A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes
Open accessAbstract
Background: In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system. Methods: Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10–14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously. Results: For the entire study period, the mean absolute relative difference (MARD) was 13.2% (95% confidence interval [CI] 12.0%–14.4%). MARD was 13.6% (95% CI 12.1%–15.4%) during week 1 and 12.7% (95% CI 11.5%–13.9%) during week 2. The mean absolute difference (MAD) for the whole study period was 19.8 mg/dL (1.1 mmol/L) (95% CI 17.8–21.8 mg/dL), including 20.5 mg/dL (1.14 mmol/L) during week 1 and 19.0 mg/dL (1.05 mmol/L) during week 2. The overall correlation coefficient was 0.96. For glucose values <72, 72–180, and >180 mg/dL (<4, 4–10, and >10 mmol/L), the MARD was 20.3% (95% CI 17.7%–23.1%), 14.7% (95% CI 13.4%–16%), and 9.6% (95% CI 8.5%–10.8%), respectively, and respective MAD values were 12.3, 17.8, and 23.6 mg/dL (0.69, 0.99, and 1.31 mmol/L). Using the 10-item visual analog scale, patients rated their experience with FreeStyle Libre as generally positive, with mean values ranging from 8.22 to 9.8. Conclusions: FreeStyle Libre had a similar overall MARD as continuous blood glucose monitoring systems in earlier studies when studied in similar at-home conditions. The overall patient satisfaction was high.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The FSL's mean absolute relative difference compared with measured capillary blood glucose levels was 13.2% (95% CI 12.0% to 14.4%).
“One study showed that the FSL’s mean absolute relative difference compared with measured capillary blood glucose levels was 13.2% (95% CI 12.0% to 14.4%) [29].”
From Wearable Technology to Quantify the Nutritional Intake of Adults: Validation Study (2020) · cites this paper as [29]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
No citation lists for this paper
No citation lists for this paper yet — checked again daily.
References 18 · Cited by 173
Research works printed but not linked: 13
- [3]Diabetes Control and Complications Trial Study Group: The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993; 329: 977– 986 8366922
- [4]Ahlén E, Pivodic A, Wedel H, et al.: Glycemic control, renal complications, and current smoking in relation to excess risk of mortality in persons with type 1 diabetes. J Diabetes Sci Technol 2016; 10: 1006– 1014 27301980 PMC5032961
- [6]American Diabetes Association. Standards of medical care for patients with diabetes mellitus. Diabetes Care 2003; 26 Suppl 1: S33– S50 12502618
- [8]Lind M, Polonsky W, Hirsch IB, et al.: Continuous glucose monitoring vs conventional therapy for Glycemic Control in adults with type 1 diabetes treated with multiple daily insulin Injections: The GOLD Randomized Clinical Trial. JAMA 2017; 317: 379– 387 28118454
- [12]Hannestad U, Lundblad A: Accurate and precise isotope dilution mass spectrometry method for determining glucose in whole blood. Clin Chem 1997; 43: 794– 800 9166233
- [13]Kos S, van Meerkerk A, van der Linden J, et al.: Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care. Clin Chem Lab Med 2012; 50: 1573– 1580 23092803
- [14]Andelin M, Kropff J, Matuleviciene V, et al.: Assessing the accuracy of continuous glucose monitoring (CGM) calibrated with capillary values using capillary or venous glucose levels as a reference. J Diabetes Sci Technol 2016; 10: 876– 884 26810924 PMC4928217
- [15]Matuleviciene V, Joseph JI, Andelin M, et al.: A clinical trial of the accuracy and treatment experience of the Dexcom G4 sensor (Dexcom G4 system) and Enlite sensor (guardian REAL-time system) tested simultaneously in ambulatory patients with type 1 diabetes. Diabetes Technol Ther 2014; 16: 759– 767 25233297 PMC4201312
- [16]Bailey T, Bode BW, Christiansen MP, et al.: The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther 2015; 17: 787– 794 26171659 PMC4649725
- [17]Armbruster D: Measurement traceability and US IVD manufacturers: the impact of metrology. Accredit Qual Assur 2009; 14: 393– 398
- [18]Bièvre PD, Kaarls R, Peiser HS, et al.: Measurement principles for traceability in chemical analysis. Accredit Quality Assurance 1996; 1: 3
- [19]Bièvre PD, Kaarls R, Peiser HS, et al.: Protocols for traceability in chemical analysis. Accredit Qual Assur 1997; 2: 270
- [20]Bièvre P, Kaarls R, Peiser HS, et al.: Protocols for traceability in chemical analysis. Accredit Qual Assur 1997; 2: 168
- Authors
- Ólafsdóttir, A. F., Attvall, S., Sandgren, U., Dahlqvist, S., Pivodic, A., Skrtic, S., Theodorsson, E., & Lind, M.
- Affiliations
- NU Hospital Group (SE)
- Sahlgrenska University Hospital (SE)
- Statistiska Konsultgruppen
- AstraZeneca (United Kingdom) (GB)
- Linköping University (SE)
- Corresponding author
- Ólafsdóttir, Arndís F.
- ORCID
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2017-03
- Volume
- 19
- Issue
- 3
- Pages
- 164-172
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC5359691
- OpenAlex
- W2597477321
- Cited by
- 173
- References
- 18
- Citations by year
- 2026: 6 · 2025: 8 · 2024: 9 · 2023: 10 · 2022: 23 · 2021: 19 · 2020: 33 · 2019: 24 · 2018: 31 · 2017: 10
- FWCI
- 21.17
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Technology & Therapeutics (publishedVersion, cc-by-nc)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc)
- PubMed Central (publishedVersion, cc-by-nc)
- PubMed (publishedVersion)
- Funders
- Novo Nordisk
- Novo Nordisk Fonden (NNF15OC0015260)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- FreeStyle Libre, flash glucose monitoring, type 1 diabetes, continuous glucose monitoring, measurement accuracy, mean absolute relative difference, patient experience, self-monitoring of blood glucose, capillary blood glucose, adults with diabetes, hypoglycemia, hyperglycemia, treatment satisfaction, cost-effectiveness
- MeSH terms
- Adolescent, Adult, Aged, Blood Glucose, Diabetes Mellitus, Type 1, Female, Humans, Male, Middle Aged, Blood Glucose Self-Monitoring, Patient Satisfaction, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review