., Place, J., Renard, E., Bruttomesso, D., Filippi, A., Avogaro, A., Arnolds, S., Benesch, C., Heinemann, L., & DeVries, O. B. O. T. A. C.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2013-08
- Volume
- 15
- Issue
- 8
- Pages
- 721-726
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2087235336
- Cited by
- 82
- References
- 0
- FWCI
- 8.02
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy and Reliability of Continuous Glucose Monitoring Systems: A Head-to-Head Comparison. https://app.cochat.ai/papers/10.1089/dia.2013.0049
- The paper
- Luijf, Y. M., Mader, J. K., Doll, W., Pieber, T., Farret, A., Place, J., Renard, E., Bruttomesso, D., Filippi, A., Avogaro, A., Arnolds, S., Benesch, C., Heinemann, L., & DeVries, O. B. O. T. A. C. (2013). Accuracy and Reliability of Continuous Glucose Monitoring Systems: A Head-to-Head Comparison. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2013.0049
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.